Dental Zirconia Ceramic

K222626 · Aidite (Qinhuangdao) Technology Co., Ltd. · EIH · Dec 6, 2022 · Dental

Device Facts

Record IDK222626
Device NameDental Zirconia Ceramic
ApplicantAidite (Qinhuangdao) Technology Co., Ltd.
Product CodeEIH · Dental
Decision DateDec 6, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2

Indications for Use

Dental Zirconia Ceramic are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.

Device Story

Dental Zirconia Ceramic is a pre-shaded, yttria-stabilized zirconia blank (ZrO2, HfO2, Y2O3, Al2O3, and inorganic pigments). Used by dental professionals or laboratories to fabricate dental restorations (crowns, bridges, veneers, inlays, onlays). Input: digital anatomical data via CAD/CAM systems or manual milling. Transformation: milling of the zirconia blank into the desired restoration shape followed by sintering to achieve final hardness and color matching natural teeth. Output: finished dental restoration. Benefit: provides durable, color-matched, metal-free dental prosthetics. Used in dental laboratory settings.

Clinical Evidence

No clinical data. Bench testing only. Performance testing conducted per ISO 6872:2015 (flexural strength, chemical solubility, radioactivity, shrinkage) and biocompatibility testing per ISO 7405:2018 and ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation).

Technological Characteristics

Material: Yttria-stabilized zirconia (ZrO2, HfO2, Y2O3, Al2O3, Fe2O3, Er2O3, Co3O4). Form factor: Pre-sintered blanks for CAD/CAM or manual milling. Flexural strength: ≥600 MPa or ≥800 MPa depending on model. Conforms to ISO 6872:2015. Non-sterile, single-use.

Indications for Use

Indicated for dental restorations including crowns, bridges, veneers, inlays, and onlays in patients requiring dental prosthetics, processed by dental professionals or laboratories.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. December 6, 2022 Aidite (Qinhuangdao) Technology Co., Ltd. % Grace Liu Consultant Shenzhen Joyantech Consulting Co., Ltd 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518000 CHINA Re: K222626 Trade/Device Name: Dental Zirconia Ceramic Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: August 31, 2022 Received: August 31, 2022 Dear Grace Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Michael E. Adjodha -S Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K22626 Device Name Dental Zirconia Ceramic Indications for Use (Describe) Dental Zirconia Ceramic are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary ## K222626 #### 1. Contact Details #### 1.1 Applicant information | Applicant Name | Aidite (Qinhuangdao) Technology Co., Ltd. | |----------------|-----------------------------------------------------------------------------------------------| | Address | No.9 Dushan Road, Economic and Technological Development Zone, Qinhuangdao City, Hebei, China | | Contact person | Chen Yingying | | Phone No. | +86-335-8357346 | | E-mail | chenyingying@aidite.com | | Date Prepared | 2022-08-25 | #### 1.2 Submission Correspondent | Image: Logo | Shenzhen Joyantech Consulting Co., Ltd<br>1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District,<br>Shenzhen, Guangdong Province, China | |-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------| | 卓远天成<br>Phone No. | +86-755-86069197 | | Contact person | Grace Liu, Field Fu | | Contact person's e-mail | grace@cefda.com, field@cefda.com | | Website | http://www.cefda.com | ### 2. Device Information | Trade name | Dental Zirconia Ceramic | |---------------------|-----------------------------------| | Common name | Dental Zirconia Ceramics | | Classification | II | | Classification name | Porcelain powder for clinical use | | Product code | EIH | | Regulation No. | 21 CFR 872.6660 | #### 3. Legally Marketed Predicate Device | Trade Name | PRETTAU®, ICE and Z-WHITE zirconia blanks (Primary Predicate) | |---------------|---------------------------------------------------------------| | 510(k) Number | K183304 | | Product Code | EIH | | Manufacturer | ZIRKONZAHN SRL | #### 4. Device Description Dental Zirconia Ceramic is pre-shaded zirconia, and it's composed of yttria-stabilized zirconia. It contains ZrO2+HfO2+Y2O3+Al2O3 and very small amount of additional inorganic pigments {4}------------------------------------------------ (Fe203+Er2O3+C03O4). The inorganic pigments generate the color on the restorations, after sintering at dental labs, that matches natural color of patient's teeth. Dental Zirconia Ceramic has six models mainly according to the flexural strength (i.e. ≥600Mpa and ≥800Mpa) and the shade uniformity in the whole blank (i.e. monolayer and multilayer). Each model also has several available shade variations. And it also offers various shapes and dimensions suitable for different milling systems. The proposed device is processed into the dental restorations such as crowns, bridges, veneers, inlays and onlays based on the anatomical rendering of the patient's teeth using CAD/CAM (computer aided design / computer aided manufacturing) method or manual milling method. The proposed device is a single-use device, provided non-sterile, and its performance conforms to ISO 6872:2015 Dentistry: Ceramic Materials. #### 5. Intended Use/Indication for Use Dental Zirconia Ceramic are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals. #### 6. Substantial Equivalence Comparison | Comparison item | Proposed Device (K222626) | Predicate Device (K183304) | Comment | |--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Manufacturer | Aidite (Qinhuangdao) Technology<br>Co., Ltd. | ZIRKONZAHN SRL | None | | Product Name | Dental Zirconia Ceramic | PRETTAU®, ICE and Z-WHITE<br>zirconia blanks | None | | Product Code | EIH | EIH | Same | | Regulation<br>Number | 21 CFR § 872.6660 | 21 CFR § 872.6660 | Same | | Classification | Class II | Class II | Same | | Prescription Use | Yes | Yes | Same | | Indications for<br>Use | Dental Zirconia Ceramic are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals. | Prettau®, Prettau® 2, Prettau® 2 Coloured, Prettau® 2 Dispersive, ICE Translucent, ICE Premium, ICE Abutment, ICE Translucent Plus, ICE Translucent Plus Coloured, ICE Translucent Plus Dispersive and Z-White are intended for the manufacturing of metal-free partial and single crowns, full arch occlusally screwed bridges, inlays, onlays, and veneers, full contour restorations as well as reduced | Similar | | | | structures in combination with veneering ceramics. The products are categorized into class 5 according to ISO 6872.<br><br>Prettau® 3, Prettau® 3 Coloured, Prettau® 3 Dispersive, Prettau® 4 Anterior®, Prettau® 4 Anterior® Coloured and Prettau® 4 Anterior® Dispersive are destined for the manufacturing of metal-free partial and single crowns, max. 3-unit bridges, inlays, onlays and veneers, full contour restorations as well as for reduced structures in combination with veneering ceramics and implant superstructures for 3-unit restorations in the anterior and posterior tooth region. The products have to be categorized as class 4 according to ISO 6872.<br><br>The products have been developed for use with Colour Liquid, ICE Zirkon Ceramics, ICE Zirkon Stains and ICE Zirkon Stains 3D. Observe the relative instructions of use when using these products. The blocks are suitable for all milling units, which are able to process presintered zirconia and which have the proper clamping device for the corresponding block. | | | Class<br>(per ISO 6872:2015) | Class 4<br>Multilayer-3D, UTC, UTM<br><br>Class 5<br>Multilayer-3D pro, SHTPC, SHTPM | Class 4<br>Prettau® 3, Prettau® 4 Anterior<br><br>Class 5<br>Prettau®, Prettau® 2, ICE group, Z-White | Same | | Composition | Based on yttria-stabilized zirconia | Based on yttria-stabilized zirconia | Similar | | Color | Color | White, Color | Similar | | | | | | | Intended User | Professional dental technicians | Professional dental technicians | Same | | Single Use | Yes | Yes | Same | | Sterile | Non-sterile | Non-sterile | Same | | Physical<br>Properties | Conform to ISO 6872:2015 | Conform to ISO 6872:2015 | Same | | Uniformity | Uniform | Uniform | Same | | Freedom from<br>extraneous<br>materials | Free from extraneous materials | Free from extraneous materials | Same | | Radioactivity | ≤ 1.0 Bq.g-1 | ≤ 1.0 Bq.g-1 | Same | | Flexural<br>strength | Multilayer-3D, UTC, UTM:<br>≥ 600 MPa | Prettau® 3, Prettau® 4 Anterior:<br>≥ 600 MPa | Similar | | | Multilayer-3D pro, SHTPC,<br>SHTPM:<br>≥ 800 MPa | Prettau®, Prettau® 2, ICE group,<br>Z-White:<br>≥ 900 MPa | | | Chemical<br>solubility | < 100 µg/cm² | < 100 µg/cm² | Same | | Linear thermal<br>expansion<br>coefficient | Multilayer-3D, Multilayer-3D pro,<br>SHTPM:<br>(10.4±0.5)×10-6 K-1<br>SHTPC:<br>(10.3±0.5)×10-6 K-1<br>UTC:<br>(10.0±0.5)×10-6 K-1<br>UTM:<br>(10.5±0.5)×10-6 K-1 | Not publicly available | Different | | | Biocompatibility | Conform to ISO 7405:2018 | | | Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same | Table 1 Substantial Equivalence Comparison {5}------------------------------------------------ {6}------------------------------------------------ Product: Dental Zirconia Ceramic The proposed device has the similar indication for use as the predicate device as well as comparable technical and biocompatibility properties and characteristics, and the minor differences don't raise any additional questions for safety and effectiveness, the proposed device is substantially equivalent to the predicate device. ### 7. Non-clinical Testing #### > Performance Testing The performance tests were performed according to ISO 6872:2015 Dentistry - Ceramic materials, and the test results showed that the proposed device meets the requirements specified in the standard (see Table 2). {7}------------------------------------------------ Product: Dental Zirconia Ceramic | Test Item | Test Results | | | | | | |-----------------------------------------------------------|----------------------------------------------------------------|--------------------------------------|--------------------------------------|--------------------------------------|--------------------------------------|--------------------------------------| | | Multilayer-3D | UTC | UTM | Multilayer-3D pro | SHTPC | SHTPM | | Uniformity | Uniform | Uniform | Uniform | Uniform | Uniform | Uniform | | Freedom from<br>extraneous<br>materials | Free from<br>extraneous<br>materials | Free from<br>extraneous<br>materials | Free from<br>extraneous<br>materials | Free from<br>extraneous<br>materials | Free from<br>extraneous<br>materials | Free from<br>extraneous<br>materials | | Radioactivity<br>(Bq.g-1) | <0.019 | <0.018 | <0.019 | <0.021 | <0.020 | <0.019 | | Flexural strength<br>(Mpa) | Incisal end: 635.60<br>Middle part: 804.40<br>Root end: 931.30 | 840.69 | 725.34 | 985.15 | 950.95 | 1014.51 | | Linear thermal<br>expansion<br>coefficient<br>(×10-6 K-1) | Incisal end: 10.4<br>Middle part: 10.3<br>Root end: 10.3 | 10.1 | 10.5 | 10.5 | 10.3 | 10.4 | | Chemical solubility<br>(µg/cm²) | 28 | 11 | 21 | 38 | 13 | 5 | | Shrinkage factor | 1.233 | 1.244 | 1.237 | 1.234 | 1.241 | 1.250 | #### Table 2 Summary of Performance Testing #### A Biocompatibility Testing The biocompatibility tests were performed according to ISO 7405:2018 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry (see Table 3), and the test results showed that the proposed device has no biocompatibility issues. | Biological Endpoint | Reference | Test Result | |------------------------------|--------------------|-------------------------------------------------------------------| | Cytotoxicity | ISO 10993-5:2009 | No cytotoxicity under the conditions of the study | | Cytotoxicity | ISO 7405:2018, 6.2 | No cytotoxicity under the conditions of the study | | Cytotoxicity | ISO 7405:2018, 6.3 | No cytotoxicity under the conditions of the study | | Skin Sensitization | ISO 10993-10:2010 | No skin sensitization under the conditions of the study | | Oral Mucosa Irritation | ISO 10993-10:2010 | No oral mucosa irritation under the conditions of the study | | Acute Systemic Toxicity | ISO 10993-11:2017 | No acute systemic toxicity under the conditions of the study | | Subchronic Systemic Toxicity | ISO 10993-11:2017 | No subchronic systemic toxicity under the conditions of the study | | Genotoxicity | ISO 10993-3:2014 | No genotoxicity under the conditions of the study | | Implantation | ISO 10993-6:2016 | No local effects under the conditions of the study | Version: A/0 {8}------------------------------------------------ The results of the non-clinical testing demonstrate that the proposed device is equivalent to the predicate device. #### 8. Clinical Testing No clinical study is included in this submission. #### 9. Conclusions The results of comparing the design specifications and non-clinical testing between the proposed device and the legally marketed predicate device (K183304) show that they are Substantially Equivalent (SE).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...