The Digital Thermometer DMT-4756 is intended to measure the human body temperature in regular mode orally, rectally or under the arm, and the device is reusable for clinical or home use on people of all ages, including children under 8 with adult supervision.
Device Story
Digital thermometer (DMT-4756) measures human body temperature via thermistor sensor. Device features foldable tip for rectal use; unfolded for oral/axillary use. Operates by measuring thermistor resistance changes processed by internal MCU. Provides 10-second predictive quick read and 30-second final reading. Used in clinical or home settings by patients or clinicians. Output displayed on integrated LCD. Includes beep alarm, unit switching, low battery detection, memory (10 readings), and auto-power off. Benefits include rapid, accurate temperature monitoring for all ages.
Clinical Evidence
Clinical testing conducted on 450 subjects across three age groups: infants (newborn to 1 year), children (>1 to 5 years), and adults (>5 years). Testing performed per ISO 80601-2-56 for all measurement sites. Results demonstrated compliance with performance requirements.
Technological Characteristics
Materials: ABS, TPE, stainless steel. Sensing: Thermistor. Energy: 3.0V CR2032 battery. Form factor: Handheld with foldable probe. Connectivity: None. Standards: ISO 80601-2-56, ASTM E1112, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for human body temperature measurement in people of all ages, including children under 8 with adult supervision. Measurement sites include oral, rectal, and axillary. No specific contraindications listed.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it, also in blue.
July 8, 2020
Joytech Healthcare Co., Ltd Mr. Yunhua Ren General Manager No. 365, Wuzhou Road, Yuhang Economic Development Zone Hangzhou City 311100 Zhejiang China
Re: K200599
Trade/Device Name: Digital Thermometer, Model DMT-4756 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 5, 2020 Received: June 8, 2020
Dear Mr. Yunhua Ren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200599
Device Name Digital Thermometer, Model DMT-4756
Indications for Use (Describe)
The Digital Thermometer DMT-4756 is intended to measure the human body temperature in regular mode orally, rectally or under the arm, and the device is reusable for clinical or home use on people of all ages, including children under 8 with adult supervision.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The word "JOYTECH" is in blue, except for the "T", which is in red. To the right of the word "JOYTECH" are some Chinese characters in blue.
# 510(k) Summary
The assigned 510(k) number is: K200599
- 1. Date Prepared: 2020.07.08
#### 2. Submitter's Identification:
Name: JOYTECH HEALTHCARE CO., LTD.
Add.:No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou city,
311100 Zhejiang,China
Contact Person: Yunhua Ren
Phone: +86-571-81957767
Fax: +86-571-81957750
Email: RENYH@SEJOY.COM
#### 3. Name of the Device:
Trade Name: Digital Thermometer Model: DMT-4756 Common Name: Digital Thermometer Classification name: Clinical Electronic Thermometer
### 4. Classification Information:
Product Code: FLL- Clinical Electronic Thermometer Device Class: II Panel: 80 Regulation number:880-2910
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The word "JOYTECH" is in blue, except for the "T", which is in red. To the right of the word "JOYTECH" are three Chinese characters in blue.
#### Predicate Device Information: 5.
The Digital Thermometer DMT-4756 is substantially equivalent to the following device:
| 510k number | model | Product code | manufacturer |
|-------------|----------|--------------|------------------------------|
| K183393 | DMT-4726 | FLL | JOYTECH HEALTHCARE CO., LTD. |
#### Intended use / Indication for Use: 6.
The Digital Thermometer DMT-4756 is intended to measure the human body temperature in regular mode orally, rectally or under the arm, and the device is reusable for clinical or home use on people of all age, including children under 8 with adult supervision.
## 7. Device Description:
The digital thermometer DMT-4756 is hand held device which can measure human body's temperature orally, axillary (under the arm), and rectally. The results can be displayed on LCD.
The flexible tip is foldable,when the flexible tip folded, DMT-4756 is used to measure human body's temperature rectally, while the flexible tip unfolded, it is used to measure temperature orally and axillary (under the arm).
The digital thermometer have several functions,such as beep alarm,unit switchable, low battery detection,memories,backlight,auto power off functions.
DMT-4756 is a predictive digital thermometer. Users only need 10 seconds for predictive quick read and 30 seconds for final temperature readings.
### 8. Substantial Equivalence Comparison:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for JOYTEC-11建拓. The text "JOYTEC" is in blue, with the "C" stylized with a red triangle. The "11建拓" is in blue and is to the right of the "JOYTEC" text.
| SE<br>Comparisons | Subject device K200599<br>DMT-4756 | Predicate device K183393<br>(Model:DMT-4726) | Comparison<br>Result |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Intended Use<br>/Indication for use | The Digital Thermometer DMT-4756 is<br>intended to measure the human body<br>temperature in regular mode orally,<br>rectally or under the arm, and the<br>device is reusable for clinical or home<br>use on people of all age, including<br>children under 8 with adult supervision | The Digital Thermometer DMT-4726 is<br>intended to measure the human body<br>temperature in regular mode orally,<br>rectally or under the arm. The device is<br>reusable for clinical or home use on<br>people of all ages, including children<br>under 8 with adult supervision. | Identical<br>(Only change<br>in model<br>number) |
| Fundamental<br>technology&<br>Operating<br>principle | A change of thermistor resistance,<br>caused by changes of temperature. The<br>resistance is measured by MCU, so<br>changes of temperature will correspond<br>to changes of resistance | A change of thermistor resistance,<br>caused by changes of temperature. The<br>resistance is measured by MCU, so<br>changes of temperature will correspond<br>to changes of resistance | Identical |
| Sensor | Thermistor | Thermistor | Identical |
| Signal processing<br>and display | -Internal firmware and local<br>LCD display | -Internal firmware and local<br>LCD display | Identical |
| Measurement<br>Site | orally, rectally or under the arm | orally, rectally or under the arm | Identical |
| Range | $32.0°C~43.9°C (89.6°F-111.0°F)$ | $32.0°C~43.9°C (89.6°F-111.0°F)$ | Identical |
| Accuracy | $±0.1°C$ between $35.5°C$ to<br>$42.0°C (±0.2°F,95.9°F-107.6°F)$ ,<br>$±0.2°C$ under $35.5°C$ or over<br>$42.0°C (±0.4°F$ under $95.9°F$ or over<br>$107.6°F)$ | $±0.1°C$ between $35.5°C$ to<br>$42.0°C (±0.2°F,95.9°F-107.6°F)$ ,<br>$±0.2°C$ under $35.5°C$ or over<br>$42.0°C (±0.4°F$ under $95.9°F$ or over<br>$107.6°F)$ | Identical |
| Display resolution | $0.1 °C/0.1 °F$ | $0.1 °C/0.1 °F$ | Identical |
| components | Sensor, buzz film, housing, stainless<br>steel cap, LCD display, measurement<br>control module. | Sensor, buzz film, housing, stainless<br>steel cap, LCD display, measurement<br>control module. | Identical |
| Memories | 10 memories | 10 memories | Identical |
| Material | ABS,TPE,Stainless steel | ABS,TPE,Stainless steel | Identical |
| Operating range | Temperature:<br>$41°F~104°F (5°C~40°C)$<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure :<br>700hPa ~ 1060hPa | Temperature:<br>$41°F~104°F (5°C~40°C)$<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure :<br>700hPa ~ 1060hPa | Identical |
| Construction | The probe is foldable into the body of the thermometer | Flexible Digital Thermometer | Different<br>(Note1) |
| Measurement method & time | 10s Predictive read<br>30s Final read. | Approximate 5 ~ 10s | Different<br>(Note 2) |
| Weight | Approx. 40 grams including battery | Approx. 23 grams including battery | Different<br>(Note3) |
| Thermometer Size | 10.6cm*3.2cm*2.5cm (L x W x H) | 13.5cm*3.4cm*1.7cm (L x W x H) | Different<br>(Note3) |
| LCD Size | 2.5cm*1.4cm | 2.7cm*2.1cm | Different<br>(Note3) |
| Color &Coding | White(R9003),<br>Green(P564C) | White(R9003),<br>Blue(P294C) | Different<br>(Note3) |
| Predictive mode | Yes | Optional | Different<br>(Note2) |
| Storage and transportation condition | Temperature:<br>-4°F~131°F (-20℃~55℃)<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure:<br>700hPa ~ 1060hPa | Temperature:<br>-4°F~131°F (-20℃~55℃)<br>Relative humidity: 15%~95%RH<br>Atmospheric Pressure:<br>700hPa ~ 1060hPa | Identical |
| Battery type | One 3.0V CR2032 battery | One 3.0V CR2032 battery | Identical |
| Biocompatibility | Comply with ISO 10993-5 and ISO 10993-10 | Comply with ISO 10993-5 and ISO 10993-10 | Identical |
| Electrical Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | Identical |
| EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | Identical |
{6}------------------------------------------------
### Analysis
From the comparison table, the subject device and predicate device(K183393) have the similar Intended use & Indications for Use, Measurement place, Scale selection, Display screen, Auto power-off while no operation & Conformance standard;
There are slightly differences between the subject device and predicate device(K183393) as follows and we have conducted the tests to verify that the differences do not raise new questions of safety and effectiveness,please see below::
| Note 1 | Add foldable<br>construction function | Performance testing shows the subject device<br>complies with standard ISO 80601-2-56. The<br>difference does not raise new questions of safety and<br>effectiveness |
|--------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Note 2 | Measure time changed<br>from approximate<br>5~10s to “10 second<br>predictive quick read,30 | Performance test including clinical test shows the<br>subject device complies with performance standard.<br>The differences do not raise new questions of safety<br>and effectiveness |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. To the right of the word JOYTECH is Chinese characters in blue.
| | second for final<br>temperature reading",<br>and predictive mode<br>changed from optional<br>to have this function | |
|--------|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Note 3 | Weight,Size,color<br>changed; | Biocompatibility test and performance bench test met<br>the requirements in the standards. The differences do<br>not raise new questions of safety and effectiveness |
## 9. Performance data
The following performance data were provided in support of the substantial equivalence determination:
Performance testing was conducted to validate and verify that Digital Thermometer, DMT-4756 met all requirements of related international standards, including electrical safety, EMC, biocompatibility, software validation and product specifications. Results of these tests demonstrate compliance to the requirements of the below consensus standards.
Electrical Safety and performance requirements:
- AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 Medical Electrical Equipment
- · ISO 80601-2-56:2017 · Medical electrical equipment Part 2-56 Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- . ASTM E1112:00(Reapproved 2011) Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
Home-used medical equipment requirements and environmental test:
- · IEC 60601-1-11:2015 General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home health care environment
Electromagnetic compatibility requirements:
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Biocompatibility Evaluation for patient contacting components:
- · ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- · ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for JOYTECH. The logo is blue and red. To the right of the word JOYTECH are three Chinese characters.
Guidance Document:
· Guidance on the content of Premarket Notifications [510(k)] Submissions for clinical electronic thermometers
The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005.
#### 10. Discussion of Clinical Tests Performed:
The clinical tests were conducted on the DMT-4756 for each measurement site. The clinical tests evaluated 450 of subjects. The thermometer was evaluated in three groups 1) infants-newborn to one year; 2) children-greater than one to five years; and 3) adults-greater than five years old.The clinical performance test protocol and data analysis was conducted in accordance with ISO 80601-2-56. The test report showed the clinical performance of the subject device complied with the requirement of ISO 80601-2-56.
#### 11. Conclusions:
Based on the performance testing,comparison and analysis in the submission, the subject digital thermometer DMT-4756 are substantially equivalent to the predicate thermometer DMT-4726(K183393).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.