K222220 · Celestray Biotech Company, LLC., · OWW · Jul 6, 2023 · General, Plastic Surgery
Device Facts
Record ID
K222220
Device Name
SpinMedix Absorbable Fibrous Membrane
Applicant
Celestray Biotech Company, LLC.,
Product Code
OWW · General, Plastic Surgery
Decision Date
Jul 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SpinMedix Fibrous Membrane is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Device Story
SpinMedix Absorbable Fibrous Membrane is a single-use, porous, polymeric membrane designed to provide a non-constricting protective layer between injured tendon tissue and surrounding tissues. It is used in clinical settings during tendon repair procedures; physicians secure the membrane over the injury site using standard fixation techniques. The device acts as a physical barrier to manage and protect the injury site. It is composed of poly(D, L-Lactide-co-Glycolide) and poly(D, L-Lactide-b-Ethylene Glycol). The membrane is resorbable, intended to be replaced by the patient's own tissue over time, potentially reducing tissue attachment and supporting healing. It is provided sterile in various sizes.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing and in vivo animal studies. Bench testing evaluated mechanical performance and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity) per ISO 10993-1. Animal studies in calcaneal tendon incision and inguinal hernia models demonstrated complete device resorption between 16-20 weeks with quiescent inflammatory response and favorable local tissue response.
Technological Characteristics
Resorbable porous membrane composed of poly(D, L-Lactide-co-Glycolide) and poly(D, L-Lactide-b-Ethylene Glycol). Dimensions range from 25mm x 25mm to 200mm x 300mm. Sterilized via Ethylene Oxide to SAL 10^-6. Non-pyrogenic (< 20EU/device). Single-use. Packaged in high-barrier double foil pouch.
Indications for Use
Indicated for management and protection of tendon injuries without substantial loss of tendon tissue.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
July 6, 2023
CelestRay Biotech Company, LLC Dr. Charles C. Han PO Box 341754 Bethesda, Maryland 20827-1754
Re: K222220
Trade/Device Name: SpinMedix Absorbable Fibrous Membrane Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWW, FTL Dated: Mav 24, 2023 Received: May 24, 2023
Dear Charles Han:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Jesse Muir -S
Jesse Muir, PhD Assistant Director DHT6C: Division of Retorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K222220
Device Name SpinMedix Absorbable Fibrous Membrane
Indications for Use (Describe)
The SpinMedix Fibrous Membrane is indicated for management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|----------------------------------------------------------|--|
|----------------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo with a circular shape that transitions from yellow at the top to red at the bottom. Inside the circle, there is a stylized design featuring a curved road or path in a gradient of orange and red. Behind the road, there is a sun-like burst of rays, also in shades of orange and yellow, suggesting warmth and brightness. The overall design conveys a sense of travel, adventure, or forward movement, possibly related to transportation or outdoor activities.
# CELESTRAY
### Biotech company, LLC
#### 5. 510(k) Summary
| Applicant information | |
|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Name: | CelestRay Biotech company, LLC |
| Applicant Address: | PO BOX 341754, BETHESDA, MD 20827-1754 |
| Phone: | 202-808-5668 |
| Fax: | -- |
| Contact Person: | Charles C. Han<br>General Manager and Partner |
| Date Prepared: | Oct 19, 2022 |
| Name of Device | |
| Device Common Name: | Soft Tissue Reinforcement Mesh |
| Device Trade Name: | SpinMedix® Absorbable Fibrous Membrane |
| Device Classification Name: | Mesh, Surgical<br>878.3300<br>Class II<br>OWW, FTL |
| Legally Marketed Devices to Which Substantial Equivalence is Claimed | |
| Predicate Device(s) | Surgi-Wrap MAST Tendon Sheet, K063648<br>MAST Biosurgery Inc. San Diego, CA<br>TAPESTRY® Biointegrative Implant, K201572<br>Embody, Inc. Norfolk, VA |
#### Description of the device
SpinMedix Absorbable Fibrous Membrane provides a non-constricting protective layer between the injured tissue and surrounding tissues and may be secured in place using standard fixation techniques. It is designed for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. SpinMedix is composed of poly (D, L-Lactide-co- Glycolide) and poly (D, L-Lactideb-Ethylene Glycol), and is provided in a variety of sizes, ranging from 25mm * 25mm to 200mm * 300mm. It is a single use, sterilized, porous, polymeric membrane in double peel packages.
{4}------------------------------------------------
#### Indication for Use
The SpinMedix Fibrous Membrane is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
| Haracteristic | Subject Device | Predicate Device | Reference Device |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SpinMedix® Absorbable<br>Fibrous Membrane | MAST Biosurgery Surgi-<br>Wrap MAST Tendon<br>Sheet (K063648) | TAPESTRY<br>Biointegrative Implant<br>(K201572) |
| | | | |
| Indication<br>for Use | The SpinMedix Fibrous<br>Membrane is indicated<br>for the management and<br>protection of tendon<br>injuries in which there<br>has been no substantial<br>loss of tendon tissue. | The Surgi-Wrap MAST<br>Tendon Sheet is<br>indicated for the<br>management and<br>protection of tendon<br>injuries in which there<br>has been no substantial<br>loss of tendon tissue.<br>The resorbable<br>Protective Film<br>minimizes tissue<br>attachment to the device<br>in case of direct contact<br>with the tissues.<br>The Surgi-Wrap MAST<br>Tendon Sheet is also<br>indicated for<br>reinforcement of soft<br>tissues repaired by suture<br>or suture anchors, during<br>tendon repair surgery,<br>including reinforcement<br>of the rotator cuff,<br>patellar, Achilles, biceps,<br>quadriceps, or other<br>tendons. Surgi-Wrap<br>MAST Tendon Sheet is | TAPESTRY® is<br>indicated for the<br>management and<br>protection of tendon<br>injuries in which there<br>has been no substantial<br>loss of tendon tissue. |
| | not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. The Surgi-Wrap MAST Tendon Sheet reinforces soft tissue and provides a remodelable scaffold that is replaced by the patients' own soft tissue. | | |
| | Design | Sheets ranging in size from 25mm x 25mm to 200mm x 300mm with thickness of 0.08mm – 0.16mm | Sheets ranging in size from 25mm x 25mm to 500mm x 500mm with thickness of 0.02mm-1.0mm |
| | Material | Poly(D,L-lactide-co-glycolide) and poly (D, L-lactide-b-ethylene glycol) | Poly(L-lactide-co-D,L-lactide) |
| | Appearance | Fibrous membrane with porous structure | Dense film without micropattern |
| | Resorbable<br>Material | Yes | Yes |
| | Sterilization | Ethylene Oxide to SAL of 10-6 | eBeam Irradiation to SAL of 10-6 |
| | Pyrogenicity | Non-pyrogenic<br>( < 20EU/device) | Non-pyrogenic |
| | Single Use | Single use only | Single use only |
#### Summary/ Comparison of Technical Characteristics
{5}------------------------------------------------
{6}------------------------------------------------
| Packaging | Dry packaged in a high-<br>barrier, double foil<br>pouch with secondary<br>sterile barrier | a Tyvek® pouch and a<br>foil pouch with<br>secondary sterile barrier | Dry packaged in a<br>high-barrier, double<br>foil pouch with<br>secondary sterile barrier |
|-----------------------|--------------------------------------------------------------------------------------------|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Storage<br>Conditions | Store at temperature<br>-4 to 86°F (-20 to 30°C) | Store at controlled room<br>temperature 59-86°F<br>(15-30°C) | Room temperature<br>/ ambient |
| Product<br>Code | OWW, FTL | FTL | OWY |
| CFR Section | 878.3300 | 878.3300 | 878.3300 |
#### Equivalence to Marketed Products
SpinMedix® Absorbable Fibrous Membrane and its predicates have been characterized for performance with mechanical strength to demonstrate substantial equivalence. Testing was conducted in accordance with to FDA's Guidance for the Preparation of a Premarket Notification Applicationfor a Surgical Mesh (March 1999).
Biocompatibility Testing for SpinMedix® Absorbable Fibrous Membrane including. cytotoxicity, sensitization, intracutaneous irritation, systemic toxicity, and genotoxicity satisfied the requirements outlined in ISO 10993- 1 (2009 & 2018) Biological Evaluation of Medical Devices Part 1 : Evaluation and testing within a risk managementprocess, and helped demonstrate that the device is substantially equivalent to its predicate and is biocompatible in accordance to ISO 10993- 1.
The in vivo animal study evaluated local tissue response, systemic toxicity, and the device resorption profile in a clinically relevant tendon model. SpinMedix® Absorbable Fibrous Membrane was found to be completely resorbed between the 16 to 20-week time points with a complete quiescence of the associated inflammatory response. The device was shown to yield favorable results in the calcaneal tendon incision model and inguinal hernia model and supports a determination of substantial equivalence.
#### Conclusion of Non-clinical test
The results of the mechanical performance, biocompatibility and animal efficacy testing demonstrated that SpinMedix® Absorbable Fibrous Membrane is substantially equivalent to its predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.