Prettau®, Prettau® 2 Coloured, Prettau® 2 Dispersive, ICE Translucent, ICE Premium, ICE Abutment, ICE Translucent Plus, ICE Translucent Plus Coloured, ICE Translucent Plus Dispersive and Z-White are intended for the manufacturing of metal-free partial and single crowns, full arch occlusally screwed bridges, inlays, onlays, and veneers, full contour restorations as well as reduced structures in combination with veneering ceramics. The products are categorized into class 5 according to ISO 6872. Prettau® 3, Prettau® 3 Coloured, Prettau® 4 Anterior®, Prettau® 4 Anterior® Coloured and Prettaw® 4 Anterior® Dispersive are destined for the manufacturing of metal-free partial and single crowns, max. 3-unit bridges, inlays, onlays and veneers, full contour restorations as well as for reduced structures in combination with veneering ceramics and implant superstructures for 3-unit restorations in the anterior and posterior tooth region. The products have to be categorized as class 4 according to ISO 6872. The products have been developed for use with Colour Liquid, ICE Zirkon Stains and ICE Zirkon Stains and ICE Zirkon Stains 3D. Observe the relative instructions of use when using these products. The blocks are suitable for all milling units, which are able to process presintered zirconia and which have the proper clamping device for the corresponding block.
Device Story
Yttria-stabilized zirconia blanks; used by professional dental technicians to manufacture dental restorations via CAD/CAM milling units. Blanks processed into crowns, bridges, inlays, onlays, veneers, or implant superstructures. Material provides metal-free, biocompatible restorative solution. Final restorations placed by dentists to restore tooth function and aesthetics. Benefits include high strength and durability.
Clinical Evidence
Bench testing only. Devices tested for physical properties per ISO 6872 and biocompatibility per ISO 10993-1. All samples met standard requirements.
Technological Characteristics
Yttria-stabilized zirconia blanks; presintered form factor for CAD/CAM milling. Categorized as ISO 6872 Class 4 or Class 5. Flexural strength: 600 MPa (Class 4) or >=900 MPa (Class 5).
Indications for Use
Indicated for dental patients requiring metal-free partial/single crowns, full arch occlusally screwed bridges, inlays, onlays, veneers, full contour restorations, reduced structures for veneering, or implant superstructures (3-unit max for Class 4 products).
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human figures.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 4, 2019
Zirkonzahn srl. Sandra Leitner Regulatory Affairs Responsible Via An der Ahr 7 Gais BZ 39030 ITALY
Re: K183304
Trade/Device Name: PRETTAU®, ICE and Z-WHITE zirconia blanks Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: January 11, 2019 Received: January 15, 2019
Dear Sandra Leitner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Digitally signed by Mary S. Mary S. Runner -S3 Runner -S3 12:39:07 -04:04
12:39:07 -04'00'
Enclosure
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183304
#### Device Name
PRETTAU®, ICE and Z-WHITE zirconia blanks
#### Indications for Use (Describe)
Prettau®, Prettau® 2 Coloured, Prettau® 2 Dispersive, ICE Translucent, ICE Premium, ICE Abutment, ICE Translucent Plus, ICE Translucent Plus Coloured, ICE Translucent Plus Dispersive and Z-White are intended for the manufacturing of metal-free partial and single crowns, full arch occlusally screwed bridges, inlays, onlays, and veneers, full contour restorations as well as reduced structures in combination with veneering ceramics. The products are categorized into class 5 according to ISO 6872.
Prettau® 3, Prettau® 3 Coloured, Prettau® 4 Anterior®, Prettau® 4 Anterior® Coloured and Prettaw® 4 Anterior® Dispersive are destined for the manufacturing of metal-free partial and single crowns, max. 3-unit bridges, inlays, onlays and veneers, full contour restorations as well as for reduced structures in combination with veneering ceramics and implant superstructures for 3-unit restorations in the anterior and posterior tooth region. The products have to be categorized as class 4 according to ISO 6872.
The products have been developed for use with Colour Liquid, ICE Zirkon Stains and ICE Zirkon Stains and ICE Zirkon Stains 3D. Observe the relative instructions of use when using these products. The blocks are suitable for all milling units, which are able to process presintered zirconia and which have the proper clamping device for the corresponding block.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 05 510(k) Summary
### 510 (k) SUMMARY
510(K) NOTIFICATION
# APPLICANT
ZIRKONZAHN SRL Via An der Ahr 7 Gais, ITALY 39030 Phone: +39 0474 066 660 Fax: +39 0474 066 661 E-mail: info@zirkonzahn.com
# CONTACT PERSON
Sandra Leitner Regulatory Affairs ZIRKONZAHN SRL Via An der Ahr 7 Gais, ITALY 39030 Phone: +39 0474 066 784 Fax: +39 0474 066 661 E-mail: sandra.leitner@zirkonzahn.com
Date Summary Prepared: November 23, 2018
# DEVICE IDENTIFICATION
| Trade/Proprietary Name: | PRETTAU®, ICE and Z-WHITE zirconia blanks |
|-------------------------|-------------------------------------------|
| Generic / Common Name: | Dental zirconia blanks |
| Regulation Number: | 872.6660 |
| Classification Name: | Porcelain powder for clinical use |
| Class: | II |
| Product Code: | EIH |
| Panel: | Dental |
{4}------------------------------------------------
### LEGALLY MARKETED PREDICATE DEVICE
| Company: | VITA Zahnfabrik H. Rauter GmbH Co. |
|----------------|------------------------------------|
| Device Name: | VITA YZ ST and VITA YZ XT |
| Product Code: | EIH |
| 510(k) Number: | K180703 |
#### INDICATIONS FOR USE
Prettau®, Prettau® 2, Prettau® 2 Coloured, Prettau® 2 Dispersive, ICE Translucent, ICE Premium, ICE Abutment, ICE Translucent Plus, ICE Translucent Plus Coloured, ICE Translucent Plus Dispersive and Z-White are intended for the manufacturing of metal-free partial and single crowns, full arch occlusally screwed bridges, inlays, and veneers, full contour restorations as well as reduced structures in combination with veneering ceramics. The products are categorized into class 5 according to ISO 6872.
Prettau® 3, Prettau® 3 Coloured, Prettau® 3 Dispersive, Prettau® 4 Anterior®, Prettau® 4 Anterior® Coloured and Prettau® 4 Anterior® Dispersive are destined for the manufacturing of metal-free partial and single crowns, max. 3-unit bridges, inlays, onlays and veneers, full contour restorations as well as for reduced structures in combination with veneering ceramics and implant superstructures for 3-unit restorations in the anterior and posterior tooth region. The products have to be categorized as class 4 according to ISO 6872.
The products have been developed for use with Colour Liquid, ICE Zirkon Ceramics, ICE Zirkon Stains and ICE Zirkon Stains 3D. Observe the relative instructions of use when using these products. The blocks are suitable for all milling units, which are able to process presintered zirconia and which have the proper clamping device for the corresponding block.
## DEVICE DESCRIPTION
The subject devices are composed of yttria-stabilized zirconia. They are available in different models that differ in form and color. The materials used for the submitted devices have a long history of safe use for same or equivalent indications.
{5}------------------------------------------------
### DISCUSSION OF NON CLINICAL TESTS
Non clinical testing was performed to determine the physical properties of the subject devices. In consideration of the International Standard ISO 6872 'Dentistry - Polymer-based crown and bridge materials', the devices were tested. All samples fulfil the requirements of the standard.
In consideration of the International Standards 10993-1 'Biological Evaluation of Medical Devices' Part 1: Evaluation and Testing, the subject devices were tested. Biocompatibility testing demonstrated that no issue of biocompatibility arises.
The results of nonclinical tests demonstrate that the devices are equivalent to the predicate devices.
# SUBSTANTIAL EQUIVALENCE DISCUSSION
The table below provides a more detailed comparison of the submitted devices and the predicate devices.
Both the Zirkonzahn devices and the predicates are made of yttria-stabilized zirconia that have a well- documented history for use in medical device applications and have been in use for many years.
The submitted devices and the predicates have similar indications for use as well as comparable technical, physical, chemical and biological properties and characteristics.
{6}------------------------------------------------
| | PRETTAU®, ICE and Z-WHITE zirconia blanks | Predicate Device:<br>VITA YZ ST and VITA YZ XT<br>K180703 | Comparison |
|---------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | ZIRKONZAHN SRL | VITA Zahnfabrik H. Rauter GmbH<br>Co. | |
| Regulation number | 872.6660 | 872.6660 | Same |
| Product Code | EIH | EIH | Same |
| Classification name | Porcelain Powder For Clinical Use | Porcelain Powder For Clinical Use | Same |
| ISO 6872 Class | Class 4<br>Prettau® 3, Prettau® 4 Anterior<br><br>Class 5<br>Prettau®, Prettau® 2, ICE group, Z-White | Class 4<br>VITA YZ XT<br><br>Class 5<br>VITA YZ ST | Same. The indications for use of the new devices as well as of the predicate devices were defined according to the recommended clinical indications in standard ISO 6872:2015. Thus, the indications for use of both are similar. |
| Composition | Based on yttria-stabilized zirconia | Based on yttria-stabilized zirconia | Similar |
| Models | Different shades and colors | Different shades and colors | Similar |
| Intended user | Professional dental technicians | Professional dental technicians | Same |
| Physical Properties | According to ISO 6872:2015 | According to ISO 6872:2015 | Same |
| Flexural Strength | Prettau® 3, Prettau® 4 Anterior: 600 MPa<br>Prettau®, Prettau® 2, ICE group, Z-White :<br>$\ge 900$ MPa | VITA YZ XT 678 MPa<br>VITA YZ ST 934 MPa | Similar |
{7}------------------------------------------------
# CONCLUSION
Based on the available information, the new devices are similar in function, composition, production technology and intended use. Therefore, we conclude that the proposed devices are substantially equivalent to the predicates.