The Bakri Essential Postpartum Balloon is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
Device Story
The Bakri Essential Postpartum Balloon is an intrauterine tamponade device used by clinicians to manage postpartum hemorrhage. It consists of a double-lumen silicone balloon catheter and a stopcock. The device is inserted into the uterus, where the balloon is inflated with up to 500 mL of fluid to apply pressure to the uterine wall, thereby providing temporary control or reduction of bleeding. The second lumen allows for drainage. It is a single-use, sterile device. The clinician monitors the patient's response to the tamponade effect to guide clinical decision-making regarding the need for further interventions. The device provides a mechanical benefit by exerting pressure to stabilize the patient during postpartum hemorrhage.
Clinical Evidence
Bench testing only. Performance testing included pouch seal strength (ASTM F88-15), bubble leak testing (ASTM F2096-11), dye penetration testing (ASTM F1929-15), and functional verification following simulated distribution and sterilization. Testing confirmed the reliability of the packaging at time zero and after a three-year shelf-life.
Technological Characteristics
Double-lumen silicone balloon catheter; 24 Fr shaft; 58 cm total length; 500 mL maximum inflation volume; polyamide female luer-lock adapter. Sterilized via ethylene oxide (EO) to a 10^-6 sterility assurance level. Packaging uses Tyvek (1073B)/polyethylene peel-open pouch. No software or electronic components.
Indications for Use
Indicated for temporary control or reduction of postpartum uterine bleeding in patients where conservative management is warranted.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 28, 2022
Cook Incorporated Ian Herrman Senior Regulatory Affairs Specialist 750 Daniels Wav. P.O. Box 489 Bloomington, IN 47402
Re: K223098
> Trade/Device Name: Bakri Essential Postpartum Balloon Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: OQY Dated: September 29, 2022 Received: September 30, 2022
Dear Ian Herrman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223098
Device Name Bakri Essential Postpartum Balloon
Indications for Use (Describe)
The Bakri Essential Postpartum Balloon is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---|
| ☑ | ☐ |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K223098 Page 1 of 3
### 510(k) SUMMARY
### Bakri Essential Postpartum Balloon
### Submitted By
| Applicant: | Cook Incorporated |
|---------------------------|----------------------------------------------------------|
| Contact: | Ian Herman |
| Address: | 750 Daniels Way<br>P.O. Box 489<br>Bloomington, IN 47402 |
| Contact Phone Number: | (812) 335-3575 x104034 |
| Contact Fax Number: | (812) 332-0281 |
| Date Prepared: | October 26, 2022 |
| <b>Device Information</b> | |
| Proprietary Name: | Bakri Essential Postpartum Balloon |
| Common Name: | Intrauterine tamponade balloon |
| Regulation Name: | Obstetric-Gynecologic Specialized Manual Instrument |
| Regulation Number: | 21 CFR 884.4530 |
### Predicate Device
Regulation Class:
Product Code:
Bakri Postpartum Balloon, manufactured by Cook Incorporated (K170622).
II
The predicate device has not been subject to a design-related recall.
### Device Description
The Bakri Essential Postpartum Balloon is composed of a double lumen silicone balloon catheter and a stopcock. The balloon catheter has a balloon with maximum inflation volume of 500 mL and a 24 Fr silicone shaft with an overall length of 58 cm, which includes the working length of 54 cm and proximal fittings. The shaft has two lumens, one designated for drainage, and the other for balloon inflation. The Bakri Essential is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
OQY, Intrauterine Tamponade Balloon
The Bakri Essential Postpartum Balloon is packaged in a Tyvek®Polyethylene Peel-Open Pouch with five individually packaged and labeled products per marketing carton, is provided sterile, and is intended for one time use.
{4}------------------------------------------------
The Bakri Essential Postpartum Balloon is a modified version of the predicate device cleared under K170622.
## Indications for Use
The Bakri Essential Postpartum Balloon is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
## Comparison to Predicate Device
A comparison of the intended use and technological characteristics of the subject and predicate device are outlined in the table below.
| General Device | Subject device | Predicate device | Comments |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Characteristics | K223098 | K170622 | |
| Product code | OQY | OQY | Same |
| Intended<br>use/Indications<br>for Use | Bakri Essential<br>Postpartum Balloon is<br>intended to provide<br>temporary control or<br>reduction of postpartum<br>uterine bleeding when<br>conservative<br>management is<br>warranted | Bakri Postpartum Balloon is<br>intended to provide<br>temporary control or<br>reduction of postpartum<br>uterine bleeding when<br>conservative management is<br>warranted.<br>&<br>Bakri Postpartum Balloon<br>with Rapid Instillation<br>Component is intended to<br>provide temporary control<br>or reduction of postpartum<br>uterine bleeding when<br>conservative management is<br>warranted. | Same. |
| Dimensions | Length: 58 cm (working<br>length of 54 cm)<br>Outer Diameter: 24 Fr<br>Balloon Volume: 500<br>mL | Length: 58 cm (working<br>length of 54 cm)<br>Outer Diameter: 24 Fr<br>Balloon Volume: 500 mL | Same |
| Materials | Balloon Catheter and<br>Shaft: Silicone<br>Female Luer-Lock<br>Adapter: Polyamide | Balloon Catheter and Shaft:<br>Silicone<br>Female Luer-Lock Adapter:<br>Polyamide | Same |
| Syringe | Syringe not included | 60 mL syringe | Different |
| Stopcock | Included | Included | Same |
{5}------------------------------------------------
| Packaging | Sterile Barrier: Tyvek<br>(1073B)/Polyethylene<br>Peel Open Pouch<br>Package.<br>5 individually packaged<br>units per Box (8" & 1/8"<br>X 2 & 3/8" X 17") | Sterile barrier: Tyvek<br>(1059B)/Polyethylene Peel<br>Open Pouch Package.<br>1 packaged unit per box<br>(8" X 1 & 7/16" X 16")<br>Inner packaging: Fitted in<br>a PETG (Polyethylene<br>Terephthalate Glycol-<br>Modified tray with PETG<br>retainer and seal | Different. |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Sterilization | EO, 10-6, single use | EO, 10-6, single use | Same |
The subject and predicate device have the same intended use, principles of operation, and technological characteristics. The only differences between the Bakri Essential and the predicate device are the packaging configuration and the removal of the syringe from the set. These differences do not raise different questions of safety and effectiveness and can be assessed by verification and validation testing.
# Summary of Non-clinical Performance Testing
The support the proposed modifications to the subject device, design control activities and a risk analysis (depicting device change, risk associated, verification method, acceptance criteria and summary of results) according to ISO 14971:2019 was performed. The Bakri Essential Postpartum Balloon was subjected to applicable testing to assure reliable design and performance under the testing parameters. As the differences in the subject device compared to the predicate device include packaging differences and the removal of the syringe from the set, the following tests were conducted to ensure the reliability of the packaging at time zero and end of the three year shelf-life:
- Pouch Seal Strength Testing: verification that the minimum seal strength shall meet ● or exceed 1.2N per 15mm when tested in accordance with ASTM F88-15
- Bubble Leak Testing: verification that packaging shows no signs of cracks, pinholes, or imperfect seals indicated by bubbles, per ASTM F2096-11
- Dye Penetration Testing: verification that no packaging seal may be breached by dye ● per F1929-15
- Additional distribution testing followed by functional verification: verification that ● device shows no signs of damage after sterilization and simulation distribution.
# Conclusion
The results of these testing described above demonstrate that the Bakri Essential Postpartum Balloon is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.