TYRO 97 (HOFOCON A) RIGID GAS PERMEABLE SPHERICAL, ASPHERIC, TORIC AND BIFOCAL CONTACT LENSES FOR DAILY WEAR
K052507 · The Lagado Corp. · HQD · Nov 2, 2005 · Ophthalmic
Device Facts
Record ID
K052507
Device Name
TYRO 97 (HOFOCON A) RIGID GAS PERMEABLE SPHERICAL, ASPHERIC, TORIC AND BIFOCAL CONTACT LENSES FOR DAILY WEAR
Applicant
The Lagado Corp.
Product Code
HQD · Ophthalmic
Decision Date
Nov 2, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Indications for Use
The TYROTM-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lens is indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism. The lenses may be disinfected only by using chemical disinfection
Device Story
TYRO-97 is a rigid gas permeable (RGP) contact lens made of hofocon A (polymer of trifluoroethyl methacrylate and silicone methacrylate). Available in spherical, aspheric, toric, and bifocal designs; clear or tinted (blue, green, gray) with optional UV-blocker. Used for daily wear to correct refractive errors in non-aphakic patients. Prescribed by eye care professionals. Lens provides oxygen permeability and refractive correction. Clinical benefit includes improved visual acuity for patients with myopia, hyperopia, presbyopia, and astigmatism. Disinfection requires chemical methods.
Clinical Evidence
Safety confirmed through clinical trials conducted per FDA 510(k) Guidance for Daily Wear Contact Lenses (May 1994). Results demonstrated safety for the intended patient population.
Technological Characteristics
Material: hofocon A (trifluoroethyl methacrylate and silicone methacrylate polymer). Water content <0.2%. Oxygen permeability: 97.98 ANSI units. Refractive index: 1.440. Hardness: D/82 Shore. Flexural strength: 3952 psi. Available in spherical, aspheric, toric, and bifocal geometries. Tinted varieties use D&C Green No. 6, CI Solvent Yellow 18, D&C Violet No. 2; UV-blocker is NORBLOC 7966.
Indications for Use
Indicated for daily wear for visual acuity correction in non-aphakic, non-diseased eyes with myopia, hyperopia, presbyopia, or corneal astigmatism.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Oxycon (wilofocon A) Rigid Gas Permeable Contact Lenses
Submission Summary (Full Text)
{1}------------------------------------------------
## 510(k) SUMMARY FOR FREEDOM OF INFORMATION
K052507
NOV - 2 2005
. .
#### . · TYRO™-97 (hofocon A) RIGID GAS PERMEABLE . SPHERICAL, ASPHERIC, TORÍC AND BIFOCAL CONTACT LENS FOR DAILY WEAR ·
| 1. Submitted by: | The Lagado Corporation<br>2890 South Tejon St<br>Englewood, CO 80110 |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | John M. Szabocsik, Ph.D.<br>Official agent<br>Szabocsik and Associates<br>203 N. Wabash, Ste 1200<br>Chicago, IL 60601<br>(312) 553-0828 |
| 2. Date prepared: | September 8, 2005 |
| 3. Device:<br>Common Name | TYRO <sup>TM</sup> -97 (hofocon A) Rigid Gas Permeable Spherical,<br>Aspheric, Toric and Bifocal Contact Lenses for Daily Wear |
| Trade Name | TYRO <sup>TM</sup> -97 (hofocon A) Rigid Gas Permeable Spherical,<br>Aspheric, Toric and Bifocal Contact Lenses for Daily Wear |
| 4. Classification | Class II (Performance Standards)<br>21 CFR 886.5916 (b) (1)<br>Rigid Gas Permeable Contact Lens for Daily Wear |
| 5. Substantial<br>Equivalence | This product is substantially equivalent to other currently marketed<br>rigid lenses, such as Oxycon (wilofocon A) Rigid Gas Permeable<br>Contact Lenses |
| 6. Device<br>Description | The TYRO <sup>TM</sup> -97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric,<br>Toric and Bifocal Contact Lenses for Daily Wear are available in<br>spherical, aspherical and toric designs in the clear untinted or blue, green<br>and gray tinted varieties with and without UV-blocker. Each is a shell of<br>the following dimensions. |
### Spherical Contact Lens:
| Horizontal Lens Size: | 6.5 mm to 11.50mm |
|-----------------------|--------------------|
| Base Curve: | 6.50mm to 9.50mm |
| Distance Powers: | +12.00D to -20.00D |
| Center Thickness: | |
| for low minus | 0.05mm to 0.30mm |
| for plus | 0.10mm to 0.70mm |
{2}------------------------------------------------
| Aspheric Lens | |
|-----------------------------------|-----------------------------------|
| Eccentricity | -1.5 to 1.5<br>Oblate to prolate |
| Peripheral Curves | 0.1 to 1.0mm |
| Toric Lens | |
| Axis | 1 to 180 degrees in 1 degree step |
| Cylinder power | 0.50 to 4.00D |
| Translating Bifocal Contact Lens: | |
| Horizontal Lens Size: | 8.00mm to 10.50mm |
| Base Curve: | 6.50mm to 8.50mm |
The lens material, hofocon A, is a polymer of trifluoroethyl methacrylate and silicone methacrylate, with no methyl methacrylate. The blue tinted lenses contain D&C Green No. 6; the green lenses contain D&C Green No 6 and CI Solvent Yellow 18; the gray lenses contain D&C Green No 6, D&C Violet No. 2, and CI Solvent Yellow 18; the blue-UV lenses contain D&C Green No 6 and a UV absorber, NORBLOC 7966. The colorants are used in quantities approved for use in contact lenses and proportions required to obtain the desired color.
+12.00D
1.00D
-20.00D
4.00D
to
to
7. Intended use:
The TYROTM-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lens is indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyopic and which may exhibit corneal astigmatism.
The lenses may be disinfected only by using chemical disinfection
Distance Power:
Add Power
{3}------------------------------------------------
: 上海
## 8. Comparison to predicate devices: see following table
# SUBSTANTIAL EQUIVALENCE
| | OXYCON (wilofocon<br>A) | TYROTM-97 (hofocon<br>A) |
|-----------------------------------------------|------------------------------|----------------------------------------|
| Water Content | <1% | <0.2% |
| Oxygen Permeability | 26x10-11<br>Fatt Units(1990) | 97.98<br>ANSI units |
| Refractive Index | 1.44 | 1.440 |
| Hardness | D/89 (Shore) | D/82 (Shore) |
| Specific Gravity | 1.25 | 1.087 |
| Residual Monomers | | no leachable monomers<br>detected |
| Wetting Angle | 23.0±2<br>(CLMA method) | 23.3°<br>(sessile drop technique) |
| Mechanical (flexural)<br>Strength | | 3952psi |
| Light Transmittance<br>clear<br>blue<br>green | 93%<br>87% | >95% T<br>>70% T<br>>70% T |
| Blue-UV | | >70% T (400-780nm)<br>0% T (200-380nm) |
#### Introduction
:
The TYRO™-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic, or presbyopic and which may exhibit corneal astigmatism.
Contained in the submission are comparisons of the product to the predicate device, information on the chemistry and manufacturing, results of toxicological tests and the report of clinical trials.
Chemistry and Manufacturing I.
> The lens material, hofocon A, is a polymer of trifluoroethyl methacrylate and silicone methacrylate, with no methyl methacrylate. The blue tinted lenses contain D&C Green No. 6; the green lenses contain D&C Green No 6 and CI Solvent Yellow 18; the gray lenses contain D&C Green No 6, D&C Violet No. 2,
7
{4}------------------------------------------------
and CI Solvent Yellow 18; the blue-UV lenses contain D&C Green No 6 and a UV absorber, NORBLOC 7966. The colorants are used in quantities approved for use in contact lenses and proportions required to obtain the desired color.
Material and lenses are manufactured in the same manner as the predicate device, and details are contained in K990895.
- II. Toxicology
The toxicological testing is summarized below.
- Agar Overlay Cytotoxicity: A.
- B. Systemic toxicity:
- C. Acute Ocular irritation:
The lens material was shown to be non-toxic in all tests.
- III. Microbiology
The material has a water content less than 1%, and is therefore exempt from the microbiological requirements for hydrophobic contact lenses.
#### IV. Clinical Studies
The safety of the TYRO™-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lenses has been confirmed through a clivical, trial for daily wear contact lens materials according to the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses (May 1994).
## OVERALL CONCLUSION OF THE CLINICAL STUDY:
The data of the clinical trials demonstrate the safety of the TYRO™-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lenses.
{5}------------------------------------------------
Public Health Service
Image /page/4/Picture/2 description: The image is a black and white emblem for the Department of Health & Human Services - USA. The emblem features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized symbol consisting of three overlapping profiles, which are meant to represent the health and well-being of individuals.
## NOV - 2 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
The Lagado Corporation c/o John M. Szabocsik, PhD Szabocsik and Associates 203 North Wabash Avenue Suite 1200 Chicago, IL 60601
Re: K052507
Trade/Device Name: TYRO-97 (hofocon A) Rigid Gas Permeable Contact Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HQD Dated: September 8, 2005 Received: September 13, 2005
Dear Dr. Szabocsik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitts varied to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
David M. Whipple
Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
September 8, 2005 :
## 510(k) NUMBER (IF KNOWN) KO 5 2 5 0 7
DEVICE NAME
TYRO™-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lens for Daily Wear
#### INDICATIONS FOR USE
The TYRO™-97 (hofocon A) Rigid Gas Permeable Spherical, Aspheric, Toric and Bifocal Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic or presbyoptic and which may exhibit corneal astigmatism.
The lenses may be disinfected only by using chemical disinfection
Prescription Use X (Part 21CFR 801 Subpart D)
OR Over-The-Counter-Use (Part 21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)_
Concurrence of CDRH, Office of Device Evaluation (ODE)
Myrabnith
10-1-44
Page 1 of 1
516 Number K052507
K.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.