Alveoair Digital Spirometer

K222525 · Roundworks Technologies Private Limited · BZG · Aug 28, 2023 · Anesthesiology

Device Facts

Record IDK222525
Device NameAlveoair Digital Spirometer
ApplicantRoundworks Technologies Private Limited
Product CodeBZG · Anesthesiology
Decision DateAug 28, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1840
Device ClassClass 2

Indications for Use

The alveoair Digital Spirometer is intended to conduct basic lung function and spirometry testing on patients aged ≥ 22 years by healthcare professionals or clinicians in any healthcare environment. The alveoair Digital Spirometer is not intended for use during patient transport.

Device Story

Portable digital spirometer; uses infrared interruption principle to measure lung function. Input: airflow from patient inhalation/exhalation through disposable turbine/mouthpiece. Turbine propeller rotation speed/direction tracked by IR sensor; firmware calculates volume and flow coordinates (liters/seconds). Data transferred via BLE to mobile app (alveoMD); uploaded to cloud (alveoFit) for parameter calculation. Used by healthcare professionals in clinics/hospitals. Output: lung function test reports displayed on mobile app. Benefits: enables standardized spirometry testing (ATS/ERS 2019) for clinical decision-making.

Clinical Evidence

Bench testing only. Device performance validated against ISO 26782:2009, ISO 23747:2015, and ATS/ERS 2019 guidelines. Biocompatibility testing performed per ISO 10993-1, -5, -10, and -23. Electrical safety and EMC testing conducted per IEC 60601-1 and IEC 60601-1-2.

Technological Characteristics

Handheld digital spirometer; infrared interruption sensing principle. Powered by 2x 1.5V AAA batteries. Connectivity: Bluetooth (BLE). Materials: Polycarbonate turbine, cellulose/paper mouthpiece. Software: Firmware for signal processing, mobile app for UI, cloud backend for parameter calculation. Compliant with ISO 26782, ISO 23747, and ATS/ERS 2019.

Indications for Use

Indicated for basic lung function and spirometry testing in patients aged ≥ 22 years. Intended for use by healthcare professionals or clinicians in clinical environments. Not for use during patient transport.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. August 28, 2023 Roundworks Technologies Private Limited % Ankur Naik Managing Director IZiel Healthcare 14, Hadapsar Industrial Estate, Hadapsar Pune, Maharashtra 411013 India Re: K222525 Trade/Device Name: alveoair Digital Spirometer Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: July 27, 2023 Received: July 27, 2023 Dear Ankur Naik: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Rachana Visaria -S Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K222525 Device Name alveoair Digital Spirometer Indications for Use (Describe) The alveoarr Digital Spirometer is intended to conduct basic lung function and spirometry testing on patients aged > 22 years by healthcare professionals or clinicare environment. The use of an alveoair Digital Spirometer is not intended for use during patient transport. Type of Use (Select one or both, as applicable) | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | |------------------------------------------------| | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary 510(K) summary of safety and effectiveness for alveoair Digital Spirometer is provided in accordance with 21 CFR 807.92. | Date: | 28 August 2023 | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter (Owner): | Prashant Patel<br>Founder & CEO<br>Roundworks Technologies Private Limited.<br>Office No. B 302, Building No. B, 3rd floor,<br>GO Square, Survey No. 249/250, Above Surya<br>Electronics, Wakad, Pune - 41104, India.<br>P: +91 7507776273<br>Email: prashant@alveo.fit | | 510(k) Contact Person: | Ankur Naik<br>Managing Director<br>IZiel Healthcare<br>14, Hadapsar Industrial Estate,<br>Hadapsar, Pune – 411013, India.<br>P: +91 72762 2555 M: +91 7069553814<br>Email: ankur.naik@izielhealthcare.com | | Device Trade Name: | alveoair Digital Spirometer | | Regulation Number: | 21 CFR 868.1840 | | Review Panel: | Anesthesiology | | Device Class: | Class II | | Product Code: | BZG | | Predicate Device: | Air Next (K183089) | | Reference device: | Spirobank G (K072979) | ### Device Description The alveoair Digital Spirometer is used to test lung function in people of all ages ≥ 22 years. It is intended to be used by healthcare professionals or clinicians in any healthcare environment. The alveoair Digital Spirometer was designed, developed, and manufactured at Roundworks Technologies Pvt Ltd. The model number is indicated below: | Model<br>Number | Model Name | Model Description | |-----------------|--------------------------------|-----------------------------------------------------------| | ALV002 | alveoair Digital<br>Spirometer | Digital Spirometer to measure lung function<br>parameters | The alveoair Digital Spirometer system includes: - alveoair Digital Spirometer . {5}------------------------------------------------ - alveoMD mobile application . - . alveofit API Cloud server backend The alveoair Digital Spirometer is intended to be used and compatible only with the flowMIR disposable turbine and cardboard mouthpiece manufactured by the Medical International Research s.r.l. The accessories are 510k cleared under K061712 and it is single-use disposable. Roundworks Technologies Pvt Ltd recommends the user to purchase the sinqle-use disposable flowMIR turbine on their own. One sample piece of the flowMIR (Ref. code: 910004) disposable turbine sensor and disposable cardboard mouthpiece is provided in the packaging. Roundworks recommends that the user purchase the same model turbine and mouthpiece from Medical International Research s.r.l. for further use. The alveoair digital spirometer is available in two different colors. The internal components, software, and function remain the same for both devices. The only difference is the color of the case; one is completely black and the other is a combination of black and white. The alveoair Digital Spirometer is used in combination with a turbine and mouthpiece. It utilizes a smartphone with a dedicated mobile application (alveoMD) and a cloud server (alveoFit) to view and store spirometer readings. This portable spirometer operates on the principle of infrared interruption. To perform a test, the user inhales and exhales air through the mouthpiece, which then flows into the turbine. The turbine's propeller rotates in both clockwise and counterclockwise directions, depending on the airflow. The firmware within the device calculates a series of volume and flow coordinates, in liters with respect to time in seconds, every time an interrupt data is received from the IR sensor. This process continues for 20 seconds or until the flow change calculated is less than 0.025 liters per second. When the patient inhales/exhales air into the spirometer during the standard or full loop tests. Once the test is completed, all coordinates are transferred to the mobile application using BLE. From there, the data is uploaded to the alveofit API Cloud server via the internet. For the alveoMD app, an internet connection is required to initiate the spirometry test. The alveoFit cloud server takes in the coordinates to calculate all lung parameters. Once the process is completed, a test report will be generated and displayed in the alveoMD mobile application. The internal program performs all calculations for measurements to meet ATS/ERS guideline standardization of spirometry 2019. # Intended Use / Indications for Use The alveoair Digital Spirometer is intended to conduct basic lung function and spirometry testing on patients aged ≥ 22 years by healthcare professionals or clinicians in any healthcare environment. The alveoair Digital Spirometer is not intended for use during patient transport. {6}------------------------------------------------ # Comparison to predicate devices # Table 1: Comparison to Predicate Device | Comparable<br>Properties | Subject Device | Predicate Device<br>(K183089) | Reference Device<br>(K072979) | Comparison<br>Results | |------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product name | alveoair Digital<br>Spirometer | Air Next | Spirobank G | Not applicable | | Manufacturer | Roundworks<br>Technologies Pvt Ltd. | NuvoAir AB | M.I.R. Medical<br>International<br>Research | Not applicable | | Classification<br>Name | Spirometer, Diagnostic | Spirometer,<br>Diagnostic | Spirometer,<br>Diagnostic | Identical | | Regulation<br>number | 21 CFR 868.1840 | 21 CFR 868.1840 | 21 CFR 868.1840 | Identical | | Product code | BZG | BZG | BZG | Identical | | Product Class | Class II | Class II | Class II | Identical | | Intended Use /<br>Indications for Use | The alveoair Digital<br>Spirometer is intended to<br>conduct basic lung<br>function and spirometry<br>testing on patients aged<br>≥ 22 years by healthcare<br>professionals or<br>clinicians in any<br>healthcare environment.<br>The alveoair Digital<br>Spirometer is not<br>intended for use during<br>patient transport. | Air Next is intended<br>to be<br>used by:<br>Healthcare<br>professionals<br>trained to perform<br>spirometry tests on<br>patients of age ≥ 5<br>years old, > 10 kg,<br>and ≥ 110 cm.<br>Air Next is intended<br>to<br>perform basic lung<br>function and<br>spirometry testing. | The Spirobank G<br>spirometer is<br>intended to be used<br>by a physician or<br>by a patient under<br>the instruction of a<br>physician or<br>paramedic. The<br>device is intended to<br>test lung function and<br>can make spirometry<br>testing in people of<br>all ages, excluding<br>infants and neonates.<br>lt can be used in any<br>setting. | Substantially<br>Equivalent | | Comparable Properties | Subject Device | Predicate Device (K183089) | Reference Device (K072979) | Comparison Results | | | | The actual diagnosis shall be done by a healthcare professional in hospital and clinical settings. | | | | Target Population | Patients aged ≥ 22 years old. | Patients of age ≥ 5 years old, > 10 kg, and ≥ 110 cm. | All ages, excluding infants and neonate. | Substantially Equivalent | | Intended User | Health care Professionals or clinicians | Trained health care Professionals. | People of all ages, excluding infants and Neonates. | Substantially Equivalent | | Prescription / Over the counter | Prescription | Prescription | Prescription | Identical | | Use Environment | Doctor's office or hospitals, and clinics | Hospital and clinical settings | It can be used in any setting | Identical | | Principle Operation | The alveoair Digital Spirometer is designed to be used in combination with a turbine and a mouthpiece. It functions as a portable spirometer, operating on the infrared interruption principle. The airflow generated by the user's breath forces a propeller to rotate inside the turbine. The direction of the turbine's rotation, whether clockwise or anti- | The Air Next is designed to work with NuvoAir disposable turbine. When performing a spirometry test, the user exhales into the turbine. The airflow generated is forcing a propeller to rotate inside the turbine. The Air Next registers the speed of the spinning propeller by counting the rotations with a digital infrared | Two different types of turbine sensors can be used with the device, one is reusable, and one is single patient disposable. A disposable mouthpiece is required in order to connect a subject to the spirometer. The flow and volume measurement sensor is a digital turbine, | Substantially Equivalent | | Comparable<br>Properties | Subject Device | Predicate Device<br>(K183089) | Reference Device<br>(K072979) | Comparison<br>Results | | | clockwise, depends on<br>the flow direction. | interruption sensor.<br>The<br>algorithm in the | based on the infrared<br>interruption<br>principle. | | | | To measure the<br>respiratory parameters<br>accurately, the<br>spirometer's sensor<br>tracks the propeller's<br>rotations and calculates<br>its spinning speed<br>concerning time. The<br>device's firmware<br>processes this data and<br>computes a series of<br>volume and flow<br>coordinates in liters per<br>second whenever an<br>interrupt data is received<br>from the IR sensor. | firmware inside the<br>Air Next device then<br>converts the rotations<br>into airflow measured<br>in liters per<br>second.<br>The device is also<br>tested<br>against B.T.P.S.<br>(body<br>temperature and<br>pressure with<br>saturated water<br>vapor) conditions as<br>prescribed by<br>ATS guidelines and<br>the results are well<br>within range | Spirometry test<br>interpretation is<br>based on the Forced<br>Vital Capacity<br>(FVC) test and is<br>based on the ATS<br>standard. | | | Technical Features | | | | | | Flow Sensor | Bidirectional digital<br>turbine with infrared<br>interruption | Bidirectional turbine<br>with<br>infrared interruption | Bidirectional turbine<br>with infrared<br>interruption | Identical | | Physical Characteristics | | | | | | Size | 112 x 55 x 21mm; | 98 x 62 x 26mm | 109 x 49 x 21mm | Substantially<br>Equivalent | | Weight | Central unit 76g<br>(including batteries) | 75g | 160g | Substantially<br>Equivalent | | Energy Type | 2 x 1.5V, AAA batteries | 2 x 1.5V, AAA<br>Alkaline | 9V Alkaline battery | Identical to<br>Predicate Device | | Comparable<br>Properties | Subject Device | Predicate Device<br>(K183089) | Reference Device<br>(K072979) | Comparison<br>Results | | Display | The touchscreen on<br>smartphone or<br>tablet | The touchscreen on<br>smartphone or<br>tablet | Touchscreen / LCD<br>and membrane | Identical to<br>Predicate Device | | Operating<br>Environment | Temp: MIN: +10°C,<br>MAX: +40°C<br>Humidity: MIN: 10% RH,<br>MAX: 95% RH | Temp: MIN: +10°C,<br>MAX: +40°C<br>RH: MIN: 10%, MAX:<br>95%<br>ALT: MAX: 2000 m | Temp: MIN: +10°C,<br>MAX: +40°C.<br>Humidity: MIN: 10%<br>RH, MAX: 95% RH | Identical | | Technical Specifications | | batteries | | | | Volume range and<br>accuracy | Up to 8L<br>±2.5% of reading or<br>±0.050 L, whichever is<br>greater | Up to 10L<br>±3% of reading or<br>±0.050 L, whichever<br>is greater | Up to 10L<br>±3% of reading or<br>±0.050 L, whichever<br>is greater | Substantially<br>equivalent as the<br>subject device is<br>compliant to the<br>requirements of<br>American Thoracic<br>Society (ATS)<br>Standardization of<br>Spirometry 2019<br>update and ISO<br>26782: 2009 | | Flow range and<br>accuracy | 0 - 14 L/s<br>±10% or 0.17 L/s | 0-15 L/s<br>±5% or 200 mL/s | 0-16 L/s<br>±5% or 200 mL/s | Substantially<br>equivalent as the<br>subject device is<br>compliant to the<br>requirements of<br>ISO 23747:2015 | | Flow resistance | <0.5 cmH2O/L/s | <0.5 cmH2O/L/s | <0.5 cmH2O/L/s | Identical | | Connectivity | Bluetooth | Bluetooth | Bluetooth of USB | Identical | | Comparable<br>Properties | Subject Device | Predicate Device<br>(K183089) | Reference Device<br>(K072979) | Comparison<br>Results | | IP Rating | IP22 | IP32 | IP22 | Identical to<br>Reference Device | | Turbine and<br>Mouthpiece | Single-use and<br>disposable | Single patient use<br>and disposable | Single patient use<br>and disposable | Identical | | Material of<br>Turbine and<br>Mouthpiece | Polycarbonate and Pure<br>cellulose or paper | Polycarbonate and<br>Pure cellulose or<br>paper | unknown | Identical to<br>Predicate Device | | Shelf life | 2 years | Unknown | Unknown | Substantially<br>equivalent | | Service Life | 5 years | 10 years | Around 10 years | Substantially<br>equivalent | | Measured Parameter | | | | | | Measured<br>Parameters | FVC, FEV0.25, FEV0.5,<br>FEV0.75, FEV1, FEV3,<br>FEV6, FEV1/FVC,<br>FEV0.5/ FVC,<br>FEV3/FVC,<br>FEV0.75/FVC, PEF,<br>FEF2575, FEF10,<br>FEF25, FEF40, FEF50,<br>FEF60, FEF75, FEF80,<br>FEF50/FVC,<br>FEV1/FEV6, FET, BEV,<br>FIVC, PIF, FIF25, FIF50,<br>FIF75, and Estimated | FVC, FEV0.75,<br>FEV1, FEV3, FEV6,<br>FEV/FVC (FER) for<br>0.75 /1/3 / 6, PEF,<br>FEF25-75,<br>FEF50/FVC,<br>FEV1/FEV6, FET,<br>FIVC, PIF, and<br>FIF25-75 (MIF25-75) | FVC, FEV0.75,<br>FEV1, FEV3, FEV6.<br>FEV/FVC (FER) for<br>0.75/1 /3 / 6, PEF,<br>FEF25-75, FEF25,<br>FEF50, FEF75,<br>FEF50/FVC,<br>FEV1/FEV6, FET,<br>Evol, FIVC, PIF,<br>FIF25-75 (MIF25-75),<br>and ELA. | Similar parameters<br>for subject,<br>predicate and<br>reference devices.<br>Differences do<br>not affect the safe<br>and effective use of<br>the subject<br>device | {7}------------------------------------------------ K222525 {8}------------------------------------------------ {9}------------------------------------------------ {10}------------------------------------------------ {11}------------------------------------------------ ### Performance data The alveoair digital spirometer complies with the following standards: # Electrical Safety and EMC: - 1) ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD). - 2) IEC 60601-1-11 Edition 2.0 2015-01 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. - 3) IEC 60601-1-2:2014+AMD1:2020 CSV Consolidated version Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. # Software: - 1) Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - 2) IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes. ### Cybersecurity: - 1) Guidance for Industry and Food and Drug Administration Staff: Content of Premarket Submissions for Management of Cybersecurity. ### Biocompatibility: - . Handheld Subject Device Contact Type and Duration: Surface Intact skin, limited duration (≤24 hr) - Turbine and mouthpiece contact the patient's gas pathway (cleared K061712) ● - 1) ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process. - 2) ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity. - 3) ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices -Part 10: Tests for skin sensitization. - 4) ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices Part 23: Tests for irritation. {12}------------------------------------------------ #### Risk Management: - 1) ISO 14971:2019 Medical devices Application of risk management to medical devices. ### Labels & IFU: - 1) ISO 15223-1 Fourth edition 2021-07 Medical devices Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements. - 2) ISO 20417 First edition 2021-04 Corrected version 2021-12 Medical devices -Information to be supplied by the manufacturer. # Performance: - 1) ISO 26782:2009 Anaesthetic and respiratory equipment -- Spirometers intended for the measurement of time-forced expired volumes in humans. - 2) ISO 23747:2015 Anaesthetic and respiratory equipment Peak expiratory flow meters for the assessment of pulmonary function in spontaneously breathing humans. - 3) ATS/ERS 2019 guideline Standardization of Spirometry 2019 Update an Official American Thoracic Society and European Respiratory Society Technical Statement. # Usability: - 1) Guidance for Applying Human Factors and Usability Engineering to Medical Devices # Conclusion All the above details collectively demonstrate that alveoair digital spirometer is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...