The SA-20 Series recorder combines ambulatory ECG Holter and BP monitor. It is intended to acquire, store and display ambulatory ECG signals from adult and pediatric patient. It is capable of measuring systolic and diastolic blood pressures of adults and pediatrics (> 12 years) over a preprogrammed period of time. These measurements are displayed and stored in the recorder. It is intended to be used by trained personnel under the direction of doctors.
Device Story
The SA-20 Series is a combined ambulatory ECG Holter and non-invasive blood pressure (NIBP) monitor. It uses an oscillometric method for blood pressure measurement and surface electrodes for ECG signal acquisition. The device records and stores ECG waveforms and BP measurements on an internal SD card. It is intended for use by trained personnel under physician direction in clinical or home settings. The device does not perform automated diagnostics; clinicians review the stored data to make clinical decisions. The system benefits patients by providing long-term ambulatory monitoring of cardiac and blood pressure parameters outside of a hospital environment.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (ANSI AAMI ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing against consensus standards (IEC 60601-2-47 for ECG, IEC 80601-2-30 for NIBP, and IEC 60601-1-11 for home healthcare). Software verification and validation were performed per FDA guidance.
Technological Characteristics
System includes recorder, NIBP cuff, and lead wires. Powered by two AA alkaline batteries. Features USB interface and SD card storage. ECG sampling frequency 25.6 ksps; 24-bit A/D conversion. Operates via oscillometric method for BP and surface electrodes for ECG. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-47, IEC 80601-2-30, and ANSI AAMI EC53.
Indications for Use
Indicated for adult and pediatric patients (>12 years for BP) requiring ambulatory ECG and/or non-invasive blood pressure monitoring. Contraindicated for patients with erratic, accelerated, or mechanically controlled irregular heart rhythms (arrhythmias); unsupervised children or patients with limited cognitive abilities (strangulation risk); and intensive care or surgical settings. ABP not for pregnant or pre-eclamptic patients.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
Vasomedical, Inc., Combined 12 Channel Ambulatory EGG and Blood Pressure Recorder, Model 2302 (K111096)
Reference Devices
Vasomedical, Inc., Combined Ambulatory EGG and Blood Pressure Recorder (K092785)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
September 25, 2023
Edan Instruments, Inc. Lavender Wang Regulatory Engineer #15 Jinhui Road, Jinsha Community, Kengzi Sub-District Pingshan District Shenzhen, Guangdong 518122 China
Re: K230184
Trade/Device Name: Holter ECG and ABP System Regulation Number: 21 CFR 870.2800 Regulation Name: Medical magnetic tape recorder Regulatory Class: Class II Product Code: DSH, MWJ, DXN Dated: August 25, 2023 Received: August 25, 2023
Dear Lavender Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer W.Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices
{2}------------------------------------------------
Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230184
Device Name
Holter ECG and ABP SystemModel: SA-20, SA-19, SA-18, SA-15, SA-16
Indications for Use (Describe)
The SA-20 Series recorder combines ambulatory ECG Holter and BP monitor.
It is intended to acquire, store and display ambulatory ECG signals from adult and pediatric patient. It is capable of measuring systolic and diastolic blood pressures of adults and pediatrics (> 12 years) over a preprogrammed period of time. These measurements are displayed and stored in the recorder.
It is intended to be used by trained personnel under the direction of doctors.
Type of Use (Select one or both, as applicable)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary Prepared in accordance with the content and format regulatory Requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc.<br>#15 Jinhui Road, Jinsha Community,<br>Kengzi Sub-District, Pingshan District,<br>Shenzhen, 518122 P.R.China.<br>Tel: +86(0755) 26858736<br>Fax: +86(0755) 26882223 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Liu Yongying |
| Preparing date: | January 16, 2023 |
| 2. Device name and<br>classification: | Trade Name: Holter ECG and ABP System<br>Common/Usual Name: SA-20 Series<br>Model: SA-20, SA-19, SA-18, SA-15, SA-16 |
| Classification Name | Product code |
|----------------------------------------------------------------------|--------------|
| 21 CFR 870.2800 Recorder, Magnetic Tape,<br>Medical | DSH |
| 21 CFR 870.2800 Electrocardiograph,<br>Ambulatory (Without Analysis) | MWJ |
| 21 CFR 870.1130 System, Measurement,<br>Blood-Pressure, Non-Invasive | DXN |
| Regulatory Class: | Class II |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 3. Predicate Device(s): | 1) Vasomedical, Inc., Combined 12 Channel Ambulatory EGG and Blood<br>Pressure Recorder, Model 2302, cleared under K111096 (Primary).<br>2) Vasomedical, Inc., Combined Ambulatory EGG and Blood Pressure<br>Recorder, cleared under K092785 (Reference).<br>3) Edan Instruments, Inc., Holter System, cleared under K151787 (Reference).<br>4) Edan Instruments, Inc., Ambulatory Blood Pressure Monitor, cleared under<br>K210139 (Reference). |
| 4. Device Description: | Holter ECG and ABP System (Model: SA-20, SA-19, SA-18, SA-15, SA-16) is<br>composed of the recorder, NIBP cuff, lead wire. The ambulatory blood pressure<br>adopts the principle of the oscillometric method for non-invasive blood pressure<br>measurement. ECG signals are obtained by placing electrodes on the patient's<br>surface to record the potential change signal of the human heart and then<br>processing the data to form an ECG waveform and store it in the Recorder to<br>monitor the ECG signal. |
{5}------------------------------------------------
#### Contraindications :
| | Do not use it on patients with erratic, accelerated, or mechanically controlled<br>irregular heart rhythms, including the patient with arrhythmias. |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Due to the strangulation risk posed by the cuff and hose, the SA-20 Series must<br>not be within reach of unsupervised children. It must not be used on<br>unsupervised patients with limited cognitive abilities. |
| | The SA-20 Series is not intended for alarm triggering monitoring purposes in<br>intensive care units. It must not be used for blood pressure monitoring purposes<br>in intensive care units or during surgery.<br>ABP is not intended for pregnant, including pre-eclamptic patients. |
| 5. Indication for Use | The SA-20 Series recorder combines ambulatory ECG Holter and BP monitor.<br>It is intended to acquire, store and display ambulatory ECG signals from adult<br>and pediatric patient. It is capable of measuring systolic and diastolic blood<br>pressures of adults and pediatrics (> 12 years) over a preprogrammed period of<br>time. These measurements are displayed and stored in the recorder.<br>It is intended to be used by trained personnel under the direction of doctors. |
# 6. Predicate Device Comparison
Comparison to the predicate devices, the subject device has the same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following tables:
| Item | Subject Device (Model: SA-20 Series ) | Predicate Device ( Model : 2302 ) | Comparison Result |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Manufacturer | Edan Instruments, Inc | Vasomedical, Inc | -- |
| K# | K230184 | K111096 | -- |
| <b>Indications for Use</b> | | | |
| Indications for Use | The SA-20 Series recorder combines ambulatory ECG Holter and BP monitor.<br>It is intended to acquire, store and display ambulatory ECG signals | Vasomedical-Biox Combined 12 Channel Ambulatory ECG and Blood Pressure Recorder, Model 2302 is a Non-Invasive device intended to acquire ambulatory 12 channel ECG | Similiar |
# Table 1: Comparison between SA-20 Series and 2302 (primary predicate device)
{6}------------------------------------------------
| | from adult and pediatric patient. It is | signals and non-invasive oscillometric | |
|------------------------------------|----------------------------------------------|---------------------------------------------------------------------|-----------|
| | capable of measuring systolic and | blood pressure signals from the upper | |
| | diastolic blood pressures of adults | body surfaces. Cardiac rhythm is | |
| | and pediatrics (> 12 years) over a | acquired via ECG signals. | |
| | preprogrammed period of time. | Vasomedical-Biox Model 1804 | |
| | These measurements are displayed | Ambulatory BP Recorder is a | |
| | and stored in the recorder. | noninvasive device intended to acquire | |
| | It is intended to be used by trained | ambulatory non-invasive oscillometric | |
| | personnel under the direction of | Blood Pressure signals from the upper | |
| | doctors. | body surfaces. This ABP Recorder | |
| | | functions exactly the same as Model | |
| | | 2301/2302 for measurement and | |
| | | recording of Blood Pressure signals. | |
| | | The Recorders are intended for adults | |
| | | and children who are over the age of | |
| | | six years. | |
| | Physical properties | | |
| System<br>Components | Recorder, blood pressure cuff, lead<br>wire. | Recorder, blood pressure cuff, lead<br>wire, and analysis software. | Different |
| Energy Used<br>and/or<br>Delivered | Two AA alkaline batteries | four AA alkaline batteries | Different |
| Interface | USB Interface Cable<br>SD card | USB Interface Cable<br>SD card | same |
| Patient<br>Connections | Blood pressure cuff<br>Patient cable | Blood pressure cuff<br>Patient cable | same |
| | Performance Specifications | | |
| Channels | 12 channels | 12 channels | same |
| Electrodes | 10 | 10 | same |
| A/D Sampling<br>Frequency | 25.6 ksps | 10000Hz | Different |
# Table 2: Comparison between SA-20 Series and 2301 (reference device)
| Item | Subject Device(Model: SA-20<br>Series ) | Reference Device (Model: 2301) | Comparison<br>Result |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacturer | Edan Instruments, Inc | Vasomedical, Inc | -- |
| K# | K230184 | K092785 | -- |
| Indications for Use | | | |
| | | | |
| | | Vasomedical-Biox Combined Ambulatory<br>ECG and Blood Pressure Recorder is a<br>Non-Invasive device intended to acquire<br>Ambulatory 3 Channel ECG signals and<br>non-invasive oscillometric Blood Pressure<br>signals from the upper body surfaces.<br>Cardiac rhythm is acquired via 3 Channel<br>ECG signals. The Recorders are intended<br>for adults and children who are over the<br>age of six years.<br>Vasomedical-Biox Ambulatory ECG CB<br>Series Analysis System Software allows<br>transfer of ECG and Blood Pressure data<br>from the Recorder to a Windows based PC<br>program via a removable and large capacity<br>storage card (SD) for the purpose of<br>creating reports and printouts. The software<br>does not perform diagnostics. Physicians<br>carry out diagnostic evaluations of this<br>data.<br>The system is only for measurement,<br>recording and display. It makes no<br>diagnosis. | Similiar |
| Indications for<br>Use | The SA-20 Series recorder<br>combines ambulatory ECG Holter<br>and BP monitor.<br>It is intended to acquire, store and<br>display ambulatory ECG signals<br>from adult and pediatric patient. It<br>is capable of measuring systolic and<br>diastolic blood pressures of adults<br>and pediatrics (> 12 years) over a<br>preprogrammed period of time.<br>These measurements are displayed<br>and stored in the recorder.<br>It is intended to be used by trained<br>personnel under the direction of<br>doctors. | | |
| Physical properties | | | |
| System<br>Components | Recorder, blood pressure cuff, lead<br>wire. | Recorder, blood pressure cuff, lead wire,<br>and analysis software. | Different |
| Energy Used<br>and/or<br>Delivered | Two AA alkaline batteries | four AA alkaline batteries | Different |
| Interface | USB Interface Cable/SD card | USB Interface Cable/SD card | same |
| Patient<br>Connections | Blood pressure cuff; Patient cable | Blood pressure cuff; Patient cable | same |
| Performance Specifications | | | |
| Channels | 3 channels | 3 channels | same |
| Electrodes | 5 or 7 | 5 or 7 | same |
| A/D Sampling<br>Frequency | 25.6 ksps | 10000Hz | Different |
| A/D | 24 bits | 10 bits | Different |
{7}------------------------------------------------
As seen in the comparison tables, the subject device and predicate devices have similar design features and performance specifications. The technological differences between the subject device and predicate
{8}------------------------------------------------
devices do not raise different questions of safety or effectiveness.
### 7. Performance Data:
### Non-clinical data:
### Electrical safety and electromagnetic compatibility (EMC)
The Holter ECG and ABP System was assessed for conformity with the relevant requirements of the following standards and complys:
- ANSI AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009(R) 2012 and A2:2010(R) 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 (Fourth Edition) Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests.
### Performance testing-Bench
Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.
- IEC 60601-2-47 Edition 2.0 2012-02 Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performanceof ambulatory electrocardiographic systems
- IEC 80601-2-30: Edition 2.0 2018-03 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- ANSI AAMI EC53:2013/(R) 2020 ECG trunk cables and patient leadwires
- IEC 60601-1-11 Edition 2.1 2020-07, Medical electrical equipment - Part 1-11 General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
#### Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
#### Clinical data: Not applicable.
#### Summary
The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.
# 8. Conclusion
{9}------------------------------------------------
The bench testing data and software verification and validation demonstrate that Holter ECG and ABP System is substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.