K080684 · Foosin Medical Supplies, Inc. · GAP · Jun 9, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080684
Device Name
WG-SURGICAL SUTURES WITH NEEDLE
Applicant
Foosin Medical Supplies, Inc.
Product Code
GAP · General, Plastic Surgery
Decision Date
Jun 9, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Nonabsorbable Silk Suture with Needle is indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures. Nonabsorbable Polypropylene Suture with Needle is indicated for use in soft tissue approximation and/or ligation including use in general cardiovascular, ophthalmic and neurological procedures. Nonabsorbable Polyester Suture with Needle is indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures. Nonabsorbable Polyamide suture with needle is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures.
Device Story
WG-Surgical Sutures with Needle consist of nonabsorbable surgical suture material (silk, polypropylene, polyester, or polyamide) attached to a stainless steel needle. Used by surgeons for soft tissue approximation and ligation in clinical settings. Devices are EO sterilized and pyrogen-free. Sutures provide mechanical closure of tissue; needle facilitates passage through tissue. Benefit is secure wound closure.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Nonabsorbable surgical sutures (silk, polypropylene, polyester, polyamide) with stainless steel needles. Braided silk; synthetic polymers for others. EO sterilized. Standalone manual surgical instruments.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in cardiovascular, ophthalmic, and neurological procedures (polyamide limited to general and ophthalmic). Prescription use only.
Regulatory Classification
Identification
Natural nonabsorbable silk surgical suture is a nonabsorbable, sterile, flexible multifilament thread composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori ) of the family Bombycidae. Natural nonabsorbable silk surgical suture is indicated for use in soft tissue approximation. Natural nonabsorbable silk surgical suture meets the United States Pharmacopeia (U.S.P.) monograph requirements for Nonabsorbable Surgical Suture (class I). Natural nonabsorbable silk surgical suture may be braided or twisted; it may be provided uncoated or coated; and it may be undyed or dyed with an FDA listed color additive.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
# 510(k) Summary
As required by section 807.92(c)
JUN - 9 2008
The Assigned 510(k) Number is:
Submitter Information:
### Manufacturer Name:
Foosin Medical Supplies Inc., Ltd. No.312, Shichang Road Weihai, Shandong, China, 264209
# Contact Person of the Submission:
Ms. Diana Hong Mr. Eric Chen Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhong Shan Nan Er Road Shanghai, China 20020 Phone: +86-21-64264467 x 152 Fax: +86-21-64264468 x 809 Email: Diana.hong(@mid-link.net eric.chen(@)mid-link.net
#### Subject Device Information
- Nonabsorbable Silk Suture with Needle 1.
# 1.1 Applicant Device Information
Trade/Proprietary Name: WG-Surgical Sutures with Needle Common Name: Nonabsorbable Silk Suture with Needle Classification Name: Suture, Nonabsorbable, Silk Device Class: II Product Code: GAP Regulation Number: 878.5030 Review Panel: General & Plastic Surgery Intended Use: Nonabsorbable Silk Suture with Needle is indicated for use in general soft tissue approximation
Shanghai Mid-Link Business Consulting Co., Ltd
{2}------------------------------------------------
and/or ligation including use in cardiovascular, ophthalmic and neurological procedures.
#### 1.2Predicate Device
#### K-number: K041514
# 1.3 Device Description
The applicant devices of nonabsorbable silk suture with needle consist of a silk surgical suture made of natural silk and a stainless steel needle. It is EO sterilized, and prygon-free. Nonabsorbable silk suture is braided.
#### 1.4 Substantially Equivalence Determination
The applicant device of nonabsorbable silk surgical suture is substantially equivalent to the predicate device.
#### 2. Nonabsorbable Polypropylene Suture with Needle
### 2.1Applicant Device Information
Trade/Proprietary Name: WG-Surgical Sutures with Needle Common Name: Nonabsorbable Polypropylene Suture with Needle Classification Name: Suture, Nonabsorbable, Synthetic, Polypropylene Device Class: II Product Code: GAW Regulation Number: 878.5010 Review Panel: General & Plastic Surgery Intended Use:
Nonabsorbable Polypropylene Suture with Needle is indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures.
#### 2.2Predicate Device
K-number: K070243
#### 2.3Device Description
The applicant devices of nonabsorbable polypropylene suture with needle consist of a silk surgical suture made of polypropylene and a stainless steel needle. It is EO sterilized, and prygon-free.
### 2.4Substantially Equivalence Determination
The applicant device of Nonabsorbable polypropylene Suture with Needle is substantially equivalent to the predicate device.
## 3. Nonabsorbable Polyester Suture with Needle
Shanghai Mid-Link Business Consulting Co., Ltd
{3}------------------------------------------------
#### 3.1Applicant Device Information
Trade/Proprietary Name: WG-Surgical Sutures with Needle Common Name: Nonabsorbable Polyester Suture with Needle Classification Name: Suture, Nonabsorbable, Synthetic, Polyethylene Devicc Class: Il Product Code: GAT Regulation Number: 878.5000 Review Panel: General & Plastic Surgery Intended Use: Nonabsorbable Polyester Suture with Needle is indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures.
#### 3.2Predicate Device
K-number: K060165
#### 3.3Device Description
The applicant devices of nonabsorbable polyester suture with needle consist of a polyester surgical suture made of polyester and a stainless steel needle. It is EO sterilized, and prygon-free.
# 3.4Substantially Equivalence Determination
The applicant device of Nonabsorbable Polyester Suture with Needle is substantially equivalent to the predicate device.
## 4.Nonabsorbable Polyamide Suture with Needle
#### 4.1Applicant Device Information
Trade/Proprietary Name: WG-Surgical Sutures with Needle Common Name: Nonabsorbable Polyamide Suture with Needle Classification Name: Suture, Nonabsorbable, Synthetic, Polyamide Devicc Class: II Product Code: GAR Regulation Number: 878.5020 Review Panel: General & Plastic Surgery Intended Use: Nonabsorbable Polyamide suture with needle is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures.
#### 4.2Predicate Device
K-number: K0060471
#### 4.3Device Description
Shanghai Mid-Link Business Consulting Co., Ltd
{4}------------------------------------------------
The applicant devices of nonabsorbable polyamide suture with needle consist of a polyamide surgical suture made of long-chain, aliphatic polymers nylon and a stainless steel needle. It is EO sterilized, and prygon-free.
# 4.4Substantially Equivalence Determination
The applicant device of nonabsorbable polyamide suture with necdle is substantially equivalent to the predicate device.
#### Shanghai Mid-Link Business Consulting Co., 1.td
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 9 2008
Foosin Medical Supplies Inc. % Shanghai Midlink Business Consulting Co., Ltd. Ms. Diana Hong Suite 8D, Zhongxin Zhongshan Mansion No. 19, Lane 999, Zhongshan No. 2 Road (S) Shanghai 20030 China
Re: K080684
Trade/Device Name: Nonabsorbable Silk Suture with Needle, Nonabsorbable Polypropylene Suture with Needle, Nonabsorbable Polyester Suture with Needle, Nonabsorbable Polyamide suture with needle Regulation Number: 21 CFR 878.5030 Regulation Name: Natural nonabsorbable silk surgical suture Regulatory Class: II Product Code: GAP, GAW, GAT, GAT, GAR Dated: May 2, 2008 Received: May 5, 2008
Dear Ms. Hansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 -- Ms. Peggy Hansen, RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{7}------------------------------------------------
K080654 1/4
510(k) Number:_
Device Name: ___ Nonabsorbable Silk Suture with Needle
Indications for Use:
Nonabsorbable Silk Suture with Needle is indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular , ophthalmic and neurological procedures.
Prescription Use √ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Johnson for San
Page 1 of 4
Division of General, Restorative, and Neurological Devices
510(k) Number.
{8}------------------------------------------------
12080684 2/4
Device Name: __ Nonabsorbable Polypropylene Suture with Needle
Indications for Use:
Nonabsorbable Polypropylene Suture with Needle is indicated for use in soft tissue approximation and/or ligation including use in general cardiovascular, ophthalmic and neurological procedures.
Prescription Use __ J (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use . (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil Rl. Ogden for mxa
Page 2 of 4
(Division Sign-Off Division of General, Restorative, and Neurological Devices
510(k) Number KD80684
{9}------------------------------------------------
1000684314
510(k) Number:***_***_
080684
Device Name:
Indications for Use:
Nonabsorbable Polyester Suture with Needle is indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures.
Prescription Use _ V (Part 21 CFR 801 Subpart D)
. AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mut. R.P. Ogle for M
St. Ogle
Page 3 of 4
Division of General, Restorative and Neurological Devices
510(k) Number
{10}------------------------------------------------
Premarket Notification 510(k) Submission—Indications for Use
Report No.: A20071012
080684 510(k) Number:
Device Name:
Indications for Usc:
Nonabsorbable Polyamide suture with needle is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures.
Prescription Use __ イ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 4 of 4
Neifredgalan fr mxm
Division of General and Neurological Devices
510(k) Number K080684
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.