Ultrasonic Endo Activation Device (Model:Actor I pro)
Applicant
Changzhou Bomedent Medical Technology Co.,Ltd
Product Code
ELC · Dental
Decision Date
Sep 8, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ultrasonic Endo Activation Device (Model:Actor I pro ) is an ultrasonic-handpiece which is intended use for root-canal cleaning and preparation. The Ultrasonic Endo Activation Device (Model:Actor I pro ) is intended for use by trained dental professionals in professional health care facilities on patients that need root-canal-treatment.
Device Story
Ultrasonic Endo Activation Device (Actor I pro) is a handheld dental instrument for root canal cleaning and preparation. Device utilizes ultrasonic cavitation to assist in cleaning. Components include ultrasonic handpiece, working tip, wrench, silicone case, charging base, power adapter, and optional wireless foot switch. Operated by trained dental professionals in professional health care facilities. Activation via ON/OFF button or optional foot switch. Device provides ultrasonic vibration (30±3 kHz) to a working tip inserted into the root canal. Healthcare providers use the device to facilitate debris removal and canal preparation, potentially improving treatment outcomes. Powered by a rechargeable Li-ion battery (3.7V, 1600mAh).
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1, IEC 80601-2-60), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), and biocompatibility (ISO 10993-5, -10, -11).
Technological Characteristics
Ultrasonic handpiece with working tip. Vibration frequency: 30±3 kHz. Power: 3.7V 1600mAh Li-ion battery. Connectivity: BLE 5.0. Sterilization: Steam sterilization for working tip, wrench, and silicone case. Compliance: IEC 60601-1, IEC 80601-2-60, IEC 60601-1-2. Biocompatibility per ISO 10993 series.
Indications for Use
Indicated for root-canal cleaning and preparation in patients requiring root-canal treatment. Intended for use by trained dental professionals in professional health care facilities.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
September 8, 2022
ChangZhou BoMedent Medical Technology Co.,Ltd Yang Chunyuan Sales Manager No.9 Changyang Road.West Taihu Science & Technology Industrial Park Changzhou, Jiangsu 213100 China
Re: K213947
Trade/Device Name: Ultrasonic Endo Activation Device (Model:Actor I pro) Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC Dated: August 11, 2022 Received: August 12, 2022
#### Dear Yang Chunyuan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213947
Device Name
Ultrasonic Endo Activation Device (Model:Actor I pro )
Indications for Use (Describe)
The Ultrasonic Endo Activation Device (Model:Actor I pro ) is an ultrasonic-handpiece which is intended use for rootcanal cleaning and preparation.
The Ultrasonic Endo Activation Device (Model:Actor I pro ) is intended for use by trained dental professionals in professional health care facilities on patients that need root-canal-treatment.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;"> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:100%;"> <span> <span style="">□</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for BOMEDENT. The logo consists of a stylized graphic to the left of the company name. The graphic is composed of two angled shapes, one in gray and the other in yellow. The company name, "BOMEDENT," is written in gray, sans-serif capital letters.
# 510(K) Summary K213947
Prepared in accordance with the requirements of 21 CFR Part 807.92
#### Prepared Date: September 6, 2022
# 1. Submitter's Information
The submitter of this pre-market notification is:
| Name: | ChangZhou BoMedent Medical Technology Co.,Ltd |
|-----------------|---------------------------------------------------------------------------------------------------|
| Address: | No.9 Changyang Road,West Taihu Science & Technology<br>Industrial Park, Changzhou, Jiangsu China. |
| Contact person: | Yang Chunyuan |
| Title: | sales manager |
| E-mail: | yang.chunyuan@bome-dent.com |
| Tel: | +86-15161150269 |
#### 2. Device Identification
| Trade/Device Name: | Ultrasonic Endo Activation Device |
|--------------------|-----------------------------------|
| Models: | Actor I pro |
| Common name: | Scaler, Ultrasonic |
| Regulation Number: | 21 CFR 872.4850 |
| Regulation Name: | Ultrasonic scaler |
| Regulation Class: | Class II |
| Panel: | Dental |
| Product Code: | ELC |
#### 3. Predicate Device
| 510(K) number: | K202906 |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | EndoPilot² |
| Manufacturer: | Schlumbohm GmbH & Co. KG |
| Common name | Dental hand instrument |
| Regulation Number: | 21 CFR 872.4850 |
| Regulation Name: | Ultrasonic Scaler |
| Regulation Class: | Class II |
| Panel: | Dental |
| Product Code: | ELC Ultrasonic scaler<br>EKX direct drive, AC-powered handpiece<br>EKR endodontic plugger, root canal<br>LQY root apex locator (unclassified) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for BOMEDENT. The logo consists of two overlapping rectangles, one gray and one yellow, followed by the text "BOMEDENT" in gray. The rectangles are oriented diagonally, with the gray rectangle slightly above and to the left of the yellow rectangle. The text is in a sans-serif font and is aligned to the right of the rectangles.
# 4. Device Description
The subject device Ultrasonic Endo Activation Device (Model: Actor I pro) is an auxiliary device for dentists to perform root canal treatment. It is mainly used to clean the root canal with the help of ultrasonic cavitation, so as to assist dentists to complete root canal treatment.
The subject device Ultrasonic Endo Activation Device (Model: Actor I pro) configuration consist of the following components:
(1) ultrasonic handpiece (2) ultrasonic working tip (3) Wrench (4) silicone case (5) charging base
(6) Power Adapter(7) wireless foot switch (Optional)
### 5. Indication for use
The Ultrasonic Endo Activation Device (Model:Actor I pro ) is an ultrasonic-handpiece which is intended use for root-canal cleaning and preparation.
The Ultrasonic Endo Activation Device (Model:Actor I pro ) is intended for use by trained dental professionals in professional health care facilities on patients that need root-canal-treatment.
### 6. Summary of the device compared to the predicate device
Compared to the predicate device, the subject device has the same intended use, similar product design, same performance as the predicate device, summarized comparison information is listed in the following table:
| SE<br>Comparisons | Subject Devices | Predicate Device<br>K202906 | Similarities/Differences |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | The Ultrasonic Endo<br>Activation Device<br>(Model:Actor I pro ) is an<br>ultrasonic-handpiece which<br>is intended use for root-<br>canal cleaning and<br>preparation.<br>The Ultrasonic Endo<br>Activation Device<br>(Model:Actor I pro ) is<br>intended for use by trained<br>dental professionals in<br>professional health care<br>facilities on patients that<br>need root-canal-treatment. | The EndoPilot2 systems<br>are dental devices which<br>combine in a single control<br>unit an endo motor to clean<br>the root canal, a dental<br>obturator to fill and<br>pressurize, an electronic<br>apex locator to assist the<br>operator to locate the file<br>tip in the root canal<br>and an ultrasonic-<br>handpiece for root-canal<br>cleaning and preparation.<br>The EndoPilot2 is intended<br>solely for use by trained<br>dental professionals in<br>professional health care<br>facilities on patients<br>that need root-canal-<br>treatment. | Similar<br>Predicate device includes<br>a single control unit an<br>endo motor to clean the<br>root canal, a dental<br>obturator to fill and<br>pressurize, an electronic<br>apex locator to assist the<br>operator to locate the file<br>tip in the root canal<br>and an ultrasonic-<br>handpiece for root-canal<br>cleaning and<br>preparation.<br>But Proposed device<br>only include ultrasonic-<br>handpiece which is<br>intended use for root-<br>canal cleaning and<br>preparation.<br>Function of proposed |
| | | | device is one of the<br>functions of predicate<br>device, there is no risk<br>arise in aspect of<br>indication for use. |
| Intermittent<br>operation | 3min | 1 min/3 min (endodontic<br>treatment) | Similar<br>Subject device only has<br>one mode, it can work<br>for 3 minutes<br>continuously.<br>No new risk raised. |
| Vibration<br>frequency | $30\pm3$ kHz(27-33 kHz) | 27 to 33 kHz | Same |
| Activation | By footswitch (optional),<br>ON/OFF button | By footswitch, ON/OFF<br>button | same |
| Patient<br>contacting<br>components | Ultrasonic working tip | Ultrasonic working tip | same |
| Power supply | Rechargeable Li-ion<br>battery<br>Capacity 1600mAh,3.7V | / | Different<br>The subject device<br>powered by a<br>rechargeable Li-ion<br>battery. The safety test<br>results show no new<br>safety risk raised. |
| Charger | Input: AC100-240V,<br>50/60Hz 0.2A<br>Output: DC5V/1A | AC: 100-240 V, 50/60 Hz<br>DC: 12 V, 1.5 A | Similar<br>testing shows no new<br>question raised |
| Bluetooth | BLE 5.0 | 4.1 Bluetooth | Different<br>Meet EMC standard and<br>FCC |
| Sterilization | Working Tip, Wrench and<br>Silicone Case are user<br>sterilized by<br>steam sterilization. | handpiece and the tool are<br>user sterilized by<br>steam sterilization | Same |
| Electrical<br>Safety | IEC 60601-1:2012<br>IEC 80601-2-60:2019 | IEC 60601-1:2012<br>IEC 80601-2-60:2019 | same |
| EMC | IEC 60601-1-2:2014 | IEC 60601-1-2:2014 | same |
| Biocompatibility | ISO 10993-5:2009<br>ISO 10993-10:2010<br>ISO 10993-11:2017 | ISO 10993-5:2009<br>ISO 10993-10:2010 | We conducted the acute<br>systemic toxicity testing<br>and pyrogen testing, test<br>results show no<br>biocompatibility risk. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for BOMEDENT. The logo consists of a stylized graphic to the left of the company name. The graphic is made up of three angled bars, two in gray and one in yellow. The company name, "BOMEDENT", is in gray, sans-serif font.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for BOMEDENT. The logo consists of two overlapping rectangles, one gray and one yellow, that are tilted to the left. To the right of the rectangles is the word "BOMEDENT" in gray, block letters. The logo is simple and modern.
All the differences don't affect substantial equivalence which is concluded after all the required testing.
#### 8. Performance Data
#### Clinical test:
Clinical testing is not required.
#### Non-clinical data
The proposed device Ultrasonic Endo Activation Device (Model: Actor I pro) complies with: Safety and performance:
IEC 60601-1:2005+A1:2012 Medical electrical equipment – Part 1: General requirements ー for basic safety and essential performance.
IEC 80601-2-60:2012 Medical electrical equipment - Part 2-60: Particular requirements for 2. the basic safety and essential performance of dental equipment
Usability
IEC 60601-1-6:2013 Medical electrical equipment - Part 1-6: General requirements for basic 3. safety and essential performance - Collateral standard: Usability
Electromagnetic Compatibility:
IEC 60601-1-2:2014 Medical electrical equipment-Part1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances-Requirements and tests
Biocompatibility:
ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation 5. and skin sensitization
6. ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
7. ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
Software Verification and Validation:
FDA software validation guidance "General Principles of Software Validation; Final Guidance for Industry and FDA Staff, Document issued on: January 11, 2002".
Software documentation for moderate level of concern per the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
Cleaning, intermediate level Disinfection, and Sterilization of the components of the subject
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for BOMEDENT. The logo consists of two angled rectangles, one gray and one yellow, followed by the text "BOMEDENT" in gray. The rectangles are positioned to the left of the text, with the yellow rectangle slightly overlapping the gray one. The text is in a sans-serif font and is aligned horizontally.
device per the FDA Guidance Document Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, AAMI TIR 30, AAMI TIR 12, ISO 17665-1, and ISO17665-2.
## 9. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is substantially equivalent to the legally marketed predicated device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.