AirSeal iFS System The ConMed AirSeal® iFS System is intended for use in diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas, to create and maintain a gas-sealed obstruction-free path of entry for endoscopic instruments, and to evacuate surgical smoke. It is indicated for use in abdominal, thoracic, and pediatric procedures where insufflation is desired to facilitate the use of various thoracoscopic and laparoscopic instruments by filling the abdominal or thoracic cavity with gas to distend it, by creating and maintaining a gas sealed obstruction-free instrument path and by evacuating surgical smoke. This instrument can also be used to insufflate the rectum and colon to facilitate endoscopic observation, diagnosis, and treatment. The trocar of the AirSeal® iFS System is indicated for use with or without visualization. AirSeal dV Solution The AirSeal Robotic Solution (ARS) is intended for use in diagnostic and/or therapeutic endoscopic procedures in conjunction with the ConMed AirSeal iFS and an Intuitive da Vinci X Cannula and Cannula Seal to distend a cavity by filling it with gas, to create and maintain a gas-sealed obstruction-free path of entry for endoscopic instruments, and to evacuate surgical smoke. It is indicated for use in abdominal, thoracic, and pediatric procedures where insufflation is desired to facilitate the use of various laparoscopic and thoracoscopic instruments by filling the abdominal or thoracic cavity with gas to distend it, by creating and maintaining a gas sealed obstruction-free instrument path and by evacuating surgical smoke. The Cannula Cap is indicated for use with or without visualization. The ARS System must be used with the Intuitive da Vinci X 8mm Instrument Cannula, an Intuitive Cannula Seal and the AirSeal iFS in AirSeal mode. When used in AirSeal mode, the Cannula Cap and Bifurcated Tube Set are designed to provide CO2 gas delivery with stable pneumoperitoneum and continuous smoke evacuation during robotic assisted laparoscopic and thoracoscopic procedures using the da Vinci Xi or da Vinci X systems. The Bifurcated Filtered Tube Set is used to connect the Cannula Cap and Cannula Seal to the AirSeal iFS.
Device Story
System comprises trocar, cannula, tube sets, and microprocessor-controlled insufflation/recirculation/filtration unit. Operates in Insufflation, AirSeal (adult/pediatric), or Smoke Evacuation modes. Inputs: CO2 gas source; outputs: regulated gas flow and smoke evacuation. Used in OR by surgeons/staff. AirSeal dV Solution integrates with Intuitive da Vinci X/Xi systems via cannula cap/bifurcated tube set to maintain stable pneumoperitoneum during robotic surgery. Microprocessor regulates pressure/flow to maintain obstruction-free instrument path. Benefits: stable cavity distension, continuous smoke evacuation, improved visualization. Healthcare providers monitor system via interface to adjust insufflation parameters; output facilitates surgical access and visibility.
Clinical Evidence
Bench testing only. Testing included packaging (ISO 11607-1), software V&V (IEC 62304), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance metrics including set pressure, initial insufflation, obturator removal, instrument insertion, and leak compensation.
Technological Characteristics
Digital insufflation pressure regulation system. Components: polycarbonate/stainless steel cannula/cap, bifurcated filtered tube sets. Connectivity: connects to robotic surgical systems. Software: microprocessor-controlled. Sterilization: sterile single-use components (cannula/cap/tubes); reusable insufflation unit.
Indications for Use
Indicated for abdominal, thoracic, and pediatric patients requiring insufflation for diagnostic/therapeutic endoscopic procedures; facilitates instrument entry and smoke evacuation. Includes rectal/colonic insufflation. ARS indicated for robotic-assisted laparoscopic/thoracoscopic procedures using da Vinci Xi/X systems.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 20, 2023
CONMED Corporation Tina Mornak Manager, Regulatory Affairs 525 French Road Utica, NY 13502
K230239 Re: Trade/Device Name: AirSeal iFS System AirSeal dV Solution Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF, GCJ Dated: August 17, 2023 Received: August 17, 2023
Dear Tina Mornak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230239
Device Name AirSeal iFS System AirSeal dV Solution
# Indications for Use (Describe)
### AirSeal iFS System
The ConMed AirSeal® iFS System is intended for use in diagnostic endoscopic procedures to distend a cavity by filling it with gas, to create and maintain a gas-sealed obstruction-free path of entry for endoscopic instruments, and to evacuate surgical smoke.
It is indicated for use in abdominal, thoracic, and pediatric procedures where insufflation is desired to facilitate the use of various thoracoscopic instruments by filling the abdominal or thoracic cavity with gas to distend it, by creating and maintaining a gas sealed obstruction-free instrument path and by evacuating surgical smoke. This instrument can also be used to insufflate the rectum and colon to facilitate endoscopic observation, diagnosis, and treatment. The trocar of the AirSeal® iFS System is indicated for use with or without visualization.
### AirSeal dV Solution
The AirSeal Robotic Solution (ARS) is intended for use in diagnostic and/or therapeutic endoscopic procedures in conjunction with the ConMed AirSeal iFS and an Intuitive da Vinci X Cannula and Cannula Seal to distend a cavity by filling it with gas, to create and maintain a gas-sealed obstruction-free path of entry for endoscopic instruments, and to evacuate surgical smoke.
It is indicated for use in abdominal, thoracic, and pediatric procedures where insufflation is dese of various laparoscopic and thoracoscopic instruments by filling the abdominal or thoracic cavity with gas to distend it, by creating and maintaining a gas sealed obstruction-free instrument path and by evacuating surgical smoke. The obturator of the Cannula Cap is indicated for use with or without visualization.
The ARS System must be used with the Intuitive da Vinci X 8mm Instrument Cannula, an Intuitive Cannula Seal and the AirSeal mode. When used in AirSeal mode, the Cannula Cap and Bifurcated Tube Set are designed to provide CO2 gas delivery with stable pneumoperioneum and continuous smoke evacuation during robotic assisted laparoscopic and thoracoscopic procedures using the da Vinci X systems. The Bifurcated Filtered Tube Set is used to connect the Cannula Cap and Cannula Seal to the AirSeal iFS.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# AirSeal iFS System
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for this 510(k).
#### l. SUBMITTER
ConMed Corporation 525 French Road Utica, NY 13502
Company Contact:
Tina Mornak Manager, Requlatory Affairs 724-518-3191 christinamornak@conmed.com
September 19, 2023 Date Prepared:
#### II. DEVICE NAME
Proprietary Name: AirSeal iFS System AirSeal dV Solution
AirSeal iFS System Model Numbers:
AirSeal dV Solution
AS-iFS1, iAS12-100, iAS12-100LPi, iAS12-120, iAS12-120LPi, iAS12-150, iAS12-150LPi, iAS5-100LP, iAS5-120LP, iAS5-75LP, iAS8-100LP, iAS8-120LP, iASB12-100, iASB12-120, iASB5-150, SEM-EVAC, SIM-TUB, ASM-EVAC, ASM-EVAC1
| Model Numbers: | ASM-EVAC1-Bi, iAS8-DV, iAS8-DVL |
|--------------------|-------------------------------------------------------------------|
| Common Name: | Insufflator, Laparoscopic; Laparoscope, General & Plastic Surgery |
| Regulation Name: | Laparoscopic Insufflator |
| Product Code: | HIF; GCJ |
| Device Class: | II |
| Regulation Number: | 21 CFR 884.1730; 21 CFR 876.1500 |
#### lll. PREDICATE/LEGALLY MARKETED DEVICE
| Predicate | |
|---------------|-----------------------------------------------------------------------|
| Device Name: | WOM 45L CORE Insufflator, Model F114 |
| Company Name: | W.O.M. World of Medicine GmBH |
| 510(k): | K063367 (Marketed by Stryker as the PneumoSure High Flow Insufflator) |
The predicate device has not been subject to a design-related recall.
# DEVICE DESCRIPTION
The ConMed AirSeal iFS System consists of the following major components: (1) a trocar, (2) a cannula, (3) tube sets, and (4) a microprocessor-controlled insufflation, recirculation
{5}------------------------------------------------
and filtration unit. The cannula, trocar and tube sets are sterile, single-use products. The AirSeal iFS System is an active medical device, nonsterile and reusable and is intended to insufflate a body cavity. The AirSeal iFS System is designed to function in one of three (3) separate modes of operation: (a) Insufflation Mode; (b) AirSeal Mode (Adult and Pediatric); or (c) Smoke Evacuation Mode.
The ConMed AirSeal dV Solution (or AirSeal Robotic Solution) consists of the following major components: (1) AirSeal® Cannula Cap with a cannula cap and an obturator (in standard and long lengths), and (2) AirSeal® Bifurcated Filtered Tube Set. The cannula cap, obturator and tube set are sterile, single-use products. The AirSeal dV Solution, in conjunction with Intuitive daVinci XiX Cannula and Cannula Seal, is operated in the Airseal Mode with the AirSeal iFS System to distend a cavity by filling it with gas, to create and maintain a gas-sealed obstruction-free path of entry for endoscopic instruments, and to evacuate surgical smoke. The ConMed AirSeal dV Solution is a sterile, single use device, primarily composed of polycarbonate, and stainless steel,
#### IV. INTENDED USE / INDICATIONS FOR USE
### AirSeal iFS System
The ConMed AirSeal® iFS System is intended for use in diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas, to create and maintain a gas-sealed obstructionfree path of entry for endoscopic instruments, and to evacuate surgical smoke.
It is indicated for use in abdominal, thoracic, and pediatric procedures where insufflation is desired to facilitate the use of various thoracoscopic and laparoscopic instruments by filling the abdominal or thoracic cavity with gas to distend it, by creating and maintaining a as sealed obstruction-free instrument path and by evacuating surgical smoke. The instrument can also be used to insufflate the rectum and colon to facilitate endoscopic observation, diagnosis, and treatment. The trocar of the AirSeal® iFS System is indicated for use with or without visualization.
# AirSeal dV Solution
The AirSeal Robotic Solution (ARS) is intended for use in diagnostic and/or therapeutic endoscopic procedures in conjunction with the ConMed AirSeal iFS and an Intuitive da Vinci X Cannula and Cannula Seal to distend a cavity by filling it with gas, to create and maintain a gassealed obstruction-free path of entry for endoscopic instruments, and to evacuate surgical smoke.
It is indicated for use in abdominal, thoracic, and pediatric procedures where insufflation is desired to facilitate the use of various laparoscopic and thoracoscopic instruments by filling the abdominal or thoracic cavity with gas to distend it, by creating and maintaining a gas sealed obstruction-free instrument path and by evacuating surgical smoke. The Cannula Cap is indicated for use with or without visualization.
The ARS System must be used with the Intuitive da Vinci X 8mm Instrument Cannula, an Intuitive Cannula Seal and the AirSeal iFS in AirSeal mode. When used in AirSeal mode, the Cannula Cap and Bifurcated Tube Set are designed to provide CO2 gas delivery with stable pneumoperitoneum and continuous smoke evacuation during robotic assisted laparoscopic and thoracoscopic procedures using the da Vinci Xi or da Vinci X systems. The Bifurcated Filtered Tube Set is used to connect the Cannula Cap and Cannula Seal to the AirSeal iFS.
#### V. COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
{6}------------------------------------------------
| Device & Predicate Device(s): | Subject Device<br>K230239 | Predicate<br>Device<br>K063367 |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | CONMED AirSeal<br>iFS System | WOM 45L CORE<br>Insufflator Fl 14 |
| Manufacturer | CONMED | WOM (World Of<br>Medicine) |
| Device Name | AirSeal iFS System | 45L CORE Insufflator<br>F1 14 |
| Regulation Number | 21 CFR 884.1730<br>21 CFR 876.1500 | 21 CFR 884.1730<br>21 CFR 876.1500 |
| Product Code | HIF<br>GCJ | HIF<br>GCJ |
| Device Class/Name | Class II/Insufflator | Class II/Insufflator |
| Fundamental Scientific Technology | Digital insufflation<br>pressure regulation<br>system | Digital Insufflation<br>pressure regulation<br>system |
| Intended Use | AirSeal® iFS System is<br>intended for use in<br>diagnostic and/or<br>therapeutic endoscopic<br>procedures to distend a<br>cavity by filling it with<br>gas, to establish and<br>maintain a path of entry<br>for endoscopic<br>instruments and to<br>evacuate surgical smoke. | The 45L CORE<br>Insufflator F1 14 (the<br>"Fl 14) is a C02<br>insufflator intended for<br>use in diagnostic and/or<br>therapeutic endoscopic<br>procedures to distend a<br>cavity by filling it with<br>gas. |
| Indications for Use | The ConMed AirSeal®<br>iFS System is intended<br>for use in diagnostic<br>and/or therapeutic<br>endoscopic procedures to<br>distend a cavity by filling<br>it with gas, to create and<br>maintain a gas-sealed<br>obstruction-free path of<br>entry for endoscopic<br>instruments, and to<br>evacuate surgical smoke.<br>It is indicated for use in<br>abdominal, thoracic, and<br>pediatric procedures<br>where insufflation is<br>desired to facilitate<br>the use of various<br>thoracoscopic and<br>laparoscopic instruments<br>by filling the abdominal<br>or thoracic cavity with<br>gas to distend it, by<br>creating and maintaining<br>a gas sealed obstruction-<br>free instrument path and<br>by evacuating surgical<br>smoke. This instrument<br>can also be used to<br>insufflate the rectum and<br>colon to facilitate<br>endoscopic observation,<br>diagnosis, and treatment. | The 45L core insufflator<br>fl 14 is a co2 insufflator<br>intended for use in<br>diagnostic and/or<br>therapeutic endoscopic<br>procedures to distend a<br>cavity by filling it with<br>gas. The high flow<br>application, the low<br>flow application and the<br>bariatric application of<br>the device are each<br>indicated for use in<br>facilitating the use of a<br>laparoscope by filling<br>the peritoneal cavity<br>with gas to distend it.<br>The low flow<br>application of the<br>device is indicated for<br>pediatric use. The<br>vessel harvesting<br>application of the fl 14<br>is indicated for use<br>during endoscopic<br>vessel harvesting<br>procedures to create a<br>cavity along the<br>saphenous vein and/or<br>radial artery during<br>endoscopic vessel<br>harvesting procedures |
| | | |
| indicated for use with or | | |
| without visualization. | | |
| | | |
| The AirSeal Robotic | | |
| Solution (ARS) is | | |
| intended for use in | | |
| diagnostic and/or | | |
| therapeutic endoscopic | | |
| procedures in conjunction | | |
| with the ConMed AirSeal | | |
| iFS and an Intuitive da | | |
| Vinci Xi or da Vinci X | | |
| Cannula and Cannula Seal | | |
| to distend a cavity by | | |
| filling it with gas, to | | |
| create and maintain a gas- | | |
| sealed obstruction-free | | |
| path of entry for | | |
| endoscopic instruments,<br>and to evacuate surgical | | |
| smoke. | | |
| | | |
| It is indicated for use in | | |
| abdominal, thoracic, and | | |
| pediatric procedures<br>where insufflation is | | |
| desired to facilitate the | | |
| use of various | | |
| laparoscopic and | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.