K171992 · Cynosure, Inc. · PKT · Sep 26, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171992
Device Name
SculpSure
Applicant
Cynosure, Inc.
Product Code
PKT · General, Plastic Surgery
Decision Date
Sep 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cynosure SculpSure laser system is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. In addition, the device is intended for non-invasive lipolysis of the submental area in individuals with a Body Mass Index (BMI) of 43 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, thighs, and submental area.
Device Story
SculpSure is a diode laser system for non-invasive lipolysis; utilizes 1060 nm infrared optical radiation to heat subcutaneous adipose tissue. System comprises a console and four applicators; applicators secured to patient via belts (body) or headgear/straps (submental). Operated by clinicians in a clinical setting. Laser energy disrupts adipocyte cells; heat-assisted mechanism reduces fat bulges. Clinicians monitor treatment; output affects fat appearance. Benefits include non-invasive fat reduction without surgery.
Clinical Evidence
Prospective, controlled study (N=57) evaluated submental fat reduction. Subjects received up to 2 treatments. Endpoints: blind photo evaluation, ultrasound-measured adipose tissue thickness, and patient satisfaction (Likert scale) at 12 weeks. Results: 93% (51/55) of post-treatment photos correctly identified; 15.2% (1.785mm) normalized fat reduction; 100% (55/55) patient satisfaction. Adverse events included swelling, pain, nodules, redness, hardness, numbness, hair loss, itching, bruising, and blisters; events were transient and mostly mild/moderate.
Technological Characteristics
Diode laser system; 1060 ±20 nm wavelength; CW pulse width. Power density: up to 1.4 W/cm2 (body), 2.35 W/cm2 (submental). Materials: patient-contacting materials compliant with ISO 10993-1. Electrical: 200-240V, 20A. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1. Connectivity: N/A.
Indications for Use
Indicated for non-invasive lipolysis of abdomen, flanks, back, and thighs (BMI ≤30) and submental area (BMI ≤43) to reduce visible fat bulges.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 26, 2017
Cynosure, Inc. Amy Tannenbaum Regulatory Affairs Specialist 5 Carlisle Road Westford, Massachusetts 01886
Re: K171992
Trade/Device Name: SculpSure Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System for Aesthetic Use Regulatory Class: Class II Product Code: PKT Dated: June 30, 2017 Received: July 3, 2017
Dear Amy Tannenbaum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-
Image /page/0/Picture/11 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white and appears to be a scanned image.
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
Image /page/1/Picture/9 description: The image shows the name "Jennifer R. Stevenson -S3" in a large, sans-serif font. The text is black and appears to be against a white background. In the background, the letters "FDA" are faintly visible in a light blue color, suggesting the image may be related to the Food and Drug Administration.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) Kpending
Device Name SculpSure™
Indications for Use (Describe)
The Cynosure SculpSure™ is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. In addition, the device is intended for non-invasive lipolysis of the submental area in individuals with a BMI of 43 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, thighs and submental area.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary for SculpSure Laser System
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| <b>807.92(a)(1) Submitter Information</b> | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Cynosure, Inc. |
| Address | 5 Carlisle Road<br>Westford, MA 01886 |
| Phone Number | (978) 256-4200 |
| Fax Number | (978) 256-6556 |
| Establishment Registration<br>Number | 1222993 |
| Contact Person | Amy Tannenbaum |
| Preparation Date | June 30, 2017 |
| <b>807.92(a)(2) Name of Device</b> | |
| Trade or Proprietary Name | SculpSure |
| Common or Usual Name | Laser |
| Classification Name | Laser for disruption of adipocyte cells for aesthetic use |
| Classification Panel | General & Plastic Surgery |
| Regulation | 21 CFR 878.5400 |
| Regulatory Class | II |
| Product Code(s) | PKT |
| <b>807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed</b> | |
| Predicate Devices | Cynosure SculpSure K171111, Zeltiq Coolsculpting K162050<br>The predicate devices have not been subject to a design-related<br>recall |
| <b>807.92(a)(4) Device Description</b> | |
| The Cynosure SculpSure is a diode laser system. The main<br>components of SculpSure are a console and four applicators that<br>deliver the laser energy to the patient. Electrically efficient<br>semiconductors generate optical radiation (1060 nm) which is<br>used to deliver laser energy to subcutaneous tissue layers. | |
| <b>807.92(a)(5) Intended Use of the Device</b> | |
| The SculpSure laser system is intended for non-invasive lipolysis<br>of the abdomen, flanks, back, and thighs in individuals with a<br>Body Mass Index (BMI) of 30 or less. In addition, the device is<br>intended for non-invasive lipolysis of the submental area in<br>individuals with a Body Mass Index (BMI) of 43 or less. The<br>device is intended to affect the appearance of visible fat bulges in | |
{4}------------------------------------------------
| the abdomen, flanks, back, thighs, and submental area. | | | |
|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------|
| 807.92(a)(6) Summary of the Technological Characteristics of the Device Compared to the<br>Predicate | | | |
| | Proposed SculpSure<br>Laser System | Cynosure SculpSure<br>Laser System | Zeltiq Coolsculpting<br>Device |
| 510(k) Number | KPending | K171111 | K162050 |
| Manufacturer | Cynosure, Inc. | Cynosure, Inc. | Zeltiq Aesthetics, Inc. |
| Lipolysis Method | Heat-assisted | Heat-assisted | Cold-assisted |
| Device Type | Diode Laser | Diode Laser | N/A |
| Wavelength | 1060 ±20 nm<br>(infrared) | 1060 ±20 nm<br>(infrared) | N/A |
| Spot Size | 4 x 6 cm2 on each of<br>the Applicator heads<br>(up to four applicators<br>per body treatment)<br>14.28 cm2 (one<br>applicator head and<br>frame used for<br>Submental treatment) | 4 x 6 cm2 on each of<br>the Applicator heads<br>(up to four applicators<br>per treatment) | 4.5 x 7 cm2 (2x)<br>approximate |
| Pulse Width (laser<br>ON time) | CW | CW | N/A |
| Power Density | Up to 1.4 W/cm2<br>(body)<br>Up to 2.35 W/cm2<br>(submental) | Up to 1.4 W/cm2 | N/A |
| Attachment to<br>patient | Belt (body treatment)<br>Headgear and straps<br>(submental treatment) | Belt (body treatment<br>only) | Belt |
| Voltage | 200-240V~, Single<br>Phase | 200-240V~, Single<br>Phase | N/A |
| Current | 20A | 20A | N/A |
Software verification and validation was performed, and it was demonstrated that the software performs as intended. Testing confirmed that the power output meets specification. Electromagnetic compatibility and electrical safety testing was performed per standards IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22 and IEC 60825-1. Results confirmed the device meets the standards. All patient contacting materials were assessed as per ISO 10993-1 and found to be biocompatible. Additional bench testing was performed to demonstrate that the submental mask provides beam uniformity to the treatment area within +/- 15% specification.
{5}------------------------------------------------
## 807.92(b)(2) Clinical tests submitted
A prospective, controlled study was conducted at 3 study centers to evaluate the safety and efficacy of Cynosure SculpSure laser for the non-invasive fat reduction of the submental area. The treatment techniques used and subsequent follow up schedule for the study were kept consistent between the three studies. 57 subjects were enrolled in the study. 23 of the subjects had a BMI of less than 30, and 34 subjects had a BMI of 30 or higher. Each subject received up to 2 treatments with SculpSure at the submental area. There were 5 patients that received 1 treatment, and the rest received 2 treatment effectiveness was assessed through blind evaluation of pre- and post- final treatment (12 week) photographs; percentage change in the adipose tissue thickness from baseline to 12 week follow up measured through ultrasound imaging; and patient satisfaction measured at the 12 week post- final treatment follow up visit through 6 point Likert scale. On an average, blind evaluators were able to identify 93% (51/55) of the post treatment photographs; ultrasound imaging at 12 week post follow up visit showed 15.2% (1.785mm) normalized fat reduction and patient satisfaction survey at 12 week post follow up visit showed 100% (55/55 subjects) satisfaction rate. All subjects that were treated in the study (N=57) were included in the safety analysis and adverse events were assessed at all visits. The side effects that were reported were swelling, pain, nodules, redness, hardness, numbness, hair loss, itching, bruising, and blisters. All events, except for one, were transient and the majority were mild or moderate in nature. There was no significant difference between the safety and efficacy results of the subjects that had a BMI below 30 as compared to the safety and efficacy results of the subjects that had a BMI of 30 or above. The study concluded that the SculpSure is safe and effective for non-invasive fat reduction at the submental area.
### 807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted
Cynosure believes that SculpSure is substantially equivalent to its predicate devices with same intended use and similar technological characteristics. The non-clinical data for SculpSure supports the safety of the device and the biocompatibility, hardware and software verification and validation demonstrate that the SculpSure should perform as intended in the specified use conditions. Additionally, the successful clinical performance of SculpSure as documented in clinical study demonstrate that the SculpSure has a safety and effectiveness profile that is similar to its predicate devices. Therefore, Cynosure is requesting clearance for treatment of the submental area.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.