Talon Transseptal Sheath

K231108 · Inari Medical, Inc. · DYB · Sep 27, 2023 · Cardiovascular

Device Facts

Record IDK231108
Device NameTalon Transseptal Sheath
ApplicantInari Medical, Inc.
Product CodeDYB · Cardiovascular
Decision DateSep 27, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The Talon Transseptal Sheath is used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation/puncture.

Device Story

Talon Transseptal Sheath is a single-lumen catheter providing a conduit for cardiovascular catheterization and angiography. It features a proximal hemostasis valve, stopcock with flush port, and a radiopaque distal marker for fluoroscopic visualization. Packaged with a dilator (compatible with 0.035" guidewires), Y-connector, and syringes (quick-release and VacLok). Used by physicians in clinical settings to facilitate access to heart chambers. The device is inserted percutaneously; the dilator aids in positioning. Output is a stable conduit for secondary devices. Benefits include enabling minimally invasive access to the left atrium and other heart chambers for diagnostic or therapeutic procedures.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), packaging integrity, dimensional inspections, compatibility testing (guidewire, dilator, Triever20, pigtail), torque, tensile, kink radius, leak, vacuum, flowrate, and radiopacity testing.

Technological Characteristics

Single-lumen catheter; Pebax 3533-7233 shaft material; 23 Fr OD, 20.7 Fr ID, 80 cm length; 90° in-plane and 30° perpendicular tip curve. Sterilized via EtO (SAL 10^-6) per ISO 11135:2014/Amd 1:2018 and AAMI TIR 28:2016. Mechanical device; no software or electronic components.

Indications for Use

Indicated for patients requiring percutaneous introduction of cardiovascular catheters and guidewires into heart chambers, including the left atrium via transseptal access.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 27, 2023 Inari Medical, Inc. Ellen Nguyen Regulatory Affairs Specialist 6001 Oak Canyon Suite 100 Irvine, California 92618 Re: K231108 Trade/Device Name: Talon Transseptal Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: August 22, 2023 Received: August 22, 2023 Dear Ellen Nguyen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely. Rachel E. Neubrander -S Rachel Neubrander, PhD Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231108 Device Name Talon Transeptal Sheath Indications for Use (Describe) The Talon Transseptal Sheath is used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation/puncture. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | <div> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### PAGE 1 OF 4 # 510(K) SUMMARY | Date prepared | September 22, 2023 | | Subject Device<br>Talon Transseptal Sheath | Primary Predicate<br>VersaCross Transseptal Sheath | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name | Inari Medical, Inc.<br>6001 Oak Canyon, Suite 100<br>Irvine, CA 92618<br>877.923.4747 | K Number | TBD | K183655 | | Contact person | Ellen Nguyen<br>Sr. Regulatory Affairs Specialist | Manufacturer | Inari Medical, Inc. | Baylis Medical Company, Inc. | | Trade name | Talon Transseptal Sheath | Regulations | 21 CFR 870.1340<br>Introducer, catheter | 21 CFR 870.1340<br>Introducer, catheter | | Common name | Catheter introducer | Product Code | DYB | DYB | | Regulation name | Catheter introducer | Indications for Use | The Talon Transseptal Sheath is used<br>for the percutaneous introduction of<br>various types of cardiovascular<br>catheters and guidewires to all heart<br>chambers, including the left atrium via<br>transseptal perforation/puncture. | The VersaCross Transseptal Sheath is<br>used for the percutaneous introduction<br>of various types of cardiovascular<br>catheters and guidewires to all heart<br>chambers, including the left atrium via<br>transseptal perforation/puncture. | | Classification number | 21 CFR 870.1340 | Target Vessel | Heart (all chambers) | Heart (all chambers) | | Product code | DYB | Sterility | SAL 10-6, EO | SAL 10-6, EO | | Regulatory class | II | Shelf-life | 6 months | Unknown | | Predicate device(s) | Baylis Medical, VersaCross Transseptal Sheath (K183655) | Guidewire<br>compatibility | 0.035" | 0.035" | | Reference device(s) | Cook, Extra Large Check-Flo Introducer (K203670)<br>Inari Medical, Triever24 (K213402) | Sheath<br>Dimensions | OD: 23 Fr<br>ID: 20.7 Fr<br>Length: 80 cm | OD: 10.5 Fr<br>ID: 8.5 Fr<br>Length: 63 cm, 81 cm | | Description | The Talon Transseptal Sheath (“Sheath”) provides a conduit for catheterization and angiography of specific heart chambers and locations. It consists of a single lumen catheter with a proximal hemostasis valve and stopcock with flush port. A radiopaque marker is positioned near the distal tip to aid with fluoroscopic visualization. The Sheath is packaged with a dilator and two 60 cc syringes. The dilator is compatible with a 0.035" guidewire and has a tapered tip which aids insertion and positioning of the Sheath. The dilator attaches to the Sheath hemostasis valve, and a Y-connector with hemostasis valve connects to the dilator luer connector. A quick-release syringe is provided to connect to the Sheath sideport, and a standard VacLok syringe is provided to connect to the Sheath flush port and dilator connector. | Shaft<br>Materials | Pebax 3533-7233 | Unknown | | Indications for Use | The Talon Transseptal Sheath is used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation/puncture. | Tip curve | 90° in plane, 30° perpendicular | 45°, 55°, 90° in plane | | Accessories<br>provided | Dilator, quick-release syringe,<br>standard VacLok syringe, Y-connector | Dilator, J-tipped guidewire | | | {4}------------------------------------------------ Summary of substantial equivalence A tabular comparison of the predicate and subject devices is provided below: Summary of substantial equivalence #### Biocompatibility ● The following biocompatibility tests were completed for the subject device: - . Cytotoxicity - Sensitization - Intracutaneous Reactivity Acute Systemic Toxicity . - Material-Mediated Pyrogenicity ● - . Hemocompatibility (Hemolysis, Complement Activation, Thromboresistance, Platelet and Leukocyte Count, and Partial Thromboplastin Time) {5}------------------------------------------------ The passing results demonstrate that the subject device and accessories meet biological safety requirements per ISO 10993-1. #### Sterilization The subject device, including its accessories, is sterilized using EtO to achieve a sterility assurance level (SAL) of 106 using a validated sterilization process in accordance with the principles of ISO 11135:2014/Amd 1:2018 and AAMI TIR 28:2016. #### Non-Clinical Testing In accordance with the Design Failure Modes and Effects Analysis, verification and validation testing were identified to support the substantial equivalence of the Talon Transseptal Sheath to the predicate device. These tests included: #### Performance Tests - Packaging Integrity ● - Visual and Dimensional Inspections ● - Guidewire and Sheath Compatibility ● - Dilator Compatibility - Test Small-Bore Connectors for Intravascular Applications - Insertion and Removal Force of Dilator - Simulated Use, Track and Transseptal Access ● - Simulated Use, Track and Triever20 Curve Compatibility - Shape Recovery Angles - Simulated Use, Track and Pigtail Compatibility - Torque Testing ● - Simulated Use, Track and Kink Radius ● - Simulated Use, Track and Torque Testing Talon ● - Simulated Use, Track and Torque Testing Talon and Triever20 ● Curve - Leak Testing Sheath and Dilator ● - Vacuum Testing . - Air Leakage During Aspiration - Tensile Testing - Particulate Evaluation - Snare Compatibility ● - Shaft Stiffness Testing ● - Tip Stiffness Testing ● - Force Transmission ● - Flowrate through Sheath - Torque to Failure ● - Radiopacity ● Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications. {6}------------------------------------------------ Animal testing was not required for the determination of substantial equivalence. Clinical testing was not required for the determination of substantial equivalence. #### Conclusion The Talon Transseptal Sheath has the same intended use/indications for use and principles of operation as the predicate. Non-clinical performance data show that the different technological characteristics between the devices do not raise any new or different questions of safety or effectiveness and support the Talon Transseptal Sheath's substantial equivalence to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%