K231131 · Shanghai Bele Medical Technology Co.,Ltd · PKT · Jun 22, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231131
Device Name
Diode Laser Body Sculpture Systems
Applicant
Shanghai Bele Medical Technology Co.,Ltd
Product Code
PKT · General, Plastic Surgery
Decision Date
Jun 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This product is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.
Device Story
Diode Laser Body Sculpture System (BL-A10) is a 1060nm hyperthermic laser lipolysis device for non-invasive body contouring. It uses infrared laser energy to target and heat adipose tissue, inducing fat cell breakdown to reduce fat bulges in abdomen, flanks, back, and thighs. The system utilizes up to four applicator heads attached to the patient via a belt, incorporating contact cooling for safety. It is intended for use by healthcare professionals in a clinical setting. The provider positions applicators on target areas; the device delivers continuous wave (CW) laser energy to heat fat cells while protecting the skin surface via contact cooling. This process reduces the number of fat cells in the treated area, resulting in body contouring benefits for the patient.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and biocompatibility (ISO 10993-1).
Technological Characteristics
1060nm (±20nm) infrared diode laser; heat-assisted lipolysis; CW pulse width; 35W peak power per applicator; 4x6 cm² spot size; contact cooling; belt-based applicator attachment; AC110-240V power. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. Biocompatibility per ISO 10993-1 (cytotoxicity, irritation, sensitization).
Indications for Use
Indicated for non-invasive lipolysis of abdomen, flanks, back, and thighs in individuals with BMI ≤ 30 to reduce visible fat bulges.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 22, 2023
Shanghai Bele Medical Technology Co.,Ltd Jeremy Li General Manager Room 402, 4F, Building 1, No.3255 Shengang Road, Songjiang District Shanghai, Shanghai 201600 China
Re: K231131
Trade/Device Name: Diode Laser Body Sculpture Systems Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System For Aesthetic Use Regulatory Class: Class II Product Code: PKT Dated: March 24, 2023 Received: April 21, 2023
Dear Jeremy Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Jianting Wang -S" in a large, clear font. The text is horizontally oriented and appears to be the primary focus of the image. The background is plain, ensuring the text is easily readable.
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K231131
Device Name Diode Laser Body Sculpture Systems
Indications for Use (Describe)
This product is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 3-510(k) summary K231131
# l Submitter
SHANGHAI BELE MEDICAL TECHNOLOGY CO.,LTD
Room 402, 4F, Building 1, NO.3255 Shengang Road, Songjiang District, Shanghai, China
### Contact Person
Jeremy Li
General manager
Phone: +86-189 6489 4730
Fax: +86-21 51012882
E-mail: jeremy@belemed.com
# II Proposed Device
| Trade Name of Device: | Diode Laser Body Sculpture Systems |
|-----------------------|------------------------------------------|
| Common name: | Low Level Laser System for Aesthetic Use |
| Regulation Number: | 21 CFR 878.5400 |
| Regulatory Class: | Class II |
| Product code: | PKT |
| Review Panel: | General & Plastic Surgery |
### III Predicate Devices
| 510(k) Number: | K182741 |
|-----------------|------------------------------------------|
| Trade name: | SculpSure |
| Common name: | Low Level Laser System for Aesthetic Use |
| Classification: | Class II |
| Product Code: | PKT |
| Manufacturer | Cynosure |
# IV Device description
The BL-A10 is a 1060nm Diode Hyperthermic Laser Lipolysis system and is the latest technology for non-invasive body contouring. It applies a 1060nm wavelength laser to target the adipose tissue to reduce fat in areas such as flank and abdomen.
The purpose of the device is to breakdown the fat cells in the target area and in doing so contour the body and reduce the number of fat cells in that area. It is designed to meet international safety and performance standards.
# V Indication for use
This product is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.
# VI Comparison of technological characteristics with the predicate devices
| Item | Proposed device | Predicate device<br>(K182741) | Discussion |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code | PKT | PKT | Identical |
| Regulation No. | 21 CFR 878.5400 | 21 CFR 878.5400 | Identical |
| Class | Class II | Class II | Identical |
| Indication for use | This product is<br>intended for non-<br>invasive lipolysis of<br>the abdomen, flanks,<br>back, and thighs in<br>individuals with a<br>Body Mass Index<br>(BMI) of 30 or less.<br>The device is<br>intended to affect the<br>appearance of visible<br>fat bulges in the<br>abdomen, flanks,<br>back, and thighs. | The Cynosure<br>SculpSureTM is<br>intended for non-<br>invasive lipolysis of<br>the abdomen, flanks,<br>back, and thighs in<br>individuals with a<br>Body Mass Index<br>(BMI) of 30 or less.<br>In addition, the<br>device is intended for<br>non-invasive lipolysis<br>of the submental<br>area in individuals<br>with a BMI of 49 or<br>less. The device is<br>intended to affect the<br>appearance of visible<br>fat bulges in the<br>abdomen, flanks,<br>back, thighs and<br>submental area.<br>When using the<br>petite mask for non-<br>invasive lipolysis of<br>the submental area,<br>the device can also<br>affect the<br>appearance of lax<br>tissue in the<br>submental area. | Equivalent |
| Laser type | Diode laser | Diode laser | Identical |
| Wavelength | 1060nm±20 nm<br>(infrared) | 1060nm±20 nm<br>(infrared) | Identical |
| Lipolysis method | Heat-assisted | Heat-assisted | Identical |
| Spot size | 4 * 6 cm² on each of<br>the Applicator heads<br>(up to four<br>applicators per body<br>treatment) | 4 * 6 cm² on each of<br>the Applicator heads<br>(up to four<br>applicators per body<br>treatment) | Equivalent |
| | | 14.28 cm² (for standard submental mask)<br>10.49 cm2 (for petite submental mask) | |
| Pulse width | CW | CW | Identical |
| Power density | 0.6~1.6W/cm² | Up to 1.4 W/ cm² (body)<br>Up to 3.06 W/ cm² (submental) | Similar |
| Attachment to patient | Belt | Belt | Identical |
| Voltage | AC110-240V, 16-4A , 50/60Hz. | AC200-240V, 50/60Hz, 20A | Equivalent |
| Peak power | 35W (per applicator) | 30W (per applicator) | Equivalent |
| Cooling | Contact cooling | Contact cooling | Identical |
{4}------------------------------------------------
{5}------------------------------------------------
# VII Non-Clinical Testing
A series of tests have been performed to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:
#### Electrical safety and electromagnetic compatibility
- IEC 60601-1 : 2005+A1 : 2012+A2 : 2020 / EN 60601-1: 2006+A1: ● 2013+A12:2014+A2:2021 Medical electrical equipment. General requirements for basic safety and essential performance
- IEC60601-1-2:2014+A1:2020 / EN 60601-1-2:2015+A1:2021 Medical electrical equipment - part 1-2 General requirements for basic - safety - and - performance -Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 60601-2-22:2019/EN 60601-2-22:2020 Medical electrical equipment - part 2-22: Particular requirements for basic essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- IEC 60825-1:2014/EN 60825-1:2014 Safety of laser products-Part 1:Equipment classification and requirements
### Biocompatibility Evaluation:
Per FDA's Biocompatibility Guidance issued on September 04, 2020 and with regard to Table A.1 Evaluation Tests for consideration in ISO, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" " the following tests performed on the material which contacts with human for Biocompatibility:
- Cytotoxicity;
- Skin irritation;
{6}------------------------------------------------
● Skin Sensitization.;
# VIII Clinical Testing
lt is not applicable.
# IX Conclusion
Base on the performance testing and validation studies that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.