K222265 · Shanghai Wonderful Opto-Electrics Tech.Co.,Ltd · PKT · Sep 23, 2022 · General, Plastic Surgery
Device Facts
Record ID
K222265
Device Name
Diosculpt
Applicant
Shanghai Wonderful Opto-Electrics Tech.Co.,Ltd
Product Code
PKT · General, Plastic Surgery
Decision Date
Sep 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DioSculpt laser system is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less.
Device Story
DioSculpt is a diode laser system for non-invasive lipolysis; delivers 1060nm infrared laser energy to subcutaneous tissue layers to disrupt adipocyte cells. Operated by clinicians in a professional setting; applicators secured to patient via belt. Device features self-contained closed-loop cooling to maintain skin surface temperatures below 41°C. Output is heat-assisted laser energy; provider monitors system to ensure safe thermal levels. Benefits include non-invasive fat reduction in specified body areas.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993), electrical safety (AAMI/ANSI ES60601-1), EMC (IEC 60601-1-2), and performance testing (IEC 60601-2-22, IEC 60825-1). Temperature testing confirmed applicator surface temperatures do not exceed 41°C.
Technological Characteristics
Diode laser system; 1060±20nm wavelength; CW pulse width; up to 1.8W/cm² energy density; 57.6W peak power per applicator. Self-contained closed-loop cooling. Belt-based patient attachment. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for non-invasive lipolysis of abdomen, flanks, back, and thighs in individuals with BMI ≤ 30.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Cynosure Sculpsure Low Level Laser System for Aesthetic Use (K171992)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
September 23, 2022
Shanghai Wonderful Opto-Electrics Tech. Co., Ltd. Lily Zhou Management Representative 2F, Building 11, Lane 1175, Tongpu Rd. Shanghai, Shanghai 200333 China
Re: K222265
Trade/Device Name: Diosculpt Regulation Number: 21 CFR 878.5400 Regulation Name: Low level laser system for aesthetic use Regulatory Class: Class II Product Code: PKT Dated: November 14, 2021 Received: July 28, 2022
Dear Lily Zhou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22265
Device Name DioSculpt
Indications for Use (Describe)
The DioSculpt laser system is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less.
Type of Use (Select one or both, as applicable)
| ☑ Prescriptive Use (Part 21 CFR 201 Subpart D) | ☐ Over-The-Counter Use (21 CFR 201 Subpart C) |
|------------------------------------------------|-----------------------------------------------|
|------------------------------------------------|-----------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for SWOT. The logo is blue and features the word "SWOT" in a stylized font. There are two lines going through the letters "S" and "W". There is a registered trademark symbol in the upper right corner of the logo.
Date Prepared: Sep. 6th, 2022
# 510(k) Summary
(As required by 21 CFR 807.92)
#### l. SUBMITTER
| Name of Sponsor: | Shanghai Wonderful Opto-Electrics Tech. Co., Ltd. |
|------------------|-------------------------------------------------------------------|
| Address: | 2F, Building 11, Lane 1175, TongPu Rd.,<br>Shanghai 200333, China |
| Contact Name: | Lily Zhou |
| Telephone No.: | 0086-021-62642623 |
| Fax No.: | 0086-021-52827988 |
| Email Address: | laser@wonderful-sh.com |
#### II. DEVICE
| Trade Name: | Diosculpt |
|---------------------------|-----------------------------------------------------------|
| Common Name: | Low Level Laser for Lipolysis |
| Model Name: | Diosculpt |
| Regulation Classification | Low Level Laser System for Aesthetic Use |
| Product Code: | PKT |
| Classification Name: | Laser for disruption of adipocyte cells for aesthetic use |
| Classification Panel: | General & Plastic Surgery |
| Device Class: | II |
| Regulation Number: | 21 CFR 878.5400 |
### III. PREDICATE DEVICE
The identified predicates within this submission are as follows:
Primary Predicate Device: The Cynosure Sculpsure Low Level Laser System for Aesthetic Use has been cleared by FDA through 510(k) No.K171992 (Decision Date-September 26, 2017).
Secondary Predicate Device: K191068, Powersculp laser lipolysis system, Wuhan Lotuxs Technology Co.,Ltd.(Decision Date-July 17, 2019)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for SWOT. The logo is blue and features the word "SWOT" in a stylized font. There are two horizontal lines that run through the word "SWOT", and there is a circle with an R in it in the upper right corner of the logo. The logo is simple and modern, and it is likely used to represent a company or organization.
#### DEVICE DESCRIPTION IV.
The DioSculpt laser lipolysis system is a diode laser system, Electrically efficient semiconductor generate optical radiation (1060nm) which is used to deliver laser energy to subcutaneous tissue layers. DioSculpt's cooling and electical systems assist in maintaining safe and comfortable skin surface temperatures.
#### V. INDICATIONS FOR USE
The DioSculpt laser system is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less.
### VI. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
The subject device in this 510(k) uses the same technology that is utilized in the predicate device. A comparison of technological characteristics is provided in the following table:
| 510(k) Number | K 222265 | K171992 | K191068 |
|------------------------------|-----------------------------------|-----------------------------------|-----------------------------------|
| Lipolysis Method | Heat-assisted | Heat-assisted | Heat-assisted |
| Device Type | Diode Laser | Diode Laser | Diode Laser |
| Wavelength | 1060±20nm<br>(infrared) | 1060±20nm<br>(infrared) | 1060±20nm<br>(infrared) |
| Pulse Width | CW | CW | CW |
| Energy Density | Up to 1.8W/cm² | Up to 1.4W/cm² | Up to 1.7W/cm² |
| Spot Size | 4*8 cm² on each<br>of Applicators | 4*6 cm² on each<br>of Applicators | 4*8 cm² on each<br>of Applicators |
| Peak power per<br>applicator | 57.6W | 33.6W | 50W |
| Cooling | Self-contained<br>closed loop | Self-contained<br>closed loop | Self-contained<br>closed loop |
| Attachment to<br>patient | Belt | Belt | Belt |
#### Discussion
The energy density of DioSculpt is 0.1W/cm² higher than the predicate device K191068, and the peak power energy is accordingly slightly higher. The higher energy may generate higher temperature to skin. To quarantee the safety, two temperature testing were conducted to measure the four applicators temperature in worst-case scenario and real operation
scenario. All testing results do not exceed maximum allowable temperature of 41°C .
The temperature testing result demonstrates the minor difference in energy and peak power per applicator does not raise safety concerns.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains a logo with the text "SWOT" in blue. The letters are stylized and connected by a blue line. There is a registered trademark symbol in the upper right corner of the logo. The logo appears to be for a company or product named SWOT.
# VII. NON-CLINICAL TESTING
#### Biocompatibility testing
The biocompatibility evaluation for DioSculpt was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
The applicators are considered tissue contacting, which is categorized as "Surface device", " Intact skin contact", and "with a limited contact duration of less than 24 hours", the following tests were evaluated:
- O Skin Sensitization
- O In Vitro Cytotoxicity
- O Intracutaneous Reactivity
#### Electrical Safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on DioSculpt. The system complies with the AAMI/ANSI ES60601-1:2005(R) 2012 and A1:2012 standards for safety and IEC 60601-1-2:2014 standard for EMC.
#### Performance Testing
The performance testing were conducted on DioSculpt. The system complied with IEC60601-2-22:2019 and IEC 60825-1:2014.
To guarantee the safety, two temperature testing were conducted to measure the four applicators temperature in worst-case scenario and real operation scenario. All testing results do not exceed maximum allowable temperature of 41℃ .
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury, either to patient or to a user of device.
## VIII. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, it is concluded, basing on performance testing, indication for use, and technology, the subject device is the same or similar to the predicate device K171992 and K191068.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.