Vertaplex HV High Viscosity Radiopaque Bone Cement; ES2 Augmentable Spinal System
Applicant
Stryker Corporation
Product Code
PML · Orthopedic
Decision Date
Mar 31, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VertaPlex® HV High Viscosity Radiopaque Bone Cement is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma). When used in conjunction with ES2® Augmentable Spinal System, VertaPlex® HV High Viscosity Radiopaque Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. VertaPlex® HV High Viscosity Radiopaque Bone Cement and the ES2® Augmentable Spinal System are for use together at spinal levels where the structural integrity of the spine is not severely compromised.
Device Story
System comprises Vertaplex® HV bone cement and ES2® Augmentable Spinal System (titanium pedicle screws with fenestrations/blocker). Cement is PMMA-based, mixed from powder/liquid components to injectable state, then cures to hardened polymer. Screws provide spinal stabilization. Used in clinical settings by surgeons for vertebral/sacral fracture fixation and tumor-related spinal stabilization. Cement injected into vertebrae/sacrum; screws implanted into pedicles. Cement augments screw anchorage in reduced-quality bone. Output is stabilized spinal structure. Benefits include pain relief from fractures and restored spinal integrity in palliative tumor cases.
Clinical Evidence
No clinical testing performed. Safety and effectiveness supported by bench testing (static/interconnection strength, pullout, cadaver validation) and clinical literature review regarding PMMA bone cement use with augmentable pedicle screw systems.
Technological Characteristics
Cement: PMMA polymer powder, monomer liquid, barium sulfate; ETO sterilized. Screws: Titanium (ASTM F136, ISO 5832); gamma sterilized; fenestrated design for cement augmentation. Testing per ASTM F1798-12, F1717-12, ASTM F1980, ISO 11607, ANSI/AAMI ST72:2011(R2016).
Indications for Use
Indicated for fixation of pathological vertebral/sacral fractures (vertebroplasty/kyphoplasty/sacroplasty) caused by osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancer, myeloma). Also indicated for spinal column integrity restoration in patients with advanced thoracic/lumbar tumors and limited life expectancy when used with ES2® Augmentable Spinal System at levels without severe structural compromise.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Predicate Devices
Kyphon Xpede Bone Cement/Medtronic HV Bone Cement (K171938)
{0}------------------------------------------------
Build Correspondence
Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
March 31, 2020
Stryker Corporation Kristi Ashton Sr Staff Regulatory Affairs Specialist 4100 E. Milham Avenue Kalamazoo, Michigan 49001
#### Re: K192818
Trade/Device Name: Vertaplex HV High Viscosity Radiopaque Bone Cement, ES2® Augmentable Spinal System Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: PML, NKB, NDN Dated: September 30, 2019 Received: October 1, 2019
## Dear Kristi Ashton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir, PhD Acting Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192818
Device Name
Vertaplex® HV High Viscosity Radiopaque Bone Cement
#### Indications for Use (Describe)
VertaPlex® HV High Viscosity Radiopaque Bone Cement is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacroplasty or sacroplasty. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
When used in conjunction with ES2® Augmentable Spinal System, Vertaplex® HV High Viscosity Radiopaque Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. Vertaplex® HV High Viscosity Radiopaque Bone Cement and the ES2® Augmentable Spinal System are for use together at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192818
Device Name ES2 ® Augmentable Spinal System
#### Indications for Use (Describe)
When used without cement, the ES2® Augmentable Spinal System is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tumor and/or trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, or lordosis), pseudarthrosis, and/or failed previous fusion.
When used in conjunction with the Vertaplex® HV High Viscosity Radiopaque Bone Cement, the ES2® Augmentable Spinal System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. ES2® Augmentable Spinal System augmented with the Vertaplex® HV High Viscosity Radiopaque Bone Cement is for use at spinal levels where the structural integrity of the spine is not severely compromised.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Number: K192818 Date Submitted: March 26, 2020
Traditional 510(k)
Stryker Instruments 4100 East Milham Ave, Kalamazoo, MI 49001 USA
Image /page/4/Picture/3 description: The image shows the text 'K192818' on the first line, followed by 'Page 1 of 5' on the second line. The text is in a simple, sans-serif font and is positioned at the top of the image. The layout suggests this is a header or identifying information for a document or a multi-page set of materials. The page number indicates that there are a total of five pages in the document.
stryker®
Instruments
# 510k Summary
## 1. Submitter
Stryker Instruments 4100 East Milham Avenue Kalamazoo, MI 49001 Phone +1 (269) 323-7700 Fax +1 (269) 324-5412
## FDA Establishment Registration Number: 1811755
## Contact:
Kristi Ashton Stryker Instruments Regulatory Affairs Sr. Staff Regulatory Affairs Specialist Phone: +1 (269) 800-1692 Kristi.Ashton@Stryker.com
Date Submitted: March 26, 2020
## 2. Subject Devices
Trade Name: Vertaplex ® HV High Viscosity Radiopaque Bone Cement Common Name: Polymethylmethacrylate (PMMA) Bone Cement Product Codes: PML, NDN Regulation: 21CFR888.3027
Trade Name: ES2® Augmentable Spinal System Common Name: Thoracolumbosacral pedicle screw system Product Codes: NKB Regulation: 21CFR888.3070
| Predicate Device (Cement) | | | |
|------------------------------------------------------|---------|--------------------|---------------|
| Kyphon Xpede Bone<br>Cement/Medtronic HV Bone Cement | K171938 | NDN, PML | 21CFR888.3027 |
| Predicate Device (Screw Systems) | | | |
| CD Horizon Fenestrated Screw Set<br>(Primary) | K171938 | MNI | 21CFR888.3070 |
| ES2® Spinal System (Secondary) | K122845 | NKB, KWP, MNH, MNI | 21CFR888.3070 |
## 3. Legally Marketed Predicate Device(s)
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a document page with the text "K192818" and "Page 2 of 5" at the top. Below that is the Stryker logo in bold black letters. Underneath the logo is a horizontal line, and below that is the word "Instruments" in bold black letters. The page appears to be part of a larger document, possibly a manual or catalog.
Stryker Instruments 4100 East Milham Ave, Kalamazoo, MI 49001 USA
# 4. Device Description
VertaPlex® HV High Viscosity Radiopaque Bone Cement is bone cement comprised of a liquid component and powder component which when mixed together polymerize to form a hardened acrylic polymer. The mixture is in an injectable state for a period of time before it cures to form a hardened structure, capable of long-term load support and bone augmentation in the treatment of symptomatic osteoporotic vertebral compression fractures.
The powder component of VertaPlex® HV High Viscosity Radiopaque Bone Cement is Ethylene Oxide (EO) sterilized and is packaged in a polyethylene/foil pouch while the liquid monomer is aseptically filled into an amber glass ampoule which is EO sterilized. The device contains the following components: Polymer powder, Monomer liquid and Barium Sulfate. Accessories to the subject cement device are compatible Stryker cement mixers such the Stryker Autoplex System and the Stryker PCD System.
ES2® Augmentable Spinal System is a gamma sterilized, titanium implantable screw with fenestrations and blocker designed as a stabilization solution for the aging spine. The subject system (screw and blocker) has been developed to provide improved anchorage of pedicle screws with and without Vertaplex® HV High Viscosity Radiopaque Bone Cement during pedicle screw augmentation in vertebrae with reduced bone quality as compared to pedicle screw augmentation in healthy bone.
# 5. Intended Use/Indications for use
## VertaPlex® HV High Viscosity Radiopaque Bone Cement:
VertaPlex® HV High Viscosity Radiopaque Bone Cement is indicated for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
When used in conjunction with ES2® Augmentable Spinal System, VertaPlex® HV High Viscosity Radiopaque Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. VertaPlex® HV High Viscosity Radiopaque Bone Cement and the ES2® Augmentable Spinal System are for use together at spinal levels where the structural integrity of the spine is not severely compromised.
# ES2® Augmentable Spinal System:
When used without cement, the ES2® Augmentable Spinal System is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the text "K192818 Page 3 of 5" at the top, followed by the Stryker logo. The Stryker logo is in black and features the company's name in a bold, sans-serif font. The logo also includes the registered trademark symbol.
Instruments
the disc confirmed by history and radiographic studies), spondylolisthesis, tumor and/or trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), pseudarthrosis, and/or failed previous fusion.
When used in conjunction with the VertaPlex® HV High Viscosity Radiopaque Bone Cement, the ES2® Augmentable Spinal System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. ES2® Augmentable Spinal System augmented with the Vertaplex® HV Radiopaque Bone Cement is for use at spinal levels where the structural integrity of the spine is not severely compromised.
#### 6. Comparison of Technological Characteristics with the Predicate Device
The Stryker Vertaplex® HV High Viscosity Radiopaque Bone Cement has not changed since it was cleared via K150582 and has the same intended use, principle of operation, and mode of action as the predicate devices. In addition, Vertaplex® HV High Viscosity Radiopaque Bone Cement and Kyphon Xpede Bone cement were previously determined to be substantially equivalent per K163032.
The Stryker ES2® Augmentable Spinal System has the same intended use as the predicate devices. The majority of the technical characteristics are similar/identical to the predicate devices K171938 CD Horizon Fenestrated Screw System and the secondary predicate K122845 ES2®Spinal System with a few exceptions such as added fenestrations and overall sterility. These differences have been found through testing, clinical literature, and a risk analysis to be insignificant and do not raise new questions of safety and effectiveness.
## 7. Non-Clinical
## Bench
Static testing and interconnection strength testing were completed according to ASTM F1798-12 and Static pullout was performed per ASTM F1717-12. In addition, cadaver validation was performed. The subject devices were tested and met all respective acceptance criteria as specified per the individual test report. The testing was conducted in as similar a manner as possible to the clinical setting using the final finished form of the device. The acceptance criteria for the subject device were similar to the acceptance criteria for the Stryker predicate device. The performance testing further demonstrates equivalence because the subject device is shown to have equivalent functional characteristics as the predicate device.
## Sterilization and Shelf Life
Vertaplex® HV Radiopaque Bone Cement will be sterilized per the same ETO methodology and dosage as was listed in submission K091606. The shelf life of 36 months for the subject cement has not been modified since submission K150582. The Vertaplex® HV Radiopaque Bone cement has been tested for bacterial endotoxins using
{7}------------------------------------------------
Image /page/7/Picture/10 description: The image shows the Stryker logo with the word "Instruments" written below it. Above the logo, the text "K192818" and "Page 4 of 5" are printed. The Stryker logo is in bold, black font, and the word "Instruments" is in a smaller, bold, black font.
Stryker Instruments 4100 East Milham Ave, Kalamazoo, MI 49001 USA
Limulus amebocyte lysate (LAL) reagent and kinetic-turbidimetric methodology per ANSI/AAMI ST72:2011(R2016). The endotoxin limits for the Vertaplex® HV Radiopaque Bone Cement are <2.15 EU/device, as recommended for devices that have the potential to unintentionally extravasate out of the vertebral body, contact the cerebrospinal fluid or the circulatory system, resulting in exposure.
The ES2® Augmentable Spinal System is sterilized by gamma irradiation in accordance with AAMI guidelines and ISO 11137-2:2013 "Sterilization of health care products -Radiation - Part 2 to achieve sterility assurance level of 10-6. Additionally, the ES2® Augmentable Spinal System has been tested for bacterial endotoxins using Limulus amebocyte lysate (LAL) reagent and kinetic-turbidimetric methodology per ANSVAAMI ST72:2011(R2016). The endotoxin limits for the ES2® Augmentable Spinal System are <20 EU/device, as recommended for devices that are not intended to contact the cerebrospinal fluid.
ES2® Augmentable Spinal System has a shelf life of 5 years based on aging studies leveraged from the Stryker Spine Thoracolumbar (TL) Portfolio. This testing was completed for all packaging assemblies of TL System (Real Time and Accelerated Aging) in accordance with ASTM F1980 and ISO 11607.
## Biocompatibility
The materials, processes and sterilization methodology for Vertaplex® HV Radiopaque Bone Cement are identical to that which was cleared in the original Vertaplex® HV submission K091606. Vertaplex® HV Radiopaque Bone Cement remains a biocompatible device.
The grade of material used by Stryker to manufacture the ES2® Augmentable Spinal System are ASTM F136 and ISO 5832 recommended. Since, the material has repetitively undergone traditional manufacturing processes and are documented as having a long history of use substantiated with supporting data from testing and literature citations, it was deemed unnecessary to conduct additional testing as allowed per FDA Guidance on ISO 10993. A biological risk assessment was completed and supported by information on the ES2® Augmentable Spinal System materials of construction, manufacturing processes/aids, gathered toxicological data on these materials, biological/chemical evaluation of equivalent devices, as well as other available literature. Based upon examination of this information, the risk assessment indicates that the risk of a toxic effect from the ES2® Augmentable Spinal System is negligible and that the devices can be considered safe for use as intended.
{8}------------------------------------------------
Stryker Instruments 4100 East Milham Ave, Kalamazoo, MI 49001 USA
Image /page/8/Picture/2 description: The image shows a logo and some text. The text at the top reads "K192818 Page 5 of 5". Below that is the Stryker logo. The word "Instruments" is at the bottom of the image.
#### 8. Clinical
No clinical testing was deemed necessary for this 510(k).
A clinical literature review has been completed to support this pre-market notification. The clinical literature review provides evidence to support the safety and effectiveness of the use of PMMA bone cement with augmentable pedicle screws systems.
#### 9. Substantial Equivalence Conclusion
Based upon the non-clinical testing completed and additional supporting documentation provided in this pre-market submission, the subject cement device and screw system have demonstrated safety and effectiveness. Based on the information provided, it can be concluded that the subject cement device. Vertaplex® HV High Viscosity Radiopaque Bone Cement, is substantially equivalent to the predicate cement devices, K171938 Kyphon Xpede Bone Cement/Medtronic HV Bone Cement; and the subject screw system, ES2® Augmentable Spinal System, is substantially equivalent to the predicate screw systems, K171938 CD Horizon Fenestrated Screw System and K122845 ES2® Spinal System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.