TORFLEX TRANSSEPTAL GUIDING SHEATH

K102948 · Baylis Medical Co., Inc. · DYB · Apr 1, 2011 · Cardiovascular

Device Facts

Record IDK102948
Device NameTORFLEX TRANSSEPTAL GUIDING SHEATH
ApplicantBaylis Medical Co., Inc.
Product CodeDYB · Cardiovascular
Decision DateApr 1, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The TorFlex™ Transseptal Guiding Sheath is used for the percutaneous introduction of various types of cardiovascular catheters to all heart chambers, including the left atrium via transseptal perforation / puncture.

Device Story

TorFlex™ Transseptal Guiding Sheath kit comprises a sheath, dilator, and J-tipped guidewire. Used by physicians in clinical settings for percutaneous introduction of cardiovascular catheters into heart chambers, including left atrium via transseptal puncture. Sheath provides torque control and flexibility; radiopaque tip enables visualization during manipulation; dilator provides support with tapered tip. Device facilitates catheterization and angiography. Benefits include safe, easy access to heart chambers.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity), mechanical testing (torque, flow rate, bend test, valve integrity, pull test, snap fit), and physical testing (surface defects, corrosion resistance). Sterilization and packaging validated per ANSI/AAMI/ISO standards.

Technological Characteristics

Kit includes sheath, dilator, and J-tipped guidewire. Features radiopaque tip for visualization. Sterilized via Ethylene Oxide per ANSI/AAMI/ISO 11135-1. Packaging validated per ANSI/AAMI/ISO 11607. Biocompatibility compliant with ISO 10993-7.

Indications for Use

Indicated for patients requiring percutaneous introduction of cardiovascular catheters into heart chambers, including the left atrium via transseptal puncture.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 6. 510(k) Summary APR - 1 2011 ## Submitter Information لر - A. Company Name: Baylis Medical Company Inc. - B: Company Address: 2645 Matheson Blvd. East Mississauga, Ontario L4W 5S4 Canada - C. Company Phone: (905) 602-4875; ext 252 - D. Company Facsimile: (905) 602-5671 - E. Contact Person: Meghal Khakhar - F. Summary Prepared on: October 04, 2010 ### Device Identification - A. Device Trade Name: TorFlex™ Transseptal Guiding Sheath - B. Device Common Name: Dilator; Sheath; J-tipped Guidewire - C. Classification Name: 1) Catheter introducer - 2) Vessel dilator for percutaneous catheterization - 3) Percutaneous catheter D. Device Class: Class II - E. Device Code: 1) DYB 2) DRE - 3) DQY ### Identification of Predicate Device The predicate device is the TorFlex™ Transseptal Guiding Sheath, which is cleared under 510(k) Premarket Notification Number K013919. Modifications to the current device include changes to the sheath design and materials, kit packaging, and device manufacturer and sterilizer. The indication for use, operation principle, and fundamental scientific technology of the original and proposed TorFlex™ Transseptal Guiding Sheath remains the same. # Device Description The TorFlex™ Transseptal Guiding Sheath kit consists of three components: a sheath, a dilator, and a J-tipped guidewire. {1}------------------------------------------------ The TorFlex™ Transseptal Guiding Sheath is designed for safe and easy catheterization and angiography of specific heart chambers and locations. The sheath provides superior torque control and is flexible. The radiopaque tip maximizes visualization of the sheath during manipulation. The dilator provides support for the sheath and has a tapered tip. ### Intended Use The TorFlex™ Transseptal Guiding Sheath is used for the percutaneous introduction of various types of cardiovascular catheters to all heart chambers, including the left atrium via transseptal perforation / puncture. ### Substantial Equivalence The proposed and 510(k)-cleared TorFlex™ Transseptal Guiding Sheath (K013919) are determined to be substantially equivalent with respect to fundamental scientific technology. This determination is based upon results from performance tests as listed below: #### i) Biocompatibility Testing Test information and results for the following biocompatibility tests demonstrate that the device is safe for its intended use: - . Cytotoxicity - . Sensitization - . Irritation or Intracutaneous Reactivity - Systemic Toxicity (Acute) ● - Hemocompatibility . - . Pyrogenecity #### ii) Mechanical Testing The following tests were performed and determined that the device is able to withstand the following physical stresses without failure: - . Torque transmission + Strength of union -Torque test - Flow rate measurement . - . Three point bend test - . Freedom from liquid leaking through hemostasis valves + valve integrity - Freedom from air leakage through hemostasis valves + valve integrity . - . Freedom from liquid leakage - . Freedom from air leakage - . Strength of union-Pull test - . Valve insertion force - . Tip transition - Snap fit . {2}------------------------------------------------ #### iii) General Physical Testing The following general physical characteristics of the device were evaluated: - Surface Defects - . Corrosion Resistance #### Bench Testing iv) Bench testing was performed to assess the compatibility of the sheath with 8F devices. #### v) Sterlization Validation Sterilization Validation for the TorFlex™ Transseptal Guiding Sheath was conducted as per ANSI/AAMI/ISO 11135-1 and FDA Guideline on Validation of the Limulus Amebocyte Lysate (LAL) Test as an End-Product Endotoxin Test. Ethylene Oxide residue levels are in compliance with ISO 10993-7. #### vii) Packaging Validation The packaging for the TorFlex™ Transseptal Guiding Sheath Kit has been validated in accordance with ANSI/AAMI/ISO 11607. The test results demonstrate that the proposed device is safe and effective and performs as per the intended use. The proposed and 510(k)-cleared TorFlex™ Transseptal Guiding Sheath (K013919) are determined to be substantially equivalent with respect to fundamental scientific technology. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Baylis Medical Co., Inc. c/o Mr. Meghal Khakhar Manager, Regulatory & Scientific Affairs 2645 Matheson Blvd. E. Mississauga, Ontario Canada L4W 5S4 APR - 1 2011 Re: K102948 Trade/Device Name: TorFlex™ Transseptal Guiding Sheath Regulatory Number: 21 CFR 870.1340 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DYB; DQY; DRE Dated: March 24, 2011 Received: March 25, 2011 Dear Mr. Khakhar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 – Mr. Meghal Khakhar Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part.803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. M. G. Hillemann Gram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure - {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): |< 10 2498 Device Name: TorFlex™ Transseptal Guiding Sheath Indications For Use: The TorFlex™ Transseptal Guiding Sheath is used for the percutaneous introduction of various types of cardiovascular catheters to all heart chambers, including the left atrium via transseptal perforation / puncture. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M. A. Killoren (Division Sign-Off) Division of Čardiovascular Devices 510(k) Number_ 1<162948 Page 1 of _1 _________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...