The XIA® 4.5 Spinal System is interior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications: • Degenerative Disc Disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) - · Spondylolisthesis - · Trauma (i.e. fracture of dislocation) - · Spinal Stenosis - · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis) - Tumor - Pseudarthrosis - · Failed Previous Fusion The Stryker Spine DIAPASON® Spinal System, Opus® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. Except for the staples, when used for posterior non-cervical pedicle screw fixation in pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the XIA® 4.5 Spinal System is intended to treat petients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Device Story
The subject devices are a collection of previously cleared spinal fixation systems (pedicle screws, rods, hooks, connectors, and laminoplasty implants) used by surgeons in the OR to provide immobilization and stabilization of spinal segments as an adjunct to fusion. The primary purpose of this submission is to establish MR Conditional labeling for these existing implants. The devices are implanted during spinal surgery to correct deformities or stabilize the spine in patients with degenerative disease, trauma, or tumors. The surgeon selects components based on the patient's anatomy and clinical needs. The devices provide mechanical support to the spinal column to facilitate bony fusion or, in specific tumor cases, to restore structural integrity. No changes were made to the device designs or materials; the submission relies on bench testing to confirm MR compatibility.
Clinical Evidence
Bench testing only. MR Compatibility testing was performed in accordance with ASTM F2503. Results demonstrate that the subject devices meet prescribed acceptance criteria and are substantially equivalent to the predicate devices.
Technological Characteristics
Multiple component spinal fixation systems including pedicle screws, rods, hooks, and connectors. Materials are consistent with previously cleared versions. MR Conditional per ASTM F2503. No changes to fundamental scientific technology.
Indications for Use
Indicated for skeletally mature patients and pediatric patients (except staples) requiring non-cervical spinal fixation as an adjunct to fusion. Treats degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal deformities (scoliosis, kyphosis, lordosis), tumor, pseudarthrosis, and failed previous fusion. Pediatric use limited to posterior approach for progressive deformities.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
EVEREST Spinal System, RANGE (MESA and DENALI) Spinal System, CASPIAN OCT (MESA Mini and DENALI Mini) Spinal System, YUKON OCT Spinal System (K181603)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 8, 2023
Stryker Spine Sierra Mertz Staff Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401
## Re: K222684
Trade/Device Name: XIA® 4.5 Spinal System, XIA® 4.5 Cortical Trajectory, XIA® 3 Spinal System, Serrato® Spinal System, XIA® Growth Rod Conversion Set, XIA® II Spinal System, XIA® Precision System, XIA® Anterior, Diapason® Spinal System, Opus™ Spinal System, Radius® Spinal System, Mantis® Spinal System, Mantis® Redux, Trio® & Trio+ Spinal Fixation System, ES2™ Spinal System, ES2™ Augmentable Spinal System, Oasys® Occipito-Cervico-Thoracic System, Nile® Proximal Fixation Spinal System, Nile® Alternative Fixation Spinal System, and Escalate® Laminoplasty System Regulation Number: 21 CFR 888.3070
Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWO. KWP, PGM, NKG, OWI, NOW Dated: May 11, 2023 Received: May 12, 2023
Dear Sierra Mertz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image contains the text "Colin O'neill -S" in a simple, sans-serif font. The text is arranged vertically, with "Colin" on the top line and "O'neill -S" on the bottom line. In the background, there is a faded, light blue watermark of the letters "FDA".
Colin O'Neill. M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
## Indications for Use
510(k) Number (if known)
#### K222684
Device Name XIA® 4.5 Spinal System
Indications for Use (Describe)
The XIA® 4.5 Spinal System is interior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications:
• Degenerative Disc Disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- · Spondylolisthesis
- · Trauma (i.e. fracture of dislocation)
- · Spinal Stenosis
- · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudarthrosis
- · Failed Previous Fusion
The Stryker Spine DIAPASON® Spinal System, Opus® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion.
Except for the staples, when used for posterior non-cervical pedicle screw fixation in pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the XIA® 4.5 Spinal System is intended to treat petients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
#### K222684
Device Name Xia® 4.5 Cortical Trajectory
## Indications for Use (Describe)
The XIA® 4.5 Spinal System is interior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications:
• Degenerative Disc Disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- · Spondylolisthesis
- · Trauma (i.e. fracture of dislocation)
- · Spinal Stenosis
- · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudarthrosis
- · Failed Previous Fusion
The Stryker Spine DIAPASON® Spinal System, Opus® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion.
Except for the staples, when used for posterior non-cervical pedicle screw fixation in pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the XIA® 4.5 Spinal System is intended to treat petients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------|
|----------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
|---------------------------------------------|
|---------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known)
#### K222684
Device Name Xia® 3 Spinal System
Indications for Use (Describe)
The Xia® 3 Spinal System is intended for use in the non-cervical spine. When used as an anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the Xia® 3 Spinal System is intended to provide additional support during fusion using auto graft or allograft in skeletally mature patients in the following acute and chronic instabilities or deformities:
• Degenerative Disc Disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- Spondylolisthesis
- · Trauma (i.e. fracture of dislocation)
- · Spinal stenosis
- · Curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudarthrosis
- · Failed previous fusion
The 5.5 mm rods from the Stryker Spine Radius™ Spinal System and 6.0 mm Vitallium rods from the Xia® Spinal System are intended to be used with the other components of the Xia® 3 Spinal System.
When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
#### K222684
Device Name Xia® 3 Serrato® Spinal System
### Indications for Use (Describe)
The Xia® 3 Spinal System is intended for use in the non-cervical spine. When used as an anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the Xia® 3 Spinal System is intended to provide additional support during fusion using auto graft or allograft in skeletally mature patients in the following acute and chronic instabilities or deformities:
· Degenerative Disc Disease ( as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- Spondylolisthesis
- · Trauma (i.e. fracture of dislocation)
- · Spinal stenosis
- · Curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudarthrosis
- · Failed previous fusion
The 5.5 mm rods from the Stryker Spine Radius™ Spinal System and 6.0 mm Vitallium rods from the Xia® Spinal System are intended to be used with the other components of the Xia® 3 Spinal System.
When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K222684
Device Name Xia® Growth Rod Conversion Set
#### Indications for Use (Describe)
he Xia® Growth Rod Conversion Set is indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The Xia® Growth Rod Conversion Set may be used with any cleared Xia® 4.5 Spinal System rod construct. The Xia® Growth Rod Conversion Set is not intended for use in conjunction with staples.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------|------------------------------------------------------------------------|
| <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K222684
Device Name Xia® II Spinal System
### Indications for Use (Describe)
The Xia® and Xia II Titanium Spinal Systems are interior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: degenerative-disc disease (DOD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion.
The 6 mm diameter rods from the DIAPASON™ Spinal System and OPUS™ Spinal System are intended to be used with the other components of the Xia II Titanium Spinal Systems. The Titanium Multi-Axial Cross Connectors are intended to be used with the other components of the Xia II Titanium Spinal Systems.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{8}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K222684
Device Name Xia® Precision System
### Indications for Use (Describe)
The Xia® Titanium Spinal System is intended for anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: degenerative-disc disease (DOD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion.
The 6 mm diameter rods from the DIAPASON™ Spinal System and OPUS™ Spinal System are intended to be used with the other components of the Xia Titanium Spinal System. The Titanium Multi-Axial Cross Connectors are intended to be used with the other components of the XiaTitanium Spinal Systems.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{9}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K222684
Device Name Xia® Anterior
### Indications for Use (Describe)
The Xia® Titanium Spinal System is interior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: degenerative-disc disease (DOD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion.
The 6 mm diameter rods from the DIAPASON™ Spinal System and OPUS™ Spinal System are intended to be used with the other components of the Xia Titanium Multi-Axial Cross Connectors are intended to be used with the other components of the Xia Titanium Spinal Systems.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{10}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K222684
Device Name DIAPASON Spinal System
### Indications for Use (Describe)
The 6mm diameter rods from the DIAPASON Spinal System are intended to be used with the components of the XIA Titanium Spinal System. As a posterior, non-pedicle screw system of the Diapason Spinal system is indicated for long and short curve scoliosis, vertebral fracture or dislocation, spondylolisthesis, degenerative-disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies), previously failed fusion and spinal tumor. When used as a pedicle screw fixation system of the non-cervical spine in skeletally mature patients, the Diapason Spinal system is indicated for one or more of the following: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).
In addition, the Diapason Spinal System is indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grade 3 and 4) at the L5-S1 joint, having fusions with autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (with pedicle placement at L3 and below) with removal of the implants after the development of a solid fusion mass.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{11}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K222684
Device Name Opus™ Spinal System
### Indications for Use (Describe)
The OPUS™ System is intended for posterior, noncervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion.
The OPUS™ Spinal System is also intended to be used in conjunction with the titanium hooks from the OSS/Diapason Spinal System and the Xia Spinal System. The OPUS Spinal System is also intended to be used in conjunction with the Multi-Acis Cross Connectors.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{12}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
#### K222684
Device Name Radius® Spinal System
### Indications for Use (Describe)
The Radius® Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation system, the Radius® Spinal system is intended to provide additional support during fusion using autograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:
- · Degenerative Disc Disease (DOD) ( defined as back pain of discogenic origin with
degeneration of the disc confimned by history and radiographic studies):
- · Spondylolisthesis;
- · Trauma (i.e. fracture or dislocation);
- · Spinal Stenosis:
- · Curvature (i.e. scoliosis, kyphosis, and/or lordosis);
- Tumor:
- · Pseudoarthorisis:
- and Failed Previous Fusion
The Radius® Spinal System can also be linked to the XIA® Titanium Spinal System via the 05.5mm to 06.0mm Radius® rod-to-rod connector.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{13}------------------------------------------------
510(k) Number (if known)
#### K222684
Device Name MANTIS® Redux Spinal System
Indications for Use (Describe)
The MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications:
· Degenerative Disc Disease ( defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
· Spondylolisthesis:
- · Trauma (i.e. fracture or dislocation); ·
- · Spinal Stenosis;
- · Curvature (i.e. scoliosis, kyphosis, and/or lordosis);
- · Tumor;
- Pseudoarthorisis; and
- Failed Previous Fusion
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{14}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
#### K222684
Device Name MANTIS® Spinal System
### Indications for Use (Describe)
The MANTIS® Spinal System is intended for percutaneous, posterior, non-cervical pedicle fixation of the spine to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications:
- · Degenerative Disc Disease ( defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
- · Spondylolisthesis:
- · Trauma (i.e. fracture or dislocation); ·
- · Spinal Stenosis;
- · Curvature (i.e. scoliosis, kyphosis, and/or lordosis);
- · Tumor;
- Pseudoarthorisis; and
- Failed Previous Fusion
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{15}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
### K222684
Device Name Trio® and Trio + Spinal Fixation Systems
#### Indications for Use (Describe)
The Stryker Spine TRIO® Spinal System is intended for posterior. noncervical pedicle and nonpedicle fixation of the spine to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications:
- · Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with
degeneration of the disc confirmed by history and radiographic studies);
- · Spondylolisthesis;
- Trauma (i.e. fracture or dislocation};
- · Spinal Stenosis; .
- · Curvature (i.e. scoliosis, kyphosis, and/or lordosis};
- · Tumor;
- · Pseudoarthorisis:
- and Failed Previous Fusion
The TRIO® Spinal Fixation System is intended to be used in conjunction with the OSS Diapason Rods, Opus Spinal System Rods. and the Multi-Axis Cross Connectors. The TRIO®+ Spinal System is intended to be used in conjunction with the OSS/Diapason Rods or Opus Rods, XIA® Pre-bent Rods, and the Multi-Axis Cross Connectors.
Type of Use (Select one or both, as applicable)
| <div> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.2222 3.77776C13.5157 4.07129 13.5157 4.54645 13.2222 4.83998L6.22222 11.8399C5.92869 12.1335 5.45353 12.1335 5.15999 11.8399L2.77778 9.45776C2.48425 9.16423 2.48425 8.68907 2.77778 8.9826C3.07131 8.68907 3.54647 8.68907 3.83999 8.9826L5.69111 10.8337L12.16 2.66665C12.4535 2.37312 12.9287 2.37312 13.2222 2.66665C13.5157 2.96018 13.5157 3.43534 13.2222 3.77776Z" fill="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="2.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{16}------------------------------------------------
510(k) Number (if known)
#### K222684
Device Name ES2TM Spinal System
Indications for Use (Describe)
The ES2™ Spinal System is intended for percutaneous, posterior, non-cervical pedicle and nonpedicle fixation of the pine to provide immobilization of spinal segment in skeletally mature patients as an adjunct to fusion for the following indications:
- · Degenerative Disc Disease (defined as back pain of discogenic origin with
degeneration of the disc confirmed by history and radiographic studies),
- · Spondylolisthesis.
- · Trauma (i.e. fracture or dislocation),
- · Spinal stenosis,
- · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis),
- · Tumor,
- Pseudoarthrosis, and
- Failed previous fusion.
The Titanium and Vitallium® rods from the Stryker Spine RADIUS®; MANTIS® and MANTIS® Redux Spinal Systems are intended to be used with the other components of the ES2TM Spinal System.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{17}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K222684
Device Name ES2 ® Augmentable Spinal System
### Indications for Use (Describe)
When used without cement, the ES2® Augmentable Spinal System is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease ( defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tumor and/or trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, or lordosis), pseudarthrosis, and/or failed previous fusion.
When used in conjunction with the Vertaplex® HV Hiscosity Radiopaque Bone Cement, the ES2® Augmentable Spinal System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. ES2® Augmented with the Vertaplex® HV High Viscosity Radiopague Bone Cement is for use at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px"> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{18}------------------------------------------------
Expiration Date: 06/30/2023
See PRA Statement below.
## Indications for Use
510(k) Number (if known)
K222684
Device Name OASYS® System
### Indications for Use (Describe)
The Stryker Spine OASYS® System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine ( Cl to C7) and the thoracic spine (TI-T3): traumatic spinal fraumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.
The Stryker Spine OASYS® System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The Stryker Spine OASYS® System can be linked to the Xia® System and Xia® 4.5 Spinal System via the rod-to-rod connectors and transition rods.
The Stryker Spine OASYS® System can also be linked to the polyaxial screws of the Xia® 3 Systems via the saddle connector.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{19}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
### K222684
Device Name NILE Proximal Fixation Spinal System
Indications for Use (Describe)
The NILE Proximal Fixation Spinal System are temporary implants for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.
The indications for use include the following applications:
1. Spinal trauma surgery, used in sublaminar, interspinous, or facet wiring techniques;
2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age and older, adult scoliosis, and spondylolisthesis:
3. Spinal degenerative surgery, as an adjunct to spinal fusions.
The NILE Proximal Fixation Spinal System may also be used in conjunction with other medical implants made of similar metals whenever 'wiring' may help secure the attachment of other implants.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{20}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
#### K222684
Device Name NILE Alternative Fixation Spinal System
#### Indications for Use (Describe)
The NILE Alternative Fixation Spinal System are temporary implants for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.
The indications for use include the following applications:
1. Spinal trauma surgery, used in sublaminar, interspinous, or facet wiring techniques;
2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age and older, adult scoliosis, and spondylolisthesis:
3. Spinal degenerative surgery, as an adjunct to spinal fusions.
The NILE Alternative Fixation Spinal System may also be used in conjunction with other medical implants made of similar metals whenever 'wiring' may help secure the attachment of other implants.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{21}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K222684
Device Name Escalate™ Laminoplasty System
Indications for Use (Describe)
The Escalate™ Laminoplasty System is intended for use in the lower cervical and upper thoracic spine (C3-T3) in laminoplasty procedures. The system is intended to hold the lamina open following a laminoplasty procedure.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{22}------------------------------------------------
| 510(k) Summary | | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Submitter: | Stryker Spine<br>2 Pearl Court<br>Allendale, NJ 07401 | |
| Contact Person : | Name: Sierra Mertz<br>Phone: (484) 408-8221<br>Email: sierra.mertz@stryker.com | |
| Date Prepared: | May 09, 2023 | |
| Trade Name: | 1. XIA® 4.5 Spinal System<br>2. XIA® 4.5 Cortical Trajectory<br>3. XIA® 3 Spinal System<br>4. Serrato® Spinal System<br>5. XIA® Growth Rod Conversion Set<br>6. XIA® II Spinal System<br>7. XIA® Precision System<br>8. XIA® Anterior<br>9. Diapason® Spinal System<br>10. Opus™ Spinal System<br>11. Radius® Spinal System<br>12. Mantis® Spinal System<br>13. Mantis® Redux<br>14. Trio® & Trio+ Spinal Fixation System<br>15. ES2™ Spinal System<br>16. ES2™ Augmentable Spinal System<br>17. Oasys® Occipito-Cervico-Thoracic System<br>18. Nile® Proximal Fixation Spinal System<br>19. Nile® Alternative Fixation Spinal System<br>20. Escalate® Laminoplasty System | |
| Common Name: | 1. XIA® 4.5 Spinal System<br>Pedicle Screw System<br>2. XIA® 4.5 Cortical Trajectory<br>Pedicle Screw System<br>3. XIA® 3 Spinal System<br>Pedicle Screw System<br>4. XIA 3® Serrato® Spinal System<br>Pedicle Screw System<br>5. XIA® Growth Rod Conversion Set<br>Growing Rod System<br>6. XIA® Spinal System (Xia II)<br>Pedicle Screw System<br>7. XIA® Precision System<br>Pedicle Screw System<br>8. XIA® Anterior<br>Pedicle Screw System<br>9. Diapason® Spinal System<br>Pedicle Screw System | |
| 510(k) Summary | | |
| | 10. Opus™ Spinal System<br>Pedicle Screw System | |
| | 11. Radius® Spinal System<br>Pedicle Screw System | |
| | 12. Mantis® Spinal System<br>Pedicle Screw System | |
| | 13. Mantis® Redux<br>Pedicle Screw System | |
| | 14. Trio® & Trio+ Spinal Fixation System<br>Pedicle Screw System | |
| | 15. ES2® Spinal System<br>Pedicle Screw System | |
| | 16. ES2® Augmentable Spinal System<br>Pedicle Screw System | |
| | 17. Oasys® Occipito-Cervico-Thoracic System<br>Occipito-Cervico-Thoracic Screw System | |
| | 18. Nile® Proximal Fixation Spinal System<br>Bone fixation cerclage | |
| | 19. Nile® Alternative Fixation Spinal System<br>Bone fixation cerclage | |
| | 20. Escalate® Laminoplasty System<br>Spinal Interlaminal Fixation Orthosis | |
| Classification: | Class II | |
| Product code and<br>Classification Name: | 1. XIA® 4.5 Spinal System<br><b>NKB</b> - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br><b>KWQ</b> - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br><b>KWP</b> - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| | 2. XIA® 4.5 Cortical Trajectory<br><b>NKB</b> - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br><b>KWQ</b> - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br><b>KWP</b> - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| | 3. XIA® 3 Spinal System<br><b>NKB</b> - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br><b>KWQ</b> - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br><b>KWP</b> - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| | 4. Xia® 3 Serrato® Spinal System<br><b>NKB</b> - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br><b>KWQ</b> - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br><b>KWP</b> - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| | 5. XIA® Growth Rod Conversion Set<br><b>PGM</b> - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070 | |
| | 6. XIA® II Spinal System<br><b>NKB</b> - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070 | |
| 510(k) Summary | | |
| | KWQ - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>KWP - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| 7. | XIA® Precision System<br>NKB - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br>KWQ - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>KWP - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| 8. | XIA® Anterior<br>NKB - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br>KWQ - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>KWP - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| 9. | Diapason® Spinal System<br>NKB - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br>KWQ - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>KWP - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| 10. | Opus Spinal System<br>NKB - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br>KWQ - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>KWP - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| 11. | Radius® Spinal System<br>NKB - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br>KWQ - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>KWP - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050<br>OSH - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070 | |
| 12. | Mantis® Spinal System<br>NKB - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br>KWQ - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>KWP - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050<br>OSH - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070 | |
| 13. | Mantis® Redux<br>NKB - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070<br>KWQ - Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>KWP - Spinal Interlaminal Fixation Orthosis, 21 CFR §888.3050 | |
| 510(k) Summary | | |
| | OSH - Thoracolumbosacral Pedicle Screw System, 21 CFR §888.3070…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.