Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG®, TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lens
K230965 · Acuity Polymers, Inc. · HQD · Jun 16, 2023 · Ophthalmic
Device Facts
Record ID
K230965
Device Name
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG®, TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lens
Applicant
Acuity Polymers, Inc.
Product Code
HQD · Ophthalmic
Decision Date
Jun 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Indications for Use
Acuity™ 100 (hexafocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lens may be disinfected using a chemical disinfection system only. TYRO™ 97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyonic and which may exhibit corneal astigmatism. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lenses may be disinfected only by using chemical disinfection.
Device Story
Lathe-cut rigid gas permeable (RGP) contact lenses; manufactured from hexafocon A or hofocon A (fluorosilicone acrylate copolymers). Lenses feature covalently bonded Tangible™ Hydra-PEG® surface coating (polyethylene glycol-based polymer) to improve wettability. Prescribed by eye care professionals for daily wear. Patients self-apply lenses. Coating reduces sessile drop contact angle to <10°, enhancing surface properties while maintaining mechanical integrity of base material. Benefits include improved comfort and visual acuity for patients with refractive errors or irregular corneas. Disinfection requires chemical systems only.
Clinical Evidence
No new clinical data provided. Safety and effectiveness are supported by clinical data from predicate RGP lenses and previously established clinical performance of the Tangible™ Hydra-PEG® coating. Bench testing included 30-day disinfection stability, accelerated aging, bioburden testing, and ISO 10993-compliant biocompatibility (cytotoxicity, ocular irritation, systemic toxicity).
Technological Characteristics
Materials: hexafocon A or hofocon A (fluorosilicone acrylate copolymers). Surface: covalently bonded Tangible™ Hydra-PEG® (polyethylene glycol-based polymer). Form factor: lathe-cut RGP contact lens. Wettability: <10° sessile drop contact angle. Disinfection: chemical only. Connectivity: none.
Indications for Use
Indicated for daily wear correction of refractive error (myopia, hyperopia, presbyopia, astigmatism) in aphakic/non-aphakic persons with non-diseased eyes. Also indicated for management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-penetrating keratoplasty, or post-refractive surgery).
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lens (K162005, K180988)
TYRO™-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lenses (K052507, K102154)
Boston XO® (hexafocon A) Rigid Gas Permeable Contact Lenses with Tangible™ Hydra-PEG (K183167)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
June 16, 2023
Acuity Polymers, Inc. Jose Chavez Director, Regulatory Affairs 1667 Lake Ave Building 59, Suite 303 Rochester, NY 14615
Re: K230965
Trade/Device Name: Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lens , TYROTM-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HOD Dated: March 31, 2023 Received: April 5, 2023
Dear Jose Chavez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
J Angelo Green -S
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K230965
#### Device Name
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lens, TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lens
#### Indications for Use (Describe)
Acuity™ 100 (hexafocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lens may be disinfected using a chemical disinfection system only.
TYRO™ 97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyonic and which may exhibit corneal astigmatism. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lenses may be disinfected only by using chemical disinfection.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b></b></span> | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG®
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K230965
#### I. SUBMITTER
| Date Prepared: | June 8th, 2023 |
|-----------------|------------------------------------------------------------------------|
| Name: | Acuity Polymers, Inc. |
| Address: | 1667 Lake Ave.<br>Building 59, Suite 303<br>Rochester, NY 14615<br>USA |
| Contact Person: | Jose Chavez<br>Director, Regulatory Affairs |
| Phone number: | (585) 458-8409 |
| Email | Jose.Chavez@Acuitypolymers.com |
#### II. DEVICE
| Trade Name: | Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG®<br>Rigid Gas Permeable Contact Lens<br>TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid<br>Gas Permeable Contact Lens |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Daily wear rigid gas permeable contact lens |
| Classification Name: | Rigid gas permeable contact lens. (21 CFR§886.5916) |
| Regulatory Class: | Class II |
| Product Code: | HQD |
{5}------------------------------------------------
#### Purpose of the 510(k) Submission:
Acuity 100™ (hexafocon A) and TYRO™-97 (hofocon A) Rigid Gas Permeable Contact Lenses, cleared under 510(k) K162005, K180988, K052507 and K102154 are modified to include Tangible™ Hydra-PEG® surface coating, which is a thin, polyethylene glycol (PEG) based polymer designed to improve the wettability of the contact lenses. Specifically, Tangible™ Hydra-PEG® treated contact lenses demonstrate a measurable improvement in the contact angle compared to untreated lenses.
## III,PREDICATE DEVICE
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® is substantially equivalent to predicate devices: Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lens 510(k) numbers K162005 and K180988.
TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® is substantially equivalent to predicate devices: TYRO™-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lenses for Daily Wear, 510(k) numbers K052507 and K102154
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® are substantially equivalent to predicate device:
Boston XO® (hexafocon A) Rigid Gas Permeable Contact Lenses with Tangible™ Hydra-PEG Rigid Gas Permeable Contact Lenses, 510(k) number K183167.
# IV. DEVICE DESCRIPTION
Acuity 100™ (hexafocon A) are available as lathe cut rigid gas permeable contact lenses for daily wear only. The lenses are manufactured from a currently marketed contact lens material, hexafocon A. comprised of a siloxanyl fluoromethacrylate copolymer (fluorosilicone acrylate monomers), tinted for visibility, and with or without UV light absorber. Non-proprietary names were assigned by the United States Adopted Names Council (USAN).
Acuity 100™ (hexafocon A) Rigid Gas Permeable Contact Lenses are treated to incorporate Tangible™ Hydra-PEG® surface coating, which is a thin, polyethylene glycol (PEG) based polymer that is covalently bonded to the surface of the contact lens and is designed to enhance the surface properties of the contact lens while retaining the mechanical properties of the underlying material. When treated with Tangible™ Hydra-PEC® , the underlying material is encapsulated in a thin layer of polymer that results in measurable improvement of wettability (sessile drop contact angle) compared to untreated lenses.
TYRO" -97 (hofocon A) Rigid Gas Permeable Contact Lenses are available as lathe cut rigid gas permeable contact lenses for daily wear only. The lenses are manufactured from currently marketed contact lens material. hofocon A. polymer of trifluoroethyl methacrylate and silicone
{6}------------------------------------------------
| Acuity Polymers<br>510(k) Premarket Notification | 510(k) Summary |
|--------------------------------------------------------------------------------------------------------|----------------|
| Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® | |
methacrylate (fluorosilicone acrylate monomers), tinted for visibility, with or without UV light absorber. Non-proprietary names were assigned by the United States Adopted Names Council (USAN).
TYRO""-97 (hofocon A) Rigid Gas Permeable Contact Lenses are treated to incorporate Tangible™ Hydra-PEG® surface coating, which is a thin, polyethylene glycol (PEG) based polymer that is covalently bonded to the surface of the contact lens and is designed to enhance the surface properties of the contact lens while retaining the mechanical properties of the underlying material. When treated with Tangible™ Hydra-PEG® .the underlying material is encapsulated in a thin laver of polymer that results in measurable improvement of wettability (sessile drop contact angle) compared to untreated lenses.
The surface properties of: Acuity 100™ (hexafocon A), and TYRO™-97 (hofocon A) materials uncoated and coated with Tangible™ Hydra-PEG® are depicted in the following table:
| | Acuity 100™ (hexafocon A) | TYRO™-97 (hofocon A) |
|-------------------------------|---------------------------|----------------------|
| Sessile Drop<br>Contact Angle | Coated: $<10°$ | Coated: $<10°$ |
| | Uncoated: 56° | Uncoated:23.3° |
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are available in the same design configurations and available parameters as the predicate devices, cleared under K162005 and K180988.
TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are available in the same design configurations and available parameters as the predicate devices, cleared under K052507 and K102154.
# V. INDICATION FOR USE
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in aphakic and non-aphakic persons with non-diseased eves. The lenses may be prescribed for daily wear in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lens may be disinfected using a chemical disinfection system only.
TYRO™-97 (hofocon A) with Tangible™ Hydra PEG® Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with nondiseased eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas
{7}------------------------------------------------
| Acuity Polymers | 510(k) Summary |
|-------------------------------|----------------|
| 510(k) Premarket Notification | |
Acuity 100<sup>TM</sup> (1-focon A) with Tranquil<sup>TM</sup> Hydro-PEG<sup>®</sup> and TYRO<sup>TM</sup> 97 (1-focon A) with Tranquil<sup>TM</sup> Hydro-PEG<sup>®</sup>
Acuity 100" (hexatocon A) with Tangible " Hydra-PEG" and TYRO" -97 (hotocon A) with Tangible " Hvdra-PEG
permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lenses may be disinfected only by using chemical disinfection.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
### ACUITY 100™ (hexafocon A) WITH TANGIBLE™ HYDRA-PEG®
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® surface coating technology is substantially equivalent to Acuity 100™ (hexafocon A) uncoated lenses (primary predicate devices cleared under K162005 and K180988 in terms of the following:
- 1. Proprietary contact lens material formulation and USAN
- 2. Intended Use- Daily wear contact lens.
- 3. Indication for Use
- 4. Lens Designs and available parameters
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® surface coating technology is substantially equivalent to Boston XO® (hexafocon A) Rigid Gas Permeable Contact Lenses with Tangible™ Hydra-PEG (secondary Predicate device cleared under K183167) in terms of the following:
- 1. USAN name
- 2. Intended use- Daily wear contact lens.
- 3. Indication for use- similar
- 4. Tangible™ HYDRA-PEG® surface coating
#### TYRO™-97 (hofocon A) WITH TANGIBLE™ HYDRA-PEG®
TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® surface coating technology is substantially equivalent to TYRO™-97 (hofocon A) uncoated lenses (primary predicate devices cleared under K052507 and K102154) in terms of the following:
- 1. Proprietary contact lens material formulation and USAN
- 2. Intended Use- Daily wear contact lens.
- 3. Indication for Use
- 4. Lens Designs and available parameters
TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® surface coating technology is substantially equivalent to Boston XO® (hexafocon A) Rigid Gas Permeable Contact Lenses with Tangible™ Hydra-PEG (Secondary Predicate device cleared under K183167) in terms of the following:
{8}------------------------------------------------
| Acuity Polymers<br>510(k) Premarket Notification | 510(k) Summary |
|---------------------------------------------------------------------------------------------------------|----------------|
| Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™ -97 (hofocon A) with Tangible™ Hydra-PEG® | |
- 1. Thermoset copolymers derived from fluorosilicone acrylate monomers.
- 2. Intended Use- Daily wear contact lens.
- 3. Indication for use- similar
- 4. Tangible™ Hydra-PEG® surface coating
The following Substantial Equivalence Matrix illustrates the production method, intended use, and material characteristics of Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses as well as the predicate devices.
| | Acuity 100™<br>(hexafocon A)<br>with Tangible™<br>Hydra-PEG®, | Acuity 100™<br>(hexafocon A) Rigid Gas<br>Permeable Contact Lens | TYRO™ -97<br>(hofocon A)<br>Rigid Gas<br>Permeable<br>Contact Lens | Boston XO®<br>(hexafocon A)<br>with Tangible™<br>Hydra-PEG | Indication for Use | Acuity 100™<br>(hexafocon A) with Tangible™<br>Hydra-PEG® Rigid Gas<br>Permeable Contact Lenses are<br>indicated for daily wear for the<br>correction of refractive error<br>(myopia, hyperopia, presbyopia<br>and/or astigmatism) in aphakic<br>and non-aphakic persons with<br>non-diseased eyes. The lenses<br>may be prescribed for daily wear<br>in otherwise non-diseased eyes<br>that require a rigid contact lens<br>for the management of irregular<br>corneal conditions such as<br>keratoconus, pellucid marginal<br>degeneration, or following<br>penetrating keratoplasty or<br>refractive (e.g., LASIK) surgery.<br><br>The lens may be disinfected<br>using a chemical disinfection<br>system only.<br><br>TYRO™-97 (hofocon A) with<br>Tangible™ Hydra-PEG® Rigid<br>Gas Permeable Contact Lens are<br>indicated for daily wear for the<br>correction of visual acuity in<br>non-aphakic persons with non-<br>diseased eyes that are myopic,<br>hyperopic or presbyopic and<br>which may exhibit corneal<br>astigmatism. Also, the lenses<br>may be prescribed in otherwise<br>non-diseased eyes that require a<br>gas permeable contact lens for<br>the management of irregular<br>corneal conditions such as<br>keratoconus, pellucid marginal<br>degeneration, or following<br>penetrating keratoplasty or<br>refractive (e.g., LASIK) surgery.<br><br>The lenses may be disinfected<br>only by using chemical<br>disinfection. | The Acuity 100™ (hexafocon<br>A) Rigid Gas Permeable<br>Contact Lenses are indicated<br>for daily wear for the<br>correction of refractive error<br>(myopia, hyperopia, presbyopia<br>and/or astigmatism) in aphakic<br>and non-aphakic persons with<br>non-diseased eyes. The lens<br>may be prescribed in spherical<br>and aspheric powers ranging<br>from -20.00 D to +20.00 D for<br>daily wear. The lenses may be<br>prescribed for daily wear in<br>otherwise non-diseased eyes<br>that require a rigid contact lens<br>for the management of irregular<br>corneal conditions such as<br>keratoconus, pellucid marginal<br>degeneration, or following<br>penetrating keratoplasty or<br>refractive (e.g., LASIK)<br>surgery.<br><br>The lens may be disinfected<br>using a chemical disinfection<br>system only.<br><br>The lenses may be stored in a<br>multipurpose solution<br>(BOSTON SIMPLUS or<br>Menicon Unique pH) solution<br>up to 30 days. | TYRO™-97 (hofocon<br>A) Rigid Gas<br>Permeable Spherical,<br>Aspheric, Toric and<br>Bifocal Contact Lens<br>is indicated for daily<br>wear for the<br>correction of visual<br>acuity in not-aphakic<br>persons with non-<br>diseased eyes that are<br>myopic, hyperopic or<br>presbyopic and which<br>may exhibit corneal<br>astigmatism.<br><br>The lenses may be<br>disinfected only by<br>using chemical<br>disinfection.<br><br>TYRO™-97 (hofocon<br>A) Rigid Gas<br>Permeable Contact<br>Lenses is indicated for<br>daily wear for the<br>correction of visual<br>acuity in non-aphakic<br>persons with non-<br>diseased eyes that are<br>myopic, hyperopic or<br>presbyopic and which<br>may exhibit corneal<br>astigmatism. Also, the<br>lenses may be<br>prescribed in<br>otherwise non-<br>diseased eyes that<br>require a gas<br>permeable contact<br>lens for the<br>management of<br>irregular corneal<br>conditions such as<br>keratoconus, pellucid<br>marginal<br>degeneration, or<br>following penetrating<br>keratoplasty or<br>refractive (e.g.,<br>LASIK) surgery.<br><br>The lenses may be<br>disinfected only by<br>using chemical<br>disinfection | The Boston XO®<br>(hexafocon A) with<br>Tangible™ Hydra-PEG<br>Contact Lens is<br>indicated for daily wear<br>for the correction of<br>refractive ametropia<br>(myopia, hyperopia,<br>astigmatism, and<br>presbyopia) in aphakic<br>and non aphakic persons<br>with non-diseased eyes.<br>Also, the lenses may be<br>prescribed in otherwise<br>non diseased eyes that<br>require a gas permeable<br>contact lens for the<br>management of irregular<br>corneal conditions such<br>as keratoconus, pellucid<br>marginal degeneration,<br>or following penetrating<br>keratoplasty or refractive<br>(e.g., LASIK) surgery.<br><br>The Boston XO®<br>(hexafocon A) with<br>Tangible™ Hydra-PEG<br>lens may be disinfected<br>using a chemical<br>disinfection system (not<br>heat) only. | | < 1% | < 1% | < 1% | < 1% | | | | | |
|-------------------------------------|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|--------------------------------------|------------------|-------------------------------------|------------------------------------------------------------------|----------------------|--|--|--|--|
| | TYRO™ -97<br>(hofocon A)<br>with Tangible™<br>Hydra-PEG® | | | | Power Range | Production<br>Method | hexafocon A<br>-20.00 D to +20.00 D<br>hofocon A<br>-20.00 D to +12.00 D | Lathe Cut | hexafocon A<br>-20.00 D to +20.00 D | Lathe Cut | hofocon A<br>-20.00 D to +12.00 D | Lathe Cut | hexafocon A<br>-20.00 D to +20.00 D | Lathe Cut | Water Content<br>(%) | | | | |
| | New Device | Predicate Device<br>(K162005, K180988) | Predicate Device<br>(K052507,<br>K102154) | Predicate Device<br>K183167 | USAN Name | hexafocon A<br>hofocon A | hexafocon A | hofocon A | hexafocon A | Wettability<br>(Sessile Drop<br>Advancing<br>Contact<br>Angle) | hexafocon A: <10°<br>hofocon A: <10° | hexafocon A: 56° | hofocon A: 23.3° | (uncoated)<br>hexafocon A: 63.4°<br>(coated)<br>hexafocon A: 10° | | | | | |
| Intended Use | Daily Wear | Daily Wear | Daily Wear | Daily Wear | Lens Type | RGP | RGP | RGP | RGP | | | | | | | | | | |
| Device Name | Acuity 100™<br>(hexafocon A) with<br>Tangible™ Hydra-PEG®<br>Rigid Gas Permeable<br>Contact Lens | Acuity 100™<br>(hexafocon A) Rigid Gas<br>Permeable Contact Lens | TYRO™-97<br>(hofocon A)<br>Rigid Gas<br>Permeable<br>Spherical,<br>Aspheric, Toric<br>and Bifocal<br>Contact Lenses<br>for Daily Wear | Boston XO®<br>(hexafocon A)<br>Rigid Gas<br>Permeable Contact<br>Lenses with<br>Tangible™ Hydra-<br>PEG Rigid Gas<br>Permeable Contact<br>Lenses | Available with<br>or without UV<br>Blocker | Yes | Yes | Yes | Yes | | | | | | | | | | |
| | TYRO™ -97<br>(Hofocon A) with<br>Tangible™ Hydra PEG®<br>Rigid Gas Permeable<br>Contact Lenses | | | | Lenses may be<br>disinfected<br>using a<br>Chemical<br>Disinfection<br>System | Yes | Yes | Yes | Yes | | | | | | | | | | |
| Includes<br>Tangible™<br>Hydra PEG® | Yes | No | No | Yes | | | | | | | | | | | | | | | |
{9}------------------------------------------------
# Acuity Polymers
510(k) Summary
510(k) Premarket Notification
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG®
{10}------------------------------------------------
# uity Polymers
510(k) Premarket Notification
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG®
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
A series of in vitro and in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses. All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols.
Test results of the non-clinical testing on the device demonstrate that:
Performance Testing- Bench
- 1. The physical properties of the lenses are stable following 30 (thirty) disinfection cycles in Boston Simplus disinfection solution at ambient temperatures.
- The physicochemical and mechanical properties of the contact lenses are unchanged after 2. the addition of Tangible™ Hydra-PEG®, with the exception of wettability (contact angle).
- 3. The surface properties of the lens are stable following 30 (thirty) days of accelerated aging.
- 4. Bioburden levels are below the acceptance criteria (<100 CFU/lens) initially and following 30 (thirty) days of storage in solution at ambient temperatures.
Performance Testing- Animal
{11}------------------------------------------------
Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG®
- 5. Acute Ocular Irritation testing per ISO 10993-23 and Acute Systemic Toxicity testing per ISO 10993-11 were performed for both the final finished subject devices.
#### Performance Testing- In Vitro
- 6. Cytotoxicity testing per ISO 10993-5 was performed for both the final finished subject devices.
#### Performance Testing -Clinical
The clinical safety and effectiveness of finished rigid gas permeable contact lenses manufactured from Acuity™ 100 (hexafocon A) and TYRO™-97 (hofocon A) Rigid Gas Permeable Contact Lenses materials have been demonstrated by predicate devices. Additionally, the clinical safety and effectiveness for contact lenses treated with Tangible™ Hydra-PEG® has been previously demonstrated.
#### VIII. CONCLUSIONS
#### Validity of Scientific Data
Laboratories under Good Laboratory Practice regulations conducted toxicology, microbiology, and shelf-life stability studies following scientific protocols. The data were determined to be scientifically valid under 21 CFR §860.7.
#### Substantial Equivalence
Information presented in this 510(k) Premarket Notification establishes that the Acuity 100™ (hexafocon A) with Tangible™ Hydra-PEG® and TYRO™-97 (hofocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact Lenses are as safe and effective as the predicate devices when used in accordance with the labeled directions for use and for the proposed indication.
#### Risks and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.