Celerity 5 HP Biological Indicator; Celerity 5 HP Challenge Pack
K223717 · STERIS Corporation · FRC · Jun 15, 2023 · General Hospital
Device Facts
Record ID
K223717
Device Name
Celerity 5 HP Biological Indicator; Celerity 5 HP Challenge Pack
Applicant
STERIS Corporation
Product Code
FRC · General Hospital
Decision Date
Jun 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
Celerity 5 HP Biological Indicator The Celerity 5 HP Biological Indicator (BI) is used for routine monitoring of the following sterilizer cycles: • Lumen, Non Lumen, Fast Non Lumen, Fast and Flexible Cycles of the V-PRO 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems • STERRAD® 100S Sterilizer (Default Cycle) • Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear® • Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear®. When used in conjunction with the Celerity® Incubator provides a fluorescent result within 5 minutes. Celerity 5 HP Challenge Pack The Celerity 5 HP Challenge Pack is intended for qualification testing of the V-PRO Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing. The Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included. The Challenge Pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
Device Story
Device consists of Geobacillus stearothermophilus spores and nutrient media in a plastic vial; used to monitor vapor-phased hydrogen peroxide sterilization cycles. After sterilization, user activates BI by puncturing cap to release media; incubated in Celerity Incubator at 55-60°C. Reporter enzyme produced by surviving spores reacts with fluorogenic substrate to produce fluorescent signal. Reader detects fluorescence within 5 minutes to confirm viability. Challenge pack version includes BI and chemical indicator in a pouch for system qualification testing. Used in hospitals to verify sterilization efficacy; provides rapid feedback to clinicians to ensure instruments are safe for patient use.
Clinical Evidence
No clinical data. Bench testing only. Studies included viable spore population verification, resistance characteristics (D-value, survival/kill times), growth inhibition, holding time assessment, signal generation, and simulated use testing in V-PRO and STERRAD sterilizers. Reduced incubation time (5-minute) was validated against 7-day growth results, demonstrating ≥97% correlation.
Technological Characteristics
Biological indicator containing G. stearothermophilus spores and nutrient media in a plastic vial. Sensing principle: enzymatic fluorescence detection. Energy source: Celerity Incubator (55-59°C). Connectivity: Standalone reader. Sterilization: Vapor-phased hydrogen peroxide. Spore population: 1.0–4.0 x 10^6 CFU/BI. Resistance: D-value ≥ 6 sec at 9.1 mg/L H2O2.
Indications for Use
Indicated for routine monitoring of specific V-PRO and STERRAD low-temperature hydrogen peroxide sterilization cycles and for qualification/requalification testing of V-PRO systems. Used by healthcare providers in clinical settings.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 15, 2023
STERIS Corporation Gregory Land Lead Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060
Re: K223717
Trade/Device Name: Celerity 5 HP Biological Indicator; Celerity 5 HP Challenge Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: May 19, 2023 Received: May 19, 2023
Dear Gregory Land:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Christopher K. Dugard -2
Clarence W. Murray, III, PhD Deputy Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K223717
Device Name
Celerity 5 HP Biological Indicator;
Celerity 5 HP Challenge Pack
Indications for Use (Describe)
Celerity 5 HP Biological Indicator
The Celerity 5 HP Biological Indicator (BI) is used for routine monitoring of the following sterilizer cycles:
• Lumen, Non Lumen, Fast Non Lumen, Fast and Flexible Cycles of the V-PRO 1, 1 Plus, maX
2, 60 and s2 Low Temperature Sterilization Systems
- · STERRAD® 100S Sterilizer (Default Cycle)
• Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear®
· Standard. FLEX. Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear®.
When used in conjunction with the Celerity® Incubator, the Incubator provides a fluorescent result within 5 minutes.
Celerity 5 HP Challenge Pack
The Celerity 5 HP Challenge Pack is intended for qualification testing of the V-PRO Low
Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing.
The Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The Challenge Pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary For Celerity 5 HP Biological Indicator And Celerity 5 HP Challenge Pack
# Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax:
# Manufacturing Facility
STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax:
# Contact
Gregory Land Lead Regulatory Affairs Specialist Phone: (440) 392-7424 Email: greg land@steris.com
Submission Date: June 13, 2023
K Number: K223717
{4}------------------------------------------------
#### 1 Device Name
### Celerity 5 HP Biological Indicator
| Trade Name: | Celerity 5 HP Biological Indicator |
|------------------------|-----------------------------------------------------------------------|
| Common/usual Name: | Biological Indicator |
| Device Classification: | Class II |
| Classification Name: | Indicator, Biological Sterilization Process [21 CFR 880.2800(a), FRC] |
#### Celerity 5 HP Challenge Pack
| Trade Name: | Celerity 5 HP Challenge Pack |
|------------------------|-----------------------------------------------------------------------|
| Common/usual Name: | Biological Indicator |
| Device Classification: | Class II |
| Classification Name: | Indicator, Biological Sterilization Process [21 CFR 880.2800(a), FRC] |
#### Predicate Device 2
### Celerity 5 HP Biological Indicator
| Proprietary Name | Celerity 20 HP Biological Indicator |
|-------------------------|---------------------------------------------|
| Common/usual Name | Biological indicator |
| Classification Name: | Indicator, Biological Sterilization Process |
| 510(k) Submitter/Holder | STERIS Corporation |
| 510(k) Number: | K183294 |
### Celerity 5 HP Challenge Pack
| Proprietary Name | Celerity 20 HP Biological Indicator |
|-------------------------|---------------------------------------------|
| Common/usual Name | Biological indicator |
| Classification Name: | Indicator, Biological Sterilization Process |
| 510(k) Submitter/Holder | STERIS Corporation |
| 510(k) Number: | K183294 |
#### Description of Device 3
### Celerity 5 HP Biological Indicator
The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indications for use. It produces an optical change (signal) that is detected by the STERIS proprietary reader, STERIS Celerity Incubator, within 5 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.
#### Celerity 5 HP Challenge Pack
The CELERITY 5 HP Challenge Pack (pack), is used by healthcare providers for qualification testing of the V-PRO Low Temperature Sterilization Systems. The pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The user places the pack into the V-PRO Sterilizer and performs a sterilization cycle. After cycle completion, the VERIFY HPU Chemical Indicator (CI) and the CELERITY 5 HP Biological
{5}------------------------------------------------
Indicator (BI) contained in the pack are retrieved. The CI is assessed for a passing color change immediately and the BI can either be immediately activated or it can be held at room temperature for a maximum of 2 hours prior to activation.
The BI is activated by sealing the vial and thus puncturing the cap to release the contained media.
The activated BI is incubated at 55-60 ℃ in the CELERITY Incubator for a final determination of viability within 5 minutes of incubation.
#### 4 Intended Use/Indications for Use
### Celerity 5 HP Biological Indicator
The Celerity 5 HP Biological Indicator (BI) is used for routine monitoring of the following sterilizer cycles:
- Lumen, Non Lumen, Fast Non Lumen, Fast and Flexible Cycles of the V-PRO 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems
- . STERRAD® 100S Sterilizer (Default Cycle)
- Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear®
- Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear®.
When used in conjunction with the Celerity® Incubator provides a fluorescent result within 5 minutes.
### Celerity 5 HP Challenge Pack
The Celerity 5 HP Challenge Pack is intended for qualification testing of the V-PRO Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing.
The Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The Challenge Pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
#### റ Summary of Technical Characteristics
#### Table 1. Biological Indicator Subject Device Comparison to the Predicate Device
| Feature | CELRITY 5 HP Biological<br>Indicator<br>(proposed) | Celerity 20 HP Biological<br>Indicator (K183294) | Comparison |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Intended Use | The Celerity 5 HP Biological<br>Indicator (BI) is used for<br>routine monitoring of the<br>following sterilizer cycles:<br>• Lumen, Non Lumen, Fast<br>Non Lumen, Fast and | The Celerity 20 HP<br>Biological Indicator is<br>intended for routine<br>monitoring of the following<br>sterilizer cycles: | Proposed device has a<br>5-minute read time |
| Feature | CELRITY 5 HP Biological<br>Indicator<br>(proposed) | Celerity 20 HP Biological<br>Indicator (K183294) | Comparison |
| | Flexible Cycles of the V-<br>PRO 1, 1 Plus, maX, maX<br>2, 60 and s2 Low<br>Temperature Sterilization<br>Systems<br>• STERRAD® 100S<br>Sterilizer (Default Cycle)<br>• Standard and Advanced<br>Cycles of the STERRAD®<br>NX® Sterilizer with or<br>without ALLClear®<br>• Standard, FLEX, Express<br>and DUO Cycles of the<br>STERRAD® 100NX®<br>Sterilizer with or without<br>ALLClear®.<br>When used in conjunction<br>with the Celerity® Incubator,<br>the Incubator provides a<br>fluorescent result within 5<br>minutes. | • Lumen, Non Lumen, Fast<br>Non Lumen, Fast and<br>Flexible Cycles of the V-<br>PRO® Low Temperature<br>Sterilization Systems<br>• STERRAD® 100S<br>Sterilizer (Default Cycle)<br>• Standard and Advanced<br>Cycles of the STERRAD®<br>NX Sterilizer with or<br>without ALLClear<br>• Standard, Flex Scope,<br>Express and DUO Cycles<br>of the STERRAD® 100NX<br>Sterilizer with or without<br>ALLClear<br>When used in conjunction<br>with the Celerity HP<br>Incubator, the Celerity 20 HP<br>BI provides a fluorescent<br>result within 20 minutes. | |
| Indicator<br>organism | Geobacillus<br>stearothermophilus | Geobacillus<br>stearothermophilus | Same |
| Mechanism of<br>action | An enzyme, which is produced<br>by the organism, reacts with a<br>fluorogenic substrate within<br>the defined nutrient media to<br>produce a fluorescent moiety. | An enzyme, which is<br>produced by the organism,<br>reacts with a fluorogenic<br>substrate within the defined<br>nutrient media to produce a<br>fluorescent moiety. | Same |
| Accessories | Automated incubator / reader | Automated incubator /<br>reader | Same |
| Viable spore<br>population | 1.0 – 4.0 x 106 spore/BI | 1.0 – 4.0 x 106 spore/BI | Same |
| Resistance | Resistance @ 9.1 mg/L H2O2:<br>• D-value ≥ 6 sec<br>• Survival Time ≥ 4 sec<br>• Kill Time ≤ 7 min | Resistance @ 9.1 mg/L H2O2:<br>• D-value ≥ 6 sec<br>• Survival Time ≥ 4 sec<br>• Kill Time ≤ 7 min | Same |
| Culture<br>Conditions | 55-59°C, media included in BI,<br>5-minute incubation time. | 55-59°C, media included in<br>BI, 20-minute incubation<br>time. | Similar |
| Feature | CELRITY 5 HP Biological<br>Indicator<br>(proposed) | Celerity 20 HP Biological<br>Indicator (K183294) | Comparison |
| Primary Packaging | Direct inoculum on plastic<br>vial, cap with recovery media. | Direct inoculum on plastic<br>vial, cap with recovery<br>media. | Same |
| Process indicator | VERIFY V-PRO Chemical<br>Indicator (K140515); magenta<br>to yellow color change. | VERIFY V-PRO Chemical<br>Indicator (K140515);<br>magenta to yellow color<br>change. | Same |
| Shelf-life | 8 months | 10 months | Similar |
{6}------------------------------------------------
{7}------------------------------------------------
# Table 2: Challenge Pack Comparison to Predicate Device
| Feature | CELERITY 5 HP Challenge pack<br>Proposed | CELERITY 20 HP Challenge pack<br>(K173488) Predicate | Comparison |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended Use<br>/ Indication<br>for Use | The CELERITY 5 HP Challenge<br>Pack is intended for<br>qualification testing of the V-<br>PRO Low<br>Temperature Sterilization<br>System following installation,<br>relocation, malfunctions or<br>major repairs and for routine<br>requalification testing.<br><br>The Challenge Pack is placed<br>in an otherwise empty<br>sterilizer chamber; a hospital-<br>defined challenge load is<br>not included.<br><br>The challenge pack is not<br>intended for routine<br>monitoring of the V-PRO<br>Sterilizers. It has been tested<br>and validated solely for use in<br>periodic testing of the<br>sterilizers. | The CELERITY 20 HP<br>Challenge Pack is intended<br>for qualification testing of the<br>V-PRO Low<br>Temperature Sterilization<br>System following installation,<br>relocation, malfunctions or<br>major repairs and for routine<br>requalification testing.<br><br>The Challenge Pack is placed<br>in an otherwise empty<br>sterilizer chamber; a hospital-<br>defined challenge load is<br>not included.<br><br>The challenge pack is not<br>intended for routine<br>monitoring of the V-PRO<br>Sterilizers. It has been tested<br>and validated solely for use in<br>periodic testing of the<br>sterilizers. | Same |
| General<br>Design | Sealed sterilization pouch<br>containing SCBI, CI and barrier<br>material. | Sealed sterilization pouch<br>containing SCBI, CI and<br>barrier material. | Same |
| Biological<br>Indicator | Celerity 5 HP Biological<br>Indicator | Celerity 20 HP Biological<br>Indicator | Similar |
{8}------------------------------------------------
| Feature | CELERITY 5 HP Challenge pack<br>Proposed | CELERITY 20 HP Challenge pack<br>(K173488) Predicate | Comparison |
|------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------|
| Chemical<br>Indicator | Celerity HP Chemical Indicator | VERIFY HPU Chemical<br>Indicator | Same - brand<br>name has<br>changed |
| Means to<br>distinguish<br>processed<br>pack from<br>unprocessed | Proposed device's internal<br>indicator is visible through the<br>pack. | Proposed device's internal<br>indicator is visible through<br>the pack. | Same |
| Required<br>accessories | CELERITY Incubator (K223715) | CELERITY HP Incubator*<br>(K171587)<br>Celerity Incubator (K213881) | Similar |
#### 6 Summary of nonclinical Tests
Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 3 below.
| Test | Description | Acceptance Criteria | Conclusion |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Viable Spore Population | Study to demonstrate that<br>the product meets its<br>specification for the spore<br>population. | Average population for each<br>lot of 1.0 x 106 - 4.0 x 106<br>CFU/BI | Pass |
| Resistance Characteristics<br>Study | Study to verify that the<br>Celerity 5 Biological<br>Indicator meets the<br>acceptance criteria for<br>resistance. Acceptance<br>criteria are designed to<br>demonstrate comparable or<br>greater resistance values<br>than the predicate device | D-Value: 2-20 secondsAll Survive: ≤ 4 secondsAll Kill: ≥ 6 minutes | Pass |
| Carrier and Pack Material<br>Growth Inhibition<br>Evaluation | Study to demonstrate that<br>the primary packaging<br>material used in the<br>manufacture of the<br>Biological Indicator (BI)<br>does not hinder, or<br>otherwise affect, the<br>outgrowth of a low<br>number of Geobacillus<br>stearothermophilus spores | All BIs inoculated with<br>Geobacillus<br>stearothermophilus spores<br>demonstrate growth | Pass |
| Holding Time Assessment | Study to demonstrate the<br>labeled hold time of the BI | 120 minutes | Pass |
| Signal Generation | Study to demonstrate<br>the PBS media | All BIs will demonstrate a<br>positive fluorescent signal | Pass |
| | contained in the cap of<br>the Celerity 5 HP<br>Biological Indicator is<br>able to generate a<br>detectable fluorescent<br>signal from viable G.<br>stearothermophilus<br>organisms. | | |
| Simulated Use Testing – V-<br>PRO | Study to demonstrate<br>that the Project Blayze<br>Biological Indicator<br>(BI) for Gaseous<br>Hydrogen Peroxide<br>Sterilization Processes<br>functions<br>appropriately when<br>processed in the<br>claimed sterilizer<br>cycles with a validated<br>load. | All BIs processed in a full<br>cycle will demonstrate a<br>negative signal<br>corresponding to a no-<br>growth result upon seven<br>day incubation. | Pass |
| Simulated Use Testing –<br>STERRAD | Study to demonstrate<br>the Project Blayze<br>Biological Indicator<br>(BI) for Gaseous<br>Hydrogen Peroxide<br>Sterilization Processes<br>functions<br>appropriately when<br>processed in<br>STERRAD®<br>Sterilizers with a<br>validated load. | All BIs processed in a<br>full cycle will<br>demonstrate a negative<br>signal corresponding to a<br>no-growth result upon<br>seven day incubation. | Pass |
| Population after ALLClear<br>Exposure | Study to demonstrate<br>ALLClear<br>preconditioning does<br>not impact BI<br>performance. | Each BI signal evaluation was<br>positive following exposure<br>to the All-Clear precondition<br>treatment<br>The mean population of each<br>BI exposed to the All-Clear<br>precondition treatment was<br>within 50-300% of the<br>unexposed control mean<br>population. | Pass |
| Reduced Incubation Time - | Study to verify Reduced<br>Incubation Time of the BI | Study will demonstrate that<br>97% or greater of the positive | Pass |
| | | samples at seven days also gave a positive signal at the claimed RIT time | |
| Reduced Incubation Time –<br>Engineering Mode | Study to verify the<br>Reduced Incubation Time<br>of 5-minute correlates to a<br>20-minute result like the<br>predicate device and 7 day<br>growth. | study will demonstrate that<br>97% or greater of the positive<br>samples at seven days also Pass | |
| Celerity 5 HP Challenge Pack | | | |
| Simulated Use Testing | Study to demonstrate<br>performance of challenge<br>pack under worst-case<br>conditions. | Demonstrate the pack shows<br>passing results in worst-case<br>load under worst-case<br>sterilization conditions | Pass |
| Dose Response Testing | Study to demonstrate the<br>challenge pack provides<br>adequate challenge to<br>perform intended use. | The pack provides an<br>equivalent or greater<br>challenge to the claimed cycle<br>than the biological model<br>under worst-case conditions. | Pass |
Table 3. Summary of Non-clinical Testing
{9}------------------------------------------------
{10}------------------------------------------------
#### 7 Conclusion
Based on the intended use, technological characteristics and the nonclinical tests performed, the subject Celerity 5 HP Biological Indicator device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate device, the CELERITY 20 HP Biological Indicator, K183294 Class II [21 CFR 880.2800(a)], product code FRC.
Based on the intended use, technological characteristics and the nonclinical tests performed, the subject Celerity 5 HP Challenge Pack device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate device, the CELERITY 20 HP Challenge Pack, K183294 Class II [21 CFR 880.2800(a)], product code FRC.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.