K222096 · Foshan Coxo Medical Instrument Co., Ltd. · ELC · Mar 23, 2023 · Dental
Device Facts
Record ID
K222096
Device Name
Endo Ultrasonic Activator
Applicant
Foshan Coxo Medical Instrument Co., Ltd.
Product Code
ELC · Dental
Decision Date
Mar 23, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For root canal cleaning. The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.
Device Story
Endo Ultrasonic Activator is a cordless, battery-powered dental handpiece for root canal cleaning. Device components include handpiece, sleeve, wrench, adapter, and tip. Operation involves CPU-driven circuitry powering a piezoelectric ceramic transducer; reverse piezoelectric effect generates ultrasonic vibration (30kHz ± 10%) in the working tip. Vibration induces cavitation and acoustic streaming effects within the root canal to facilitate cleaning. Used by qualified dental practitioners in clinical settings. Output is mechanical ultrasonic vibration of the tip. Device is cordless, powered by a 3.7V 1200mAh Li-ion battery. Healthcare providers use the device to perform root canal irrigation and cleaning, potentially improving treatment outcomes through enhanced debris removal.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993-5, ISO 10993-10), sterilization validation (ISO 17665-1), electrical safety (IEC 60601-1, IEC 80601-2-60), EMC (IEC 60601-1-2), and performance testing per ISO 18397:2016.
Technological Characteristics
Cordless handpiece; piezoelectric ceramic transducer; 3.7V 1200mAh Li-ion battery; 30kHz ± 10% operating frequency; <100 μm tip excursion. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 80601-2-60, ISO 18397, IEC 62133-2 (battery), ISO 10993-5/10 (biocompatibility), ISO 17665-1 (sterilization).
Indications for Use
Indicated for root canal cleaning in hospital, clinic, or dental office settings by qualified practitioners.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
Ultrasonic Endo Activation Device (Model: Actor I pro) (K213947)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 23, 2023
Foshan COXO Medical Instrument Co.,Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui. No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 CHINA
Re: K222096
Trade/Device Name: Endo Ultrasonic Activator Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC Dated: February 24, 2023 Received: February 24, 2023
## Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -2
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222096
Device Name Endo Ultrasonic Activator
Indications for Use (Describe) For root canal cleaning. The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|_ | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________
## 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
- 1. Date of Preparation:2023/02/24
- 2. Sponsor Identification
### Foshan COXO Medical Instrument Co., Ltd.
BLDG 4, District A, Guangdong New Light Source Industrial Base, South of Luocun Avenue, Nanhai District, Foshan, 528226 Guangdong, China
Contact Person: Yongjian Zheng Position: Legal Representative Tel: +86-13702544788 Fax: +86-0757-81800058 Email: 13702544788@163.com
- Designated Submission Correspondent 3.
Mr. Ray Wang
### Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China
Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name:Endo Ultrasonic Activator Common Name: Scaler,Ultrasonic
Regulatory Information Classification Name: Scaler,Ultrasonic Classification: 2 Product Code: ELC Regulation Number: 872.4850 Review Panel: Dental
Indication For Use Statement:
For root canal cleaning.
The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.
Device description:
The Endo Ultrasonic Activator is mainly used for root canal cleaning.It mainly contain handpiece, sleeve, wrench, adapter and tip. The ultrasonic frequency generated by the CPU is driven by the drive circuit to drive the piezoelectric ceramic transducer, using the reverse piezoelectric effect to produce the ultrasonic vibration, motivating the working tip to produce the resonance, and using the cavitation and sound flow effect generated by the ultrasonic vibration to wash the root canal swing.
#### న. Identification of Predicate Device(s)
Predicate Device 510(k) Number: K053555 Product Name: Ultrasonic Scaler(Model: UDS-L) Manufacturer: Guilin Woodpecker Medical Instrument Company, Limited
- Non-Clinical Test Conclusion 6.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ANSI/AAMI ES60601-1:2005 + A1, Medical Electrical Equipment Part 1: General requirements > for basic safety and essential performance.
- > EC 60601-1-2:2014, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Tests
- IEC 80601-2-60:2019 Medical electrical equipment Part 2-60: Particular requirements for the >
{5}------------------------------------------------
basic safety and essential performance of dental equipment
- > ISO 18397:2016 Dentistry-Powered scaler
- > IEC 62133-2 Edition1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- > ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- > ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- > ISO 17665-1:2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
- Substantially Equivalent (SE) Comparison 8.
| Item | Proposed Device | Predicate Device<br>K053555<br>(Model:UDS-L) | Reference Device<br>K213947 | Remark |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Device Name | Endo Ultrasonic<br>Activator | Ultrasonic Scaler | Ultrasonic<br>Endo<br>Activation Device | / |
| Classification<br>Regulation | 872.4850 | 872.4850 | 872.4850 | SAME |
| Classification<br>Panel | Dental | Dental | Dental | SAME |
| Product Code | ELC | ELC | ELC | SAME |
| Common Name | Scaler,Ultrasonic | Scaler,Ultrasonic | Scaler,Ultrasonic | SAME |
| Indication<br>for<br>use | For root canal<br>cleaning.<br>The instrument<br>must only be<br>used in hospital<br>environments,<br>clinics or dental<br>offices, by<br>qualified<br>practitioners. | 1. Removing supra<br>and sub gingival<br>calculus deposits and<br>stains from the teeth<br>2. Periodontal pocket<br>lavage with<br>simultaneous<br>ultrasonic tip<br>movement<br>3. Preparing,cleaning<br>and irrigating root | The Ultrasonic Endo<br>Activation Device<br>(Model:Actor I pro ) is an<br>ultrasonic-handpiece<br>which is intended use for<br>root canal cleaning and<br>preparation.<br>The Ultrasonic Endo<br>Activation Device<br>(Model:Actor I pro ) is<br>intended for use by | SAME |
| | | canals<br>4. Retrograde<br>preparation of root<br>canals | trained dental<br>professionals in<br>professional health care<br>facilities on patients that<br>needroot-canal-treatment. | |
| Input | Adapter<br>AC100-240V<br>50/60Hz | Main Unit 24V 50Hz<br>1.3A | AC100-240V, 50/60Hz | Analysis(1) |
| Input Power | 10VA | 38VA | Not Available | Analysis(2) |
| Power supply | Battery 3.7V<br>1200mAh(Li-ion<br>Battery) | Power source 220V<br>50HZ | Rechargeable Li-ion<br>Battery Capacity<br>1600mAh,3.7V | Analysis(3) |
| Operating<br>frequency | 30kHz $\pm$ 10% | 28kHz $\pm$ 3kHz | 30 $\pm$ 3kHz(27-33 kHz) | Analysis(4) |
| Primary tip<br>vibration<br>excursion | <100 μ m | 1 μ m-100 μ m | Not Available | SAME |
| Operation mode | Non-continuous | Continuous | Continuous | Analysis(5) |
| Classified of<br>protection<br>against Electric<br>Shock | Class II | Class II | Not Available | SAME |
| Protection<br>against Electric<br>Shock | Type B applied<br>part | Type BF applied part | Not Available | Analysis(6) |
| Degree of<br>Protection | IPX0 | General equipment<br>IPX0 | Not Available | SAME |
| Operating<br>temperature | +5°C - +40°C | +5°C - +40°C | Not Available | SAME |
| Operating<br>humidity | 20% - 80% | 30%-75% | Not Available | Similar |
| Atmospheric<br>pressure | 86kPa -106kPa | 70kPa -106kPa | Not Available | Similar |
| Storage<br>temperature | -10°C - +55°C | -20°C - +55°C | Not Available | Similar |
| Storage<br>humidity | ≤93% | 10%-93% | Not Available | Similar |
| Atmospheric<br>pressure | 50kPa -106kPa | 70kPa -106kPa | Not Available | Similar |
| Cordless | Yes | No | Yes | Analysis(7) |
| Device type | Handpiece | Unit | Handpiece | Analysis(7) |
| Biocompatibility | ISO10993-5<br>ISO10993-10 | Not Available | ISO 10993-5:2009<br>ISO 10993-10:2010 | Analysis(8) |
| | | | ISO 10993-11:2017 | |
| Sterilization | ISO 17665-1 | Not Available | Not Available | Analysis(8) |
| Electrical Safety | IEC 60601-1<br>IEC 80601-2-60 | Not Available | IEC 60601-1:2012<br>IEC 80601-2-60:2019 | Analysis(8) |
| EMC | IEC60601-1-2 | Not Available | IEC 60601-1-2:2014 | Analysis(8) |
| Dental | ISO 18397 | Not Available | Not Available | Analysis(8) |
#### Table 1 Comparison of Technology Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
#### Analysis :
#### Analysis(1):
There is difference on the input of proposed device is adapter, the input of predicate device is main unit, however, both of them could meet the electricity safety standard IEC 60601-1, the difference does not affect the substantial equivalence in effectiveness and safety.
#### Analysis(2):
There is difference on the input power, however, both of them could meet the electricity safety standard IEC 60601-1, the difference does not affect the substantial equivalence in effectiveness and safety.
#### Analysis(3):
The proposed device is powered by rechargeable 3.7V Li-ion battery, and the predicate device is powered by AC power source, there is difference on the power supply mode. In principle, no matter what kind of power supply mode, piezoelectric ceramic transducer is driven electrically to drive the working tip to vibrate, and the intended use of root canal cleaning is achieved by using acoustic flow effect and thermal effect generated by ultrasonic. In terms of type, the subject device the transducer through battery power, while the transucer of the predicate device is powered by AC power source, and it may need to drive the transducer through a transformer. But the vibration generated by the current driven transducer and the ultrasonic are essentially the same. So the difference would not affect its effectiveness.The proposed device comply with electrical safety standard IEC 60601-1, and its Li-ion battery comply with the battery safety standard IEC 62133-2, so the power supply mode difference would not affect its safety.
In addition, powered by battery is a very common power supply mode. For example, the Reference Device K213947 we determined uses the battery to supply power.
This reference device K213947 is consistent on the indication for use, technological characteristics and other relevant information with the subject device.
By comparing with the reference device, we can find that the intended use of the subject device is consistent with the reference device. Both the subject device and the reference device are based on the principle of ultrasonic vibration for root canal cleaning. And the reference device, like the subject device, is also a cordless, battery-powered device.
And the reference device is also use the 3.7V Li-ion battery which is same with the subject device. Thus, powered by a battery is a very mature technology.
#### Analysis(4):
The proposed devices and the predicate device are different operation frequency. Operation frequency of
{8}------------------------------------------------
the subject device is only slightly different with the predicate device.
And we have also identified the reference device K213947 which is consistent with the operation frequency of the subject device.
We believe that the difference of the operating frequency of the subject device and predicate will not affect the substantial equivalence of safety and effectiveness.
#### Analysis(5):
The proposed devices and the predicate device are difference operation mode. The operating mode of the proposed devices are comply with the 60601-1, so this could not effect the effectiveness and safety.
#### Analysis(6):
The proposed devices and the predicate device are difference Protection against Electric Shock.The Protection against Electric Shock of the proposed devices are comply with the 60601-1, so this could not effect the effectiveness and safety.
#### Analysis(7):
There are some design differences between the subject device and the predicate device, for example, cordless and hanpiece. However, we believe that the design differences will not affect the safety and efficacy of the subject device.
First, in principle, the unit of the predicate device is connected to a power supply to power the CPU, and the software is built into the unit.
The unit and the handpiece through the cable connection. This cable is used to power the handpiece and transmit signals to the handpiece, and thus realize the functions of the product.
The subject device, the software is built into the handpiece. After battery power, it drives the transducer to control the CPU and set the function.
No matter how the power supply and signal transmission, the power supply ultimately drives the CPU to work and realize various functional settings of the product. The electrical structure is consistent in essence, only the appearance changes and the connection mode is not consistent, leading to the different selection of CPU, which will not affect the product to achieve the final intended use. So from a principle point of view we are the same as the predicate device, just the form of power supply and the signal transmission have changed.
Because of the connection mode changes, we introduced software validation to prove that functionality of the subject device can be fully implemented. And we introduced ISO 18397 testing to prove that our performance implementation meets the requirements.
In addition, the reference device K213947 we determined is also a cordless handpiece device, and both the subject device and the reference device are based on the principle of ultrasonic vibration for root canal cleaning, which shows that this design form is a very mature technology and will not affect the safety and effectiveness of the subject device.
#### Analysis(8):
We did not obtain relevant test information for the predicate device, We discuss here to make it clear that these minor differences are not concerns for the equivalency of the proposed device and predicate device.
{9}------------------------------------------------
And will not affect the performance of safety and effectiveness of the proposed device.
a. We performed applicable tests to support biocompatibility, including Cytotoxicity, Sensitization and Irritation.
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. Biocompatibility testing of in vitro cytotoxicity was performed to verify the equivalent safety of the materials that are in direct contact with mucosal membrane and teeth in the mouth, and skin including tip and sleeve.
ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin
Sensitization Biocompatibility testing for Irritation and Skin Sensitization was performed to verify the equivalent safety of the materials that are in direct contact with mucosal membrane and teeth in the mouth, and skin including tip and sleeve.
b.Sterilization validation testing is performed in accordance with ISO 17665-1:2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices. But Reprocessing methods also meet the FDA Guidance Document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
c.Electrical safety testing per IEC 60601-1 and IEC 80601-2-60 required testing.
d.Electromagnetic Compatibility testing per IEC 60601-1-2 required testing.
e. The functions of Endo Ultrasonic Activator were verified according to ISO 18397: 2016 Dentistry -Powered scaler. Performance test performed on proposed device aims to support the substantial equivalence of the proposed device to the predicate device. Test's setup and execution was in accordance with applicable standards. Results of the testing demonstrate the comparable performance (eg: Operating frequency) between proposed device and predicate device.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Ultrasonic Scaler cleared under K053555.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.