The Pangea Femur Plating System is indicated for the internal fixation of femur bone fractures and osteotomies in normal and osteopenic bone, including: Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures Non-unions, malunions and deformities Periprosthetic fractures. The Pangea Fibula Plating System is indicated for the internal fixation of fibula bone fractures and osteotomies in normal and osteopenic bone, including: Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures Non-unions, malunions, and deformities Periprosthetic fractures. The Pangea Tibia Plating System is indicated for the internal fixation of tibia bone fractures and osteotomies in normal and osteopenic bone, including: Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures Non-unions, malunions, and deformities Periprosthetic fractures. The Pangea Humerus Plating System is indicated for the internal fixation of humerus bone fractures and osteotomies in normal and osteopenic bone, including: Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures Non-unions, malunions, and deformities Periprosthetic fractures.
Device Story
Pangea Plating Systems (Femur, Fibula, Tibia, Humerus) consist of anatomically pre-contoured metallic bone plates; used for internal fixation and stabilization of fractures and osteotomies. Plates are implanted by surgeons in clinical/OR settings. System includes locking and non-locking screws, cable plugs, and washers. Devices provide mechanical support to bone segments to facilitate healing. No software or electronic components involved.
Clinical Evidence
No clinical data. Bench testing only. Performance supported by dynamic construct testing and MRI compatibility assessments per ASTM standards.
Technological Characteristics
Plates manufactured from Ti6Al4V ELI (ASTM F136). Anatomically pre-contoured; lengths 68 mm to 441 mm. Compatible with locking and non-locking screws, cable plugs, and washers. Sterile or non-sterile. Mechanical fixation device; no software or energy source.
Indications for Use
Indicated for internal fixation and stabilization of femur, fibula, tibia, and humerus bone fractures and osteotomies in patients with normal or osteopenic bone, including diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures, non-unions, malunions, deformities, and periprosthetic fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
VariAx 2 One-Third Tubular Plating System (K151879)
PeriPRO Femur and Variable Angle Fixation (K222381)
Submission Summary (Full Text)
{0}------------------------------------------------
August 18, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Stryker GmbH Danese Joiner-Fox Associate Manager Regulatory Affairs 325 Corporate Drive Mahwah. New Jersey 07430
Re: K231262
Trade/Device Name: Pangea Femur Plating System, Pangea Fibula Plating System, Pangea Tibia Plating System, Pangea Humerus Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: May 1, 2023 Received: May 1, 2023
Dear Danese Joiner-Fox:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali-S
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K231262
Device Name Pangea Femur Plating System
## Indications for Use (Describe)
Pangea Femur Plating System:
The Pangea Femur Plating System is indicated for the internal fixation of femur bone fractures and osteotomies in normal and osteopenic bone, including:
- · Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures
- · Non-unions, malunions and deformities
- · Periprosthetic fractures
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known)
K231262
Device Name Pangea Fibula Plating System
## Indications for Use (Describe)
Pangea Fibula Plating System:
The Pangea Fibula Plating System is indicated for the internal fixation of fibula bone fractures and osteotomies in normal and osteopenic bone, including:
- · Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures
- · Non-unions, malunions, and deformities
- · Periprosthetic fractures
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known)
K231262
Device Name Pangea Tibia Plating System
## Indications for Use (Describe)
Pangea Tibia Plating System:
The Pangea Tibia Plating System is indicated for the internal fixation of tibia bone fractures and osteotomies in normal and osteopenic bone, including:
- · Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures
- · Non-unions, malunions, and deformities
- · Periprosthetic fractures
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known)
K231262
Device Name Pangea Humerus Plating System
## Indications for Use (Describe)
Pangea Humerus Plating System:
The Pangea Humerus Plating System is indicated for the internal fixation of humerus bone fractures and osteotomies in normal and osteopenic bone, including:
- · Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures
· Non-unions, malunions, and deformities
- · Periprosthetic fractures
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
Stryker GMBH Pangea Femur Plating System, Pangea Fibula Plating System, Pangea Tibia Plating System, Pangea Humerus Traditional 510(k) Premarket Notification Plating System
## K231262 510(k) Summary
| Proprietary Name: | Pangea™ Femur Plating System<br>Pangea™ Fibula Plating System<br>Pangea™ Tibia Plating System<br>Pangea™ Humerus Plating System |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Plate, Fixation, Bone and Screw, Fixation, Bone |
| Regulation Description: | 21 CFR 888.3030: Single/Multiple component metallic bone<br>fixation appliances and accessories |
| Regulation Number: | 21 CFR 888.3030 |
| Product Code: | HRS, HWC |
| Device Class: | Class II |
| Sponsor: | Stryker GMBH<br>Bohnackerweg l<br>2545 Selzach / Switzerland |
| Contact Person: | Danese Joiner-Fox<br>Associate Manager Regulatory Affairs<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (475) 333-4452 |
| Date Prepared: | 05/01/2023 |
| Primary Predicate: | AxSOS 3 Ti System (K200398/K181091/ K153379/ K123964) |
| Additional Predicate: | SPS Small Fragment Set The Stryker® Plating System<br>(K060798)<br>VariAx Fibula (K081284)<br>VariAx 2 One-Third Tubular Plating System (K151879)<br>PeriPRO Femur and Variable Angle Fixation (K222381) |
| Description | This traditional 510(k) submission is being supplied to the U.S.<br>FDA to gain clearance to market the Pangea Femur Plating<br>System, Pangea Fibula Plating System, Pangea Tibia Plating<br>System, and Pangea Humerus Plating System. This submission<br>encompasses multiple systems that have similar intended use. |
{7}------------------------------------------------
Stryker GMBH Pangea Femur Plating System, Pangea Fibula Plating System, Pangea Tibia Plating System, Pangea Humerus Plating System Traditional 510(k) Premarket Notification
> All plates are manufactured from Ti6Al4V ELI (ASTM F136), are anatomically pre-contoured. and are available in different sizes and left/right versions. The plates are provided sterile or non-sterile and range in length from 68 mm to 441 mm. The plates allow for the use of locking and non-locking screws and are used with cable plugs and washers.
#### Indications for Use
### Pangea Femur Plating System
The Pangea Femur Plating System is indicated for the internal fixation and stabilization of femur bone fractures and osteotomies in normal and osteopenic bone, including:
· Diaphyseal, metaphyseal, epiphyseal, extra- and intraarticular fractures
- Non-unions, malunions, and deformities
- · Periprosthetic fractures
#### Pangea Fibula Plating System
The Pangea Fibula Plating System is indicated for the internal fixation and stabilization of fibula bone fractures and osteotomies in normal and osteopenic bone, including:
· Diaphyseal, metaphyseal, epiphyseal, extra- and intraarticular fractures
- Non-unions, malunions, and deformities
- · Periprosthetic fractures
#### Pangea Tibia Plating System
The Pangea Tibia Plating System is indicated for the internal fixation and stabilization of tibia bone fractures and osteotomies in normal and osteopenic bone, including:
· Diaphyseal, metaphyseal, epiphyseal, extra- and intraarticular fractures
- Non-unions, malunions, and deformities
- Periprosthetic fractures
#### Pangea Humerus Plating System
The Pangea Humerus Plating System is indicated for the internal fixation and stabilization of humerus bone fractures and osteotomies in normal and osteopenic bone, including:
- · Diaphyseal, metaphyseal, epiphyseal, extra- and intraarticular fractures
- Non-unions, malunions, and deformities
- Periprosthetic fractures
### Summary of Technologies
{8}------------------------------------------------
Stryker GMBH Pangea Femur Plating System, Pangea Fibula Plating System, Pangea Tibia Plating System, Pangea Humerus Plating System Traditional 510(k) Premarket Notification
> A comparison of the systems demonstrated that the subject Pangea Femur Plating System. Pangea Fibula Plating System. Pangea Tibia Plating System, Pangea Humerus Plating System is substantially equivalent to the AxSOS 3 Ti (K181091/ K200398), SPS Small Fragment Set The Stryker® Plating System (K060798), VariAx 1/3 tubular (K151879), VariAx Fibula (K081284), and PeriPRO Femur and Variable Angle Fixation (K222381) in regard to intended use, material, design, and operational principles.
Non-Clinical Testing: The following non-clinical laboratory testing, and performance assessments were made in support of substantial equivalence:
- Dynamic construct testing
Assessments performed to establish compatibility with a magnetic resonance environment:
- Magnetically Induced Displacement per ASTM F2052
- Magnetically Induced Torque per ASTM F2213
- Image Artifacts per ASTM F2119
Tests performed to establish compatibility with a magnetic resonance environment:
- . RF Heating per ASTM F2182
Clinical testing was not required for this submission.
The subject device Pangea Femur Plating System, Pangea Fibula Plating System, Pangea Tibia Plating System, and Pangea Humerus Plating System are substantially equivalent to the previously cleared predicate device AxSOS 3 Ti (K200398/K181091/ K153379/ K123964). Except for the modifications described in this submission, the subject devices are identical to the predicate device, and the performance data and analyses demonstrate that:
· any differences do not raise new questions of safety and effectiveness as established with performance testing; and
• the subject devices are at least as safe and effective as the legally marketed predicate devices
Clinical Testing:
Conclusion:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.