The Altaviz Needle Kit II is a convenience kit that includes: A 29G injection needle that is a sterile medical device for single use intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. Either an 18G or a 19G 5 micron filter needle that is a sterile medical device for single use intended to draw up fluids into a syringe. The Altaviz Needle Kit II is indicated for intravitreal use.
Device Story
Convenience kit containing two sterile, single-use needles: one 29G injection needle and one 18G or 19G 5-micron filter needle. Used in hospital or surgical settings by clinicians. Filter needle draws medication into syringe; 29G needle injects/withdraws fluids from body. Kit tray (PETG) secures components during transport. No device modification or re-sterilization occurs during kit assembly. Benefits include clinical convenience by providing pre-selected, compatible components for intravitreal procedures.
Clinical Evidence
Bench testing only. Testing included particulate matter analysis per USP <788> and USP <789>, intraocular irritation testing per ISO 10993-10, ethylene oxide residual assessment per ISO 10993-7, and endotoxin testing per FDA guidance for single-use intraocular ophthalmic devices. All tests passed.
Technological Characteristics
Convenience kit tray: PETG. 29G needle: stainless steel, polypropylene. 18G/19G filter needle: stainless steel, polycarbonate or ABS. Manual operation. Sterile, single-use components. No electronic or software components.
Indications for Use
Indicated for intravitreal use in patients requiring injection or withdrawal of fluids below the skin surface.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the full name "U.S. Food & Drug Administration" in blue text.
Altaviz, LLC % Sean Griffin President Allied Regulatory Consulting 1540 Keller Parkway, Suite 108 #170 Keller. Texas 76248
# Re: K231261
Trade/Device Name: Altaviz Needle Kit II Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI, GAA Dated: May 1, 2023 Received: May 2, 2023
# Dear Sean Griffin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Courtney Evans -S
Digitally signed by Courtney Evans -S Date: 2023.07.27 12:34:46 -04'00'
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231261
Device Name Altaviz Needle Kit II
#### Indications for Use (Describe)
The Altaviz Needle Kit II is a convenience kit that includes:
A 29G injection needle that is a sterile medical device for single use intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.
Either an 18G or a 19G 5 micron filter needle that is a sterile medical device for single use intended to draw up fluids into a syringe.
The Altaviz Needle Kit II is indicated for intravitreal use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: overline;">☒</span> Proprietary Use (Part 21 CFR 601 Subpart D) | ☐ Compassionate Use (21 CFR 601 Subpart E) |
|-----------------------------------------------------------------------------------------------|--------------------------------------------|
|-----------------------------------------------------------------------------------------------|--------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K231261
This 510(k) Summary has been prepared in accordance with 21 CFR 807.92.
### Applicant
The name and address of the Applicant is:
Altaviz LLC 13766 Alton Parkway, Suite 143 Irvine, CA 92618
Application Correspondent:
Sean Griffin President Allied Regulatory Consulting 1540 Keller Parkway, Suite 108 #170 Keller, TX 76248 Phone: (817) 805-8392
Date Prepared: July 27, 2023
# Device
Device Subject to this 510(k):
| Trade Name: | Altaviz Needle Kit II |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Needle Convenience Kit |
| Product Code: | FMI (Needle, Hypodermic, Single Lumen; 21CFR Section 880.5570)<br>GAA (Needle, Aspiration and Injection, Disposable; 21CFR Section 878.4800) |
| Classification: | II |
### Predicate Device
| 510(k) Number | Device |
|---------------|------------------------------------------------------------|
| K22681 | Altaviz Needle Kit |
| K970370 | TSK Steriject 29G Hypodermic Needle (Additional Predicate) |
{4}------------------------------------------------
# Device Description
The Altaviz Needle Kit II is a convenience kit with a tray that is made of polyethylene terephthalate glycol (PETG) and a coated paper lid. The original version of this kit, the Altaviz Needle Kit, was cleared on December 5, 2022 (K222681), and the only difference between the Altaviz Needle Kit and the Altaviz Needle Kit II is the 29G injection needle included.
The tray of the Altaviz Needle Kit II is designed to securely store two separately packaged and previously sterilized needles: a single-use 29G injection needle and either a sterile 18G or 19G 5 micron filter needle. The tray lid provides protection and helps secure the needles during transport. No changes to the primary packaging of the needles or additional sterilization occurs prior to inclusion in the Altaviz Needle Kit II.
The 29G needle is a sterile hypodermic single lumen needle, for single use consisting of a stainless steel tube that is sharpened at one end and at the other end joined to a female luer connector (hub) made of polycarbonate designed to be connected with a male connector (nozzle) of a piston syringe.
The 18G or 19G 5 micron filter needle is a sterile medical device for single use consisting of a stainless steel tube that is sharpened at one end and at the other end joined to a female luer connector (hub) made of polycarbonate or ABS (respectively) designed to be connected with a male connector (nozzle) of a piston syringe. The filter needle is used to draw up fluids into a syringe.
### Indications for Use
The Altaviz Needle Kit II is a convenience kit that includes:
A 29G injection needle that is a sterile medical device for single use intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.
Either an 18G or a 19G 5 micron filter needle that is a sterile medical device for single use intended to draw up fluids into a syringe.
{5}------------------------------------------------
The Altaviz Needle Kit II is indicated for intravitreal use.
# Physical and Performance Characteristics
The Altaviz Needle Kit II is a convenience kit of currently distributed and routinely used needles that have been combined for the convenience of the end user for use in the injection of intravitreal fluids. The Altaviz Needle Kit II components (needles) are identical to the separately distributed versions and have not been changed. The included 29G injection needle is commonly used to inject intravitreal fluids and the included 5 micron filter needle is commonly used to draw up medication into a syringe prior to injection.
| Features and<br>Characteristics | Proposed Altaviz<br>Needle Kit II | Primary Predicate<br>Altaviz Needle Kit | Comparison of Proposed<br>Device and Predicate |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K231261 | K222681 | |
| Device Code | FMI, GAA | FMI, GAA | Same as Predicate |
| Intended Use | Used to inject<br>fluids into, or<br>withdraw fluids<br>from, parts of the<br>body below the<br>surface of the skin. | Used to inject fluids<br>into, or withdraw<br>fluids from, parts of<br>the body below the<br>surface of the skin | Same as Predicate |
| Indications for<br>Use | The Altaviz Needle<br>Kit II is a<br>convenience kit<br>that includes:<br><br>A 29G injection<br>needle that is a<br>sterile medical<br>device for single<br>use intended to<br>inject fluids into,<br>or withdraw fluids<br>from, parts of the<br>body below the<br>surface of the skin.<br><br>Either an 18G or a<br>19G 5 micron filter<br>needle that is a | The Altaviz Needle<br>Kit II is a<br>convenience kit that<br>includes:<br><br>A 29G injection<br>needle that is a<br>sterile medical<br>device for single<br>use intended to<br>inject fluids into, or<br>withdraw fluids<br>from, parts of the<br>body below the<br>surface of the skin.<br><br>Either an 18G or a<br>19G 5 micron filter<br>needle that is a | Same as Predicate<br>Additional testing to<br>intraocular acceptance<br>criteria for the TSK 29G<br>needle has been performe |
| | sterile medical<br>device for single<br>use intended to<br>draw up fluids into<br>a syringe.<br><br>The Altaviz Needle<br>Kit II is indicated<br>for intravitreal use. | sterile medical<br>device for single<br>use intended to<br>draw up fluids into<br>a syringe.<br><br>The Altaviz Needle<br>Kit II is indicated<br>for intravitreal use. | |
| Mechanism of<br>Action | Manual | Manual | Same as Predicate |
| Where Used | Hospital, surgical<br>setting | Hospital, surgical<br>setting | Same as Predicate |
| Kit Components | TSK Steriject<br>29G Hypodermic<br>Needle<br><br>Either an 18G or<br>a 19G 5 micron<br>filter needle | Terumo K-Pack II<br>29G Thin Wall<br>Needle<br><br>Either an 18G or a<br>19G 5 micron<br>filter needle | Same as Predicate with<br>exception of supplier of<br>injection needle |
| Packaging/<br>How Supplied | A convenience kit<br>containing<br>previously<br>sterilized devices | A convenience kit<br>containing<br>previously<br>sterilized devices | Same as Predicate |
| Materials of<br>Construction | Needle kit tray:<br>polyethylene<br>terephthalate<br>glycol (PETG)<br><br>29G Needle:<br>stainless steel and<br>polypropylene<br><br>18G or 19G filter<br>needle: stainless<br>steel and<br>polycarbonate or<br>ABS (respectively) | Needle kit tray:<br>polyethylene<br>terephthalate<br>glycol (PETG)<br><br>29G Needle:<br>stainless steel and<br>polypropylene<br><br>18G or 19G filter<br>needle: stainless<br>steel and<br>polycarbonate or<br>ABS (respectively) | Same as Predicate |
| Sterility | Contains Sterile<br>(previously<br>sterilized) devices.<br>No additional<br>sterilization<br>performed prior to<br>including in the<br>Altaviz Needle Kit<br>II | Contains Sterile<br>(previously<br>sterilized) devices.<br>No additional<br>sterilization<br>performed prior to<br>including in the<br>Altaviz Needle Kit<br>II | Same as Predicate |
## Comparison of Technological Characteristics with the Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
## Substantial Equivalence Discussion
The subject device is substantially equivalent to the predicate device when evaluating intended use and technological characteristics. There are no differences between the subject device and the predicate device with respect to intended use. The subject device is specifically indicated for intravitreal use while the standard hypodermic needle has a broad indication for all parts of the body. To support intravitreal use, ethylene oxide residuals, endotoxin, and particulate matter (see Performance Data) were tested to ensure compliance with intraocular requirements. An evaluation of biocompatibility for intravitreal injection was also conducted to ensure that the proposed device is safe for intravitreal injections.
The technological characteristics of the 29G injection needle included in the subject device are substantially equivalent to the additional predicate device in that they are identical devices. The 18G or 19G 5 micron filter included in the Altaviz Needle Kit II is a previously marketed class 1 device provided with the 29G injection needle for convenience of the user.
# Performance Data
Particulate testing of the kit components (needles) according to USP <788> Particulate Matter in Injections and USP <789> Particulate Matter in Ophthalmic Solutions was performed and passed. In addition, intraocular irritation testing according to ISO 10993-10 was performed and passed. Ethylene oxide residuals were assessed per ISO 10993-7. Endotoxins were also tested as recommend in FDA guidance Endotoxin Testing Recommendations for Single-Use Intraocular Ophthalmic Devices.
#### Conclusion
The Altaviz Needle Kit II is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.