Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems
K223458 · Philips Medical Systems Nederland B.V. · LNH · Apr 6, 2023 · Radiology
Device Facts
Record ID
K223458
Device Name
Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems
Applicant
Philips Medical Systems Nederland B.V.
Product Code
LNH · Radiology
Decision Date
Apr 6, 2023
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
MR Image Reconstruction
Deep learning based reconstruction
—
Significantly better SNR and sharpness compared to predicate
—
—
Reader evaluation by ABR board certified radiologists
>1 (ABR board certified radiologists)
Indications for Use
Philips Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. This MR system enables trained physicians to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the tissue and the MR scan technique applied, and presence of contrast agents. The use of contrast imaging applications should be performed consistent with the approved labeling for the contrast agent. The trained clinical user can adjust the MR scan parameters to customize image appearance, accelerate image acquisition, and synchronize with the patient's breathing or cardiac cycle. The systems can use combinations of images to produce physical parameters, and related derived images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assust diagnosis, and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters and must be controlled and validated by the clinical user. In addition, the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body and extremities. MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated by the clinical user.
Device Story
Philips MR systems (1.5T/3.0T) utilize magnetic resonance phenomenon to acquire cross-sectional/spectroscopic images of head, body, and extremities. Systems include magnets, transmit body coils, gradient coils, and patient support. New R12 software introduces 'Precise Image', a deep learning-based reconstruction technique to increase SNR, improve sharpness, and reduce ringing artifacts. Operated by trained physicians/technicians in clinical settings. Output consists of MR images/spectra displayed on video screens for diagnostic interpretation and therapy planning. Precise Image enhances image quality, potentially aiding diagnostic accuracy. System supports MR fluoroscopy for interventional guidance. Clinical users control scan parameters and validate physical parameter accuracy.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing, including technical verification of the Precise Image reconstruction technique and a reader evaluation study by ABR board-certified radiologists. Radiologists assessed SNR, sharpness, CNR, and diagnostic quality of images with/without pathology, confirming equivalence and performance improvements over predicate technology.
Technological Characteristics
1.5T and 3.0T MR systems with 60cm/70cm bores. Components: magnet, transmit body coil, gradient coil/amplifier, RF amplifier, patient support. Software R12 includes deep learning-based 'Precise Image' reconstruction. Connectivity: standard MR system interfaces. Standards: IEC 60601-2-33, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, ISO 14971, IEC 62366-1, IEC 62304.
Indications for Use
Indicated for diagnostic imaging of head, body, or extremities in patients requiring cross-sectional MR images, spectroscopic images, or spectra. Used for interventional and minimally invasive procedure guidance. Contraindicated for use with non-MR safe/conditional instrumentation in the MR environment.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Achieva, Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K213583)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 6, 2023
Philips Medical Systems Nederland B.V. % Ioana Ulea Senior Regulatory Affairs Specialist Veenpluis 6 5684 PC Best NETHERLANDS
Re: K223458
Trade/Device Name: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LNI Dated: March 6, 2023 Received: March 6, 2023
Dear Ioana Ulea:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
D. Ray Kerlick
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K223458
Device Name
Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems
## Indications for Use (Describe)
Philips Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device.
This MR system enables trained physicians to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin.
Image appearance is determined by many different physical properties of the tissue and the MR scan technique applied, and presence of contrast agents. The use of contrast imaging applications should be performed consistent with the approved labeling for the contrast agent.
The trained clinical user can adjust the MR scan parameters to customize image appearance, accelerate image acquisition, and synchronize with the patient's breathing or cardiac cycle.
The systems can use combinations of images to produce physical parameters, and related derived images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assust diagnosis, and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters and must be controlled and validated by the clinical user.
In addition, the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body and extremities. MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated by the clinical user.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
## 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | Apr. 05, 2023 | |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 6, 5684 PC, Best, The Netherlands<br>Establishment Registration Number: 3003768277 | |
| Primary Contact Person: | Ioana Ulea<br>Senior Regulatory Affairs Specialist<br>Phone: +31 618345875<br>E-mail: ioana.ulea@philips.com | |
| Secondary Contact Person | Jan van de Kerkhof<br>Associate Director Regulatory Affairs<br>Phone: +31 613300542<br>E-mail: jan.van.de.kerkhof@philips.com | |
| Device Name: | Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems | |
| Classification: | Classification name: Magnetic Resonance Diagnostic Device (MRDD)<br>Classification Regulation: 21CFR 892.1000<br>Classification Panel: Radiology<br>Device Class: Class II<br>Primary Product Code: 90LNH<br>90LNI | |
| Primary Predicate Device: | Trade name: Achieva, Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (R11)<br>Manufacturer: Philips Medical Systems Nederland B.V.<br>510(k) Clearance: K213583<br>Classification Regulation: 21CFR 892.1000<br>Classification name: Magnetic Resonance Diagnostic Device (MRDD)<br>Classification Panel: Radiology<br>Device class: Class II<br>Product Code: 90LNH<br>90LNI | |
| | Trade name: MR 5300 and MR 7700 R11 MR Systems | |
| Secondary<br>Predicate Device: | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) Clearance: | K223442 |
| | Classification<br>Regulation: | 21CFR 892.1000 |
| | Classification name: | Magnetic Resonance Diagnostic Device<br>(MRDD) |
| | Classification Panel: | Radiology |
| | Device class | Class II |
| | Product Code: | 90LNH<br>90LNI |
| Reference<br>Device | Trade name: | SIGNA Premier |
| | Manufacturer: | GE Medical Systems, LLC |
| | 510(k) Clearance: | K193282 |
| | Classification<br>Regulation: | 21CFR 892.1000 |
| | Classification name: | Magnetic Resonance Diagnostic Device<br>(MRDD) |
| | Classification Panel: | Radiology |
| | Device class | Class II |
| | Product Code: | 90LNH<br>90LNI |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
| Device | The proposed Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition,<br>MR 5300 and MR 7700 MR Systems R12 are 60 cm and 70 cm bore 1.5<br>and 3.0 Tesla (1.5T and 3.0T) Magnetic Resonance Diagnostic Devices,<br>hereafter to be known as Ingenia, Ingenia CX, Ingenia Elition, Ingenia<br>Ambition, MR 5300 and MR 7700 MR Systems. |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description: | This bundled abbreviated 510(k) submission, is prompted by the<br>introduction of a new optional software feature called Precise Image<br>contained in software R12 for the proposed Ingenia, Ingenia CX, Ingenia<br>Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems, as<br>compared to our legally marketed primary predicate device Achieva,<br>Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems<br>(R11) (K213583) and the secondary predicate device MR 5300 and MR<br>7700 R11 MR Systems (K223442). |
| | Precise Image is a deep learning based reconstruction technique<br>designed to increase signal-to-noise ratio (SNR), increase sharpness and<br>decrease ringing artefacts from MR images. |
| | The introduction of Precise Image only required updates to the MR System<br>Software. |
| | The proposed Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition,<br>MR 5300 and MR 7700 MR Systems are intended to be marketed with |
| | the following pulse sequences and coils that are previously cleared by FDA: |
| | 1. mDIXON (K102344)<br>2. SWIp (K131241)<br>3. mDIXON-Quant (K133526)<br>4. MRE (K140666)<br>5. mDIXON XD (K143128)<br>6. O-MAR (K143253)<br>7. 3D APT (K172920)<br>8. Compatible System Coils |
| | The accessories to be used with the proposed device Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems have not changed compared to the primary predicate device and secondary predicate device and can be found in the Instructions for Use accompanying the device: |
| | System coils PPU Sensor for wireless physiology Pediatric PPU Sensor FlexTrak trolleys (FlexTrak / HA FlexTrak II) Acoustic Hood MR Elastography |
| | When Philips MRI system is used in combination with the Philips MR-RT or MR-OR solutions, the user is referred to the dedicated MR-RT and MR-OR Instructions for Use for information on additional accessories that may apply: |
| | Flextrak OR MR-RT CouchTop RT CouchTop XD |
| Indications for Use: | The indications for Use statement provided below for the proposed device is identical to the one of the primary and secondary predicate devices. The intended use is also not impacted by the introduction of Precise Image feature. |
| | Philips Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. |
| | This MR system enables trained physicians to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin. |
| | Image appearance is determined by many different physical properties of the tissue and the anatomy, the MR scan technique applied, and presence of contrast agents. |
| | The use of contrast agents for diagnostic imaging applications should be<br>performed consistent with the approved labeling for the contrast agent. |
| | The trained clinical user can adjust the MR scan parameters to customize<br>image appearance, accelerate image acquisition, and synchronize with<br>the patient's breathing or cardiac cycle. The systems can use<br>combinations of images to produce physical parameters, and related<br>derived images. Images, spectra, and measurements of physical<br>parameters, when interpreted by a trained physician, provide information<br>that may assist diagnosis and therapy planning. The accuracy of<br>determined physical parameters depends on system and scan parameters<br>and must be controlled and validated by the clinical user. |
| | In addition, the Philips MR systems provide imaging capabilities, such as<br>MR fluoroscopy, to guide and evaluate interventional and minimally<br>invasive procedures in the head, body and extremities. MR Interventional<br>procedures, performed inside or adjacent to the Philips MR system, must<br>be performed with MR Conditional or MR Safe instrumentation as selected<br>and evaluated by the clinical user for use with the specific MR system<br>configuration in the hospital. The appropriateness and use of information<br>from a Philips MR system for a specific interventional procedure and<br>specific MR system configuration must be validated by the clinical user. |
| Design Features/<br>Fundamental<br>Scientific<br>Technology: | Just as the primary and secondary predicate devices, the proposed<br>Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and<br>MR 7700 MR Systems are based on the principle that certain atomic<br>nuclei present in the human body will emit a weak relaxation signal when<br>placed in a strong magnetic field and excited by a radio signal at the<br>precession frequency. The emitted relaxation signals are analyzed by the<br>system and a computed image reconstruction is displayed on a video<br>screen.<br><br>The principal technological components (magnet, transmit body coil,<br>gradient coil, gradient amplifier, RF amplifier and patient support) of the<br>proposed Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR<br>5300 and MR 7700 MR Systems are unchanged compared to primary<br>predicate (K213583) and the secondary predicate device (K223442). |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue letters.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.
| No. | Recognition Number | Standard Number and Date | Standard Name |
|-----|--------------------|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 12-295 | IEC60601-2-33<br>Ed. 3.2:2010 +<br>Amd 1:2013 +<br>Amd 2:2015 | Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis |
| 2 | 19-4 | ANSI / AAMI ES60601-<br>1:2005/(R)2012<br>and A1:2012 | Medical Electrical Equipment -<br>Part 1: General Requirements<br>For Basic Safety And Essential<br>Performance (IEC 60601-1:2006, MOD). |
| 3 | 19-8 | IEC60601-1-2<br>Ed. 4.0:2014 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests |
| 4 | 5-89 | IEC 60601-1-6<br>Ed. 3.1:2010 +<br>Amd 1:2013 | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability |
| 5 | 5-76 | IEC 60601-1-8<br>Ed. 2.1:2006 +<br>Amd 1:2012<br>(Ed.2.1) | Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems |
| 6 | 5-125 | ISO 14971 Third<br>Edition 2019 | Medical devices – Application of risk management to medical devices. |
| 7 | 5-114 | IEC 62366-1 Ed.<br>1.0:2015 | Medical devices – Part 1: Application of usability engineering to medical devices |
| 8 | 13-79 | IEC 62304 Ed.<br>1.1:2015 | Medical device software - Software life cycle processes. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out. The overall impression is clean and professional, reflecting the brand's image.
| review, reproducibility of Precise in comparison to predicate device<br>reconstruction technique can be considered established. Additionally, a<br>reader evaluation by ABR board certified radiologists was performed,<br>where following properties have been analyzed as part of the comparison:<br>signal-to-noise ratio (SNR), artifact level, sharpness, and contrast-to-noise<br>ratio (CNR). Furthermore, the radiologists have assessed the quality of the<br>visualization of abnormalities and pathologies in case they were present<br>in the images as well as if the images were of sufficient quality for<br>diagnostic purposes. The review evaluation shows that the proposed<br>device is assessed as equivalent for diagnosis and holds significantly<br>better SNR and sharpness compared to the predicate reconstruction<br>technology, also in the presence of (subtle) abnormalities and pathology. | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The verification and/or validation test results demonstrate that the<br>proposed Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR<br>5300 and MR 7700 MR Systems meet the acceptance criteria and are<br>adequate for the intended use. | |
| The risk management activities show that all risks are sufficiently<br>mitigated, that no new risks are introduced, and that the overall residual<br>risks are acceptable. | |
| Therefore, the proposed Ingenia, Ingenia CX, Ingenia Elition, Ingenia<br>Ambition, MR 5300 and MR 7700 MR Systems are substantially<br>equivalent to the legally marketed primary predicate device Achieva,<br>Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems<br>(R11) (K213583, 16/05/2022) and the secondary predicate device MR<br>5300 and MR 7700 R11 MR Systems (K223442) in terms of safety and<br>effectiveness. | |
| Summary of<br>Clinical Data: | With the proposed Ingenia, Ingenia CX, Ingenia Elition, Ingenia<br>Ambition, MR 5300 and MR 7700 MR Systems the indications for use<br>remain unchanged and there were no technological characteristics relative<br>to the primary and secondary predicate device that would require clinical<br>testing. |
| Substantial<br>Equivalence: | The proposed Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition,<br>MR 5300 and MR 7700 MR Systems and the legally marketed primary<br>predicate device Achieva, Ingenia, Ingenia CX, Ingenia Elition, and<br>Ingenia Ambition MR Systems (R11) (K213583, 16/05/2022) and the<br>secondary predicate device MR 5300 and MR 7700 R11 MR Systems<br>(K223442), have the same indications for use with respect to the following:<br>Providing cross-sectional images based on the magnetic resonance phenomenon Interpretation of the images is the responsibility of trained physicians Images can be used for interventional and treatment planning purposes |
| Conclusion: | The proposed Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition,<br>MR 5300 and MR 7700 MR Systems are substantially equivalent to the<br>legally marketed primary predicate device Achieva, Ingenia, Ingenia CX,<br>Ingenia Elition, and Ingenia Ambition MR Systems (R11) (K213583) and<br>the secondary predicate device MR 5300 and MR 7700 R11 MR Systems (K223442) in terms of safety and effectiveness. |
| | (K223442), in terms of design features, fundamental scientific technology,<br>indications for use, and safety and effectiveness. |
| | Additionally, substantial equivalence is demonstrated with non-clinical<br>performance (verification and validation) tests, which complied with the<br>requirements specified in the international and FDA-recognized<br>consensus standards and device-specific guidance. |
| | The results of these tests demonstrate that the proposed Ingenia, Ingenia<br>CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR<br>Systems meet the acceptance criteria and are adequate for the intended<br>use. |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.