The Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; shoulder, foot/ankle, knee, hand/wrist and elbow. Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hand Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Device Story
Fix2Lock is an absorbable bone fixation screw system used in orthopedic surgery to reattach soft tissues (ligaments, tendons, articular capsules) to bone. The system comprises an anchor (composed of PLGA/beta-TCP or PEEK), non-absorbable UHMWPE suture, needle, driver shaft, and handle. Operated by surgeons in clinical settings, the device is inserted into bone to secure soft tissue via the attached suture. The mechanical fixation provides stability during the healing process. The device is provided sterile for single use.
Clinical Evidence
Bench testing only. Performance verified against ASTM F543 (metallic bone screws), ASTM F2026 (PEEK), and ASTM F2848 (UHMWPE). Sterilization, packaging, and shelf-life validated per ISO 11135, ISO 11607, and ASTM F1980. Bacterial endotoxin testing performed per USP <85> and <161>.
Technological Characteristics
Materials: PLGA/beta-TCP (anchor), PEEK (Biocombi self-punching anchor), UHMWPE (suture). Standards: ASTM F2026, ASTM F2848, ASTM F543. Form factor: Bone fixation screw with driver shaft/handle. Sterilization: Ethylene oxide (EtO). Single-use, sterile.
Indications for Use
Indicated for soft tissue-to-bone fixation using suture in shoulder, foot/ankle, knee, hand/wrist, and elbow orthopedic procedures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
May 31, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Osteonic Co., Ltd. % Sanglok Lee Manager Wise Company Inc. #507, #508, 166 Gasandigital 2-ro, Geumcheon-gu Seoul. 08507 Korea, South
# Re: K231326
Trade/Device Name: Fix2Lock(Biocomposite medial, lateral, Biocombi Self Punching) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: April 1, 2023 Received: May 8, 2023
# Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sara S. Thompson -S
For
Yu-Chieh Chiu, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231326
#### Device Name
Fix2Lock (Biocomposite medial, lateral, Biocombi Self Punching)
Indications for Use (Describe)
The Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; shoulder, foot/ankle, knee, hand/wrist and elbow.
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Osteonic, a medical company. The logo features a geometric design of nested squares in blue, positioned to the left of the company name. The text "OSTEONIC" is written in a bold, sans-serif font, with the tagline "The Medical Explorer" appearing in a smaller font size beneath it.
## The assigned 510(k) Number: K231326
#### Date of Submission: 2023.04.01
#### 01. Applicant
OSTEONIC Co., Ltd. 405Ho, 505-2Ho, 505-3Ho, 508Ho, 902Ho, 1004Ho, 1005Ho, 1201Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea, 102Ho, 103Ho, 24, Digital-ro 27-gil, Guro-gu, Seoul, Korea TEL: +82-2-6902-8456 FAX: +82-2-6902-8401 Email: dakham@osteonic.com
#### 02. Submission Correspondent
Sanglok, Lee Wise COMPANY Inc. 08503, #507, #508, 166 Gasan digital 2-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
#### 03. Subject Device Identification
Trade Name: Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi Self Punching) Common Name: Bioabsorbable bone anchor Classification Name: fastener, fixation, non-biodegradable, soft tissue Classification Product Code: MAI Subsequent Product Code: MBI Panel: Orthopedic Regulation Number: 21 CFR 888.3030 Device Class: II
#### 04. Predicate device
510(k) Number: K202806 Device Name: Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi Self Punching) Manufacturer: Osteonic Co., Ltd.
### 05. Reference device
510(k) Number: K192032 Device Name: Fixone Biocomposite Small Anchor Manufacturer: AJU Pharm Co., Ltd..
#### 06. Device Description
The Fix2Lock (Bioabsorbable bone anchor) is an absorbable bone fixation screw that fixes soft tissues such as ligament, tendon, and the articular capsules to bone, and is used in orthopedic surgery. This product consists of an absorbable and/or non-absorbable suture and driver shaft and handle.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Osteonic. The logo consists of a blue square with a white outline and a white line running through the middle. To the right of the square is the word "OSTEONIC" in bold, dark gray letters. Below the word "OSTEONIC" is the phrase "The Medical Explorer" in a smaller, lighter gray font.
#### 07. Indication for use
The Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; shoulder, foot/ankle, knee, hand/wrist and elbow.
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral
ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Osteonic. The logo consists of a blue square with a smaller square inside, and the word "OSTEONIC" in bold, black letters. Below the word "OSTEONIC" is the phrase "The Medical Explorer" in a smaller, lighter font.
#### 08. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device with the following standards:
#### ▪ Material
- ASTM F2026: 2017 Standard specification for polyetheretherketone(PEEK) polymers for surgications - ASTM F2848: 2017 Standard Specification for Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns and Surgical Implants
#### ▪ Mechanical performance
- ASTM F543: 2013 Standard specification and test methods for metallic medical bone screws
- = Sterilization, shelf-life and packaging for sterile product
- ISO 11135:2014, Sterilization of health-care products Ethylene oxide Requirements for the development validation and routine control of a sterilization and routine control of a sterilization process for medical devices
- ISO 11138-1:2006, Sterilization of health care products Biological indicators Part 1: General requirements
- ISO 11138-2:2009, Sterilization of health care products Biological indicators Part 2: Biological indicators for ethylene oxide sterilization processes
- ISO 11140-1:2014, Sterilization of health care products Chemical indicators Part 1: General requirements
- ISO 11737-1:2018 Sterilization of medical devices Microbiological methods- Part 1: Estimation of microorganisms on products
- ISO 11737-2:2009 Sterilization of medical devices Microbiological methods- Part 2: Tests of sterility performed in the validation of a sterilization process
- ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system
- ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1980:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials.
- ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- Bacterial Endotoxin
- USP <85> Bacterial Endotoxin Test
- USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests
### 09. Comparison of Technological Similarities and Differences
| Product<br>Name | SUBJECT Device | Predicate Device | REFERENCE Device | Equivalence<br>Discussion |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Product code | MAI | MAI, | MAI | Same |
| Regulatory<br>class | Class II | Class II | Class II | Same |
| Regulation<br>Number | 21 CFR 888.3030 | 21 CFR 888.3030 | 21 CFR 888.3030 | Same |
| Intended use | The Fix2Lock is intended use for<br>fixation of soft tissue to bone,<br>using suture, in the following<br>procedure;<br>shoulder, foot/ankle, knee,<br>hand/wrist and elbow.<br>Shoulder: Rotator Cuff Repair, | Fix2Lock(Biocomposite medial,<br>lateral, Biocombi Self<br>Punching)(K202806) | Fixone Biocomposite Small<br>Anchor (K192032) | Same |
| | Bankart Repair, SLAP Lesion<br>Repair, Biceps Tenodesis,<br>Acromio-Clavicular Separation<br>Repair, Deltoid Repair, Capsular<br>Shift or Capsulolabral<br>Reconstruction<br>Foot/Ankle: Lateral Stabilization,<br>Medial Stabilization, Achilles<br>Tendon Repair, Metatarsal<br>Ligament Repair, Hallux<br>Valgus reconstruction, digital<br>tendon transfers, Mid-foot<br>reconstruction<br>Knee: Medial Collateral Ligament<br>Repair, Lateral Collateral<br>Ligament Repair, Patellar Tendon<br>Repair, Posterior Oblique<br>Ligament Repair, Iliotibial Band<br>Tenodesis<br>Hand/Wrist: Scapholunate<br>Ligament Reconstruction, Carpal<br>Ligament Reconstruction,<br>Repair/Reconstruction of<br>collateral<br>ligaments, Repair of Flexor and<br>Extensor Tendons at the PIP, DIP,<br>and MCP joints for all digits,<br>digital tendon transfers<br>Elbow: Biceps Tendon<br>Reattachment, Ulnar or Radial<br>Collateral Ligament<br>Reconstruction | Shoulder: Rotator Cuff Repair,<br>Bankart Repair, SLAP Lesion<br>Repair, Biceps Tenodesis,<br>Acromio-Clavicular Separation<br>Repair, Deltoid Repair,<br>Capsular Shift or Capsulolabral<br>Reconstruction<br>Foot/Ankle: Lateral<br>Stabilization, Medial<br>Stabilization, Achilles Tendon<br>Repair, Metatarsal Ligament<br>Repair, Hallux<br>Valgus reconstruction, digital<br>tendon transfers, Mid-foot<br>reconstruction<br>Knee: Medial Collateral<br>Ligament Repair, Lateral<br>Collateral Ligament Repair,<br>Patellar Tendon Repair,<br>Posterior Oblique<br>Ligament Repair, Iliotibial<br>Band Tenodesis<br>Hand/Wrist: Scapholunate<br>Ligament Reconstruction,<br>Carpal Ligament<br>Reconstruction,<br>Repair/Reconstruction of<br>collateral<br>ligaments, Repair of Flexor and<br>Extensor Tendons at the PIP,<br>DIP, and MCP joints for all<br>digits, digital tendon transfers<br>Elbow: Biceps Tendon<br>Reattachment, Ulnar or Radial<br>Collateral Ligament<br>Reconstruction | Shoulder: Rotator Cuff Repair,<br>Bankart Repair, SLAP Lesion<br>Repair, Biceps Tenodesis,<br>Acromio-Clabicular Separation<br>Repair, Deltoid Repair, Capsular<br>Shift or Capsulolabral<br>Reconstruction<br>Foot/Ankle: Lateral<br>Stabilization, Medial<br>Stabilization, Achilles Tendon<br>Repair, Metatarsal Ligament<br>Repair, Hallux<br>Valgus reconstruction, digital<br>tendon transfers, Mid-foot<br>reconstruction<br>Knee: Medial Collateral<br>Ligament Repair, Lateral<br>Collateral Ligament Repair,<br>Patellar Tendon Repair,<br>Posterior<br>Oblique Ligament Repair,<br>Iliotibial Band Tenodesis<br>K192032<br>006_510(k) Summary<br>Page 2 / 4<br>Hand/Wrist: Scapholunate<br>Ligament Reconstruction,<br>Carpal Ligament<br>Reconstruction,<br>Repair/Reconstruction of<br>collateral ligaments, Repair of<br>Flexor and Extensor Tendons at<br>the PIP, DIP, and MCP joints for<br>all digits, digital<br>tendon transfers<br>Elbow: Biceps Tendon<br>Reattachment, Ulnar or Radial<br>Collateral Ligament<br>Reconstruction<br>Hip: Capsular repair, Acetabular<br>Labral repair | |
| | Bone fixation screw that ties soft | Bone fixation screw that ties soft | The Fixone Biocomposite Small<br>Anchor is intended for reattaching<br>soft tissue to bone with sutures. | |
| Operating<br>Principles | tissues such as ligament, tendon,<br>and the articular capsules to bone. | tissues such as ligament, tendon,<br>and the articular capsules to<br>bone. | | Same |
| Material | Anchor:<br>-Biocomposite medial, lateral:<br>PLGA+ $\beta$ -TCP<br>-Biocombi self punching:<br>PLGA+ $\beta$ -TCP, PEEK<br>Suture: UHMWPE | Anchor:<br>-Biocomposite medial, lateral:<br>PLGA+ $\beta$ -TCP<br>-Biocombi self punching:<br>PLGA+ $\beta$ -TCP, PEEK<br>Suture: UHMWPE | PLGA+ $\beta$ -TCP | Same |
| Structure | This product consists of an<br>absorbable and/or non-absorbable<br>anchor, a non-absorbable suture,<br>needle and driver shaft and handle. | This product consists of an<br>absorbable and/or non-<br>absorbable anchor, a non- | Driver, Anchor, Suture and needle | Same |
| | | absorbable suture and driver | | |
| | | shaft and handle. | | |
| Product Size | Anchor diameter: Ø 2.6 to 4.8 mm<br>Anchor length: 10.0 to 16.0mm | Anchor diameter: Ø 2.6 to 6.5 mm<br>Anchor length: 10.0 to 20.8mm | Anchor diameter : Ø 3.0mm<br>Length : 11.9mm | Similar |
| Sterilization | Sterile (EtO sterilization) | Sterile (EtO sterilization) | Sterile (EtO sterilization) | Same |
| Single Use/<br>Reuse | Single use | Single use | Single use | Same |
| Packaging | 1 EA / BOX | 1 EA / BOX | - | Same |
| Shelf -life | 3Years | 5Years | - | Same |
#### Table 1: Substantial Equivalence Comparison
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Osteonic. The logo consists of a blue geometric shape on the left and the word "OSTEONIC" in bold, dark gray letters on the right. Below "OSTEONIC" is the tagline "The Medical Explorer" in a smaller, lighter gray font.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Osteonic. The logo consists of a blue geometric shape resembling a stylized diamond or square with internal lines, positioned to the left of the company name. The text "OSTEONIC" is written in bold, dark lettering, and below it, in smaller font, is the tagline "The Medical Explorer."
The Product Size of subject device is similar with the predicate device. It is not exactly same but the is within the range of unmodified device and the safety was evaluated as the performance bench test.
### 10. Substantially Equivalent Conclusion
Based on above, the subject device, Fix2Lock (Biocomposite medial, lateral, Biocombi Self punching), is determined to be Substantially Equivalent (SE) to the predicate device(Unmodified device), Fix2Lock(Biocomposite medial, lateral, Biocombi Self pundhing), (K202806) in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.