BD L-CATH SINGLE AND DUAL LUMENT CATHETERS, MIDLINE CATHETER
K091670 · Becton, Dickinson & CO · LJS · Jul 1, 2009 · General Hospital
Device Facts
Record ID
K091670
Device Name
BD L-CATH SINGLE AND DUAL LUMENT CATHETERS, MIDLINE CATHETER
Applicant
Becton, Dickinson & CO
Product Code
LJS · General Hospital
Decision Date
Jul 1, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BD L-Cath™ single and dual lumen peripherally inserted central catheters intended use shall be for the administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously. The BD L-Cath™ Midline™ Catheter intended use shall be for the administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
Device Story
BD L-Cath™ series (Single Lumen, Dual Lumen, Midline) are intravascular catheters used for fluid/medication/nutrient administration, blood sampling, and monitoring blood pressure/temperature. Devices feature plastic bushing component connecting tubing to female luer lock adapter. Modification involves higher molecular weight polycarbonate (PC Makrolon RX2530-1118) for bushing. Intended for clinical use; prescription only. No changes to geometry, performance, or fundamental technology compared to predicates.
Clinical Evidence
No clinical data. Substantial equivalence supported by nonclinical testing of the modified bushing material and comparison of technological characteristics.
Technological Characteristics
Intravascular catheters with polycarbonate (PC Makrolon RX2530-1118) bushing component. Linear polycarbonate formulation with higher molecular weight. Same geometry and performance as predicate. No electronic components or software.
Indications for Use
Indicated for patients requiring intravenous administration of fluids, medications, and nutrients; blood sampling; and monitoring of blood pressure and temperature.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{1}------------------------------------------------
BD Seament Name
K091670
1 of 2
JUL - 1 2009
Image /page/0/Picture/3 description: The image shows the BD logo. The logo consists of a circular emblem on the left and the letters "BD" on the right. The emblem features a stylized sun with rays emanating from the top half and a stylized figure below. The letters "BD" are in a bold, sans-serif font.
Helping all people live healthy lives
# SPECIAL PREMARKET NOTIFICATION [510(K)] Summary- (Bundled Submission)
Submitter:
Becton Dickinson Infusion Therapy Systems, Inc.
Specialist, Regulatory Affairs
9450 South State Street
Sandy, UT 84070
Justice Alder
801-565-2662
801-565-2749
June 2, 2009
Address:
Contact Person:
Telephone Number: Fax Number:
Date Summary Prepared:
Trade Names:
BD L-Cath™ Single Lumen BD L-Cath™ Dual Lumen BD L-Cath™ Midline
Intravascular Catheter (IV Catheter)
Common Name:
Classification Name:
Catheter, Intravascular, Therapeutic, Long- Term Greater Than 30 Days (LJS/FOZ)
Peripherally Inserted Central Catheter (PICC);
Unmodified/Predicate Device: (K925979) Dual lumen L-Cath® Catheter System (K920755) L-Cath Peel Away Systems Catheter
### Description of the device:
Becton Dickinson obtained the L-Cath™ catheter product lines through the purchase of Luther Medical Inc, a California based company, in January 1999. Luther Medical obtained market clearance from the US Food and Drug Administration (FDA) during 1993 for the L-Cath™ peripherally inserted central catheters through the premarket notification [510(k)] process. Becton Dickinson currently markets the L-Cath single lumen, dual lumen, and midline catheters. These catheters have a plastic bushing material that is used for the purpose of catheter assembly and/or extension tubing with a female luer lock adapter.
The primary modification addressed in this 510(k) submission relates to a material modification in the bushing component. The modifications to the devices pertain to a
:_1 ℃ 1 × /1 × .. T .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{2}------------------------------------------------
modified bushing material that is known as PC Makrolon RX2530-1118, which has been tested by Becton Dickinson's supplier and meets FDA & ISO standards. The bushing material is used in all three types of catheters to join the tubing to the female luer lock adapter. The material modification affects only the bushing that is seated within the luer adapters of the BD L-Cath™ PICC and BD L-Cath™ Midline products. The formulation modification is a linear polycarbonate with the same end groups, formulation, additives, and colorants. The only difference being a higher molecular weight for the modified material.
Intended Use of the L-Cath™ single lumen, and BD L-Cath™ dual lumen catheter devices: [21 CFR Part 880---- GENERAL HOSPITAL AND PERSONAL USE DEVICES: General Hospital and Personal Use Therapeutic Devices Sec. 880.59701
The BD L-Cath™ single and dual lumen peripherally inserted central catheters intended use shall be for the administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
# Intended Use of the BD L-Cath™ Midline catheter devices:
The BD L-Cath™ Midline™ Catheter intended use shall be for the administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
#### Technological Characteristics Comparison:
The L-Cath products have the same technological characteristics with no changes in part geometry or performance, and have the same formulation, additives, and colorants as the current device material. The only difference being a higher molecular weight in the formulation change which is not evident to the end user.
### Nonclinical Tests Support Substantial Equivalence:
The following qualities of the three devices are the same and support substantial equivalence: Indications for use, fundamental scientific technology, gauge sizes of the devices, visualization of blood return and potential for blood exposure during use.
#### Conclusions from Nonclinical Tests:
The intended use and technological characteristics of the BD L-Cath™ products described in this submission have not been altered and therefore have remained the same since receiving their original 510(k) approval from the FDA.
The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. A determination of
1 /1 /1 /1 /1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Becton Dickinson and Company C/O Ms. Justice Alder Regulatory Affairs Specialist Becton Dickinson Infusion Therapy Systems, Incorporated 9450 South State Street Sandy, Utah 84070
Re: K091670
Trade/Device Name: BD L-Cath™ Midline™ Catheter BD L-Cath™ Single Lumen BD L-Cath™ Dual Lumen
Peripherally Inserted Central Catheters (PICC)
Regulation Number: 21 CFR 880.5970
Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II
Product Code: LJS, FOZ
Dated: June 2, 2009
Received: June 9, 2009
Dear Ms. Alder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Alder
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Anthony D. Wathenfer
Susan Runner, D.D.S., M. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
KO91670 1 of 1
# Indications for Use
#### 510(k) Number (if known): K091670
Device name: BD L-Cath™ Midline™ Catheter
Indications for Use: The intended use shall be for the administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Charles R. Hoppe
Page 1 of
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number. K091670
{6}------------------------------------------------
K091670
, of /
# Indications for Use
510(k) Number (if known): K091670 K091670
Device name: BD L-Cath™ Single Lumen/ BD L-Cath™ Dual Lumen Peripherally Inserted Central Catheters (PICC)
Indications for Use: The intended use shall be for the administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Charles R
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of
510(k) Number: K091670
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.