The Ambulatory Blood Pressure Monitor is capable of measuring systolic and diastolic blood pressures, and pulse of adults and pediatrics (> 12 years) over a preprogrammed period of time. These measurements are stored in the monitor and can be transferred to analysis system via wire or wireless transmission. It is intended for use as an aid or adjunct to diagnosis and treatment.
Device Story
The SA Series Ambulatory Blood Pressure Monitor (ABPM) consists of an electronics module, cuff, and 'Smart ABPM View' analysis software. The device is worn by patients in home or clinical settings to measure systolic/diastolic blood pressure and pulse via the oscillometric method with step deflation. Measurements are stored on the device and transferred to a PC via USB or Bluetooth for clinician review. The software allows clinicians to edit, display, and analyze the recorded data. By providing longitudinal blood pressure data, the device assists physicians in diagnosing and managing hypertension or other cardiovascular conditions, potentially improving patient outcomes through better-informed treatment decisions.
Clinical Evidence
Clinical investigation conducted in accordance with ISO 81060-2:2018+A1:2020. The study validated the safety and effectiveness of the SA series ABPM in both resting and ambulatory states. Results confirmed the device meets clinical accuracy requirements for its intended population.
Technological Characteristics
Oscillometric measurement with step deflation. Connectivity via USB or Bluetooth. Software-based analysis on Windows (XP/7/8/10). Biocompatibility per ISO 10993-1, -5, -10. Safety compliance with ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 80601-2-30.
Indications for Use
Indicated for measuring systolic/diastolic blood pressure and pulse in adults and pediatric patients (>12 years) over a preprogrammed period. Intended as an aid/adjunct to diagnosis and treatment.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Suntech Medical, Inc., Oscar 2 NIBP Monitor (K151520)
Reference Devices
A&D MEDICAL DIV, TM-2430 Ambulatory Blood Pressure Monitors (K992808)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
October 1, 2021
Edan Instruments, Inc. Ying Dai Regulatory Engineer #15 Jinhui Road, Jinsha Community, Kengzi Sub-District, Pingshan District Shenzhen, Guangdong 518122 China
Re: K210139
Trade/Device Name: Ambulatory Blood Pressure Monitor, Models: SA-10, SA-05, SA-08 and SA-09 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 26, 2021 Received: August 31, 2021
Dear Ying Dai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210139
Device Name
Ambulatory Blood Pressure Monitor, Models: SA-10, SA-05, SA-06, SA-08 and SA-09
Indications for Use (Describe)
The Ambulatory Blood Pressure Monitor is capable of measuring systolic and diastolic blood pressures, and pulse of adults and pediatrics (> 12 years) over a preprogrammed period of time. These measurements are stored in the monitor and can be transferred to analysis system via wire or wireless transmission. It is intended for use as an aid or adjunct to diagnosis and treatment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block;">☒</div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="display:inline-block;">☐</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Traditional 510K for SA Series Ambulatory Blood Pressure Monitor
## 510(k) Summary
## Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc.<br>#15 Jinhui Road, Jinsha Community, Kengzi Sub-District, Pingshan District,<br>Shenzhen, 518122 P.R.China.<br>Tel: +86-755-2685 6469 Fax: +86-755-2689 8330 | | | | | | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
| Contact person: | Ying DAI | | | | | | |
| Preparing date: | Jan 5th, 2021 | | | | | | |
| 2. Device name and<br>classification: | Trade name: Ambulatory Blood Pressure Monitor, Model: SA-10, SA-05,<br>SA-06, SA-08 and SA-09<br>Common/Usual Name: Noninvasive blood pressure measurement system Classification Name Product code 21 CFR 870.1130<br>System, Measurement,<br>Blood-Pressure, Non-Invasive/ DXN Regulatory Class: Class II | | | | | | |
| 3. Predicate Device(s): | Suntech Medical, Inc., Oscar 2 NIBP Monitor, cleared under K151520<br>(Primary)<br>A&D MEDICAL DIV, TM-2430 Ambulatory Blood Pressure Monitors, cleared under K992808 (reference) | | | | | | |
| 4. Device Description: | The SA Series ABPM (Ambulatory Blood Pressure Monitor), primarily composed of Ambulatory Blood Pressure Monitor (ABPM) and ABPM Analysis software (Smart ABPM View), is designed to measure systolic and diastolic blood pressures, and pulse of adults and pediatrics (> 12 years) over a preprogrammed period of time.<br>The monitor inflates and deflates the cuff on the upper arm to measure blood pressures and pulse by the oscillometric method and stores the measurement.<br>The stored measurements are transferred into analysis software installed in generally used PC via wire transfer or Bluetooth.<br>The monitor is carried around by patients. Patients are requested to come back to medical treatment site after recording time. The systolic and diastolic blood pressures, and pulse data recorded in monitor will be transferred into PCs and then analyzed, displayed and edited by Analysis software. | | | | | | |
{4}------------------------------------------------
### Traditional 510K for SA Series Ambulatory Blood Pressure Monitor
## 5. Indication for Use
The Ambulatory Blood Pressure Monitor is capable of measuring systolic and diastolic blood pressures, and pulse of adults and pediatrics (> 12 years) over a preprogrammed period of time. These measurements are stored in the monitor and can be transferred to analysis system via wire or wireless transmission. It is intended for use as an aid or adjunct to diagnosis and treatment.
## 6. Predicate Device Comparison
| Item | <Subject Device> | <Predicate Device> | Comparison<br>on Result |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Manufacturer/<br>K# | Current Submission | K151520 | --- |
| Indications for Use | | | |
| Indication for<br>use | The Ambulatory Blood Pressure<br>Monitor is capable of measuring<br>systolic and diastolic blood<br>pressures, and pulse of adults and<br>pediatrics (> 12 years) over a<br>preprogrammed period of time.<br>These measurements are stored in<br>the monitor and can be transferred<br>to analysis system via wire or<br>wireless transmission. It is<br>intended for use as an aid or<br>adjunct to diagnosis and treatment. | The Oscar 2, Model 250 system is a non-invasive<br>oscillometric ambulatory blood pressure monitor<br>that is intended to be used with AccuWin Pro, a<br>PC-based computer program for the recording<br>and displaying of up to 250 measurements of<br>systolic and diastolic blood pressure and heart<br>rate. It is intended for use as an aid or adjunct to<br>diagnosis and treatment when it is necessary to<br>measure an adult and pediatric (> 3yrs.) patient's<br>systolic and diastolic blood pressures over an<br>extended period of time. The system is only for<br>measurement, recording, and display. It makes no<br>diagnoses.<br>Optionally, The Model 250 will provide a derived<br>ascending aortic blood pressure waveform and a<br>range of central arterial indices. These<br>measurements are provided noninvasively<br>through the use of a brachial cuff.<br>It is to be used on those patients where<br>information related to ascending aortic blood<br>pressure is desired but the risks of cardiac<br>catheterization procedure or other invasive<br>monitoring may outweigh the benefits (excludes<br>pediatric subjects).<br>Bluetooth, wireless connectivity will be offered<br>as an option. | Different |
| | | | |
| Intended<br>Patient<br>Population | adults and pediatrics (>12yrs.) | adult and pediatric (>3yrs.) patients | Different |
| Intended<br>Application<br>Environment | Physician's office, clinic, research<br>center(under supervision of<br>physician) and patient home<br>environment | Physician's office, clinic, research center(under<br>supervision of physician) and patient home<br>environment | Same |
| Measurement | Systolic and Diastolic Blood<br>Pressure<br>Heart Rate | Systolic and Diastolic Blood Pressure<br>Heart Rate<br>Central Blood Pressure (Sp/Dp/MAP/Pp)<br>Central Augmentation Index & Pressure, and<br>Index at HR 75 | Different |
| | | Physical properties | |
| System<br>Components | Electronics Module (EDAN SA<br>Series)<br>EDAN NIBP Cuff sleeve<br>Software Disc<br>Operator's Manual<br>Cables<br>Carrying pouch | Electronics Module (Oscar 2, Model 250)<br>SunTech BP Cuff<br>Software download<br>Operator's Manual<br>Cables<br>Pouch | Same |
| Interface | USB Interface Cable<br>Wireless-Bluetooth | USB Interface Cable<br>Wireless-Bluetooth | Same |
| | | Software Section | |
| Software<br>Operating<br>Platform | Windows XP, Windows 7,<br>Windows 8, or Windows 10 | Microsoft Windows® 7, 8 | Different |
| PC Software | Smart ABPM View | AccuWin Pro | Different |
| NIBP<br>Algorithm | NIBP iCUFS | SunTeck Oscar 2 | Different |
| CBP<br>Algorithm | N/A | SphygmoCor® XCEL | Different |
| Performance Specifications | | | |
| Operation<br>Principle | Oscillometric with step deflation | Oscillometric with step deflation | Same |
| Blood Pressure<br>Range | Systolic: 25 to 290 mmHg (3.3 to<br>38.6 kPa)<br>Diastolic: 10 to 250 mmHg (1.3 to<br>33.2 kPa) | Systolic: 40-260 mmHg<br>Diastolic: 25-200 mmHg | Different |
| Accuracy of | Mean error: ±5 mmHg (±0.67 kPa) | Blood Pressure results meet or exceed | Same |
| Blood Pressure | Standard deviation: Max. 8 mmHg<br>(1.07 kPa) | ANSI/AAMI/ISO 81060-2:2013 standards for<br>non-invasive accuracy: ±5 mmHg mean error & 8 mmHg standard deviation. | |
| Heart Rate<br>Range | 40-240 bpm | 40-200 bpm | Different |
| Error of Heart<br>Rate | Heart Rate accurate within ±3 bpm<br>or ±3.5%, whichever is greater. | Heart Rate accurate within ±2% or ±3 bpm,<br>whichever is greater. | Cleared<br>under<br>K992808 |
| Environmental Specifications | | | |
| Temperature | | | |
| Operating | +5 ℃ (+41 F) ~ +40 ℃ (+104 F) | +10 ℃ (+50 °F) ~ +50 ℃ (+122 F) | |
| Transport/Storage | -20 ℃ (-4 F) ~ +55 ℃ (+131 F) | -20 ℃ ~ +70 ℃ | |
| Humidity | | | Different |
| Operating | 15%RH~95%RH Non-Condensing | 20%RH~95%RH Non-Condensing | |
| Transport/Storage | 15%RH~95%RH Non-Condensing | 15%RH~95%RH Non-Condensing | |
{5}------------------------------------------------
## Traditional 510K for SA Series Ambulatory Blood Pressure Monitor
006_6.0_510(k)_Summary
{6}------------------------------------------------
## Traditional 510K for SA Series Ambulatory Blood Pressure Monitor
The subject and predicate device have same general intended use, similar design features and performance specifications. The technological differences between the subject and predicate device do not raise different questions of safety or effectiveness.
## 7. Performance Data:
## Clinical test:
The subject devices have been undertaken clinical investigation in accordance with ISO 81060-2:2018+A1:2020 to validate the clinical safety and effectiveness within its intended use. From the investigation with qualified subject distribution, SA series ambulatory blood pressure monitor with its accessories are concluded to be safe and effective in both resting state and ambulatory monitoring within its intended use.
## Non-clinical test:
The SA series Ambulatory Blood Pressure Monitor complies with:
- (1) ANSUAAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009(R)2012 and A2:2010/(R)2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- (2) IEC 60601-1-2 Edition 4.0 2014-02, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- (3) IEC 60601-1-11 Edition 2.0 2015-01, Medical electrical equipment Part 1-11 General requirements for basic
{7}------------------------------------------------
#### Traditional 510K for SA Series Ambulatory Blood Pressure Monitor
safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- (4) IEC 80601-2-30 Edition 1.1 2013-07 Medical electrical equipment Part 2-30 Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- (5) ISO 81060-2 Third edition 2018-11 Non-invasive sphygmomanometers Part 2: Clinical investigation of intermittent automated measurement type [including: Amendment 1 (2020)]
- (6) ISO 14971 Second edition 2007-03-01 Medical devices Application of risk management to medical devices The following biocompatibility standards are complied with on the subject device:
The following biocompatibility standards are complied with on the subject device:
(7) ISO 10993-1: 2018. ISO 10993-5:2009 and ISO 10993-10:2010
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
## Summary
The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate devices.
## 8. Conclusion
Verification and validation testing has been conducted on the SA series Ambulatory Blood Pressure Monitor. This premarket notification submission demonstrates that SA series Ambulatory Blood Pressure Monitor is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.