The SpiroSphere is a compact device to measure inspiratory and expiratory lung function parameters in adults and children aged 4 years and older. It can be used by physicians in the office or hospital. The SpiroSphere ECG is a compact device to measure inspiratory and expiratory lung function parameters in adults and children aged 4 years and older. With the option ECG, electrocardiographic measurements can be made under resting conditions. For this purpose a 12-channel surface electrocardiogram can be measured and recorded. The acquired ECG can be displayed on the screen or printed on paper. The interpretation software is intended to support the physician in evaluation of the resting ECG in terms of morphology and rhythm. Automatic interpretation of the ECG is not possible for pediatric subjects with an age below 16 years and for pacemaker subjects. It is not intended for intra-cardial use. A qualified physician has to reassess all ECG measurements. An interpretation by SpiroSphere ECG is only significant if it is considered in connection with other clinical findings. ECG interpretation statements made by the SpiroSphere ECG represent partial qualitative and quantitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements. The SpiroSphere ECG is not intended for use in an EMS environment (Emergency Medical Services Environment). The minimum age for ECG application is 4 years. It can be used by physicians in the office or hospital. With the CardioSphere, electrocardiographic measurements can be made under resting conditions. For this purpose a 12-channel surface electrocardiogram can be measured and recorded. The acquired ECG can be displayed on the screen or printed on paper. The interpretation software is intended to support the physician in evaluation of the resting ECG in terms of morphology and rhythm. Automatic interpretation of the ECG is not possible for pediatric subjects with an age below 16 years and for pacemaker subjects. It is not intended for intra-cardial use. A qualified physician has to reassess all ECG measurements. An interpretation by CardioSphere is only significant if it is considered in connection with other clinical findings. ECG interpretation statements made by the CardioSphere represent partial qualitative and quantitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements. The CardioSphere is not intended for use in an EMS environment (Emergency Medical Services Environment). The minimum age for ECG application is 4 years. It can be used by physicians in the office or hospital.
Device Story
SpiroSphere/CardioSphere are compact devices for pulmonary function testing and 12-channel resting ECG. Spirometry uses pneumotach transducer for flow/volume measurement. ECG option uses wireless amplifier (Bluetooth) to acquire surface signals. Devices used in office/hospital by physicians. Output includes lung function parameters (FEV1, FVC, etc.) and ECG morphology/rhythm interpretation. Interpretation software supports physician evaluation; requires qualified physician reassessment. Provides partial qualitative/quantitative cardiovascular data; not for therapy/drug administration decisions. Benefits include integrated diagnostic assessment of pulmonary and cardiovascular status.
Clinical Evidence
Bench testing only. No clinical data. Performance testing confirmed spirometry accuracy/precision per ATS/ERS 2005 standards. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. Software verification/validation conducted per IEC 62304.
Technological Characteristics
Pneumotachograph (Lilly type) for spirometry; 12-channel surface ECG amplifier. Connectivity: Bluetooth 2.1+EDR, USB, Wifi, 3G, Ethernet. Power: Li-Ion battery or power supply (Main Unit); AA/NiMH battery (ECG amplifier). Dimensions: 8.0 x 9.3 x 2.1 cm. Software: Hannover ECG System (HES) for interpretation. Standards: IEC 60601-1, IEC 60601-2-25, IEC 60601-1-2, ISO 14971, IEC 62304, IEC 62366-1.
Indications for Use
Indicated for measurement of inspiratory and expiratory lung function parameters in adults and children aged 4 years and older. ECG option indicated for 12-channel surface ECG measurement under resting conditions in patients aged 4 years and older. Contraindicated for intra-cardial use; automatic interpretation not for pediatric patients <16 years or pacemaker patients.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 28, 2024
eResearchTechnology GmbH % Prithul Bom Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K223629
Trade/Device Name: SpiroSphere, SpiroSphereECG, CardioSphere Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: Class II Product Code: BTY, DPS Dated: June 5, 2023 Received: June 5, 2023
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223629
Device Name SpiroSphere
Indications for Use (Describe)
The SpiroSphere is a compact device inspiratory and expiratory lung function parameters in adults and children aged 4 years and older.
It can be used by physicians in the office or hospital.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Beneficial Use (Part 31, GER 331, Subpart D) </div> |
|--------------------------------------------------------------------------------|
| <div> <span></span>On-Site Corrective Action (31 GER 331, Subpart G) </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223629
Device Name SpiroSphere ECG
### Indications for Use (Describe)
The SpiroSphere is a compact device inspiratory and expiratory lung function parameters in adults and children aged 4 years and older.
With the option ECG, electrocardio measurements can be made under resting conditions. For this purpose a 12channel surface electrocardiogram can be measured and recorded. The acquired ECG can be displayed on the screen or printed on paper. The interpretation software is intended to support the physician in evaluation of the resting ECG in terms of morphology and rhythm. Automatic interpretation of the ECG is not possible for pediatric subjects with an age below 16 years and for pacemaker subjects. It is not intended for intra-cardial use.
A qualified physician has to reassess all ECG measurements. An interpretation by SpiroSphere ECG is only significant if it is considered in connection with other clinical findings. ECG interpretation statements made by the SpiroSphere ECG represent partial qualitative and quantitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements. The SpiroSphere ECG is not intended for use in an EMS environment (Emergency Medical Services Environment).
The minimum age for ECG application is 4 years.
It can be used by physicians in the office or hospital.
Type of Use (Select one or both, as applicable)
| <span> <b> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> |
|----------------------------------------------------------------------------------------------------------------------|
| <span> <b> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223629
Device Name CardioSphere
### Indications for Use (Describe)
With the CardioSphere, electrocardio measurements can be made under resting conditions. For this purpose a 12channel surface electrocardiogram can be measured and recorded. The acquired ECG can be displayed on the screen or printed on paper. The interpretation software is intended to support the physician in evaluation of the resting ECG in terms of morphology and rhythm. Automatic interpretation of the ECG is not possible for pediatric subjects with an age below 16 years and for pacemaker subjects. It is not intended for intra-cardial use.
A qualified physician has to reasurements. An interpretation by CardioSphere is only significant if it is considered in connection with other clinical findings. ECG interpretation statements made by the CardioSphere represent partial qualitative and quantitative information on the patient's cardition and no therapy or drugs can be administered based solely on the interpretation statements. The CardioSphere is not intended for use in an EMS environment (Emergency Medical Services Environment).
The minimum age for ECG application is 4 years.
It can be used by physicians in the office or hospital.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"></span> | |
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for ERT. The logo consists of a circular graphic made up of three overlapping shapes in the colors green, light blue, and dark blue. To the right of the graphic is the text "ERT" in a dark blue sans-serif font, with the registered trademark symbol above the "T". Below the text "ERT" is the tagline "Confidence at every turn" in a smaller, lighter font.
# GENERAL INFORMATION
#### 1 Type of Submission
Traditional 510(k) Submission Date 510(k) summary prepared: 21APR2023
#### 2 Submitter
eResearchTechnology GmbH Name: Address: Sieboldstrasse3 D-97230 Estenfeld Germany
### Contact person in Germany: (Official Correspondent) Address:
## Kristin Feld
eResearchTechnology GmbH Sieboldstrasse 3, 97230 Estenfeld Germany
+49 9305 720 6377 +49 9305 720 70 kristin.feld@clario.com
Contact person in the U.S .: Todd Kisner
Address
Phone:
FAX:
E-mail
eResearchTechnology 1818 Market St Suite 2600 Philadelphia, PA 19103
Phone/Fax: E-mail
+1 609 779 5935 todd.kisner@clario.com
#### 3 Establishment Registration Number
3008505660
#### 4 Common Name or Classification Name
Predictive pulmonary-function value calculator (CFR 868.1890, Product Code BTY) Electrocardiograph (CFR 870.2340, Product Code DPS)
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for ERT. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circle is divided into three sections, colored light green, light blue, and dark blue. Below the text "ERT" is the tagline "Confidence at every turn."
- 5 Trade Names SpiroSphere SpiroSphere ECG CardioSphere
- 6 Device Classification Class II
#### 7 Classification Panel
73 Anaesthesiology Part 868 Code BTY (primary) 74 Cardiovascular Part 870 Code DPS (secondary)
#### 8 Reason for Premarket Notification
New option (ECG) to a cleared device (SpiroSphere K173937)
#### 9 Legally Marketed Predicate and Reference Devices
Predicate Device: K202754 Code BTY, DPS Masterscope, Masterscope ECG, Masterscope CT, Masterscope WSSU
Reference Devices:
K173937 Code BTY SpiroSphere
K183369 Code DPS COR12
#### 10 Predicate Device Company
eResearchTechnology GmbH
## Reference Device Companies
eResearchTechnology GmbH (SpiroSphere) Corscience GmbH & Co. KG (COR12)
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circular graphic is divided into three sections, colored light green, dark blue, and light blue. Below the text "ERT" is the tagline "Confidence at every turn" in a smaller font.
#### Device Description 11
### SpiroSphere
SpiroSphere is a compact spirometry device. Its Sensor Unit is batterypowered. The Main Unit can be powered by battery or power supply. The SpiroSphere / SpiroSphere ECG is used to measure inspiratory and expiratory lung function parameters in adults and children 4 years and older. The measured data is saved to the device and can be read out at any time.
# SpiroSphere ECG
SpiroSphere ECG is a compact spirometry device. Its Sensor Unit is battery-powered. The Main Unit can be powered by battery or power supply. The SpiroSphere / SpiroSphere ECG is used to measure inspiratory and expiratory lung function parameters in adults and children 4 years and older. The measured data is saved to the device and can be read out at any time.
With the ECG option (subject of this 510(k)), 12-channel surface electrocardiogram can be measured and recorded.
### CardioSphere
The Main Unit can be powered by battery or power supply. The Main Unit is wirelessly connected to an ECG amplifier via Bluetooth.
With the ECG option (subject of this 510(k)), 12-channel surface electrocardiogram can be measured and recorded.
A printer can be connected and data (e.g. reports, screenshots) can be printed. Moreover, it is possible to transfer data by USB. Wifi. 3G. and Ethernet connections.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for ERT. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circular graphic is divided into three sections, with the top section being light green and the bottom two sections being dark blue. Below the text "ERT" is the tagline "Confidence at every turn".
Pulmonary function assessments
- Slow Spirometry .
- Forced Spirometry .
- Flow-Volume loop and Volume-Time tracing, pre/post tests ●
- Trending capabilities ●
Cardiovascular assessments
- 12 Lead Electrocardiogram ●
#### 12 Intended Use/Indications for Use - SpiroSphere
The SpiroSphere is a compact device to measure inspiratory and expiratory lung function parameters in adults and children aged 4 years and older.
It can be used by physicians in the office or hospital.
#### 13 Intended Use/Indications for Use - SpiroSphere ECG
The SpiroSphere ECG is a compact device to measure inspiratory and expiratory lung function parameters in adults and children aged 4 years and older.
With the option ECG, electrocardiographic measurements can be made under resting conditions. For this purpose a 12-channel surface electrocardiogram can be measured and recorded. The acquired ECG can be displayed on the screen or printed on paper. The interpretation software is intended to support the physician in evaluation of the resting ECG in terms of morphology and rhythm. Automatic interpretation of the ECG is not possible for pediatric subjects with an age below 16 years and for pacemaker subjects. It is not intended for intra-cardial use.
A qualified physician has to reassess all ECG measurements. An interpretation by SpiroSphere ECG is only significant if it is considered in connection with other clinical findings. ECG interpretation statements made by the SpiroSphere ECG represent partial qualitative and quantitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation
{10}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for ERT. The logo consists of a green and blue abstract shape on the left, followed by the letters "ERT" in blue. Below the letters, the text "Confidence at every turn" is written in a smaller font.
statements. The SpiroSphere ECG is not intended for use in an EMS environment (Emergency Medical Services Environment).
The minimum age for ECG application is 4 years.
lt can be used by physicians in the office or hospital.
#### 14 Intended Use/Indications for Use - CardioSphere
With the CardioSphere, electrocardiographic measurements can be made under resting conditions. For this purpose a 12-channel surface electrocardiogram can be measured and recorded. The acquired ECG can be displayed on the screen or printed on paper. The interpretation software is intended to support the physician in evaluation of the resting ECG in terms of morphology and rhythm. Automatic interpretation of the ECG is not possible for pediatric subjects with an age below 16 years and for pacemaker subjects. It is not intended for intra-cardial use.
A qualified physician has to reassess all ECG measurements. An interpretation by CardioSphere is only significant if it is considered in connection with other clinical findings. ECG interpretation statements made by the CardioSphere represent partial qualitative and quantitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements. The CardioSphere is not intended for use in an EMS environment (Emergency Medical Services Environment).
The minimum age for ECG application is 4 years.
It can be used by physicians in the office or hospital.
#### 15 Required Components
Power Supply (medical grade) Instructions for Use
SpiroSphere and SpiroSphere ECG: Pneumotach with Mouthpiece (ERT PT) Nose Clips
SpiroSphere ECG and CardioSphere: ECG Unit/amplifier ECG electrodes
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circular graphic is divided into three sections, colored light green, light blue, and dark blue. Below the text "ERT" is the tagline "Confidence at every turn" in a smaller font.
# 16 Summary Table of Comparison
Comparison tables with Predicate and Reference devices: MasterScope (K202754), COR12 (K183369), and SpiroSphere (K173937)
| Pulmonary / Spirometry Function Comparison | | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device<br>MasterScope (K202754) | Reference Device<br>SpiroSphere (K173937) | Subject Device | |
| | | | SpiroSphere ECG | SpiroSphere |
| Intended Use /<br>Indications for Use | MasterScope is a medical<br>device to measure inspiratory<br>and expiratory lung function<br>parameters.<br>With the option ECG a 12-<br>channel surface<br>electrocardiogram (ECG) can<br>be measured and recorded. It is<br>not intended for intracardial use.<br>Automatic interpretation of the<br>ECG is not possible for pediatric<br>patients with an age below 16<br>years and for pacemaker<br>patients.<br>A qualified physician has to<br>reassess all<br>MasterScope/MasterScope<br>ECG measurements. An<br>interpretation by<br>MasterScope/MasterScope<br>ECG is only significant if it is<br>considered in connection with<br>other clinical findings.<br>ECG interpretation statements<br>made by the | The SpiroSphere is a<br>compact device to measure<br>inspiratory and expiratory<br>lung function parameters in<br>adults and children aged 4<br>years and older.<br>It can be used by physicians<br>in the office or hospital. | The SpiroSphere ECG is a<br>compact device to<br>measure inspiratory and<br>expiratory lung function<br>parameters in adults and<br>children aged 4 years and<br>older.<br>With the option ECG,<br>electrocardiographic<br>measurements can be<br>made under resting<br>conditions. For this<br>purpose a 12-channel<br>surface electrocardiogram<br>can be measured and<br>recorded. The acquired<br>ECG can be displayed on<br>the screen or printed on<br>paper. The interpretation<br>software is intended to<br>support the physician in<br>evaluation of the resting<br>ECG in terms of<br>morphology and rhythm.<br>Automatic interpretation of<br>the ECG is not possible for | The SpiroSphere is a<br>compact device to<br>measure inspiratory and<br>expiratory lung function<br>parameters in adults<br>and children aged 4<br>years and older.<br>It can be used by<br>physicians in the office<br>or hospital. |
| MasterScope/MasterScope<br>ECG represent partial<br>quantitative information on the<br>patient's cardiovascular<br>conditions and no therapy or<br>drugs can be administered<br>based solely on the<br>interpretation statements.<br>It can be used by physicians in<br>the office or hospital.<br>The MasterScope spirometry<br>and ECG application is intended<br>to measure adults and children<br>aged 4 years and older. The<br>patients must be able to<br>understand and perform<br>instructions of the physician. | pediatric subjects with an<br>age below 16 years and<br>for pacemaker subjects. It<br>is not intended for intra-<br>cardial use.<br>A qualified physician has<br>to reassess all ECG<br>measurements. An<br>interpretation by<br>SpiroSphere ECG is only<br>significant if it is<br>considered in connection<br>with other clinical findings.<br>ECG interpretation<br>statements made by the<br>SpiroSphere ECG<br>represent partial<br>qualitative and quantitative<br>information on the<br>patient's cardiovascular<br>condition and no therapy<br>or drugs can be<br>administered based solely<br>on the interpretation<br>statements. The<br>SpiroSphere ECG is not<br>intended for use in an<br>EMS environment<br>(Emergency Medical<br>Services Environment).<br>The minimum age for ECG<br>application is 4 years.<br>It can be used by<br>physicians in the office or<br>hospital. | Identical | Identical | |
| Patient population | 4 years and older | 4 years and older | 4 years and older | |
| Operation principle | Measurement of inspiratory and<br>expiratory flows and volumes<br>with pneumotach transducer<br>Calculation of lung function<br>parameters<br>Results are displayed and<br>stored, they can be printed and<br>exported | Measurement of inspiratory<br>and expiratory flows and<br>volumes with pneumotach<br>transducer<br>Calculation of lung function<br>parameters<br>Results are displayed and<br>stored, they can be printed<br>and exported | Identical | Identical |
| Measurements | FEV1, FVC, PEF, FEF25-75,<br>VC, IC, ERV and others acc. to<br>ATS* | FEV1, FVC, PEF, FEF25-<br>75, VC, IC, ERV and others<br>acc. to ATS* | Identical | Identical |
| *ATS conformity<br>(criteria) | 2005 ATS/ERS Spirometry<br>Standards | 2005 ATS/ERS Spirometry<br>Standards | Identical | Identical |
| Fundamental<br>scientific technology | Pneumotachograph, pressure to<br>flow conversion technique<br>(Lilly Type Pneumotachograph) | Pneumotachograph,<br>pressure to flow conversion<br>technique<br>(Lilly Type<br>Pneumotachograph) | Identical | Identical |
| User | Physicians in the office or<br>hospital | Physicians in the office or<br>hospital | Identical | Identical |
| | Laptop | Main Unit | Identical to SpiroSphere<br>(K173937) | Identical to SpiroSphere<br>(K173937) |
| Components | Sensor Unit | Sensor Unit | Identical | Identical |
| | ERT PT | ERT PT | Identical | Identical |
| Screen Display | Depends on used Laptop | 162 x 122 mm | Identical to SpiroSphere<br>(K173937) | Identical to SpiroSphere<br>(K173937) |
| Interface | Bluetooth<br>USB<br>Wifi/Ethernet | Bluetooth<br>USB<br>Wifi/Ethernet<br>3G | Identical to SpiroSphere<br>(K173937) | Identical to SpiroSphere<br>(K173937) |
| Energy type | Li-Ion Battery and Power Supply | Li-Ion Battery and Power<br>Supply | Identical | Identical |
{12}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circular graphic is divided into three sections, colored blue, green, and dark blue. Below the text "ERT" is the tagline "Confidence at every turn".
{13}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic made up of three curved shapes in blue and green, followed by the letters "ERT" in bold blue font. Below the letters, the text "Confidence at every turn" is written in a smaller blue font.
{14}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic with three overlapping leaves in different shades of green and blue. To the right of the graphic is the text "ERT" in a bold, blue font. Below the text is the tagline "Confidence at every turn" in a smaller, lighter font.
| ECG Function Comparison | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device MasterScope<br>(K202754) | Subject Device | |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.