TSK STERIJEKT PREMIUM DISPOSABLE HYPODERMIC NEEDLE

K970370 · Tsk Laboratory, Japan · FMI · Apr 10, 1997 · General Hospital

Device Facts

Record IDK970370
Device NameTSK STERIJEKT PREMIUM DISPOSABLE HYPODERMIC NEEDLE
ApplicantTsk Laboratory, Japan
Product CodeFMI · General Hospital
Decision DateApr 10, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

To inject fluids into, or withdraw fluids from parts of the body below the surface of the skin.

Device Story

TSK STERUEKT is a sterile, single-use hypodermic needle; consists of stainless steel beveled cannula bonded to plastic hub; compatible with standard luer lock or luer slip syringe tips. Used by clinicians for fluid injection or withdrawal from body tissues. Device provides physical conduit for fluid transfer; ensures patient safety through sterile, disposable design; prevents cross-contamination.

Clinical Evidence

Bench testing only. Results: Protector removal and needle bond strength met specifications. Biocompatibility testing passed for sensitization, intracutaneous reactivity, hemolysis, systemic toxicity, and cytotoxicity. LAL pyrogen test passed.

Technological Characteristics

Stainless steel cannula; plastic hub; luer lock/luer slip compatible; sizes 14G-31G; sterile, single-use; removable plastic protector.

Indications for Use

Indicated for patients requiring fluid injection or withdrawal via hypodermic needle; applicable for all ages and genders requiring subcutaneous, intramuscular, or intravenous access.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Submission Summary (Full Text)

{0} Attachment VII 510 (k) SUMMARY K970370 TSK STERUEKT Premium Disposable Hypodermic Needle APR 10 1997 Manufacturer: Tochigi Seiko KK 2-1-5 Hirayanagi-cho Tochigi-shi, Tochigi Ken 328 Japan Telephone: +81 282 27 8426 Fax: +81 282 27 1943 510 (k) submitted by: Norman Shane Johnson Distribution Manager TSK Laboratory, Japan 2-1-5 Hirayanagi-cho Tochigi-shi, Tochigi Ken 328 Japan Telephone: +81 282 27 8426 Fax: +81 282 27 1943 Date Summary Prepared: January 10, 1997 Device Trade Name: TSK STERUEKT Premium Disposable Hypodermic Needle Common Name: Hypodermic Needle Classification Name: Hypodermic Single Lumen Needle (per 21CFR 880.5570) Predicate Device: Becton Dickinson & Company, 'B-D PrecisionGlide' needles Description of Device: TSK STERUEKT is a sterile Hypodermic Single Lumen Needle, packaged individually for single patient use. It consists of a stainless steel beveled cannula bonded to a plastic hub. Sizes include 14G-31G and various lengths. The needle is intended to fit any luer lock or luer slip syringe tip. Intended Use: To inject fluids into, or withdraw fluids from parts of the body below the surface of the skin. Safety and Efficacy: The following tests were performed to show the safety and efficacy of the device. | TEST | RESULT | | --- | --- | | Protector Removal | Meets Specification | | Needle Bond Strength | Meets Specification | | Sensitization | Pass | | Intracutaneous | Pass | | Hemolysis | Pass | | Systemic Toxicity | Pass | | LAL Pyrogen Test | Pass | | Cytotoxicity | Pass | Substantial Equivalence: The TSK Premium Disposable Hypodermic Needle and the predicate device, the B-D PrecisionGlide Needle, have been shown to be substantially equivalent based on the following: - identical intended use - disposable sterile hypodermic needle, individually packaged. - equivalent materials - stainless steel cannula and plastic hub. - equivalent design - metal cannula with bonded plastic hub; removable plastic protector on needle tip; sufficient physical integrity for the intended use; fits Luer Lock or Luer Slip hypodermic tips. - similar size range - TSK 14G-31G; B-D 16G-30G. - needle tip selection - both brands are available with several bevels.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...