Harvest Dental HD Gum Strip is indicated for adding texture, form and color to the gingival area of a dental prosthesis, which includes crowns, bridges and full and partial prostheses. This composite strip is bondable to polymethylmethacrylate (PMMA), metals and ceramics (zirconia).
Device Story
Harvest Dental HD Gum Strip is a light-cured, textured, malleable composite resin strip; used by dental lab technicians to add gingival texture, form, and color to dental prostheses (crowns, bridges, full/partial). Input: manual application of strip to dentine/bone-like substructure of prosthesis; includes gingival vein features. Output: finished gingival surface on dental restoration. Operation: strip cut to size, applied, trimmed, finished, and light-cured using standard dental light-curing device. Includes optional paste/liquid modifiers for repair and enhanced coloration. Benefit: provides an alternative to manual build-up of shaded composite paste; improves efficiency and consistency in modeling natural-looking gingival tissues. Final restoration approved by licensed dental practitioner before patient use.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (cytotoxicity, sensitization, oral mucosa irritation) per ISO 7405 and ISO 10993-1, and physical/mechanical property testing per ISO 10477 (flexural strength 77 MPa, exceeding the 50 MPa requirement).
Technological Characteristics
Light-cured composite resin; materials: urethane methacrylate, dimethacrylate, trimethacrylate, silica. Form factor: textured, malleable strip. Curing: standard dental light-curing device. Standards: ISO 10477 (polymer-based crown/bridge materials), ISO 7405 (biocompatibility), ISO 10993-1 (biological evaluation).
Indications for Use
Indicated for adding texture, form, and color to the gingival area of dental prostheses (crowns, bridges, full/partial prostheses) in patients requiring gingival restoration. Bondable to PMMA, metals, and zirconia.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 16, 2023
Harvest Dental Products, LLC Colleen Boswell Regulatory Affairs Consultant 905 Columbia Street Brea, California 92821
Re: K231389
Trade/Device Name: Harvest Dental HD Gum Strip Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI, EBF Dated: May 20, 2023 Received: July 25, 2023
Dear Colleen Boswell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K231389
Device Name
Harvest Dental HD Gum Strip
Indications for Use (Describe)
Harvest Dental HD Gum Strip is indicated for adding texture, form and color to the gingival area of a dental prosthesis which includes crowns, bridges and full and partial prostheses. This composite strip is bondable to polymethylmethacrylate (PMMA), metals and ceramics (zirconia).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below "HarvestDental" is the text "LABORATORY PRODUCTS" in a smaller, gray font.
# 510(k) Summary
## 1. Submitter:
2.
3.
Harvest Dental Products, LLC 905 Columbia Street Brea, California 92821
| Contact Person: | Colleen Boswell |
|--------------------------|----------------------------------------------------------------------------------------------------------------------|
| Telephone Number: | (714) 674-7400 |
| Fax Number: | (714) 674-7402 |
| Date Prepared: | July 25, 2023 |
| <b>Device:</b> | |
| Name of Device: | Harvest Dental HD Gum Strip |
| Common Name: | Composite resin for gingival area of dental prosthesis |
| Classification Name: | Tooth Shade Resin Material, per 21 CFR 872.3690<br>Denture, Relining, Repairing, Rebasing Resin, per 21 CFR 872.3760 |
| Device Class: | II |
| Product Code: | EBF<br>EBI |
| <b>Predicate Device:</b> | |
| Primary Predicate: | Gradia Gum, GC America, Inc., K033808, Product Code EBF & EBI |
#### 4. Device Description
Harvest Dental HD Gum Strip is a bondable, light-cured composite for adding texture and color to the gingival area of a dental prosthesis which includes crowns, bridges and full and partial prostheses. This composite strip is bondable to polymethylmethacrylate (PMMA), metals and ceramics (zirconia). It is provided in a textured, malleable strip meant to be applied to a dentine/bone-like substructure. The strip is an alternative to the method of using a shaded composite paste to manually build-up and model visible gingival surfaces to look similar to natural gingival tissues. The Harvest Dental HD Gum Strip includes features such as gingival veins formed into its surface as well as being available in various shades of coloration. The strip can be easily cut, applied to the gingival surface of the restoration, then trimmed and finished with appropriate features such as gingival line and interdental papilla. Gum Strip Modifiers (paste versions of the same composite) are available in various shades and used for additional surface features and gingival repairs, and Liquid Modifiers (stains, which are a low-viscosity color-concentrated form of the same composite) are available in several shades and used for enhanced coloration resembling the natural gingiva. The Harvest Dental HD Gum Strip is bonded to the dental prosthesis
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features the words "HarvestDental" in a stylized font, with "Harvest" in gray and "Dental" in a darker shade. Above the text is a green leaf design, with one large leaf and a smaller leaf below it. Underneath the company name, the words "LABORATORY PRODUCTS" are written in smaller, gray font.
utilizing a bonding adhesive provided with the strip and modifiers and, after any final detailing and coloring, is to be cured utilizing a typical light-curing device.
The Harvest Dental HD Gum Strip is to be used by dental lab technicians and approved by licensed dental practitioners before being provided to the patient.
## 5. Statement of Intended Use:
Harvest Dental HD Gum Strip is indicated for adding texture, form and color to the gingival area of a dental prosthesis, which includes crowns, bridges and full and partial prostheses. This composite strip is bondable to polymethylmethacrylate (PMMA), metals and ceramics (zirconia).
## 6. Summary of Technological Characteristics with the Predicate Device
The technological characteristics of the subject Harvest Dental HD Gum Strip are similar to the predicate device, Gradia Gum (K033808). There are no substantial technical or functional differences between Harvest Dental HD Gum Strip and the predicate device in terms of chemical composition, function and intended use. Both are composite resins used to reproduce the gingival area (gum) on dental prostheses. See Table 1 below for technological characteristics and comparisons of the composite resin for the gingival area of dental prostheses.
| Element | Harvest Dental HD Gum<br>Strip<br>(Proposed Device) | Gradia Gum<br>(Primary Predicate) | Comparison |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Harvest Dental Products,<br>LLC | GC America, Inc. | N/A |
| 510(k) | K231389 | K033808 | N/A |
| Target Users | Certified Dental<br>Professionals and<br>Technicians | Certified Dental<br>Professionals and<br>Technicians | Same |
| Common Name | Composite resin for<br>gingival area of dental<br>prostheses | Composite resin for<br>gingival area of dental<br>prostheses | Same |
| Device | Harvest Dental HD Gum | Gradia Gum is a light-cured | Similar - Harvest |
| Description | Strip is a bondable, light-<br>cured composite for adding<br>texture and color to the<br>gingival area of a dental<br>prosthesis which includes<br>crowns, bridges and full and<br>partial prostheses. This<br>composite strip is bondable<br>to polymethylmethacrylate | composite resin used for<br>the reproduction of<br>gingival tissues by means of<br>basic or multiple layering<br>techniques. The material is<br>applied using a build-up<br>technique, color adjusted<br>using modifiers and<br>textured. | Dental HD Gum<br>Strip is provided<br>in a strip form<br>rather than a<br>paste form for<br>ease of<br>application and<br>consistent<br>thickness. |
| Element | Harvest Dental HD Gum<br>Strip<br>(Proposed Device) | Gradia Gum<br>(Primary Predicate) | Comparison |
| | (PMMA), metals and<br>ceramics (zirconia). It is<br>provided in a textured,<br>malleable strip meant to be<br>applied to a dentine/bone-<br>like substructure. The strip<br>is an alternative to the<br>method of using a shaded<br>composite paste to<br>manually build-up and<br>model visible gingival<br>surfaces to look similar to<br>natural gingival tissues. The<br>Harvest Dental HD Gum<br>Strip includes features such<br>as gingival veins formed<br>into its surface as well as<br>being available in various<br>shades of coloration. The<br>strip can be easily cut,<br>applied to the gingival<br>surface of the restoration,<br>then trimmed and finished<br>with appropriate features<br>such as gingival line and<br>interdental papilla. Gum<br>Strip Modifiers (paste<br>versions of the same<br>composite) are available in<br>various shades and used for<br>additional surface features<br>and gingival repairs, and<br>Liquid Modifiers (stains,<br>which are a low-viscosity<br>color-concentrated form of<br>the same composite) are<br>available in several shades<br>and used for enhanced<br>coloration resembling the<br>natural gingiva. The<br>Harvest Dental HD Gum<br>Strip is bonded to the | | |
| Element | Harvest Dental HD Gum<br>Strip<br>(Proposed Device) | Gradia Gum<br>(Primary Predicate) | Comparison |
| | bonding adhesive provided<br>with the strip and modifiers<br>and, after any final detailing<br>and coloring, is to be cured<br>utilizing a typical light-<br>curing device. The <b>Harvest</b><br><b>Dental HD Gum Strip</b> is to<br>be used by dental lab<br>technicians and approved<br>by licensed dental<br>practitioners before being<br>provided to the patient. | | |
| Indications For<br>Use | <b>Harvest Dental HD Gum<br/>Strip</b> is indicated for adding<br>texture, form and color to<br>the gingival area of a dental<br>prosthesis which includes<br>crowns, bridges and full<br>and partial prostheses.<br>This composite strip is<br>bondable to<br>polymethylmethacrylate<br>(PMMA), as well as metals<br>and ceramics (zirconia). | Gradia Gum is a light-cured<br>composite resin specifically<br>used for the reproduction<br>of the gum on crown and<br>bridge prostheses. | Same - Indicated<br>to reproduce the<br>gingival area<br>(gum) on dental<br>prostheses. |
| Curing<br>Mechanism | Light-Cure | Light-Cure | Same |
| Technique | Manual application | Manual application, build-<br>up | Similar - <b>Harvest<br/>Dental HD Gum<br/>Strip</b> does not<br>require any<br>build-up. |
| Material Type | Urethane methacrylate,<br>dimethacrylate,<br>trimethacrylate, silica | Urethane dimethacrylate<br>(UDMA), dimethacrylate,<br>trimethacrylate, silica | Same |
| Flexural<br>Strength<br>(ISO 10477) | 77 MPa<br>> 50 MPa | 124Mpa<br>> 50 MPa | Same - Passed<br>ISO 10477<br>requirement |
| Element | Harvest Dental HD Gum<br>Strip<br>(Proposed Device) | Gradia Gum<br>(Primary Predicate) | Comparison |
| Vickers<br>Hardness | N/A<br>No direct masticatory<br>forces on device | 38Hv | Not defined for<br>Subject device. |
| Occlusal Wear<br>(vs bovine's<br>enamel, after<br>200,000 times) | N/A<br>No occlusal contact<br>intended | 170μ | Not defined for<br>Subject device. |
| Storage<br>Conditions | 5-30°C,<br>protect against direct<br>sunlight | Store in a cool place away<br>from direct sunlight | Similar |
#### Table 1: Comparison of Subject and Predicate Devices
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features the words "HarvestDental" in a stylized font, with "Harvest" in gray and "Dental" in green. Above the text is an image of two green leaves. Below the text is the phrase "LABORATORY PRODUCTS" in smaller, gray font.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features the words "HarvestDental" in a stylized font, with "Harvest" in gray and "Dental" in green. Above the text is an image of two green leaves. Below the text is the phrase "LABORATORY PRODUCTS" in smaller, gray font.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the words "HarvestDental" in gray. Below that, the words "LABORATORY PRODUCTS" are written in a smaller, gray font.
# 7. Performance Data
# Biocompatibility Testing
The biocompatibility evaluation for Harvest Dental HD Gum Strip was conducted in accordance with ISO 7405:2018 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry, Annex A, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process" as recognized by FDA. The biocompatibility testing included the following tests:
- 1. Cytotoxicity
- 2. Sensitization
- 3. Oral Mucosa Irritation
The biocompatibility testing conducted demonstrates adequate biocompatibility for Harvest Dental HD Gum Strip.
# ISO 10477 Testing
Testing according to ISO 10477:2020 Dentistry. Polymer-based crown and bridge materials was performed on the Harvest Dental HD Gum Strip is applied directly to a crown (or bridge) and becomes part of that device. As compared to the predicate device, it is substantially equivalent to the device and met the physical/mechanical properties of the standard.
#### Clinical Studies
No human clinical testing was conducted to support substantial equivalence.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the words "HarvestDental" in gray. Below "HarvestDental" are the words "LABORATORY PRODUCTS" in a smaller, gray font.
- 8. Conclusion as to Substantial Equivalence
The similarities in chemical composition, function, performance testing and intended use of Harvest Dental HD Gum Strip with the legally marketed predicate device, Gradia Gum (K033808) support substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.