K201572 · Embody, Inc. · OWY · Oct 19, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201572
Device Name
TAPESTRY Biointegrative Implant
Applicant
Embody, Inc.
Product Code
OWY · General, Plastic Surgery
Decision Date
Oct 19, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TAPESTRY® Biointegrative Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Device Story
TAPESTRY® Biointegrative Implant is a resorbable, non-constricting, protective flat sheet designed for tendon injuries. Composed of 70% poly(D,L-lactide) (PDLLA) and 30% type I bovine collagen. Used as a physical barrier between tendon and surrounding tissues to manage and protect injury sites. Applied by surgeons during open, laparoscopic, endoscopic, or arthroscopic procedures; secured using standard fixation techniques. Optional polyethylene insertion sleeve facilitates application and orientation, then discarded. Device promotes dense collagenous fibrous connective tissue ingrowth. Provided sterile, non-pyrogenic, for single-use. Surgeon determines if hydration with sterile isotonic solution is required prior to placement.
Clinical Evidence
No human clinical data. Evidence includes bench testing (chemical composition, purity, density, porosity, strength) per FDA surgical mesh guidance and biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity) per ISO 10993-1. In vivo animal study in a calcaneal tendon incision model evaluated local tissue response, systemic toxicity, and resorption profile; device resorbed between 52-72 weeks with quiescent inflammatory response.
Technological Characteristics
Composite implant: 70% PDLLA, 30% type I bovine collagen. Non-crosslinked. Resorbable flat sheet. Porosity >85%. Density <0.016g/cm3. Sizes 20mm x 25mm to 70mm x 50mm. Sterilization: Electron Beam (SAL 10^-6). Standalone use. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for management and protection of tendon injuries without substantial loss of tendon tissue.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Rotation Medical Inc. Collagen Tendon Sheet (K112423)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 16, 2021
Embody, Inc. % Scott Bruder CEO, Bruder Consulting and Venture Group Bruder Consulting and Venture Group 4211 Monarch Way, Ste. 500 Norfolk, Virginia 23508
Re: K201572
Trade/Device Name: TAPESTRY Biointegrative Implant Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OWY
Dear Mr. Bruder:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 19, 2020. Specifically, FDA is updating this SE Letter as an administrative correction to the product codes removing FTM as a secondary product code.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Deborah Fellhauer, OHT4: Office of Surgical and Infection Control Devices, Deborah.Fellhauer@fda.hhs.gov.
Sincerely,
Min Zhang-S
for Deborah Fellhauer Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized eagle. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 19, 2020
Embody, Inc. % Scott Bruder CEO, Bruder Consulting and Venture Group Bruder Consulting and Venture Group 4211 Monarch Way, Ste. 500 Norfolk, Virginia 23508
Re: K201572
Trade/Device Name: TAPESTRY Biointegrative Implant Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWY, FTM Dated: June 10, 2020 Received: June 11, 2020
Dear Dr. Bruder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201572
Device Name TAPESTRY® Biointegrative Implant
Indications for Use (Describe)
TAPESTRY® Biointegrative Implant is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue .
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
| Device Trade Name: | TAPESTRY® Biointegrative Implant | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------|--|
| Device Common Name | Tendon Protector | |
| Device Class: | II | |
| Classification Name: | Mesh, Surgical | |
| Regulation No .: | 878.3300 | |
| Product Code: | OWY | |
| Predicate Device: | Rotation Medical Inc. Collagen Tendon Sheet, K112423 | |
| Reference Device: | Surgi Wrap MAST Tendon Sheet, K063648 | |
| Owner/Submitter: | Embody, Inc.<br>4211 Monarch Way, Suite 500<br>Norfolk, VA 23508<br>(757) 777-5674 | |
| Regulatory Contact: | Scott Bruder, MD, PhD<br>Founder and CEO, Bruder Consulting & Venture Group<br>scott@bruderconsulting.com<br>Tele: 201.874.9701 | |
| Date Prepared: | October 13, 2020 | |
### DEVICE DESCRIPTION
The TAPESTRY® Biointegrative Implant device is composed of collagen and poly(D,L-lactide). It is designed to function as a non-constricting, protective layer between the tendon and surrounding tissues. TAPESTRY® is conformable and designed for easy placement between the tendon and surrounding tissue and may be secured in place using standard fixation techniques. TAPESTRY® is provided sterile, non-pyrogenic, for single-use only, in a variety of sizes, ranging from 20mm x 25mm to 70mm x 50mm. TAPESTRY® is available with or without a co-packaged polyethylene Insertion Sleeve, which is used to maintain the implant's orientation and to facilitate easy application onto the tendon. The Insertion Sleeve is discarded after use and not implanted. TAPESTRY® is designed for stand-alone use. At the discretion of the surgeon, TAPESTRY® may be hydrated with sterile isotonic solution.
Preclinical studies of TAPESTRY® showed dense collagenous fibrous connective tissue ingrowth into and around the scaffolding.
### INTENDED USE/INDICATIONS FOR USE
TAPESTRY® is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
{5}------------------------------------------------
#### SUMMARY/COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
TAPESTRY® and its predicate have the same intended use/indications for use, and basic similarities in design, materials, and technological characteristics and are substantially equivalent. A comparison of the subject and predicate device are provided in the table below.
| haracteristic | Subject Device<br>Embody<br>TAPESTRY® | Predicate Device<br>Rotation Medical Collagen Tendon<br>Sheet |
|-----------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Material<br>Composition | Highly aligned composite implant<br>(70% PDLLA, 30% type I bovine<br>collagen) non crosslinked | Highly oriented implant, (100% type I<br>bovine collagen), crosslinked |
| Form | Resorbable flat sheet | Resorbable flat sheet |
| Color/<br>Appearance | White to off white with blue<br>orientation lines on unaligned side | White to off white with a blue perimeter |
| Size | Multiple sizes from 2.0cm x 2.5cm to<br>7.0cm x 5.0cm | 1.5x2cm, 2x2.5cm,2.5x3cm |
| Density | Low (<0.016g/cm3) | Low (0.3g/cm3) |
| Porosity/Void<br>Fraction | >85% porosity | 85-90% porosity |
| Intraoperative<br>Characteristics | Open and laparoscopic/endoscopic and<br>arthroscopic procedures.<br>Can be sutured | Open and laparoscopic/endoscopic and<br>arthroscopic procedures.<br>Can be sutured |
| Pyrogenicity | Non-pyrogenic<br>(< 1 EU/device) | Non-pyrogenic<br>(< 0.5 EU/ml) |
| Sterilization<br>Method | Electron Beam to SAL of 10-6 | Ethylene Oxide to SAL of 10-6 |
| Single Use<br>/Reuse | Single use only | Single use only |
| How Supplied | Dry packaged in a high-barrier, double<br>foil pouch with secondary sterile<br>barrier | Lyophilized, packaged in sterile<br>packaging |
| Storage<br>Conditions | Room temperature/ambient | Room temperature/ambient |
TAPESTRY® and its predicate have been characterized for chemical composition, purity, density, porosity, and strength to demonstrate substantial equivalence. Testing was conducted in accordance with to FDA's Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh.
Biocompatibility Testing for TAPESTRY® including, cytotoxicity, sensitization, intracutaneous irritation, systemic toxicity, pyrogenicity, and genotoxicity satisfied the requirements outlined in ISO 10993-1 Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process, and helped demonstrate that the device is substantially equivalent to its predicate and is biocompatible in accordance to ISO 10993-1.
{6}------------------------------------------------
TAPESTRY® Biointegrative Implant Premarket Notification - K201572
The in vivo animal study evaluated local tissue response, systemic toxicity, and the device resorption profile in a clinically relevant tendon model. TAPESTRY® was found to be completely resorbed between the 52 to 72-week time points with a complete quiescence of the associated inflammatory response. The device was shown to yield favorable results in the calcaneal tendon incision model and supports a determination of substantial equivalence.
#### REFERENCE DEVICE
The Surgi-Wrap MAST Tendon Sheet is a legally marketed device that has additional applicability for establishing substantial equivalence, having the same intended use and composed of poly(D,Llactide.
#### CONCLUSION
The results of the mechanical performance, biocompatibility and animal efficacy testing demonstrated that TAPESTRY® is substantially equivalent to its predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.