The Concert Medical Hands-Free Syringe (HFS) is indicated for general fluid irrigation/infiltration.
Device Story
Infiltration pump for infusion of fluids (e.g., regional anesthetics); consists of sterile 20 mL single-use syringe, reusable motor housing, and foot pedal. User fills syringe, locks into motor housing; motor housing gear mates with syringe plunger rack. Foot pedal controls infusion (green) or aspiration (yellow) via DC motor. Max flow rate 0.5 mL/sec; overpressure safety feature stops plunger at 15 psi. LED indicators (green/yellow/red) provide real-time status. Used in clinical settings by healthcare professionals; enables two-handed needle placement without assistant for plunger manipulation. Benefits include hands-free control and pressure-limited infusion.
Clinical Evidence
Bench testing only. Performance testing included foot pedal/plunger control functionality, overpressure cutoff, flow rate measurement, reuse functionality, and electrical safety/EMC testing. Includes Infusion Pump Safety Case and Human Factors study. No clinical data presented.
Technological Characteristics
Battery-powered (non-replaceable AAA) DC motor infusion pump. Components: 20 mL single-use plastic syringe, reusable motor housing, foot pedal. Max flow rate 0.5 mL/sec. Overpressure safety limit: 15 psi. LED status indicators. Connectivity: None. Sterilization: Single-use components are sterile.
Indications for Use
Indicated for general fluid irrigation/infiltration in patients requiring infusion of fluids such as regional anesthetics.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Psi-Tec Syringe Infusion Pump and Accessories (K980738)
Macosta Medical B-Smart Nerve Block Injection Pressure Manometer (K031128)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a symbol that resembles a stylized caduceus or a representation of human profiles, suggesting health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 1, 2016
Concert Medical, LLC % Ms. Pamela Papineau, RAC Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, Massachusetts 01432
Re: K153599
Trade/Device Name: Concert Medical Hands-Free Syringe (HFS) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: FRN Dated: August 8, 2016 Received: August 9, 2016
Dear Ms. Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153599
#### Device Name
Concert Medical Hands-Free Syringe (HFS)
Indications for Use (Describe)
The Concert Medical Hands-Free Syringe (HFS) is indicated for general fluid irrigation/infiltration.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Appendix C – 510(k) Summary (revised)
# General Information
| Date Prepared: | August 26, 2016 |
|----------------------------|--------------------------------------------------------------------------------------------------|
| Owner's Name: | Concert Medical, LLC |
| Address: | 77 Accord Park Drive<br>Norwell, MA 02061 |
| Owner Contact Person: | Timothy S. Powers, V.P. Operations/Regulatory |
| Telephone Number: | 781-261-7407 |
| Regulatory Contact Person: | Pamela Papineau, RAC |
| Telephone Number: | 978-772-3552 |
| Subject Device Name: | Concert Medical Hands-Free Syringe (HFS) |
| Trade Name: | Concert Medical Hands-Free Syringe (HFS) |
| Common/Usual Name: | Infusion Pump |
| Product Code: | FRN |
| Classification Name: | Infusion Pump<br>21 CFR 880.5725; Class II |
| Predicate Device: | Psi-Tec Syringe Infusion Pump and Accessories |
| Trade Name: | Psi-Tec Syringe Infusion Pump and Accessories |
| Common/Usual Name: | Infusion Pump |
| Product Code: | FRN |
| Classification Name: | Infusion Pump<br>21 CFR 880.5725; Class II |
| Premarket Notification: | K980738; Byron Medical Psi-Tec Syringe Infusion Pump and<br>Accessories<br>SE date June 16, 1998 |
| Primary Reference Device: | KleinTouch Pump |
| Trade Name: | KleinTouch Pump |
| Common/Usual Name: | Infusion Pump |
| Product Code: | FRN |
| Classification Name: | Infusion Pump<br>21 CFR 880.5725; Class II |
| Premarket Notification: | K123822; HK Surgical KleinTouch Pump<br>SE date September 27, 2013 |
# Indications for Use
The Concert Medical Hands-Free Syringe is indicated for general fluid irrigation/infiltration.
{4}------------------------------------------------
# Device Description
The Concert Medical Hands-Free Syringe (HFS) is an infiltration pump used to infuse fluids (such as regional anesthetics) into the body. The HFS consists of a sterile, single-use plastic syringe (20 mL capacity) fitted with a custom plunger, a reusable motor housing, and a reusable foot pedal. The user fills the syringe with the fluid of choice, then locks the filled syringe into the motor housing, where the specially designed syringe plunger rack mates with the custom plunger housing gear. The male luer fitting of a sterile, single-use needle set (not supplied by Concert Medical) is attached to the standard female luer fitting on the HFS syringe. The foot pedal connector is plugged into the receptacle on the motor housing, then the foot pedal is placed in a convenient location on the floor. The motor housing contains non-replaceable AAA batteries, which power a small DC motor which in turn controls the movement of the syringe plunger. The foot pedal consists of a color-coded rubber bar; when the user applies light foot pressure to the appropriate half of the foot pedal (green = infuse; yellow = aspirate), the motor activates the plunger movement mechanism to move the plunger forward (infuse) or backward (aspirate). The motor control is designed to deliver fluid at a maximum flow rate of 0.5 mL/sec. The motor is designed with an overpressure safety feature, which causes the plunger movement to stop if the pressure in the syringe exceeds 15 psi regardless of whether the "infuse" portion of the foot pedal is depressed. The motor housing has three LED indicator lights for real-time display of the syringe plunger action: a green light illuminates when the syringe is infusing, a yellow light illuminates when the syringe is aspirating, and a red light illuminates if the plunger movement has stopped because the 15 psi pressure limit has been reached. The foot pedal control provides the user with two-handed control of the needle placement without the need for a second person to manipulate the syringe plunger for infusion or aspiration.
# Substantial Equivalence
The Concert Medical HFS is substantially equivalent to the Byron Medical Psi-Tec Syringe Infusion Pump and Accessories, which was cleared in K980738. The Psi-Tec Syringe Infusion Pump consists of a pneumatic-driven syringe pump used to irrigate or infuse fluids contained in a reservoir consisting of a sterile, disposable 10 mL piston syringe with a standard luer connector that is attached to a sterile, disposable tubing set for delivery of the fluid contained in the syringe. As for the Concert Medical Hands-Free Syringe, infusion/aspiration using the Psi-Tec Syringe Infusion Pump is controlled via a foot pedal. Flow rates for the Psi-Tec Syringe Infusion Pump range from 0 - 600 mL/min. The maximum infusion pressure of the Concert Medical Hands-Free Syringe (15 psi) is equivalent to that of the Psi-Tec Syringe Infusion Pump (14.8 psi). The Macosta Medical B-Smart Nerve Block Injection Pressure Manometer (K031128), which is used in conjunction with a standard infiltration anesthesia needle set to limit infusion pressure to a maximum of 15 psi, is cited as a reference device for the maximum 15 psi infusion pressure of the Concert Medical HFS.
The HK Surgical KleinTouch Pump, which was cleared in K123822, is cited as a reference device for the Concert Medical Hands-Free Syringe. The KleinTouch Pump is indicated for use as an infiltration pump used to cause a flow of fluid from an IV bag into a patient in a manner controlled manually by a health care professional. The KleinTouch Pump consists of a powered (battery or AC), reusable peristaltic roller pump that achieves fluid movement via force exerted
{5}------------------------------------------------
by three rollers on the plastic tubing connecting the fluid reservoir (IV bag) to the infusion needle. The KleinTouch Pump is used with the sterile, single-use KT20 KleinTouch Tubing. The user activates the KleinTouch Pump via a foot pedal. The KleinTouch Pump can be used with 12g - 20g needles; infusion flow rates of 50 - 1000 mL/min are available depending on the selected needle gauge.
This 510(k) includes a second reference device, the Macosta Medical B-Smart Nerve Block Injection Pressure Manometer (K031128); the B-Smart device is used in conjunction with a standard infiltration anesthesia needle set to limit infusion pressure to a maximum of 15 psi.
# Non-Clinical Performance Testing
Performance data demonstrated that the Concert Medical HFS has met pre-determined acceptance criteria and is substantially equivalent to the predicate device. The risks associated with the new device were found acceptable when evaluated in accordance with ISO 14971:2007. Performance testing included in this 510(k) consists of foot pedal / syringe plunger control functionality, overpressure cutoff functionality, maximum flow rate testing, flow rate measurement, limited reuse functionality testing of the motor control unit and foot pedal, and electrical safety/EMC testing. This submission also includes an Infusion Pump Safety Case and the results of a Human Factors study.
# Conclusion
The Concert Medical Hands-Free Syringe (HFS) meets all pre-determined acceptance testing criteria performed to confirm substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.