K230083 · Medovate Limited · FRN · Oct 31, 2023 · General Hospital
Device Facts
Record ID
K230083
Device Name
SAFIRA
Applicant
Medovate Limited
Product Code
FRN · General Hospital
Decision Date
Oct 31, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medovate SAFIRA system is intended for use by trained clinicians to administer local anaesthetic below a specified pressure threshold to a target nerve bundle for regional anaesthesia.
Device Story
SAFIRA is a battery-powered, motorized syringe driver for regional anesthesia procedures. It consists of a sterile, single-use 20mL syringe (polypropylene) and a reusable driver/operator. Clinicians fill the syringe with local anesthetic, lock it into the driver, and attach a needle set. The operator (foot pedal or palm-held) controls a DC motor that advances or retracts the syringe plunger. The device features an overpressure safety mechanism that stops plunger movement if pressure exceeds 17psi ±3psi. Used in clinical settings, it allows controlled infusion of anesthetic to target nerve bundles. The device benefits patients by providing a consistent, pressure-limited delivery of anesthetic, reducing risks associated with manual injection pressures.
Clinical Evidence
No clinical data. Evidence consists of bench-top system verification testing and a human factors study to validate design changes, including the new palm operator, NRFit syringe compatibility, and updated pressure limit. Performance met pre-determined acceptance criteria.
Indicated for trained clinicians to administer local anesthetic below a specified pressure threshold to a target nerve bundle for regional anesthesia.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 31, 2023
Medovate Limited % Pamela Papineau Delphi Medical Device Consulting, Inc. 5 Whitcomb Ave Ayer, Massachusetts 01432
Re: K230083
Trade/Device Name: Safira Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: FRN Dated: September 29, 2023 Received: September 29, 2023
Dear Pamela Papineau:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Jake K. Digitally signed by Jake K Lindstrom -S Lindstrom -S Date: 2023.10.31 16:23:58 -04'00"
Jake Lindstrom, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230083
Device Name SAFIRA
Indications for Use (Describe)
The Medovate SAFIRA system is intended for use by trained clinicians to administer local anaesthetic below a specified pressure threshold to a target nerve bundle for regional anaesthesia.
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------|
| <div> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### General Information 1
Owners Name
Contact Person 1:
Email address:
Email address:
Contact Person 2:
Telephone Number:
Telephone Number:
Medovate Limited The Workplace, Camboro Business Park Girton, Cambridge, CB3 0QH Alan Finnerty (Technology Director) +44 (0)1223 901991 alan.finnerty@medovate.co.uk Stuart Thomson (Managing Director) +44 (0)1223 901991 stuart.thomson@medovate.co.uk
| Date Prepared: | October 30, 2023 |
|-----------------------|------------------|
| Device Trade Name: | SAFIRA |
| Common/Usual Name: | Infusion Pump |
| Product Code: | FRN |
| Classification Name: | Infusion Pump |
| Device Regulation: | 21 CFR 880.5725 |
| Device Classification | Class II |
### Predicate Device:
Product Code:
Common/Usual Name:
Classification Name:
Device Classification
Device Regulation
Medovate Limited SAFIRA; K153599 Infusion Pump FRN Infusion Pump 21 CFR 880.5725 Class II
{5}------------------------------------------------
#### Indications for Use: 1.1
The Medovate SAFIRA system is intended for use by trained clinicians to administer local anaesthetic below a specified pressure threshold to a target nerve bundle for regional anaesthesia.
#### Device Description: 1.2
The Medovate SAFIRA device is an infiltration pump used to infuse local anesthetic drugs into the body as part of a Regional Anaesthesia procedure.
The SAFIRA device consists of a sterile, single-use plastic Syringe (20mL capacity) fitted with a custom collar and plunger, a non-sterile reusable Driver, and a non-sterile reusable Operator. SAFIRA is a battery powered motorized syringe driver for use in delivering anaesthetic via injection during a Peripheral Nerve Block (PNB) Procedure.
The user fills the syringe with the fluid of choice, then locks the filled Syringe into the Driver, where the specially designed Syringe Plunger rack mates with the Driver Gear. When the SAFIRA Syringe is attached to the Driver, the collar locks the Syringe to the Driver housing, and the Plunger engages with the Driver Gear to advance or retract the Syringe Plunger.
The female Luer or NRFit connection of a sterile, single-use needle set (not supplied by Medovate) is attached to the standard male Luer or NRFit fitting on the SAFIRA Syringe. The Operator (Foot Pedal or Palm Operator) connector is plugged into the receptacle on the Driver, ready for use.
The Driver contains non-replaceable AAA batteries which power a small DC motor which in turn controls the movement of the Syringe Plunger.
The Operator consists of 2 color-coded controls; when the user applies pressure to the appropriate control (green = infuse, yellow = aspirate), the motor activates the Plunger movement mechanism to move the Plunger forwards (infuse/inject) or backwards (aspirate).
The Driver is designed to deliver fluid at a maximum flow rate of 0.5mL/sec. The motor is designed with an overpressure safety feature, which causes the Plunger movement to stop if the pressure in the syringe exceeds 17psi ±3psi regardless of whether the "infuse" control of the operator is depressed.
{6}------------------------------------------------
#### Substantial Equivalence: 1.3
The proposed SAFIRA device described in this 510(k) is substantially equivalent to the predicate SAFIRA device cleared in 510(k) K153599.
Substantial equivalence has been demonstrated by following the FDA Guidance: "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]", and accompanying text in the guidance.
The following table demonstrates a summary of the Substantial Equivalence between the Predicate Unmodified SAFIRA device (cleared in 510(k) K153599), and the proposed SAFIRA device.
| Criteria | Predicate Device:<br>Unmodified SAFIRA<br>(As per 510(k) K153599) | Proposed Device:<br>SAFIRA<br>(K230083) | Comments |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory<br>Equivalence | | | |
| FDA<br>Classification<br>Name | Infusion Pump | Infusion Pump | Identical |
| Product Code | FRN | FRN | Identical |
| Regulation | 21 CFR 880.5725 | 21 CFR 880.5725 | Identical |
| Device Class | II | II | Identical |
| Classification<br>Panel | General Hospital Devices | General Hospital Devices | Identical |
| Common / Usual<br>Name | Infiltration Pump | Infiltration Pump | Identical |
| Manufacturer | Medovate Ltd | Medovate Ltd | Identical |
| Intended<br>Operator | Physician or another<br>qualified medical<br>professional | Physician or another<br>qualified medical<br>professional | Identical |
| Criteria | Predicate Device:<br>Unmodified SAFIRA | Proposed Device:<br>SAFIRA | Comments |
| | (As per 510(k) K153599) | (K230083) | |
| Intended Use | Syringe infusion pump<br>operated via a foot pedal<br>for the infusion of non-IV<br>fluids into the body. | Syringe infusion pump<br>operated via an operator<br>for the infusion of non-IV<br>fluids into the body. | Substantially Equivalent<br>The wording "Foot Pedal" has<br>been replaced by "Operator"<br>as the SAFIRA Palm<br>Operator has been added to<br>the proposed SAFIRA device.<br>The principle of operation,<br>indications for use and<br>intended purpose remain<br>unchanged. |
| Contraindicated<br>for IV Fluid<br>Delivery | Yes | Yes | Identical |
| Indications for<br>Use | The Concert Medical<br>Hands-Free Syringe is<br>indicated for general fluid<br>irrigation / infiltration. | The Medovate SAFIRA<br>System is intended for use<br>by trained clinicians to<br>administer local<br>anaesthetics below a<br>specified pressure<br>threshold to a target nerve<br>bundle for regional<br>anaesthesia. | Substantially Equivalent<br>The wording of the proposed<br>SAFIRA device has been<br>amended to be more specific<br>and provide an Indication for<br>Use which is in line with the<br>use of the device.<br>The use of the proposed<br>SAFIRA remains unchanged<br>from the predicate<br>unmodified SAFIRA device,<br>and both devices have the<br>same medical intention. |
| Criteria | Predicate Device:<br>Unmodified SAFIRA<br>(As per 510(k) K153599) | Proposed Device:<br>SAFIRA<br>(K230083) | Comments |
| Design and<br>Technology<br>Equivalence | | | |
| Sterile Device? | Sterile:<br>Syringe<br><br>Non-Sterile:<br>Driver<br>Operator | Sterile:<br>Syringes<br><br>Non-Sterile:<br>Driver<br>Operators | Identical |
| Single Use<br>Device? | Single Use:<br>Syringe<br><br>Reusable:<br>Driver<br>Operator | Single Use:<br>Syringes<br><br>Reusable:<br>Driver<br>Operators | Identical |
| Fundamental<br>Scientific<br>Technology | Non-IV fluids infused<br>into the body via a<br>needle/cannula through<br>the application of pressure<br>exerted on the syringe<br>plunger through motor-<br>driven motion. | Local anaesthetic fluids<br>infused into the body via<br>a needle/cannula through<br>the application of pressure<br>exerted on the syringe<br>plunger through motor-<br>driven motion. | Equivalent<br>SAFIRA has restricted its use<br>to Local Anesthetic in line<br>with its Indications for Use.<br>The Fundamental Technology<br>of the predicate unmodified<br>and proposed SAFIRA<br>devices remains unchanged. |
| Pump Type | Piston-Driven Syringe | Piston-Driven Syringe | Identical |
| Infusion Fluid<br>Reservoir | Syringe (20ml), filled by<br>user | Syringe (20ml), filled by<br>user | Identical |
| | Predicate Device:<br>Unmodified SAFIRA | Proposed Device:<br>SAFIRA | |
| Criteria | (As per 510(k) K153599) | (K230083) | Comments |
| Compatible<br>Needle Sets | B. Braun Stimuplex A<br>22G x 2"<br>B. Braun Ultra<br>22G x 3-1/8"<br>Pajunk Sonoflex Stim<br>22G x 50mm<br>Life-Tech EchoBright<br>22G x 50mm | Needle Gauge: >22G<br>Needle Length: <120mm | Minor Difference<br>The unmodified predicate<br>SAFIRA device limited its<br>use to 4 specific regional<br>nerve block needles.<br>The proposed SAFIRA device<br>limits use to a range of needle<br>sizes which include the<br>previously specified needles. |
| Power Source | Battery | Battery | Identical |
| Materials | | | |
| Fluid Path<br>Materials | Syringe Body:<br>Polypropylene | Syringe Body:<br>Polypropylene | Identical |
| | Syringe Body Lubricant:<br>Silicone oil | Syringe Body Lubricant:<br>Silicone oil | |
| | Seal:<br>Isoprene Synthetic rubber | Seal:<br>Isoprene Synthetic rubber | |
| | Note:<br>Materials are identical to<br>those contained in the<br>Dragon Heart Medical<br>Devices syringes cleared<br>in K042547 | Note:<br>Materials are identical to<br>those contained in the<br>Dragon Heart Medical<br>Devices syringes cleared<br>in K042547 | |
| Criteria | Predicate Device:<br>Unmodified SAFIRA<br>(As per 510(k) K153599) | Proposed Device:<br>SAFIRA<br>(K230083) | Comments |
| Performance<br>Equivalence | | | |
| Maximum<br>Infusion Flow<br>Rate | 0.5ml/sec (30mL/min)<br>using 22G needle set | 0.5ml/sec (30ml/min)<br>using compatible needle<br>set | Identical |
| Maximum<br>Infusion Pressure | 13psi (±2psi) | 17psi (±3psi) | Difference to predicate.<br>The pressure limit has been<br>increased, however there is no<br>impact on safety and no new<br>risks have been introduced. |
| User Interface<br>Equivalence | | | |
| User Interface | User fills the sterile<br>syringe with infusion<br>fluid, places the syringe in<br>the pump mechanism<br>(Driver), attaches the<br>sterile infusion tubing and<br>needle (not provided by<br>Medovate) and attaches<br>the Foot Pedal to the<br>pump mechanism.<br>User initiates fluid<br>infusion or aspiration by<br>pressing on the Foot<br>Pedal. | User fills the sterile<br>syringe with local<br>anaesthetic, places the<br>Syringe in the pump<br>mechanism (Driver),<br>attaches the sterile<br>infusion tubing and<br>needle (not provided by<br>Medovate) and attaches<br>the Operator to the pump<br>mechanism.<br>User initiates fluid<br>infusion or aspiration by<br>pressing on the Operator<br>control. | Difference to predicate.<br>'Foot Pedal' replaced by<br>'Operator' due to the<br>additional Palm Operator<br>variant which has been added<br>to the proposed SAFIRA<br>device.<br>The principle of operation<br>remains the same. |
| Criteria | Predicate Device:<br>Unmodified SAFIRA<br>(As per 510(k) K153599) | Proposed Device:<br>SAFIRA<br>(K230083) | Comments |
| Syringe<br>Connection | Luer Syringe<br>(ISO 80369-7) | Luer Syringe<br>(ISO 80369-7)<br>NRFit Syringe<br>(ISO 80369-6) | Minor Difference to Predicate<br><br>An NRFit Syringe has been<br>developed and added to the<br>proposed SAFIRA device.<br><br>The introduction of the NRFit<br>Syringe to the SAFIRA<br>device has not affected device<br>use and allows the user /<br>facility to choose a syringe to<br>match the needle type they<br>already use for regional<br>blocks. |
## Table 1: General Substantial Equivalence Comparison
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
#### Non-Clinical Performance Testing: 1.4
Performance data demonstrated that SAFIRA has met the pre-determined acceptance criteria and is substantially equivalent to the predicate SAFIRA device. The risks associated with the proposed SAFIRA device were found to be acceptable when evaluated in accordance with ISO 14971:2019.
Performance testing included in this 510(k) application consists of system verification testing to verify the changes made to the SAFIRA device, including the injection pressure limit, inclusion of the NRFit Syringe and Palm Operator. This 510(k) submission includes an Infusion Pump Safety Case and the results of a Human Factors Study to validate the above changes.
#### 1.5 Conclusion:
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The SAFIRA is substantially equivalent to the predicate unmodified SAFIRA cleared under K153599 with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.