K161633 · Sm Eng Co., Ltd. · GAR · Mar 2, 2017 · General, Plastic Surgery
Device Facts
Record ID
K161633
Device Name
REXLON, REXSIL
Applicant
Sm Eng Co., Ltd.
Product Code
GAR · General, Plastic Surgery
Decision Date
Mar 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
REXLON is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue. REXSIL is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
Device Story
REXLON and REXSIL are sterile, nonabsorbable surgical sutures used for soft tissue approximation and ligation. REXLON is a monofilament polyamide (Nylon 6 and 6,6); REXSIL is a braided silk suture (fibroin protein from Bombyx mori) coated with silicone. Both are available with or without stainless steel (STS 304) needles. Used by clinicians in surgical settings. Sutures provide mechanical tissue closure; performance verified against USP standards for diameter, tensile strength, and needle attachment. Biocompatibility confirmed per ISO 10993. Sterilized via ethylene oxide. Benefits include secure tissue approximation and visibility (dyed black/blue or natural).
Clinical Evidence
Bench testing only. No clinical data. Testing included USP 35 <861> (diameter), <871> (needle attachment), <881> (tensile strength), <71> (sterility), <85> (bacterial endotoxin), and <151> (pyrogen). Biocompatibility testing performed per ISO 10993-1, -3, -4, -5, -6, -7, -10, and -11. Shelf life supported by real-time and accelerated aging studies.
Indicated for general soft tissue approximation and/or ligation, including ophthalmic procedures. Contraindicated for cardiovascular or neural tissue. No specific age or gender restrictions.
Regulatory Classification
Identification
Nonabsorbable polyamide surgical suture is a nonabsorbable, sterile, flexible thread prepared from long-chain aliphatic polymers Nylon 6 and Nylon 6,6 and is indicated for use in soft tissue approximation. The polyamide surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. monograph for nonabsorbable surgical sutures; it may be monofilament or multifilament in form; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized symbol. The symbol consists of three human profiles facing right, stacked on top of each other, with flowing lines extending from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 2, 2017
SM ENG CO., Ltd % Mr. Sanglok Lee Wise Company, Inc. #306, 3 Daerim-ro 27ga-gil Yeongdeungpo-gu, 07413 KR
Re: K161633
Trade/Device Name: REXLON®, REXSIL® Regulation Number: 21 CFR 878.5020 Regulation Name: Nonabsorbable polyamide surgical suture Regulatory Class: Class II Product Code: GAR, GAP Dated: January 24, 2017 Received: January 31, 2017
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
#### 510(k) Number (if known)
#### K161633
Device Name
REXLON
Indications for Use (Describe)
REXLON' is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the nollection of informating data sources, gather and needed and complete of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
# 510(k) Number (if known)
## K161633
Device Name
REXSIL
Indications for Use (Describe)
REXSIL is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (1/14)
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review institutions, search existing data sources, gather and maintain the data needed and complete and reviouslion of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{7}------------------------------------------------
# 510(k) Summary
# The assigned 510(k) Number: K161633
# Date of Submission: January 24, 2017
# Applicant
Company name: SM ENG CO., LTD Address: 46, Nakdong-daero 1302beon-gil, Sasang-gu, Busan, Korea TEL: +82 51 305 8016 FAX: +82 51 305 8021 Email: leesg@sm-eng.net
# Submission Correspondent
Sanglok, Lee Wise COMPANY Inc. #1005, 11-19, Gamasan-ro 27-gil, Guro-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
# Subject Device Information
#### 1. Nylon Monofilament Suture With or Without Needle
#### 1.1 Proposed Device Identification
Proprietary Name: REXLON Common Name: Nonabsorbable Polyamide Suture With or Without Needle Device Class: Class II Regulation Number: 21 C.F.R. 878.5020 Product Code: GAR Device Classification Name: Suture, Nonabsorbable, Synthetic, Polyamide
#### 1.2 Indication for use
REXLON is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
{8}------------------------------------------------
# 1.3 Predicate devices
Predicate device 01 510(k) Number: K080684 Device Name: WG-Surgical Sutures with Needle Classification Name: Suture, Nonabsorbable, Synthetic, Polyamide Manufacturer: FOOSIN MEDICAL SUPPLIES INC.LTD.
Predicate device 02 510(k) Number: K982646 Device Name: Reli Classification Name: SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE Manufacturer: AILEE CO LTD.
# 1.4 Device Description
REXLON is a sterilized nonabsorbable monofilament surgical suture made out of polyamide used in general soft tissue approximation and/or ligation with or without needle made out of Stainless Steel STS 304.
REXLON suture is a nonabsorbable, sterile surgical monofilament suture composed of the long-chain aliphatic polymers Nylon 6 and Nylon 6,6. REXLON sutures are not coated. The sutures are dyed black (Logwood) or blue (FD&C Blue No.2) to enhance visibility in tissue. The suture is also available undyed (Natural).
REXLON suture meets all requirements established by the United States Pharmacopoeia (U.S.P.) for nonabsorbable surgical suture. Available suture size is as below.
- Dyed suture Black (Logwood): USP 11-0 ~ USP 1
- i Dyed suture Blue (FD&C Blue No.2): USP 8-0 ~ USP 2
- Undyed suture (Natural): USP 8-0 ~ USP 1
# 1.5 Non-Clinical Test Conclusion
USP 35 <861> SUTURES - DIAMETER
USP 35 <871> SUTURES - NEEDLE ATTACHMENT
USP 35 <881> TENSILE STRENGTH
USP Nonabsorbable Surgical Suture
ASTM F88-09. Standard Test Method for Seal Strength of Flexible Barrier Materials;
ASTM F1929-98(2004) Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
USP <71> STERILITY TEST
USP 36, <151>; Pyrogen Test
ISO 11135-1:2007 Sterilization of health care products- Ethylene oxide-Part 1: Requirements for development, validation and routing control of a sterilization process for medical devices:
USP <85> Bacterial Endotoxin Limit;
ISO 10993, Biological Evaluation of Medical Devices: Including:
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (ISO 10993-1:2009)
Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity (ISO 10993-3:2014)
Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood (ISO 10993-4:2002, including Amd 1:2006)
Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (ISO 10993-5:2009)
{9}------------------------------------------------
Biological evaluation of medical devices - Part 6: Tests for local effects after implantation (ISO 10993-6:2007)
Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals (ISO 10993-7:2008)
Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (ISO 10993-10:2010)
Biological evaluation of medical devices - Part 11: Tests for systemic toxicity (ISO 10993-11:2006)
REXLON were evaluated in accordance with the listed tests above.
Real-time and accelerated aging stability testing was performed to support shelf life of REXLON. Testing confirmed that the sutures conform to USP requirements for nonabsorbable sutures, are biocompatible per ISO 10993-1, and are stable over the proposed shelf life
# 1.6 Substantially Equivalent Conclusion
| | Table 1: Substantial Equivalence Comparison | |
|--|---------------------------------------------|--|
| | | |
| Property | Proposed device<br>REXLON | Predicate device 01<br>WG-Surgical Sutures<br>with Needle | Predicate device 02<br>Reli |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K161633 | K080684 | K982646 |
| Product Code | GAR | GAR | GAR |
| Regulation No, | 21 C.F.R. 878.5020 | 21 C.F.R. 878.5020 | 21 C.F.R. 878.5020 |
| Class | II | II | II |
| Sterile | Yes | Yes | Yes |
| Single Use | Yes | Yes | Yes |
| Intended use | Indication for use is in<br>general soft tissue<br>approximation and/or<br>ligation, including use in<br>ophthalmic procedures, | Indication for use is in<br>general soft tissue<br>approximation and/or<br>ligation, including use in<br>ophthalmic procedures. | Indication for use as non<br>absorbable sutures is in<br>general soft tissue<br>approximation and/or<br>ligation, including use in |
| | but not for use in<br>cardiovascular or neural<br>tissue. | | cardiovascular,<br>ophthalmic, neural<br>tissue. |
| Configuration | Nonabsorbable<br>Polyamide Suture With<br>Needle | Nonabsorbable<br>Polyamide Suture With<br>Needle | Nonabsorbable<br>Polyamide Suture With<br>Needle |
| | | Suture | |
| Raw Suture Material | polyamide surgical<br>suture from long-chain<br>aliphatic polymers Nylon<br>6 and Nylon 6,6 | polyamide surgical<br>suture from long-chain<br>aliphatic polymers Nylon<br>6 and Nylon 6,6 | polyamide surgical<br>suture from long-chain<br>aliphatic polymers Nylon<br>6 and Nylon 6,6 |
| Suture Manufacturer | Ashaway Line & Twine<br>Mfg. Co. | Pearsalls Limited | Ashaway Line & Twine<br>Mfg. Co. |
| Color (Colorant) | Dyed suture Black<br>(Logwood), Blue<br>(FD&C Blue No.2) and<br>Undyed suture (Natural) | Dyed suture Black<br>(Logwood), Blue &<br>phthalocyanine blue<br>(FD&C Blue No.2) and<br>Undyed suture (Natural) | Dyed suture Black<br>(Logwood), Blue<br>(FD&C Blue No.2) and<br>Undyed suture (Natural) |
| Coating material | Not used | Not used | Not used |
| Absorbable/Nonabsorb<br>able | Nonabsorbable | Nonabsorbable | Nonabsorbable |
| Braided/Monofilament | Monofilament | Monofilament | Monofilament |
| Suture Size | Available suture sizes<br>are standard according<br>to USP requirements<br>Available suture size is<br>as below.<br>#1 Dyed suture Black<br>(Logwood): USP 11-<br>0~USP 1<br>#2 Dyed suture Blue<br>(FD&C Blue No.2): USP<br>8-0~USP 2<br>#3 Undyed suture<br>(Natural): USP 8-0~USP<br>1 | Available suture sizes<br>are standard according<br>to USP requirements | Available suture sizes<br>are standard according<br>to USP requirements |
| | | | |
| Length of Suture | 15cm, 20cm, 30cm,<br>35cm, 40cm, 45cm,<br>50cm 60cm, 70cm,<br>75cm, 80cm, 100cm,<br>150cm, 200cm | Unknown | 15cm, 20cm, 30cm,<br>45cm, 50cm |
| Diameter of Suture | The suture diameters of<br>proposed device comply<br>with the diameter<br>requirement listed in<br>U5P 35 <861> Diameter. | The suture diameters of<br>proposed device comply<br>with the diameter<br>requirement listed in<br>U5P 35 <861> Diameter. | Meet the requirements<br>defined in the USP |
| Tensile strength | The tensile strengths of<br>proposed device comply<br>with the tensile<br>requirement listed in<br>U5P 35 <881> Tensile<br>Strength | The tensile strengths of<br>proposed device comply<br>with the tensile<br>requirement listed in<br>U5P 35 <881> Tensile<br>Strength | Meet the requirements<br>defined in the USP |
| Needle Attachment | The bond between<br>suture and needle of the<br>applicant device meet<br>the requirements defined<br>in USP 35 <871>. | The bond between<br>suture and needle of the<br>applicant device meet<br>the requirements defined<br>in USP 35 <871>. | Meet the requirements<br>defined in the USP |
| Needle | | | |
| Material | Stainless Steel | Stainless Steel | Stainless Steel |
| Needle type | Reverse Cutting, Taper<br>point, Spatula | Taper, cutting, spatual,<br>blund ,taper cut,<br>diamond, premimum<br>cutting and square | Spatula, taper point,<br>Blunt point |
{10}------------------------------------------------
{11}------------------------------------------------
The proposed device, REXLON, is determined to be Substantially Equivalent (SE) to the predicate device, WG-Surgical Sutures with Needle (K080684) and Reli (K982646) in respect of safety and effectiveness.
{12}------------------------------------------------
# 2. Nonabsorbable Silk Suture With or Without Needle
#### 2.1 Proposed Device Identification
Proprietary Name: REXSIL Common Name: Nonabsorbable Braided Silk Suture With or Without Needle Device Class: Class II Regulation Number: 21 C.F.R. 878.5030 Product Code: GAP Device Classification Name: Suture, Nonabsorbable, Silk
#### 2.2 Indication for use
REXSIL is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
#### 2.3 Predicate devices
Predicate device 01 510(k) Number: K080684 Device Name: WG-Surgical Sutures with Needle Classification Name: SUTURE, NONABSORBABLE, SILK Manufacturer: FOOSIN MEDICAL SUPPLIES INC.LTD
Predicate device 02 510(k) Number: K982646 Device Name: Reli Classification Name: SUTURE, NONABSORBABLE, SILK Manufacturer: AILEE CO LTD.
#### 2.4 Device Description
REXSIL is a sterilized nonabsorbable silk surgical suture made out of polyamide used in general soft tissue approximation and/or ligation with or without needle made out of Stainless Steel STS 304.
REXSIL suture is a nonabsorbable, sterile, surgical suture composed of an organic protein call fibroin. This protein is derived from the domesticated species Bombyx mori (b. More) of the family Bombycidae. REXSIL sutures are processed to remove the natural waxes and gums. REXSIL suture is dyed black (Logwood) and coated with silicone. REXSIL suture is also available undyed (Natural).
REXSIL suture meets all requirements established by the United States Pharmacopoeia (U.S.P.) for nonabsorbable surqical suture. Available suture size is as below.
- Dyed suture Black (Logwood): USP 8-0 ~ USP 2
- Undyed suture (Natural): USP 7-0 ~ USP 2
### 2.5 Non-Clinical Test Conclusion
USP 35 <861> SUTURES - DIAMETER USP 35 <871> SUTURES - NEEDLE ATTACHMENT USP 35 <881> TENSILE STRENGTH USP Nonabsorbable Surgical Suture ASTM F88-09. Standard Test Method for Seal Strength of Flexible Barrier Materials; ASTM F1929-98-98 Standard Test Method for Detecting Seal Leaks in Porous Medical
{13}------------------------------------------------
Package by Dye Penetration
USP <71> STERILITY TEST
USP <151>; Pyrogen Test
ISO 11135-1:2007 Sterilization of health care products- Ethylene oxide- Part 1: Requirements for development, validation and routing control of a sterilization process for medical devices;
USP <85> Bacterial Endotoxin Limit;
ISO 10993, Biological Evaluation of Medical Devices:
Includina:
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (ISO 10993-1:2009)
Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity (ISO 10993-3:2014)
Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood (ISO 10993-4:2002, including Amd 1:2006)
Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (ISO 10993-5:2009)
Biological evaluation of medical devices - Part 6: Tests for local effects after implantation (ISO 10993-6:2007)
Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals (ISO 10993-7:2008)
Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (ISO 10993-10:2010)
Biological evaluation of medical devices - Part 11: Tests for systemic toxicity (ISO 10993-11:2006)
REXSIL were evaluated in accordance with the listed tests above.
Real-time and accelerated aging stability testing was performed to support shelf life of REXSIL. Testing confirmed that the sutures conform to USP requirements for nonabsorbable sutures, are biocompatible per ISO 10993-1, and are stable over the proposed shelf life
# 2.6 Substantially Equivalent Conclusion
| Property | Proposed device<br>REXSIL | Predicate device 01<br>WG-Surgical Sutures<br>with Needle | Predicate device 02<br>Reli |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| 510(k) Number | K161633 | K080684 | K982646 |
| Product Code | GAP | GAP | GAP |
| Regulation No, | 21 C.F.R. 878.5030 | 21 C.F.R. 878.5030 | 21 C.F.R. 878.5030 |
| Class | II | II | II |
| Sterile | Yes | Yes | Yes |
| Single Use | Yes | Yes | Yes |
| Configuration | Nonabsorbable Silk<br>Suture and Needle | Nonabsorbable Silk<br>Suture and Needle | Nonabsorbable Silk<br>Suture |
| Suture | | | |
| Raw Suture Material | Silk suture | Silk suture | Silk suture |
| Suture Manufacturer | Ashaway Line & Twine<br>Mfg. Co. | Pearsalls Limited | Ashaway Line & Twine<br>Mfg. Co. |
| Coating material | Silicone coated | Silicone and wax coated | Silicone coated |
| Color (Colorant) | Dyed suture Black<br>(Logwood) and Undyed<br>suture (Natural) | Dyed suture Black<br>(Sulphol or Logwood)<br>and Undyed suture<br>(Natural) | Dyed suture Black<br>(Logwood) and Undyed<br>suture (Natural) |
| Absorbable/Nonabsorb<br>able | Nonabsorbable | Nonabsorbable | Nonabsorbable |
| Braided/Monofilament | Braided | Braided | Braided |
| Suture Size | Available suture sizes<br>are standard according<br>to USP requirements<br>Available suture size is<br>as below.<br>#1 Dyed suture Black<br>(Logwood): USP 8-0 ~ | Available suture sizes<br>are standard according<br>to USP requirements | Available suture sizes<br>are standard according<br>to USP requirements |
| | USP 2<br>#2 Undyed suture<br>(Natural): USP 7-0 ~<br>USP 2 | | |
| Length of Suture | 30cm, 35cm, 40cm,<br>45cm, 50cm, 55cm,<br>60cm, 70cm, 75cm,<br>100cm, 110cm, 150cm,<br>180cm, | Unknown | Unknown |
| Diameter of Suture | The suture diameters of<br>proposed device comply<br>with the diameter<br>requirement listed in<br>USP 35 <861><br>Diameter. | The suture diameters of<br>proposed device comply<br>with the diameter<br>requirement listed in<br>USP 35 <861><br>Diameter. | Meet the requirements<br>defined in the USP |
| Tensile strength | The tensile strengths of<br>proposed device comply<br>with the tensile<br>requirement listed in<br>USP 35 <881> Tensile<br>Strength | The tensile strengths of<br>proposed device comply<br>with the tensile<br>requirement listed in<br>USP 35 <881> Tensile<br>Strength | Meet the requirements<br>defined in the USP |
| Needle Attachment | The bond between<br>suture and needle of the<br>applicant device meet<br>the requirements defined<br>in USP 35 <871>. | The bond between<br>suture and needle of the<br>applicant device meet<br>the requirements defined<br>in USP 35 <871>. | Meet the requirements<br>defined in the USP |
| Needle | | | |
| Material | Stainless Steel | Stainless Steel | Stainless Steel |
| Needle type | Reverse Cutting, Taper<br>point, Spatula | Taper point, Reverse<br>Cutting, Conventional<br>cutting, Taper cutting,<br>Spatula, Blunt point | Spatula, taper point |
{14}------------------------------------------------
{15}------------------------------------------------
The proposed device, REXSIL, is determined to be Substantially Equivalent (SE) to the predicate device, WG-Surgical Sutures with Needle (K080684) and Reli (K982646) in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.