L-Cath™ Single and Dual Lumen Catheters, L-Cath™ Midline Catheters
K230669 · Argon Medical Devices · LJS · Nov 30, 2023 · General Hospital
Device Facts
Record ID
K230669
Device Name
L-Cath™ Single and Dual Lumen Catheters, L-Cath™ Midline Catheters
Applicant
Argon Medical Devices
Product Code
LJS · General Hospital
Decision Date
Nov 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The L-Cath Single and Dual Lumen Catheters and the L-Cath Midline Catheters are indicated for short or long term peripheral access to the central venous system for the administrations, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
Device Story
L-Cath catheters are sterile, single-use polyurethane devices designed for peripheral access to the central venous system. They function as an enclosed fluid pathway for administering fluids, medications, and nutrients, sampling blood, and monitoring physiological parameters. Used in clinical settings by trained healthcare professionals, the catheters are inserted into the venous system (PICC or midline placement) following institutional protocols. The device includes accessories like a trim tool, tape measure, and stylet; some versions are supplied in a kit with an introducer. The output is a direct venous access point, allowing clinicians to deliver therapies or monitor patient status. Benefits include reliable, long-term vascular access for patients requiring intravenous support. The subject device introduces a material change to the Luer Lock assembly (PEBAX 7433) compared to the predicate.
Clinical Evidence
No clinical or animal testing was required. Substantial equivalence was demonstrated through bench testing, including dimensional analysis, pressure burst, radiopacity, pull force, and luer integrity (ISO 80369-7). Biocompatibility testing included systemic toxicity (ISO 10993-17) and genotoxicity (ISO 10993-3).
Indicated for short or long-term peripheral access to the central venous system for fluid, medication, and nutrient administration, blood sampling, and intravenous blood pressure and temperature monitoring. Contraindicated in patients with device-related infections, thrombosis in the insertion vessel or pathway, insufficient body size, or known allergies to device materials.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.
November 30, 2023
Argon Medical Devices Scott Bishop Director, Regulatory Affairs 1445 Flat Creek Road Athens. Texas 75751
Re: K230669
Trade/Device Name: L-Cath™ Single and Dual Lumen Catheters, L-Cath™ Midline Catheters Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS, FOZ Dated: October 31, 2023 Received: October 31, 2023
Dear Scott Bishop:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming
product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230669
Device Name
L-Cath™ Single and Dual Lumen Catheters, L-Cath™ Midline Catheters
Indications for Use (Describe)
The L-Cath Single and Dual Lumen Catheters and the L-Cath Midline Catheters are indicated for short or long term peripheral access to the central venous system for the administrations, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## K230669 510(k) Summary
November 30, 2023 Date Prepared:
| Company: | Argon Medical Devices, Inc.<br>1445 Flat Creek Road<br>Athens, Texas 75751 USA<br>Facility Registration number: 1625425 |
|----------|-------------------------------------------------------------------------------------------------------------------------|
| Contact: | Scott Bishop<br>Director, Regulatory Affairs<br>Phone: 469.430.0546<br>Email: Scott.Bishop@argonmedical.com |
| Device Trade Name: | L-Cath™ Single and Dual Lumen Catheters, L-Cath™ Midline<br>Catheters |
|-------------------------|-------------------------------------------------------------------------|
| Device Common Name: | Catheter, Intravascular, Therapeutic, Long-Term Greater Than<br>30 Days |
| Regulation Name: | Percutaneous, implanted, long-term intravascular catheter |
| Primary Product Code: | LJS |
| Secondary Product Code: | FOZ |
| Regulation: | 21 CFR 880.5970 |
| Classification: | Class II |
| Review Panel: | General Hospital |
| Predicate Device(s): K091670 | BD L-Cath™ Midline Catheter<br>BD Cath™ Single Lumen<br>BD Cath™ Dual Lumen<br>Peripherally Inserted Central Catheters PICC |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description of the Device: | The L-Cath™ Single and Dual Lumen and L-Cath™ Midline Catheters are sterile, single use devices, sold as standalone devices which includes the following accessories:<br>• Trim Tool<br>• Tape measure<br>• Stylet (include in the final catheter assembly as applicable)<br>• Polyurethane catheter |
{6}------------------------------------------------
These catheters are also sold as a component in a Basic Kit (catheter and introducer packaged together) which includes the following:
- BD Introsyte Autoguard Introducer or Splittable Introducer ●
- Stylet (include in the final catheter assembly as applicable)
- Trim Tool
- Tourniquet
- Tape measure
#### Indication for The L-Cath Single and Dual Lumen Catheters and the L-Cath Midline Catheters are Use: indicated for short or long term peripheral access to the central venous system for the administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
### Substantial Equivalence:
There is no change of intended use or fundamental scientific technology between the proposed modified and predicate device. The proposed modified device has the same indication for use as the predicate, K091670.
| | Predicate Device<br>L-Cath™ Single and Dual Lumen and L-<br>Cath™ Midline Catheter | Modified Proposed Device<br>(Subject of This Submission) &<br>Discussion |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Argon Medical Devices, Inc | SAME |
| FDA Clearance | K091670 | K230669 |
| Class | II | SAME |
| Device Classification Name | Percutaneous, implanted, long-term<br>intravascular catheter. | SAME |
| Regulation | 21 CFR 880.5970 | SAME |
| Product Code | LJS/FOZ | SAME |
| Technological Characteristics:<br>Technological characteristics of the Subject device are substantially equivalent to those of the cited predicate<br>device with respect to intended use, indications for use, target patient population, operating principle,<br>fundamental scientific technology, packaging configurations, sterility assurance level and method of sterilization.<br>The differences of the Subject device from the Predicate device are limited to the materials used in the Luer Lock<br>assembly. Additionally, contraindications have been added based on review of similar devices.<br>The following table provides a summary comparison between the subject and predicate device. | | |
| Intended Use | The L-Cath Single and Dual Lumen<br>Catheters and the L-Cath Midline Catheters<br>are indicated for short or long term<br>peripheral access to the central venous<br>system for the administration of fluids,<br>medications, and nutrients; the sampling of<br>blood; and monitoring blood pressure and<br>temperature intravenously. | SAME |
| Principle of Operation | The L-Cath Single and Dual Lumen<br>Catheters and the L-Cath Midline Catheters<br>are a family of surgically Invasive, long-<br>term, polyurethane catheters that are in<br>direct contact with the central circulatory<br>system.<br>• For proper use, clinicians must be familiar<br>with and trained in the placement,<br>maintenance, and use of PICC and/or<br>midline catheters. Their use should be<br>preceded by an established institutional<br>protocol and accepted professional<br>standards and guidelines and performed by<br>persons trained in the procedure and<br>knowledgeable of the inherent risks.<br>• PICC catheters shall have the distal tip<br>dwelling in the lower one third of the<br>superior vena cava to the junction of the<br>superior vena cava and the right atrium.<br>Lower extremity placement shall have the<br>distal tip dwelling in the thoracic inferior<br>vena cava above the level of the<br>diaphragm.<br>• Midline catheters shall have the distal tip<br>residing in the basilic, cephalic, or brachial<br>vein at or below the axillary level and distal<br>to the shoulder. Infusates should be at or<br>near normal serum osmolality.<br>• Follow sterile technique per institutional<br>policy. Sterile technique and proper skin<br>preparation are essential for proper use.<br>The use of maximal barrier precautions<br>during catheter insertion is recommended<br>to help reduce the risk of catheter-related<br>infection, including the use of sterile gown,<br>mask, cap, sterile gloves, and large sterile<br>drapes to cover the body. Observe<br>Standard/Universal precautions for all<br>patients to reduce the potential exposure<br>to bloodborne pathogens. | SAME |
| Mechanism of Action | The polyurethane PICC and Midline<br>Catheter Systems transport medicines and<br>nutritional therapies into the bloodstream<br>of the patient by providing an enclosed<br>fluid pathway that connects other<br>catheters and then creates a seal at the<br>connections and junctions with the other<br>enclosed fluid pathways (catheters). | SAME |
| Indication for Use | The L-Cath Single and Dual Lumen<br>Catheters and the L-Cath Midline Catheters<br>intended use shall be for the administration<br>of fluids, medications, and nutrients; the<br>sampling of blood; and monitoring blood<br>pressure and temperature intravenously. | SAME |
| Contraindication | There are no Contraindications | The devices are contraindicated whenever there is presence of device related infections or presence of thrombosis in the intended insertion vessel or catheter pathway. The devices are contraindicated when the patient's body size is insufficient to accommodate the size of the device. The patient is known or is suggested to be allergic to materials contained in the device. Clinical assessment of the patient must be completed to ensure no contraindications exist. |
| Single Use | Yes | SAME |
| Supplied Sterile | Yes, EO | SAME |
| Device Description | All Argon L-Cath PICC and Midline polyurethane catheters are sold as standalone devices which includes the following accessories: Trim Tool Tape measure Stylet (include in the final catheter assembly as applicable) Polyurethane catheter These catheters are also sold as a component in a Basic Kit (catheter and introducer packaged together) which includes the following: BD Introsyte Autoguard Introducer Stylet (include in the final catheter assembly as applicable) Trim Tool Tourniquet Tape measure | SAME |
| PICC & Midline Catheter<br>Outer Diameter | Single Lumen: 16, 18, or 20-gauge<br>Dual Lumen: 16, 18, and 20-gauge<br>Small vein access (SVA) Single Lumen: 26 and 28-gauge<br>Midline: 18 and 20-gauge | SAME |
| PICC & Midline Catheter<br>Length | Single Lumen: 60 cm<br>Dual Lumen: 60 cm<br>SVA Single Lumen: 25 and 30 cm<br>Midline: 20 cm | SAME |
| Luer Lock Material | Gamma Stable Polycarbonate | PEBAX 7433 SA01 MED<br>NATURAL/CLEAR<br><br>• Biocompatibility and<br>performance testing<br>demonstrated that the difference<br>between the subject and<br>predicate devices does not raise<br>new or different questions to<br>safety and effectiveness |
| Catheter Material | Polyurethane | SAME |
| Strain Relief Material | Silicon | SAME |
| Tubing Material | Polyurethane | SAME |
| Stylet Material | 304V Stainless Steel<br>ABS | SAME |
| Y Site Assembly Material | Makrolon…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.