K161447 · Orthovita, Inc. · MQV · Oct 6, 2016 · Orthopedic
Device Facts
Record ID
K161447
Device Name
HydroSet XT
Applicant
Orthovita, Inc.
Product Code
MQV · Orthopedic
Decision Date
Oct 6, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hydroset XT Mixing and Delivery System is intended to be used for the delivery of Hydroset XT to an orthopedic surgical site. HydroSet XT is a self-setting calcium phosphate cement indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, craniofacial, posterolateral spine, and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bone. HydroSet XT is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. HydroSet XT cured in-situ provides an open void/gap filler that can augment provisional hardware (e.g. k-wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support medium and is not intended to provide structural support during the healing process.
Device Story
HydroSet XT is a self-setting calcium phosphate bone cement. The system includes pre-filled liquid and powder syringes, a threaded plunger, and delivery accessories (cannula, luer adapter). Used in orthopedic surgical sites by surgeons to fill non-structural bony voids or gaps. The cement is mixed and delivered via manual plunger advancement or rotation of the threaded plunger. Once delivered, it cures in-situ to act as a temporary void filler and provisional hardware augment (e.g., supporting bone fragments during surgery). It does not provide structural support during the healing process. The device is provided sterile for single use.
Clinical Evidence
Bench testing only. Performance testing evaluated working time (injectability and torque), compression, and setting time (wet field penetration resistance). Bacterial endotoxin testing (BET) performed per USP 39 <85> and Ph. Eur. 8.0 2.6.14 to ensure pyrogenicity limits (<20 EU/Device).
Technological Characteristics
Self-setting calcium phosphate cement. Components: pre-filled liquid and powder syringes. Delivery mechanism: manual or threaded plunger. Sterile, single-use. No electronic components or software.
Indications for Use
Indicated for patients requiring filling of bony voids or gaps in the skeletal system (extremities, craniofacial, posterolateral spine, pelvis) resulting from surgery or traumatic injury. Indicated only for non-load-bearing defects not intrinsic to bony structure stability. Not for structural support during healing.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 6, 2016
Orthovita, Inc. % Ms. Lia Gonzalez Senior Regulatory Affairs Specialist Stryker Corporation 2 Pearl Court Allendale, New Jersey 07401
Re: K161447
Trade/Device Name: HydroSet XT Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV, FMF Dated: September 1, 2016 Received: September 2, 2016
Dear Ms. Gonzalez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Mark N. Melkerson -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The letters are uniformly sized and spaced, contributing to the legibility of the name.
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K161447
Device Name HydroSet XT
## Indications for Use (Describe)
The Hydroset XT Mixing and Delivery System is intended to be used for the delivery of Hydroset XT to an orthopedic surgical site.
HydroSet XT is a self-setting calcium phosphate cement indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, craniofacial, posterolateral spine, and pelvis). These defects may be surgically created or osseous defects created from traumatic injury to the bone. HydroSet XT is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure.
HydroSet XT cured in-situ provides an open void/gap filler that can augment provisional hardware (e.g. k-wires, plates, screws) to help support bone fragments during the surgical procedure. The cured cement acts only as a temporary support medium and is not intended to provide structural support during the healing process.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:12px;"> <span style="font-family:Arial;"> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size:12px;"> <span style="font-family:Arial;"> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESSBELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMBnumber."
{3}------------------------------------------------
Traditional 510(k) Premarket Notification
| 510(k) Summary: HydroSet XT (K161447) | |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Orthovita, Inc.<br>45 Great Valley Parkway<br>Malvern, PA 19355 |
| Contact Person | Lia Gonzalez<br>Sr. Regulatory Affairs Specialist<br>Stryker Spine<br>2 Pearl Court<br>Allendale, NJ 07401<br>Phone: 201-749-8699<br>E-mail: lia.gonzalez@stryker.com |
| Date Prepared | September 1, 2016 |
| Trade Name | HydroSet XT |
| Common Name | Resorbable calcium salt bone void filler device |
| Proposed Class | Class II |
| Classification Name and<br>Number | Filler, Bone Void, Calcium Compound<br>21 CFR §888.3045<br>Syringe, Piston<br>21 CFR §880.5860 |
| Product Code | MQV, FMF |
| Predicate Devices | Predicate Device:<br>Stryker Injectable Cement(K060061)<br>ETEX MIXING AND DELIVERY SYSTEM(K141245) |
| Device Description | HydroSet XT is a self-setting calcium phosphate cement. The device includes a pre-<br>filled liquid syringe and a pre-filled powder syringe utilized for mixing and delivery of<br>the cement. The device also includes the necessary components to prepare and deliver<br>the cement (syringe cap, luer adapter, luer cap, threaded plunger, and cannula). After<br>preparation, the cement may be delivered by traditional plunger advancement or by<br>rotating the threaded plunger to deliver the cement. |
| Indications for Use | The Hydroset XT Mixing and Delivery System is intended to be used for the<br>delivery of Hydroset XT to an orthopedic surgical site.<br><br>HydroSet XT is a self-setting calcium phosphate cement indicated to fill bony voids or<br>gaps of the skeletal system (i.e. extremities, craniofacial, posterolateral<br>spine, and pelvis). These defects may be surgically created or osseous defects<br>created from traumatic injury to the bone. HydroSet XT is indicated only for bony<br>voids or gaps that are not intrinsic to the stability of the bony structure.<br><br>HydroSet XT cured in-situ provides an open void/gap filler that can augment<br>provisional hardware (e.g. k-wires, plates, screws) to help support bone fragments<br>during the surgical procedure. The cured cement acts only as a temporary support<br>medium and is not intended to provide structural support during the healing process. |
| Summary of the<br>Technological<br>Characteristics | The subject device and the predicate device have similar technological characteristics<br>as the mixed powder and liquid components are identical and combine chemically in<br>identical fashion. Both devices are provided sterile and for single use only. Any<br>differences in mixing method do not raise new questions of safety or effectiveness. |
| Summary of Non-Clinical<br>Performance Testing | Performance testing was completed for HydroSet XT to determine its suitability for<br>use. The following tests were performed to evaluate the performance of HydroSetXT:<br>Working Time (Injectability and Torque) |
| 510(k) Summary: HydroSet XT (K161447) | |
| | Compression Setting Time (Wet Field Penetration Resistance) |
| | Bacterial endotoxin testing (BET) as specified in USP 39 <85>, Ph. Eur. 8.0 2.6.14 is<br>used for pyrogenicity testing to achieve the Endotoxin limit of <20EU/Device. Based<br>on an endotoxin limit of 20 EU/Device and an extraction volume of 40 mlthe<br>Maximum Valid Dilution (MVD) will be a 1:50 dilution. |
| Conclusions | Based upon a comparison of intended use, technological characteristics and device<br>performance in the non-clinical tested listed above, HydroSet XT has demonstrated<br>substantial equivalence to its predicate device. |
{4}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.