SafePath Suturing System - Silk Suture The SafePath Suturing System is intended for use in placement of a silk suture in the skin and subcutaneous tissues. SafePath Suturing System - Polyamide (Nylon) Suture The SafePath Suturing System is intended for use in placement of a nylon suture in the skin and subcutaneous tissues. SafePath Suturing System - Polypropylene Suture The SafePath Suturing System is intended for use in placement of a polypropylene suture in the skin and subcutaneous tissues.
Device Story
SafePath Suturing System consists of a hand-held, manually operated suturing device pre-loaded with a sterile, nonabsorbable surgical suture (silk, nylon, or polypropylene) swaged to a stainless steel reverse cutting needle. Device used by clinicians to place sutures in skin and subcutaneous tissues. Upon manual actuation, device places a single stitch through tissue; needle is automatically captured and reset for subsequent stitches. System facilitates tissue approximation. Benefits include simplified, efficient suture placement compared to manual needle handling. Device is single-patient use.
Clinical Evidence
Bench testing only. Performance evaluated against USP requirements for nonabsorbable sutures, including diameter, needle attachment, and tensile strength. Biocompatibility testing included cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity, and endotoxicity. All tests met or exceeded established criteria.
Technological Characteristics
Nonabsorbable sutures: Silk (braided, silicone-coated), Nylon (monofilament 6,6), Polypropylene (monofilament). All meet USP standards. Needles: 1/2 circle, 26 mm, reverse cutting, stainless steel. Manual hand-held suturing device mechanism for needle capture and reset. Sterilized for single-patient use.
Indications for Use
Indicated for placement of silk, nylon, or polypropylene sutures in skin and subcutaneous tissues for soft tissue approximation.
Regulatory Classification
Identification
Natural nonabsorbable silk surgical suture is a nonabsorbable, sterile, flexible multifilament thread composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori ) of the family Bombycidae. Natural nonabsorbable silk surgical suture is indicated for use in soft tissue approximation. Natural nonabsorbable silk surgical suture meets the United States Pharmacopeia (U.S.P.) monograph requirements for Nonabsorbable Surgical Suture (class I). Natural nonabsorbable silk surgical suture may be braided or twisted; it may be provided uncoated or coated; and it may be undyed or dyed with an FDA listed color additive.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
October 18, 2024
SafePath Medical, Inc. Howard Schrayer Regulatory Consultant 8 Lookout Hilton Head Island, South Carolina 29928
Re: K222944
Trade/Device Name: SafePath Suturing System-Silk Suture, SafePath Suturing System Polyamide Suture (Nylon), SafePath Suturing System- Polypropylene Suture Regulation Number: 21 CFR 878.5030 Regulation Name: Natural Nonabsorbable Silk Surgical Suture Regulatory Class: Class II Product Code: GAP, GAR, GAW Dated: May 22, 2023 Received: May 23, 2023
Dear Howard Schrayer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tek N.
Lamichhane -S
Digitally signed by
Tek N. Lamichhane -S
Date: 2024.10.18
13:01:04 -04'00'
Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222944
### Device Name
SafePath Suturing System - Silk Suture SafePath Suturing System – Polyamide (Nylon) Suture SafePath Suturing System - Polypropylene Suture
Indications for Use (Describe)
SafePath Suturing System - Silk Suture
The SafePath Suturing System is intended for use in placement of a silk suture in the skin and subcutaneous tissues.
SafePath Suturing System - Polyamide (Nylon) Suture The SafePath Suturing System is intended for use in placement of a nylon suture in the skin and subcutaneous tissues.
SafePath Suturing System - Polypropylene Suture
The SafePath Suturing System is intended for use in placement of a polypropylene suture in the skin and subcutaneous tissues.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) SUMMARY – K222944 (Per 21 CFR 807.92)
# General Company Information
| Name: | SafePath Medical, Inc. |
|-----------------------------------|------------------------------------------------------|
| Contact: | Howard Schrayer<br>Regulatory Affairs Consultant |
| Address: | 21 Water Street - 5th Floor<br>Amesbury, MA 01913 |
| Telephone: | 978-697-3267 |
| Date Prepared | October 16, 2024 |
| <b>General Device Information</b> | |
| Product Name: | SafePath Suturing System – Silk Suture |
| Classification: | Silk Suture<br>Product code: GAP – Class II |
| Regulation: | 21 CFR 878.5030 |
| Product Name: | SafePath Suturing System – Polyamide (Nylon) Suture |
| Classification: | Nylon Suture<br>Product code: GAR – Class II |
| Regulation: | 21 CFR 878.5020 |
| Product Name: | SafePath Suturing System -Polypropylene Suture |
| Classification: | Polypropylene Suture<br>Product code: GAW – Class II |
| Regulation: | 21 CFR 878.5010 |
| Classification: | Suture Passer<br>Product code: GAB - Class I |
| Regulation: | 21 CFR 878.4800 |
| <b>Predicate Devices</b> | |
| Primary Predicate | SafePath Suturing System (with silk suture |
| SafePath Medical | |
suture passer) [510(k) K180701] and
{6}------------------------------------------------
# Secondary Predicate Devices
| SM ENG CO Ltd. | REXLON® Nylon Suture<br>[510(k) K161633] |
|----------------|---------------------------------------------------|
| SM ENG CO Ltd. | REXLENE® Polypropylene Suture<br>[510(k) K173747] |
## Description
The silk suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of the organic protein, fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. The silk suture is silicone coated, braided and dyed (black) with logwood extract. The silk suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP size 2-0. The suture is provided as a 36 in. Jength, swaged on a single-armed 1½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
The nylon suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of long chain aliphatic polymers - nylon 6,6. SafePath nylon sutures are dyed with Logwood extract, FDA approval number §73.1410. The nylon suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 2-0 and 3-0. The suture is provided as a 36 in. length, swaged to a singlearmed ½ circle. 26 mm reverse cutting needle. The suture is pre-loaded in a handheld, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
The polypropylene suture provided with the SafePath Suturing System is a monofilament synthetic non-absorbable, sterile surgical suture composed of an isotactic crystalline stereoisomer of polypropylene, a synthetic linear polyolefin. The molecular formula is (C3H6)n. The Safepath suture is dyed with [Phthalocyaninato(2-)] copper, FDA approval number §74.3045. The polypropylene suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 2-0 and 3-0. The suture is provided as a 36 in. length, swaged to a single-armed ½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
Finished devices may be packaged individually, or in multi-unit cartons or procedure packs. The SafePath Suturing System is designed to place a suture in skin and subcutaneous tissue.
{7}------------------------------------------------
# Intended Use / Indications for Use
### SafePath Suturing System - Silk Suture
The SafePath Suturing System is intended for use in placement of a silk suture in the skin and subcutaneous tissues.
### SafePath Suturing System - Polyamide (Nylon) Suture
The SafePath Suturing System is intended for use in placement of a nylon suture in the skin and subcutaneous tissues.
### SafePath Suturing System - Polypropylene Suture
The SafePath Suturing System is intended for use in placement of a polypropylene suture in the skin and subcutaneous tissues.
#### Substantial Equivalence
A table comparing the SafePath device with a predicate silk suture is provided below. This submission supports the position that the SafePath Suturing System is substantially equivalent to previously cleared devices, including those listed above. A number of predicate devices list the same clinical use.
### SUBSTANTIAL EQUIVALENCE INFORMATION
#### Predicate Devices
### SafePath Medical, Inc.
SafePath Suturing System - K180701 REXLON Nylon Suture - K161633 REXLENE Polypropylene Suture K173747 SafePath Suturing System K222944
## Similarities and Differences
Silk suture is indicated for use in skin and subcutaneous tissue and other soft tissue applications, excluding use in cardiovascular and neural tissue
REXLON is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
REXLENE is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
SafePath silk suture is indicated for use in skin and subcutaneous tissue.
SafePath nylon suture is indicated for use in skin and subcutaneous tissue.
SafePath polypropylene suture is indicated for use in skin and subcutaneous tissue.
{8}------------------------------------------------
Regulation: 21 CFR 878.5030 Regulation: 21 CFR 878.5020 Regulation: 21 CFR 878.5010
Product code: GAP - Class II Product code: GAR - Class II Product code: GAW - Class II
Suture is provided with and without preattached (swaged) needle
Suture implant component is nonabsorbable.
Device is provided sterile for single-patientuse.
Suture is provided without a suture passer component
Silk suture is dyed black with logwood extract and coated with silicone for lubricity
Nylon suture is dyed black with logwood extract
Polypropylene suture is dyed blue with ([Phthalocyaninato (2-)] Copper)
Suture material is manufactured by Ashaway Line & Twine Mfg. Co.
Suture is provided with swaged, stainless steel needle
Suture passes all required tests for USP non-absorbable suture material and needle attachment
Silk suture is braided
Nylon suture is monofilament
Polypropylene suture is monofilament
Regulation: 21 CFR 878.5030 Regulation: 21 CFR 878.5020 Regulation: 21 CFR 878.5010
Product code: GAP - Class II Product code: GAR - Class II Product code: GAW - Class II
Suture is provided with pre-attached (swaged) needle
Suture implant component is nonabsorbable.
Device is provided sterile for single-patientuse.
Suture is pre-loaded in manually operated suturing device
Silk suture is dyed black with logwood extract and coated with silicone for lubricity
Nylon suture is dyed black with logwood extract
Polypropylene suture is dyed blue with ([Phthalocyaninato (2-)] Copper)
Suture material is manufactured by Ashaway Line & Twine Mfg. Co.
Suture is provided with swaqed, stainless steel needle
Suture passes all required tests for USP non-absorbable suture material and needle attachment
Silk suture is braided
Nylon suture is monofilament
Polypropylene suture is monofilament
{9}------------------------------------------------
# Performance Testing - Non-Clinical
# Physical Testing
Non-clinical testing has been performed in accordance USP (United States Pharmacopeia) Nonabsorbable Sutures - Diameter, Sutures- Needle Attachment, and Tensile Strength. In all cases, the USP criteria were met or exceeded.
# Biocompatibility Testing
The biocompatibility evaluation for the SafePath Suturing System (with silk suture and the suture passer) was referenced from the previously cleared SafePath submission K180701. For the nylon and polypropylene sutures, the safety evaluation was conducted based on the following considerations: (i)The nylon and polypropylene sutures are made from identical raw materials as the predicate device. (ii) The packaging and sterilization conditions are the same as those used for the predicate device. (iii) The manufacturing information from raw materials to the packaging was leveraged from the primary predicate device (K180701) for the silk suturing system.
A risk assessment was conducted on the raw materials and manufacturing processes for the nylon and polypropylene sutures to evaluate the safety of the device. The additional safety information was leveraged from the previously cleared K180701 submission.
For those tests conducted, the battery of testing included the following evaluations:
- Cytotoxicity
- Sensitization ●
- Irritation ●
- Systemic Toxicity ●
- Hemocompatibility ●
- Pyrogenicity
- Endotoxicity ●
The SafePath Suturing System met the test criteria for each of the studies.
{10}------------------------------------------------
# Conclusions
SafePath Medical, Inc. believes that the information provided establishes that similar legally marketed devices including the referenced predicate devices, have been used for the same clinical application as the SafePath Medical SafePath Suturing System.
The safety and performance of the subject devices have been evaluated and found to be substantially equivalent to the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.