Bakri Postpartum Balloon, Bakri Postpartum Balloon with Rapid Instillation Component
Applicant
Cook Incorporated
Product Code
OQY · Obstetrics/Gynecology
Decision Date
Mar 31, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Bakri® Postpartum Balloon is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted. Bakri® Postpartum Balloon with Rapid Instillation Components is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
Device Story
Intrauterine tamponade device; consists of dual-lumen silicone balloon catheter, stopcock, and 60 mL syringe; Rapid Instillation (RI) configuration adds dual check valve and polymer tubing for saline bag connection. Used in clinical settings by physicians to manage postpartum hemorrhage. Balloon inserted into uterus; inflated with saline to apply pressure to uterine wall; provides tamponade effect to control bleeding. RI components allow continuous saline instillation without disconnecting syringe. Output is mechanical pressure on uterine tissue; aids in hemorrhage control; potentially avoids more invasive surgical interventions.
Clinical Evidence
Bench testing only. Testing included inflated balloon reliability, dimensional evaluation, integrity, distal tip tensile testing, sideport drainage/deflation reliability, radial pressure testing (100 mm Hg for 3 hours), balloon inflation capacity (500 mL), simulated use, and component tensile testing. Biocompatibility testing was not required as materials are identical to the predicate or do not contact the patient.
Technological Characteristics
Dual-lumen silicone balloon catheter; 500 mL maximum balloon volume. RI configuration includes polymer tubing and dual check valve. Sterile, single-use. Mechanical tamponade principle. No electronic components or software.
Indications for Use
Indicated for temporary control or reduction of postpartum uterine bleeding in patients where conservative management is warranted.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2017
Cook Incorporated Naomi Funkhouser, MBA Regulatory Product Specialist 750 Daniels Way, P.O. Box 489 Bloomington, IN 47402
Re: K170622
> Trade/Device Name: Bakri® Postpartum Balloon, Bakri® Postpartum Balloon with Rapid Instillation Components Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: OOY Dated: February 28, 2017 Received: March 1, 2017
Dear Naomi Funkhouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Charles Viviano -S
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170622
Device Name Bakri® Postpartum Balloon Bakri® Postpartum Balloon with Rapid Instillation Components
Indications for Use (Describe)
Bakri® Postpartum Balloon is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
Bakri® Postpartum Balloon with Rapid Instillation Components is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif; font-size: 10pt;"></span> <span style="font-family: Symbol;">☑</span> <span style="font-family: Arial, sans-serif; font-size: 10pt;"> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-family: Arial, sans-serif; font-size: 10pt;"></span> <span style="font-family: Symbol;">☐</span> <span style="font-family: Arial, sans-serif; font-size: 10pt;"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
## Bakri® Postpartum Balloon with Rapid Instillation Components 21 CFR §884.4530 Date Prepared: March 29, 2017
| Submitted By: | |
|-----------------------|------------------------|
| Applicant: | Cook Incorporated |
| Contact: | Naomi Funkhouser |
| Applicant Address: | 750 Daniels Way |
| | P.O. Box 489 |
| | Bloomington, IN 47402 |
| Contact Phone Number: | (812) 339-2235 x104371 |
| Contact Fax Number: | (812) 332-0281 |
### Device Information:
| Trade Names: | Bakri <sup>®</sup> Postpartum Balloon<br>Bakri <sup>®</sup> Postpartum Balloon with Rapid Instillation Components |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| Regulation Name: | Obstetrics-gynecology Specialized Manual Instrument |
| Classification Name: | Obstetrics/Gynecology |
| Regulation: | 21 CFR §884.4530 |
| Regulation Class: | II |
| Product Code: | OQY, Intrauterine Tamponade Balloon |
### Device Description:
The proposed devices are offered in two configurations, the Bakri Postpartum Balloon (henceforth referred to as Bakri) consisting of a double lumen silicone balloon catheter, a stopcock, and a 60 mL Luer-tip syringe and the Bakri Postpartum Balloon with Rapid Instillation Components (henceforth referred to as Bakri RI) configuration which additionally includes a dual check valve and polymer tubing. The tubing is fitted with a Luer lock-type connector to the dual check valve and a non-vented bag spike with a vented cap that connects to a saline bag (not included in the device).
The balloon catheter common to both configurations is designed as a dual lumen silicone shaft with 500 mL balloon. The proximal end of the catheter has a lighthouse adaptor designed to accept a drainage bag connector and a side arm which is fitted with a Luer Lock-type adaptor to
{4}------------------------------------------------
which a stopcock is attached and subsequently a Luer tip syringe or a dual check valve can be connected. The distal end of the catheter has two sideports spaced 180° opposite of each other, and the tip is rounded to a smooth finish. The polymer tubing of the Bakri RI rapid instillation components is 180 cm in length. Both the Bakri and the Bakri RI is supplied sterile and intended for one time use.
## Indications for Use:
Bakri® Postpartum Balloon is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
Bakri® Postpartum Balloon with Rapid Instillation Components is intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted.
## Predicate Device:
The predicate device, the Bakri® Postpartum Balloon manufactured by Cook Incorporated (K062438), is a medical device intended to provide temporary control or reduction of postpartum uterine bleeding when conservative management is warranted. It is a tamponade balloon catheter constructed with silicone material and a 500 mL (max) balloon adhered to a dual lumen catheter used for drainage and balloon inflation. The device is sold with a stopcock and 60 mL syringe. The predicate device has not been subject to a design-related recall.
## Comparison to Predicate Device:
The proposed Bakri and the balloon catheter of the Bakri RI as compared to the predicate device. Bakri Postpartum Balloon (K062438), are identical in terms of intended use, principles of operation, basic technological characteristics, materials of construction, and nearly identical in dimension and design. The Bakri RI includes Rapid Instillation Components which consist of a dual check valve and tubing. The Rapid Instillation components allow the physician fill the balloon without disconnecting and reconnecting the syringe at each reiteration of saline instillation into the balloon. The technological differences as a result of the changes to the balloon catheter and the addition of the rapid instillation components do not raise any different questions of safety and effectiveness.
{5}------------------------------------------------
# Test Data:
The following tests were performed to demonstrate that the modifications to the balloon catheter of both the Bakri and Bakri RI and the addition of the rapid instillation component have met applicable design and performance requirements to support a determination of substantial equivalence. Biocompatibility testing was deemed unnecessary, since the materials used in the device are identical to the predicate, and all new materials do not directly or indirectly contact the patient.
- . Inflated Balloon Reliability, Dimensional Evaluation, and Integrity -Verification that when subjected to maximum labeled volume, the inflated balloon and components of the proposed device do not show signs of leakage and/or rupture, and balloon dimensions meet the clinical and anatomical requirements.
- Tensile Test of the Distal Tip Verification that when subjected to maximum load . requirements, the distal tip satisfies the minimum load requirements.
- Sideport Drainage and Deflation Reliability- Verification that the proposed device . will drain approximately 200 mL from the sideport in less than or equal to 1 minute and will have less than or equal to 30 mL of water left inside the balloon catheter after 15 minutes of drainage.
- Radial Pressure Verification that the balloon can withstand a radial pressure of . 100 mm Hg for up to 3 hours without leakage or rupture.
- Balloon Inflation Testing Verification that the balloon can be inflated to the . minimal volume of 500 mL with the Rapid Instillation Components.
- Simulated Use and Leak Test Verification that the device performs as intended and . does not leak during normal use.
- Tensile Testing Verification that the Spike/Tubing, MLLA/Tubing, and Tubing . withstands normal use forces.
In conclusion, the results of these tests support a determination of substantial equivalence of the balloon catheter common to the two configurations of the subject device to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.