K230316 · International Life Sciences · OWW · May 26, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230316
Device Name
FlexBand; FlexPatch; FlexBand Plus
Applicant
International Life Sciences
Product Code
OWW · General, Plastic Surgery
Decision Date
May 26, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FlexBand®, FlexPatch®, and FlexBand® Plus are intended for use in surgical procedures for reinforcement of soft tissue where weakness exists. FlexBand®, FlexPatch®, and FlexBand® Plus are also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. FlexBand®, FlexPatch®, and FlexBand® Plus is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to the bone, provide mechanical strength for the tendon repair. The products reinforce soft tissue and provides a degradable scaffold that is incorporated into the patient's own tissue.
Device Story
Knitted surgical mesh made from ARTELON® (degradable polycaprolactone-based polyurethane urea) fibers; used for soft tissue reinforcement in orthopedic surgeries; provides degradable scaffold for tissue incorporation; does not replace native tissue mechanical strength. FlexBand® Plus includes pre-loaded sutures for surgical usability. Supplied sterile; cut to size by surgeon; implanted during tendon/ligament repair procedures. Acts as adjunct to primary suture/fixation repair; shares load on primary repair; supports healing via scaffold integration. Used in clinical operating room settings by surgeons.
Clinical Evidence
Clinical literature review provided regarding ligament reinforcement procedures. Literature indicates device effectiveness for proposed ligament indications. Bench testing included suture retention and tensile strength testing. Routine endotoxin (LAL) testing performed per lot.
Technological Characteristics
Knitted mesh; material: ARTELON® (degradable polycaprolactone-based polyurethane urea); form factor: mesh strip/patch; sterilization: E-Beam (min 25 kGy); non-electronic; no software.
Indications for Use
Indicated for reinforcement of soft tissue where weakness exists and reinforcement of soft tissues repaired by suture/fixation during tendon and ligament repair surgery (e.g., rotator cuff, patellar, Achilles, biceps, quadriceps, collateral, spring, deltoid, ulnar ligaments). Not intended to replace normal body structure or provide full mechanical strength for tendon/ligament repair.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
May 26, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
International Life Sciences Tiffini Wittwer Vice President Quality and Regulatory Affairs 2150 Northwest Parkway SE Suite G Marietta, Georgia 30076
## Re: K230316
Trade/Device Name: FlexBand®; FlexPatch®; FlexBand® Plus Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: QWJ, FTL, OWW Dated: April 26, 2023 Received: April 27, 2023
## Dear Tiffini Wittwer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Yu-chieh Chiu -S
Yu-Chieh Chiu, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230316
### Device Name
FlexBand®, FlexPatch®, and FlexBand® Plus
## Indications for Use (Describe)
FlexBand®, FlexPatch®, and FlexBand® Plus are intended for use in surgical procedures for reinforcement of soft tissue where weakness exists.
FlexBand®, FlexPatch®, and FlexBand® Plus are also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments.
FlexBand®, FlexPatch®, and FlexBand® Plus is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to the bone, provide mechanical strength for the tendon repair. The products reinforce soft tissue and provides a degradable scaffold that is incorporated into the patient's own tissue.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary: K230316
# FlexBand®/ FlexPatch® / FlexBand® Plus
#### I. Submitter Information
| Submitter: | International Life Sciences (dba Artelon) | |
|----------------|--------------------------------------------------------|--|
| Address: | 2150 Northwest Pkwy SE<br>Suite G<br>Marietta GA 30076 | |
| Telephone: | +1 (800) 610-3446 | |
| Contact: | Tiffini Wittwer, MPH | |
| Date Prepared: | May 23, 2023 | |
#### Device Information II.
| Trade Name(s): | FlexBand® / FlexPatch® / FlexBand® Plus |
|-----------------------|-------------------------------------------------------------------|
| Common Name: | Surgical Mesh |
| Classification: | Class II |
| Regulation: | 21 CFR 878.3300 |
| Classification Name: | Mesh, Surgical, Absorbable, Orthopedics, Reinforcement of Ligamen |
| Classification Panel: | Orthopedic |
QWJ, OWW, FTL Product Codes:
#### Predicate Device Information III.
The FlexBand® / FlexPatch® / FlexBand® Plus devices described in this submission are substantially equivalent to the following predicate devices:
The FlexBand® / FlexPatch® / FlexBand® Plus (K192112)
BioBridge™ Collagen Matrix (K151083)
#### IV. Device Description
The FlexBand® / FlexPatch® / FlexBand® Plus products are knitted mesh made from ARTELON® fibers. ARTELON® fiber is made of degradable polycaprolactone-based polyurethane urea. The construction permits the mesh to be cut into any desired shape or size without unraveling. FlexBand® Plus devices have suture attached to each end of the knitted mesh strip. The pre-loaded suture is intended to aid in usability in the operating room. The devices are supplied sterile, one product per package in double layer peel pouch packaging.
#### V. Indication for Use
{4}------------------------------------------------
FlexBand®, FlexPatch®, and FlexBand Plus® are intended for use in surgical procedures for reinforcement of soft tissue where weakness exists.
FlexBand®, FlexPatch®, and FlexBand Plus® are also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments.
FlexBand®, FlexPatch®, and FlexBand Plus® is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendons, medial collateral ligament, lateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the tendon repair. The products reinforce soft tissue and provides a degradable scaffold that is incorporated into the patient's own tissue.
#### Comparison of Technological Characteristics With the Predicate Devices VI.
FlexBand®, FlexPatch®, and FlexBand Plus® devices have the same principles of operation as the predicate devices (K192112 and K151083). They are all implants intended for the reinforcement of soft tissues that are repaired by suture or other fixation devices. The devices are used in various surgical procedures where soft tissue reinforcement is needed.
FlexBand®, FlexPatch®, and FlexBand Plus® products are similar to the BioBridge predicate (K151083) in the following ways:
- Intended use, to share load placed on the primary repair and provide structural scaffold . for torn or damaged soft tissue;
- Indication for Use
- Radiation sterilization (E-Beam method, minimum 25 kGy) ●
- Biodegradable
- May be cut to size to meet surgeons' preference ●
There have been no changes to the FlexBand®, FlexPatch®, and FlexBand Plus® products since the prior 510(k) clearance. Therefore, FlexBand®, FlexPatch®, and FlexBand Plus® products are the same as the FlexBand®, FlexPatch®, and FlexBand Plus® predicate (K192112) in the following ways:
- . Material
- Knit patterns and design. .
- Product sizes and packaging configuration
#### VII. Performance Data
{5}------------------------------------------------
## Mechanical testing
- Suture Retention Testing ●
- Tensile Strength Testing ●
The tests listed above provide objective evidence that the FlexBand®, FlexBatch®, and FlexBand Plus® devices provide adequate mechanical properties for use in ligament soft tissue reinforcement. The performance data risk / benefit analysis concluded that the differences do not affect the safety and effectiveness of the FlexBand®, FlexPatch®, and FlexBand Plus® device in relation to the predicate.
Routine endotoxin (LAL) testing is performed on each production lot to monitor endotoxin levels.
# Clinical Literature
Clinical literature was provided on the subject device involving ligament reinforcement procedures. The reviewed literature shows the device is effective when used in the proposed ligament indication.
# VIII. Conclusion
Based on the above information the FlexBand®, FlexPatch®, and FlexBand Plus® products are substantially equivalent to the soft tissue reinforcement predicate devices:
The FlexBand® / FlexPatch® / FlexBand® Plus (K192112)
```
BioBridge™ Collagen Matrix (K151083)
```
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.