The KLS Martin Level One Rib Fixation System is in the stabilization and rigid fixation of rib fractures in the chest wall including reconstructive procedures, trauma, or planned osteotomies in patients 18 years of age or older.
Device Story
The KLS Martin Level One Rib Fixation System provides rigid bone fixation for thoracic rib fractures. The system consists of pre-contoured PEEK plates and titanium locking screws. It is used by surgeons in clinical settings to stabilize rib fractures, reconstruct chest wall defects, or fixate bone following planned osteotomies. The device functions by mechanically securing bone segments to promote healing. The healthcare provider selects appropriate plate sizes and secures them to the ribs using the provided instrumentation and locking screws. The system benefits patients by providing structural stability to the chest wall, facilitating recovery from trauma or surgical intervention.
Clinical Evidence
No clinical data was required. Substantial equivalence was determined based on non-clinical bench testing, including comparative pull-out system testing against the Synthes MatrixRIB Fixation System (K081623), and biological safety risk assessments conducted in compliance with ISO 10993-1:2018.
Technological Characteristics
System comprises PEEK (ASTM F2026) plates (2-3 mm thickness) and Ti-6Al-4V (ASTM F136) 2.3 mm x 7 mm multidirectional locking screws. Mechanical fixation principle. Non-active, implantable device.
Indications for Use
Indicated for stabilization and rigid fixation of rib fractures in the chest wall, including reconstructive procedures, trauma, or planned osteotomies in patients 18 years of age or older.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
March 29, 2023
KLS-Martin L.P. Pam Martin Regulatory Affairs Project Manager 11201 Saint Johns Industrial Pkwy S Jacksonville, Florida 32246
Re: K222397
Trade/Device Name: KLS Martin Level One Rib Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: February 24, 2023 Received: March 1, 2023
Dear Pam Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir -S
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K222397
Device Name
KLS Martin Level One Rib Fixation System
Indications for Use (Describe)
The KLS Martin Level One Rib Fixation System is indicated for use in the stabilization and rigid fixation of rib fractures in the chest wall including reconstructive procedures, trauma, or planned osteotomies in patients 18 years of age or older.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Page 1 of 2
| 510(k) #: K222397 | 510(k) Summary | | | | Prepared on: 2023-03-28 |
|------------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------|---------------------|-------------------------|
| Contact Details | | | 21 CFR 807.92(a)(1) | | |
| Applicant Name | | KLS-Martin L.P. | | | |
| Applicant Address | | 11201 Saint Johns Industrial Pkwy S Jacksonville FL 32246 United States | | | |
| Applicant Contact Telephone | | 800-625-1557 | | | |
| Applicant Contact | | Ms. Melissa Bachorski | | | |
| Applicant Contact Email | | rapm_na@klsmartin.com | | | |
| Correspondent Name | | KLS-Martin L.P. | | | |
| Correspondent Address | | 11201 Saint Johns Industrial Pkwy S Jacksonville FL 32246 United States | | | |
| Correspondent Contact Telephone | | 800-625-1557 | | | |
| Correspondent Contact | | Ms. Pam Martin | | | |
| Correspondent Contact Email | | pmartin@klsmartin.com | | | |
| Device Name | | | 21 CFR 807.92(a)(2) | | |
| Device Trade Name | | KLS Martin Level One Rib Fixation System | | | |
| Common Name | | Single/multiple component metallic bone fixation appliances and accessories | | | |
| Classification Name | | Plate, Fixation, Bone | | | |
| Regulation Number | | 888.3030 | | | |
| Product Code | | HRS | | | |
| Legally Marketed Predicate Devices | | | | 21 CFR 807.92(a)(3) | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | | Product Code |
| K153482 | KLS Martin Thoracic Plating System | | | | HRS |
| K151983 | KLS Martin Longitudinal Sternal Stabilization (LSS) System | | | | JDQ |
| K081623 | Synthes MatrixRIB Fixation System | | | | HRS |
| Device Description Summary | | | | 21 CFR 807.92(a)(4) | |
The KLS Martin Level One Rib Fixation System is comprised of PEEK plates and titanium locking screws intended to provide rigid fixation of bone in the thoracic anatomy. The PEEK plates are pre-contoured to accommodate patient anatomy and are offered in plate thicknesses of 2 mm - 3 mm. The PEEK plates are compatible with the Ti-6Al-4V (ASTM F136) 2.3 mm x 7 mm multidirectional locking screws offered in the system. The plates are manufactured from PEEK (ASTM F2026). The system includes the necessary instrumentation
{4}------------------------------------------------
# Intended Use/Indications for Use
The KLS Martin Level One Rib Fixation System is in the stabilization and rigid fixation of rib fractures in the chest wall including reconstructive procedures, trauma, or planned osteotomies in patients 18 years of age or older.
### Indications for Use Comparison
The subject device differs from the primary predicate, KLS Martin Thoracic Plating System (K153482), in indications for use. The addition of the Transitional Adolescent B patient population to the subject device indications does not constitute a new intended use because no special considerations are being qiven, or required, for the Category B adolescents compared to adults. The anatomical area of use for the subject device is exclusive to the ribs, whereas the primary predicate device is cleared for the chest wall including sternum.
## Technological Comparison
### Similarities to Predicate:
Both the subject and primary predicate devices are intended for stabilization of bone in the thoracic anatomy. The design / shape of the PEEK rib plates is similar to the rib plates evaluated in the subject and primary predicate devices are fixated with titanium screws.
### Differences from Predicate:
The subject device plates are made from PEEK whereas the plates are made from CP Titanium. The subject device is limited for use in the rib, whereas the primary predicate device was cleared for use in the chest wall, including sternum.
#### Reference Devices:
The KLS Martin LSS Plating System (K151983) has been included as a reference docompatibility data for the subject device. The KLS Martin Level One Rib Fixation System includes implants manufactured from the same materials as the reference device, PEEK The PEEK plates and titanium screws are identical composition, undergo identical manufacturing processes, and have the same permanent body contact duration as those cleared in K151983 - KLS Martin LSS Plating System.
The Synthes MatrixRIB Fixation System (K081623) has been included as a reference device to compare mechanical performance against the subject device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
#### Non-Clinical Performance Data:
Comparative head-to-head bench testing was conducted to determine whether the subject device performs equivalent to or better than the primary predicate device in pull-out system testing was compared to the reference device, Synthes MatrixRlB Fixation System (K081623). The testing met all predetermined acceptance criteria and the results demonstrate that the subject device's performance is substantially equivalent to both the primary predicate device and the reference device. Biological safety risk assessments in compliance with ISO 10993-1 :2018 were completed on the subject devices and concluded the devices are biocompatible and appropriate for their intended use.
#### Clinical Performance Data:
Clinical testing was not necessary for the determination of substantial equivalence.
#### Conclusions:
The KLS Martin Level One Rib Fixation System has the same intended use and similar technological characteristics as the primary predicate device. Technological differences have been addressed through performance data from the predicate and reference devices. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
### 21 CFR 807.92(a)(5)
### 21 CFR 807.92(a)(5)
### 21 CFR 807,92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.