K230971 · Medartis AG · HRS · Jun 2, 2023 · Orthopedic
Device Facts
Record ID
K230971
Device Name
APTUS® 3.5 TriLock Straight Plates
Applicant
Medartis AG
Product Code
HRS · Orthopedic
Decision Date
Jun 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The APTUS 3.5 TriLock Straight Plates are intended for the fixation of fractures, osteotomies and non-unions of the scapula, olecranon, radius, ulna, foot, distal tibia and fibula.
Device Story
APTUS 3.5 TriLock Straight Plates are metallic bone fixation implants; used for stabilization of fractures, osteotomies, and non-unions in specified anatomical locations (scapula, olecranon, radius, ulna, foot, distal tibia, fibula). Plates feature screw holes accommodating locking (TriLock) and non-locking (cortical) screws and K-wires. Device is used by orthopedic surgeons in clinical/surgical settings. Plates are manufactured from titanium alloy (ASTM F136); provided sterile or non-sterile for end-user sterilization. Surgeon selects appropriate plate size (11 designs, 2-16 holes) and secures to bone using compatible screws to provide rigid fixation, facilitating bone healing.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated via non-clinical bench testing, including mechanical testing per ASTM F382, and reliance on previously cleared biocompatibility, sterilization, and packaging data from predicate K193633.
Technological Characteristics
Materials: Titanium alloy (ASTM F136). Design: Straight plates with 11 configurations (2-16 holes), 32-200 mm length. Principle: Locking/non-locking bone fixation. Sterilization: Moist heat (non-sterile) or X-ray beam (sterile). Connectivity: None (mechanical implant).
Indications for Use
Indicated for fixation of fractures, osteotomies, and non-unions of the scapula, olecranon, radius, ulna, foot, distal tibia, and fibula.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
June 2, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
Medartis AG % Kevin Thomas, Phd Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K230971
Trade/Device Name: APTUS® 3.5 TriLock Straight Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: April 3, 2023 Received: April 5, 2023
Dear Dr. Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali-S
Shumaya Ali, M.P.H. Assistant Director DHT6B: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230971
Device Name
APTUS 3.5 TriLock Straight Plates
### Indications for Use (Describe)
The APTUS 3.5 TriLock Straight Plates are intended for the fixation of fractures, osteotomies and non-unions of the scapula, olecranon, radius, ulna, foot, distal tibia and fibula.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><div>☒</div><div>Prescription Use (Part 21 CFR 801 Subpart D)</div></div> | <div style="display:flex; align-items:center;"><div>☐</div><div>Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Medartis AG APTUS 3.5 TriLock Straight Plates April 3, 2022
### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Medartis AG<br>Hochbergerstrasse 60E<br>CH-4057 Basel, Switzerland<br>Telephone +41 61 633 34 34<br>Fax +41 61 633 34 00 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Claudia De Santis, Head of Regulatory Affairs |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone +1 858-792-1235<br>Fax +1 858-792-1236<br>Email kthomas@paxmed.com<br>flarson@paxmed.com |
### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | APTUS 3.5 TriLock Straight Plates |
|------------------------|-----------------------------------------------------------------------------|
| Common Name | Plate, fixation, bone |
| Regulation Number | 21 CFR 888.3030 |
| Regulation Name | Single/multiple component metallic bone fixation appliances and accessories |
| Regulatory Class | Class II |
| Product Code | HRS |
| Classification Panel | Orthopedic |
| Reviewing Office | Office of Health Technology 6 (Orthopedic Devices) |
| Reviewing Division | Division of Health Technology 6 C (Restorative, Repair and Trauma Devices) |
### PREDICATE DEVICE INFORMATION
Primary Predicate Device K193633, APTUS® Ankle Trauma System 2.8/3.5, Medartis AG
Additional Predicate Devices K082807, Synthes (USA) 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications, Synthes (USA) K011335, Synthes One-Third Tubular DCL Plate, Synthes (USA) K110908, APTUS® Foot 3.5 System, Medartis AG K193639, APTUS® Foot 2.8-3.5 System, Medartis AG
{4}------------------------------------------------
The APTUS 3.5 TriLock Straight Plates are intended for the fixation of fractures, osteotomies and non-unions of the scapula, olecranon, radius, ulna, foot, distal tibia and fibula.
### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to obtain marketing clearance for additional plate designs to expand the range of the Medartis APTUS® Ankle Trauma System 2.8/3.5, previously cleared under K193633. The subject device APTUS 3.5 TriLock Straight Plates are available in eleven (11) designs with 2, 3, 4, 5, 6, 7, 8, 10, 12, 14, or 16 screw holes. The plates have an overall length ranging from 32 mm to 200 mm. The subject device plates include screw holes designed to accommodate appropriately sized bone screws and K-wires presently marketed as part of the APTUS System. The subject device plates are compatible with screws and K-wires previously cleared in K193633, K110908, and K092038.
The subject device plates are manufactured from titanium alloy conforming to ASTM F136, and are provided non-sterile or sterile to the end user.
### PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: biocompatibility referenced from K193633; moist heat sterilization (to be performed by the end user) referenced from K193633; X-ray beam sterilization, packaging, and sterile barrier shelf life referenced from K193633; and mechanical testing according to ASTM F382. Clinical data were not provided in this submission.
## EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device, and the additional predicate devices listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements (IFUS) and the technological characteristics of the subject device, the primary predicate device, and the additional predicate devices.
The Indications for Use Statements (IFUS) for the subject device and the predicate devices listed above include language for fixation of fractures, osteotomies, and non-unions of various bones. The IFUS for the predicate devices K110908 and K193639 do not reference non-unions. The IFUS for the subject device and predicate device K193639 include language that includes use in the foot. The IFUS for the additional predicate devices K082807 and K011335 include language referring to fixation of the pelvis that is not relevant to the subject device. The IFUS for K082807 and K011335 also include language referring to use in osteopenic bone; this language is not included in the IFUS for the subject device. Additionally, the IFUS of the subject device does not list specifically adult patients and pediatric patients as in the predicate device K082807. These minor differences among the IFUS do not impact substantial equivalence because all IFUS express equivalent intended use for internal fixation of bones in various anatomic locations.
The primary predicate K193633 is in support of substantial equivalence in terms of similar plate designs; compatible screw designs; identical plate materials; identical materials for the device-specific (Class II) instruments and accessories; the same sterilization, packaging, and shelf life for product provided sterile to the end user; and the same sterilization for devices provided non-sterile to the end user.
{5}------------------------------------------------
The plates from the subject device and the primary predicate device K193633 and the additional predicate device K082807 have the same technological characteristics, have similar design characteristics, and include screw holes to accommodate locking and non-locking screws.
The subject device plates are compatible with Medartis screws previously cleared in K110908, including 3.5 TriLock (locking) and cortical (non-locking) screws. The subject device plates also are compatible with Medartis K-Wires previously cleared in K092038, K193633, and K193639.
K082807 and K011335 are in support of substantial equivalence in terms of similar Indications for Use, comparable plate designs (use with locking and non-locking screws), and a similar range of physical dimensions (overall width, overall length). K011335 is also in support of substantial equivalence in terms of comparative mechanical testing.
All subject device final finished components (plates, device-specific instrument templates, and device-specific accessories) are manufactured in the same facilities using identical materials and identical manufacturing processes as used for the primary predicate device K193633, and therefore, are substantially equivalent to these previously-cleared devices regarding biocompatibility.
The subject device components that are provided sterile to the end user are packaged using the same materials, are sterilized by the same method, and have the same sterile barrier shelf life as the Medartis devices in K193633. The subject device components that are provided non-sterile also are packaged using the same materials, and are to be sterilized by the same methods as the Medartis devices in K193633.
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: biocompatibility referenced from K193633; moist heat sterilization (to be performed by the end user) referenced from K193633; X-ray beam sterilization, packaging, and sterile barrier shelf life referenced from K193633; and mechanical testing according to ASTM F382.
# CONCLUSION
The subject device, the primary predicate device, and the additional predicate devices have the same intended use, have similar technological characteristics, and are made of identical or similar materials. The subject device, the primary predicate, and the additional predicate devices encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{6}------------------------------------------------
| Subject Device | Indications for Use Statement |
|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APTUS 3.5 TriLock Straight Plates<br>Medartis AG | The APTUS 3.5 TriLock Straight Plates are intended for the fixation of fractures, osteotomies and non-unions of the<br>scapula, olecranon, radius, ulna, foot, distal tibia and fibula. |
| Primary Predicate Device | |
| K193633<br>APTUS® Ankle Trauma System 2.8/3.5<br>Medartis AG | APTUS® Ankle Trauma System 2.8/3.5 is indicated for fractures, osteotomies, malunions and non-unions of the distal tibia and fibula. |
| Additional Predicate Devices | |
| K082807<br>Synthes (USA) 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System<br>with Expanded Indications<br>Synthes (USA) | Synthes 3.5 mm Locking Compression Plate (LCP) System:<br>The Synthes 3.5 mm Locking Compression Plate (LCP) System is indicated for fixation of fractures, osteotomies and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula, particularly in osteopenic bone for adult patients.<br>These plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones in pediatric patients. |
| | Synthes 4.5 mm Locking Compression Plate (LCP) System:<br>The Synthes 4.5 mm Locking Compression Plate (LCP) System is indicated for fixation of various long bones, such as the humerus, femur and tibia and for use in fixation of peri-prosthetic fractures, osteopenic bone and fixation of non-unions or malunions in adult patients.<br>These plates are also indicated for fracture fixation of diaphyseal and metaphyseal areas of long bones. in pediatric patients. |
| K011335<br>Synthes One-Third Tubular DCL Plate<br>Synthes (USA) | Synthes One-Third Tubular Plate is intended for fixation of fractures, osteotomies, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, and fibula, particularly in osteopenic bone as a part of the Synthes Small Fragment DCL System. |
| K110908<br>APTUS® Foot 3.5 System<br>Medartis AG | The APTUS® Foot 3.5 System is indicated for fractures and osteotomies of the calcaneus. |
| K193639<br>APTUS® Foot 2.8-3.5 System<br>Medartis AG | The APTUS® Foot System is intended for use in small bones, in particular in fractures, osteotomies and arthrodesis of the tarsals, metatarsals and phalanges. |
# Substantial Equivalence – Indications for Use Statement
{7}------------------------------------------------
| Features / Comparisons | Subject Device | Primary Predicate Device<br>K193633 | Additional Predicate Device<br>K082807 | Additional Predicate Device<br>K011335 | Additional Predicate Device<br>K110908 | Additional Predicate Device<br>K193639 |
|---------------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Product Code | APTUS 3.5 TriLock Straight Plates | APTUS® Ankle Trauma System<br>2.8/3.5 | Synthes (USA) 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications | Synthes One-Third Tubular DCL Plate | APTUS® Foot 3.5 System | APTUS® Foot 2.8-3.5 System |
| Reason for Predicate/Reference Device | Medartis AG | Medartis AG | Synthes (USA) | Synthes (USA) | Medartis AG | Medartis AG |
| | HRS | HRS, HWC, HTN | HRS, HWC | HRS | HRS, HWC | HRS, HWC, PLF |
| | Not applicable | Similar intended use and IFUS; Plate designs, including use with locking and non-locking screws; 3.5 compatible screws; compatible K-wires; identical materials; moist heat sterilization (by end user); x-ray beam sterilization, packaging, sterile barrier shelf life; biocompatibility | Similar intended use and IFUS; Plate designs, including use with locking and non-locking screws; similar range of plate dimensions; similar material; | Similar intended use and IFUS; Plate designs, including use with locking and non-locking screws; similar range of plate dimensions; similar material; comparison mechanical testing | Similar intended use and IFUS; Plate designs, including use with locking and non-locking screws; 3.5 compatible screws | Similar intended use and IFUS; Plate designs, including use with locking and non-locking screws; 3.5 compatible screws |
| Plate Description | Straight plates with 3.5 TriLock locking; scalloped edge design; rounded ends; 11 sizes | Anatomic plate designs; Multiple designs and sizes; Screw holes accommodate 3.5 TriLock (locking) and non-locking (cortex) screws | Straight plates with locking holes; LCP fixed-angle locking holes; 9 sizes | Locking technology; LCP fixed-angle locking holes | Anatomic plate designs; Multiple designs and sizes; Screw holes accommodate 3.5 TriLock (locking) and non-locking (cortex) screws | Anatomic plate designs; Multiple designs and sizes; Screw holes accommodate 3.5 TriLock (locking) and non-locking (cortex) screws |
| | Image: Plate | | Image: Plate | Image: Plate | | |
| Plate Materials | Ti-6Al-4V alloy, ASTM F136 | Ti-6Al-4V alloy, ASTM F136<br>Unalloyed titanium, ASTM F67 | Stainless steel, commercially pure titanium | Stainless steel, commercially pure titanium | Unalloyed titanium, ASTM F67 | Ti-6Al-4V alloy, ASTM F136<br>Unalloyed titanium, ASTM F67 |
| Screw Holes | 2, 3, 4, 5, 6, 7, 8, 10, 12, 14, and 16 holes | Various, 7 - 24 holes | 3, 4, 5, 6, 7, 8, 9, 10, and 12 holes | Various, 2-12 holes | 12 or 13 holes | Various |
| Plate Thickness | 2, 3, and 4 hole plates: 2.4 mm<br>5-16 hole plates: 2.2. mm | Various thicknesses,<br>1.6 mm - 30.5 mm | 1 mm | Not provided in 510(k) Summary | 1.4 mm to 2.0 mm | 1.6 mm to 2.5 mm |
| Plate Lengths | Various lengths,<br>32 mm - 200 mm | Various lengths,<br>50 mm - 260 mm | 3-10 hole plates: 33 mm - 117 mm<br>12 hole plates: 141 mm | 28 mm - 148 mm | Various overall dimensions | Various lengths,<br>25 mm to 92 mm |
| Plate Width | 10.2 mm in the widest part | Various widths | 9.8 mm | Not provided in 510(k) Summary | Various overall dimensions | Various widths,<br>12 mm to 29 mm |
| Sterility | Provided non-sterile and sterile | Provided non-sterile and sterile | Not provided in 510(k) Summary | Provided non-sterile and sterile | Provided non-sterile | Provided non-sterile |
| Sterilization | Non-sterile: to be sterilized by moist heat<br>Sterile: X-ray beam sterilization | Non-sterile: to be sterilized by moist heat<br>Sterile: X-ray beam sterilization | Not provided in 510(k) Summary | | Non-sterile: to be sterilized by moist heat | Non-sterile: to be sterilized by moist heat |
Substantial Equivalence – Technological Characterist
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.