K223044 · Shen Zhen Rong Feng Technology Co., Ltd. · FLL · May 10, 2023 · General Hospital
Device Facts
Record ID
K223044
Device Name
Digital Thermometer (Model QT001)
Applicant
Shen Zhen Rong Feng Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
May 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Digital Thermometer QT001 is intended to measure the human body temperature in armpit, orally or rectally, and the device is reusable for clinical or home use on people of all ages.
Device Story
Handheld digital thermometer; measures human body temperature via armpit, oral, or rectal sites. Input: thermistor sensor (Type 503F) resistance changes based on temperature. Processing: MCU converts resistance to frequency/voltage; calculates temperature; internal firmware manages logic. Output: temperature displayed on LCD; includes backlight color-coded fever alerts (green/yellow/red). Used in clinical or home settings by patients or clinicians. Benefits: provides accurate, reusable temperature monitoring for all ages.
Clinical Evidence
Bench testing only. Device complies with ASTM E1112-00(2018), ISO 80601-2-56:2017/AMD 1:2018, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5, and ISO 10993-10. No clinical data presented.
Technological Characteristics
Materials: ABS, TPE, stainless steel. Sensing: Thermistor (Type 503F). Energy: 3.0V CR1225 battery. Form factor: Handheld with flexible tip. Connectivity: Standalone. Software: Firmware-based measurement and display logic. Standards: ISO 80601-2-56, ASTM E1112-00, IEC 62304.
Indications for Use
Indicated for measurement of human body temperature in armpit, orally, or rectally. Suitable for all ages; clinical or home use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 10, 2023
Shen Zhen Rong Feng Technology Co., Ltd Jett Lee Regulation Manager 3/F R Building ShaSi Industrial Park, Shajing Town BaoAn District Shenzhen, Guangdong China
Re: K223044
Trade/Device Name: Digital Thermometer (Model QT001) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: April 9, 2023 Received: April 10, 2023
Dear Jett Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223044
Device Name Digital Thermometer (Model: OT001)
Indications for Use (Describe)
The Digital Thermometer OT001 is intended to measure the human body temperature in armpit, or rectally, and the device is reusable for clinical or home use on people of all ages.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: sans-serif;"> <input disabled="disabled" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <input checked="checked" disabled="disabled" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K223044
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
### Summary Prepared Date: 25 April 2023
#### 1. Submitter Information
Sponsor Company Name: SHEN ZHEN RONG FENG TECHNOLOGY CO.,LTD
Address: 3/F R building ShaSi Industrial Park,Shajing Town,BaoAn district,ShenZhen city,GuangDong
Province.China
Phone: +86-13715151187
Contact Person (including title): Hong Xiao Lan (Engineer)
E-mail: 844352259@qq.com
# Application Correspondent: SHEN ZHEN RONG FENG TECHNOLOGY CO.,LTD
Address: 3/F R building ShaSi Industrial Park,Shajing Town,BaoAn district,ShenZhen city,GuangDong Province.China
- � Contact Person: Iris Fung
- � Tile: Requlation Manager
- Tel: +86-13211147965 �
- � Email: jianda-lee@foxmail.com; mdc-fs@foxmail.com;
# 2. Subject Device Information
Type of 510(k) submission: Traditional Common Name: Clinical Electronic Thermometer Trade Name: Digital thermometer (Model: QT001) Classification Name: Clinical electronic thermometer Review Panel: General Hospital Product Code: FLL Regulation Number: 21 CFR 880.2910 Requlation Class: II
# 3. Predicate Device Information
Sponsor: JOYTECH HEALTHCARE CO., LTD. Trade Name: Digital Thermometer
{4}------------------------------------------------
510(k) number: K200599 Review Panel: General Hospital Product Code: FLL Regulation Number: 21 CFR 880.2910 Regulation Class: II
# 4. Device Description
The digital thermometer QT001 is hand held device which can measure human body's temperature at the site of armpit, orally or rectally, the device is reusable for clinical or home use on people of all ages. The results can be displayed on LCD. The QT001 has only one operating mode: direct mode.
The digital thermometer measures temperatures of human body by placing the probe tip into the armpit, oral, rectum directly with the measuring time about 3 minutes.
The digital thermometer consists of a temperature sensor, low power consumption integrated circuit(IC),LCD display and buzzer. The resistance of sensor changes with temperature and the IC converts the resistance to frequency and calculates the temperature according to the relation of resistance and frequency. The calculated temperature is displayed on the LCD.
The thermistor sensor(Type:503F) reacts to temperature. In different temperature, the thermistor has a different resistance value. The essence of temperature measurement is actually measuring the resistance of the sensor, usually is to change the resistance into voltage or current and other analog signals, and digital signals, and then the processor converts the signals to the corresponding temperatures.
QT001 has several functions, such as beep alarm, unit switchable, low battery detection, memories, backlight, auto power off functions.
# 5. Indication for use
The Digital Thermometer QT001 is intended to measure the human body temperature in armpit, orally or rectally, and the device is reusable for clinical or home use on people of all ages.
# 6. Comparison to Predicate Device
A comparison of key technological characteristics between the subject devices and predicate device was listed as below:
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | SHEN ZHEN RONG FENG<br>TECHNOLOGY CO., LTD | Joytech Healthcare Co., Ltd | -- |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| 510 (k) Number | K223044 | K200599 | -- |
| Product Name | Digital Thermometer | Digital Thermometer | -- |
| Models | QT001 | DMT-4756 | -- |
| Intended Use | The Digital Thermometer QT001<br>is intended to measure the<br>human body temperature in<br>armpit, orally or rectally, and the<br>device is reusable for clinical or<br>home use on people of all ages. | The Digital Thermometer DMT-<br>4756 is intended to measure the<br>human body temperature in regular<br>mode orally, rectally or under the<br>arm, and the device is reusable<br>for clinical or home use on people<br>of all age, including children under 8<br>with adult supervision. | Similar<br>There is minor<br>difference, but the<br>user manual have<br>been indicated in<br>the instructions that<br>children should not<br>be allowed to<br>perform the test<br>unsupervised<br>It will not affect<br>safety and<br>effectiveness. |
| Fundamental<br>technology&<br>Operating<br>principle | A change of thermistor<br>resistance,<br>caused by changes of<br>temperature. The resistance is<br>measured by MCU, so changes<br>of temperature will correspond to<br>changes of resistance | A change of thermistor resistance,<br>caused by changes of temperature.<br>The resistance is measured by<br>MCU, so changes of temperature<br>will correspond to changes of<br>resistance | Identical |
| Sensor | Thermistor | Thermistor | Identical |
| Signal<br>processing<br>and display | Internal firmware and local LCD<br>display | Internal firmware and local LCD<br>display | Identical |
| Measurement<br>Site | armpit, orally or rectally | orally, rectally or under the arm | Identical |
| Measuring<br>Range | $32.0~42.9 °C (89.6°F ~ 109.2°F)$ | $32.0°C~43.9°C (89.6°F~111.0°F)$ | Different(Note 1) |
| Accuracy | $±0.1°C,34.0°C-42.0°C (±0.2°F,93.2°F -107.6°F )± 0.2 °C under34.0 °C or over42.0°C( ± 0.4°F under 93.2°F or over107.6°F )$ | $±0.1°C between 35.5°C to42.0°C(±0.2°F,95.9°F-107.6°F),±0.2°C under 35.5°C or over42.0°C(±0.4°F under 95.9°F or over107.6°F )$ | Different(Note 2) |
| Display<br>resolution | $0.1 °C/0.1 °F$ | $0.1 °C/0.1 °F$ | Identical |
| Components | Sensor, buzz film, housing,<br>stainless steel cap, LCD display,<br>measurement control module. | Sensor, buzz film, housing,<br>stainless steel cap, LCD display,<br>measurement control module. | Identical |
| Sensor type | Thermopile | Thermopile | Identical |
| Material | ABS, TPE, Stainless steel | ABS, TPE, Stainless steel | Identical |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| Memories | For storing the last measured<br>value | 10 memories | Different(Note 3) |
| Power Supply | One 3.0 V CR1225 battery | One 3.0V CR2032 battery | Identical |
| Operating range | Temperature: 5°C-40°C (41°F -<br>104°F )<br>Relative humidity:15%RH-<br>95%RH<br>Atmospheric pressure: 70kPa-<br>106kPa | Temperature:<br>41°F ~104°F (5°C ~40°C )<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure :<br>700hPa ~ 1060hPa | Different(Note4) |
| Construction | Flexible tip | The probe is foldable into the body<br>of the thermometer | Different(Note5) |
| Measurement<br>method & time | Approximate 3 minutes | 10s Predictive read<br>30s Final read. | Different(Note6) |
| Thermometer<br>Size | 138mm x 25mm x 12mm (L x W x<br>H) | 10.6cm*3.2cm*2.5cm(L x W x H) | Different(Note7) |
| Weight | Approx. 10 grams including<br>battery | Approx. 40 grams including battery | Different(Note7) |
| LCD Size | 2.0cm*0.7cm | 2.5cm*1.4cm | Different(Note7) |
| Color &Coding | White(R9003), Green(P294C) | White(R9003),<br>Green(P564C) | Different(Note7) |
| Predictive mode | NO | Yes | Different(Note6) |
| Storage<br>and<br>transportation<br>condition | Temperature:-20℃~55℃(-<br>4°F ~ 131°F )<br>Relative humidity:15%RH-<br>95%RH<br>Atmospheric pressure: 50kPa-<br>106kPa | Temperature:<br>-4°F ~131°F (-20°C~55°C)<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure:<br>700hPa ~ 1060hPa | Different(Note4) |
| Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | Identical |
| EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | Identical |
| Biocompatibility | Complied with ISO 10993-5 and<br>ISO 10993-10 | Complied with ISO 10993-5 and<br>ISO 10993-10 | Identical |
| Performance | Complied with ISO 80601-2-<br>56:2017/AMD 1:2018 and ASTM<br>E1112:00(2018). | Complied with ISO 80601-2-<br>56:2017 | Different(Note8) |
| Backlight<br>warning function | 1.When measured temperature is<br>35.9℃,the LCD without<br>backlight<br>2.When measured temperature<br>≥36℃ and ≤37.1℃, the LCD<br>shows green backlight<br>3.When measured temperature<br>≥37.2℃ and ≤37.4℃, the LCD<br>shows yellow backlight.<br>4.When measured temperature<br>>=37.5°C and ≤42.9°C, the LCD | / | Different(Note9) |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| | shows red backlight.<br>The laboratory accuracy error no<br>greater than 0.3 °C | | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
### Comparison discussion:
### Note 1
The subject device and predicate device have different measurement range, but the measurement range of subject device meets the requirements of ASTM E1112:00 and ISO 80601-2-56. The difference does not raise any new or different safety and effectiveness questions. .
# Note 2
The subject device and predicate device have different accuracy, but the accuracy of the subject device meets the requirements of ASTM E1112:00 and ISO 80601-2-56. The difference does not raise any new or different safety and effectiveness questions.
#### Note 3
The memory function difference: The software verification were conducted to demonstrate the performance of the function as intended. The memory function does not affect accuracy of measurement; so such difference does not impact the performance of subject device.
#### Note 4
Operating range, Storage and transportation condition difference: The subject device has been tested and the test results met the requirements of IEC60601-1 and ISO 80601-2-56. The difference does not raise any new or different safety and effectiveness questions.
# Note 5
Construction difference: Performance testing shows the subject device complies with standard ISO 80601-2-56. The difference does not raise any new questions of safety and effectiveness.
# Note 6
Measurement time, predictive mode difference: Performance test was conducted in accordance with ISO 80601-2-56. The test results demonstrated the subject device complies with performance standard. The differences do not raise any new questions of safety and effectiveness.
# Note 7
Weight, size, color difference: Biocompatibility test and performance bench test met the requirements of ISO 10993-5 and ISO 10993-10. The differences do not raise any new questions of safety and effectiveness.
# Note 8
Performance testing shows the subject device complies with standard ISO 80601-2-56:2017/AMD 1:2018 and ASTM E1112-00 (2018). The difference does not raise any new questions of safety and effectiveness.
{8}------------------------------------------------
# Note 9
Backlight warning function does not affect accuracy of measurement; the software verification and validation were conducted to demonstrate the performance of the function as intended. So such difference does not raise any new safety and effectiveness questions.
# 7. Test Summary
# Non-clinical test:
Digital Thermometer conforms to applicable standards that include:
- � ASTM E1112-00(2018) Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
- � IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- � ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity,
- � ISO 10993-10:2010. Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- � IEC 62304:2006+AMD1:2015 Medical Device Software - Software Life Cycle Processes
- � IEC60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- � ISO 80601-2-56:2017/AMD 1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
# 8. Conclusion
Based on comparison of the intended use, technological characteristics, applicable safety standards, verification and validation testing, the differences between subject device and the predicate device do not raise new issues of safety and effectiveness. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.