K221979 · Shenzhen Imdk Medical Technology Co., Ltd. · DQA · Oct 16, 2022 · Cardiovascular
Device Facts
Record ID
K221979
Device Name
Pulse Oximeter (Model C101A2, C101B1, C101A3)
Applicant
Shenzhen Imdk Medical Technology Co., Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Oct 16, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
Fingertip Pulse Oximeter (Model:C101A2,C101B1,C101A3) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
Device Story
Portable, non-invasive fingertip pulse oximeter; models C101A2, C101B1, C101A3. Uses photoelectric oxyhemoglobin inspection technology; emits red (660nm) and infrared (905nm) light through finger tissue; measures light absorption via photosensitive element. Hardware/software processes signals to calculate SpO2 and pulse rate; displays PLETH curve, SpO2, PR, and battery status on OLED screen. Used in clinical environments for spot checks; operated by healthcare providers. Not for continuous monitoring, life support, or low perfusion conditions. Reusable; battery-powered (2 AAA). Provides immediate numerical and graphical feedback to clinicians for patient assessment.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing and biocompatibility evaluation (cytotoxicity, irritation, sensitization per ISO 10993).
Technological Characteristics
Materials: ABS shell, silicon clip pad, PC button. Sensing: Photoelectric (660nm red, 905nm IR). Power: 2 AAA alkaline batteries. Display: OLED. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 80601-2-61. Reusable, non-sterile. Software: Embedded, no changes from predicate.
Indications for Use
Indicated for adult patients in hospitals for non-invasive spot checking of functional oxygen saturation (SpO2) and pulse rate (PR).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 16, 2022
Shenzhen IMDK Medical Technology CO., Ltd. % Boyle Wang Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No. 161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China
Re: K221979
Trade/Device Name: Pulse Oximeter (Model C101A2, C101B1, C101A3) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: September 16, 2022 Received: September 16, 2022
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221979
Device Name Pulse Oximeter(Model:C101A2,C101B1,C101A3)
Indications for Use (Describe)
Fingertip Pulse Oximeter (Model:C101A2,C101B1,C101A3) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
Type of Use (Select one or both, as applicable)
| <div> <input checked="true" type="checkbox"/> <span>Research Use (21 CFR 201.3 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------------------------|
| <div> <input type="checkbox"/> <span>Over-The-Counter Use (21 CFR 201.3 Subpart G)</span> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K221979
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
# 1.0 Submitter's Information
| Name: | Shenzhen IMDK Medical Technology Co., Ltd. |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Address: | C Zone, 10F, Building 16, Yuanshan Industrial B Area,<br>Gongming Street, Guangming District, Shenzhen,<br>Guangdong, P.R. China |
| Tel: | 86-755-27155684 |
| Contact: | Yuan Xia |
| Registration Number: | 3015007456 |
### Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------|-----------------------------------------------------------------------|
| Name: | Shanghai Truthful Information Technology Co., Ltd. |
| Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai<br>200120 China |
| Tel: | +86-21-50313932 |
| Email: | Info@truthful.com.cn |
Date of Preparation: Sep.27,2022
# 2.0 Device Information
| Trade name: | Pulse Oximeter |
|----------------------|------------------------|
| Common name: | Pulse Oximeter |
| Classification name: | Oximeter |
| Model(s): | C101A2, C101B1, C101A3 |
| Production code: | DQA |
| Regulation number: | 21 CFR 870.2700 |
| Classification: | Class II |
| Panel: | Cardiovascular |
## 3.0 Predicate Device Information
| Manufacturer: | Shenzhen IMDK Medical Technology Co., Ltd |
|---------------|-------------------------------------------|
| Trade name: | Pulse Oximeter, Model: C101H1 |
{4}------------------------------------------------
510(k) number: K173123
### 4.0 Indication for Use Statement
Fingertip Pulse Oximeter (Model: C101A2, C101B1, C101A3) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
#### 5.0 Device Description
SpO2 is based on the absorption of pulse blood oxygen to red and infrared light by means of finger sensor and SpO2 measuring unit. The light-electronic transducer in finger sensor converts the pulse red and infrared light modulated by pulse blood oxygen into electrical signal, the signal is processed by hardware and software of the unit. The PLETH curve and numeral value of SpO2 will be obtained. The pulse oximeter, model C101A2, C101B1, C101A3 is designed for spot checking of the pulse oxygen saturation and pulse rate for adults in a clinic environment. This medical device can be reused. Not for continuously monitoring.
The device is not for life-supporting or life-sustaining, not for implant.
The device is not provided sterile and is NOT a reprocessed single-use device.
The device is a spot-check pulse oximeter and does not include alarms.
The device does not support the measurement in the condition of low perfusion.
The device is not intended for life-supporting or life-sustaining.
The device is reusable and does not need sterilization.
# 6.0 Technological Characteristics
#### Principle of Operation
There is no change in the principle of operation as part of this submission from the previous clearance under K173123. The module still utilizes the same principles of operation for pulse oximetry governed by the following principles:
1. Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
2. The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
#### Mechanism of Action for Achieving the Intended Effect
There is no change to the Mechanism of Action of the Pulse Oximeter (Model: C101A2, C101B1, C101A3) from the previous clearance K173123- Pulse Oximeter (Model: C101H1):
A mathematic formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and
{5}------------------------------------------------
Oxyhemoglobin (Hbo2) in red and near-infrared zones. Operation principle of the instrument: Photoeletric Oxyhemoglobin Inspection Technology adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm near infrared light) can be focused on a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the oximeters display through process in electronic circuits and microprocessor.
# 7.0 Summary of Technological Characteristics of the Subject Device Compared to the Predicate Device
The subject device, Pulse Oximeter (Model: C101A2, C101B1, C101A3) and the predicate device Pulse Oximeter (Model: C101H1), have the following key similarities:
- · Both devices have the same intended use
- · Both devices are indicated for the same patient population
- Both devices have the same principle of operation and mechanism of action
The subject device and the primary predicate device have the following differences: The appearance of the Subject device is different with that of the Predicate device. But this difference does not affect the basic safety and essential performance, the subject and the predicate are the same and substantially equivalent.
#### 8.0 Non-clinical Testing
In this current submission, just add four model C101A2, C101B1,C101A3 to the Legally marketed predicate device K173123.
As there were no hardware, software or performance changes made to the subject device when compared to the primary predicate device, no additional non-clinical testing was considered necessary to support the substantial equivalence.
#### Performance Bench Testing
As there are no hardware or software changes as part of this submission from the previous clearance, no performance bench testing was included as part of this submission.
#### Biocompatibility Testing
As the color of enclosure changes to the patient contacting materials as part of this submission from the previous clearance, biocompatibility testing was included as part of this submission.
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA for Industry and Food and Drug Administration Staff "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing within a risk management process". The biocompatibility
{6}------------------------------------------------
testing included the following tests:
- Cytotoxicity;
- Irritation;
- · Sensitization.
Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning
As there were no changes to the hardware or software as part of this submission from the previous clearance, no Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning testing was included with this submission.
Software Verification and Validation Testing
As there are no software changes as part of this submission from the previous clearance, no software testing was included as part of this submission.
#### 9.0 Clinical Test Conclusion
As the subject device utilizes the same monitoring technology as the predicate device, additional testing was not considered necessary to support the substantial equivalence.
#### 10.0 Technological Characteristic Comparison Table
{7}------------------------------------------------
# Table1-General Comparison
| Item | Subject Device<br>K221979 | Predicated Device<br>K173123 | Remark |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Name | Pulse Oximeter | Pulse Oximeter | -- |
| Manufacturer | Shenzhen IMDK Medical Technology<br>Co., Ltd. | Shenzhen IMDK Medical Technology<br>Co., Ltd. | -- |
| Product Code | DQA | DQA | Same |
| Regulation No. | 21CFR 870.2700 | 21CFR 870.2700 | Same |
| Class | Class II | Class II | Same |
| Model | C101A2, C101B1, C101A3 | C101H1 | -- |
| Intended<br>Use/Indication<br>for Use | Fingertip Pulse Oximeter (Model:<br>C101A2, C101B1, C101A3) is a<br>non-invasive device intended for<br>spot checking of functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate (PR). This<br>portable device is indicated for use<br>in adult patients in hospitals. | Fingertip Pulse Oximeter C101H1 is a<br>non-invasive device intended for spot<br>checking of functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate (PR). This<br>portable device is indicated for use in<br>adult patients in hospitals. | Same |
| Principle | The device displays numerical<br>values for functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate by<br>measuring the absorption of red and<br>infrared (IR) light passing through<br>perfused tissue. Changes in the<br>absorption caused by the pulsation<br>of blood in the vascular bed are<br>used to determine oxygen saturation<br>and pulse rate. | The device displays numerical values<br>for functional oxygen saturation of<br>arterial hemoglobin (SpO2) and pulse<br>rate by measuring the absorption of red<br>and infrared (IR) light passing through<br>perfuse tissue. Changes in the<br>absorption caused by the pulsation<br>of blood in the vascular bed are used to<br>determine oxygen saturation and pulse<br>rate. | Same |
| Applied<br>Population | Adults in a clinic environment | Adults in a clinic environment | Same |
| Application sites | Finger | Finger | Same |
| Display Type | OLED | OLED | Same |
| Display Content | Display the SPO2, PR, pulse bar<br>graph, pulse wave and battery<br>power status | Display the SPO2, PR, pulse bar<br>graph, pulse wave and battery power<br>status | Same |
| Contacting<br>material | Shell: ABS<br>Clip pad: Silicon<br>Button: PC | Shell: ABS<br>Clip pad: Silicon<br>Button: PC | Same |
| Overall<br>Dimension | C101A2: 60*36*33mm<br>C101B1: 66*36*33mm<br>C101A3: 58*36*33mm | 60*36*35mm | -- |
# Table 2 Performance Comparison
| Item | Subject Device<br>K221979 | Predicate Device<br>K173123 | Remark |
|-------------------------|-----------------------------|-----------------------------|--------|
| LED wavelength | Red= 660 nm; Infrared=904nm | Red= 660 nm; Infrared=904nm | Same |
| Power source | 2 AAA alkaline batteries | 2 AAA alkaline batteries | Same |
| Display data | SpO2%; PR | SpO2%; PR | Same |
| SpO2 Measuring<br>Range | 0%-100% | 0%-100% | Same |
| SpO2 Resolution | 1% | 1% | Same |
| SpO2 Accuracy | 70~100%, ±3%; | 70~100%, ±3%; | Same |
{8}------------------------------------------------
510(k) Summary
| PR Measuring<br>Range | 30-240BPM | 30-240BPM | Same |
|-----------------------|----------------------------------------|----------------------------------------|------|
| PR Resolution | 1 bpm | 1 bpm | Same |
| PR Accuracy | ±1 bpm or ±1%, whichever is<br>greater | ±1 bpm or ±1%, whichever is<br>greater | Same |
# Table 3 Safety Comparison
| Item | Subject Device<br>K221979 | Predicate Device<br>K173123 | Remark |
|-------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------|--------|
| Electrical Safety | Comply with IEC 60601-1<br>IEC 60601-1-11 | Comply with IEC 60601-1<br>IEC 60601-1-11 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
| Performance | ISO 80601-2-61 | ISO 80601-2-61 | Same |
| Biocompatibility | Cytotoxicity per ISO 10993-5;<br>Irritation per ISO 10993-10;<br>Sensitization per ISO 10993-10 | Comply with ISO 10993-1, FDA Guidance | Same |
#### 11.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the subject device is substantially equivalent to the legally marketed predicated device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.