K221875 · Qianjiang Kingphar Medical Packaging & Printing Co., Ltd. · FRG · Feb 13, 2023 · General Hospital
Device Facts
Record ID
K221875
Device Name
Sterilization Pouch and Roll
Applicant
Qianjiang Kingphar Medical Packaging & Printing Co., Ltd.
Product Code
FRG · General Hospital
Decision Date
Feb 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: · Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. · Ethylene Oxide (EO) with a concentration of 800 mg/L at 55°C (131°F) and 40% to 90% relative humidity for 6 hour. Aeration time of 7 days at 20℃). The sterilization pouch and roll are made with medical compound film. The sterilization pouch and roll maintains the sterility of the enclosed devices for up to 24 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam). Ethylene oxide sterilization indicator color change from pink before sterilization to yellow after sterilization. The steam sterilization indicator color change from blue before sterilization.
Device Story
Sterilization Pouch and Roll consist of medical-grade paper and compound film; used by healthcare workers to enclose medical devices for terminal sterilization. Input: medical devices placed inside pouch/roll; sealed via heat. Transformation: device undergoes Steam (132°C, 4 min) or EO (800 mg/L, 55°C, 6 hrs) sterilization. Output: sterilized, sealed package maintaining sterility for 24 months. External chemical ink indicators provide visual verification of exposure to sterilization process (Steam: blue to dark grey; EO: pink to yellow). Benefits: provides effective barrier to maintain sterility of instruments until use. Used in clinical settings; manual operation by healthcare staff.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via ISO 11135 (EO) and ISO 17665-1 (Steam) achieving SAL 10^-6. Biocompatibility confirmed per ISO 10993. Package integrity, seal strength, and microbial barrier efficacy verified via ASTM standards (F2251, F1140, F1929, F88, F2096) and DIN 58953-6.
Technological Characteristics
Materials: medical grade paper (60g/m²) and medical compound film (52um). Configuration: heat-sealed pouches and rolls. Chemical indicators: ink-based color change (Steam: blue to dark grey; EO: pink to yellow). Sterilization: Steam (132°C) and EO (800 mg/L). Shelf life: 5 years. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for healthcare workers to enclose medical devices for terminal sterilization via Steam or Ethylene Oxide (EO) processes. Maintains sterility of enclosed devices for up to 24 months post-sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
SterileRight Sterilization Pouch and Roll (K212338)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date February 13, 2023. The text is written in a clear, sans-serif font. The date is written in the standard US format, with the month first, followed by the day and year. The text is black and the background is white.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
Qianjiang Kingphar Medical Packaging & Printing Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China
Re: K221875
Trade/Device Name: Sterilization Pouch and Roll Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: January 17, 2023 Received: January 17, 2023
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anne D. Talley -S for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221875
Device Name Sterilization Pouch and Roll
The Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: · Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
· Ethylene Oxide (EO) with a concentration of 800 mg/L at 55°C (131°F) and 40% to 90% relative humidity for 6 hour. Aeration time of 7 days at 20℃).
The sterilization pouch and roll are made with medical compound film. The sterilization pouch and roll maintains the sterility of the enclosed devices for up to 24 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam). Ethylene oxide sterilization indicator color change from pink before sterilization to yellow after sterilization. The steam sterilization indicator color change from blue before sterilization.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
The maximum validated pouch load is 1.1 lbs (0.5kg); The maximum validated roll load is 1.1 lbs (0.5kg).
| Model | Item | Specification Size (WXL) | | Content/Max Load(lbs) | | |
|------------------------------------|--------|--------------------------|------------|-----------------------|---------|--------------|
| | NO. | Inch | mm*mm | Metal | Plastic | Gauze/Linens |
| | P-001 | 2.25"x3.5" | 57x90 | 0.02 | 0.01 | 0.003 |
| | P-002 | 2.25"x4" | 57x100 | 0.03 | 0.02 | 0.004 |
| | P-003 | 2.25"x5" | 57x130 | 0.04 | 0.03 | 0.005 |
| | P-004 | 2.75"x10" | 70x260 | 0.09 | 0.08 | 0.07 |
| | P-005 | 3.5"x5.25" | 90x135 | 0.10 | 0.09 | 0.08 |
| | P-006 | 3.5"x6.5" | 90x165 | 0.12 | 0.11 | 0.1 |
| | P-007 | 3.5"x10" | 90x260 | 0.14 | 0.13 | 0.11 |
| PLANE-01<br>Sterilization<br>Pouch | P-008 | 51/3"x10" | 135x260 | 0.48 | 0.46 | 0.45 |
| | P-009 | 51/3"x11 | 135x280 | 0.50 | 0.47 | 0.46 |
| | P-010 | 51/3"x13" | 135x330 | 0.55 | 0.53 | 0.50 |
| | P-011 | 7.5"x14" | 190x360 | 0.60 | 0.58 | 0.55 |
| | P-012 | 10"x14.5" | 250x370 | 1.06 | 1.05 | 0.62 |
| | P-013 | 10.25"x16" | 260x410 | 1.08 | 1.07 | 0.64 |
| | P-014 | 12"x17.7" | 300x450 | 1.10 | 1.09 | 0.66 |
| ROLL-01<br>Sterilization<br>Roll | R-101 | 2" | 50mm×200m | 0.02 | 0.01 | 0.003 |
| | R-102 | 3" | 75mm×200m | 0.09 | 0.05 | 0.004 |
| | R-103 | 4" | 100mm×200m | 0.14 | 0.13 | 0.11 |
| | R-104 | 6" | 150mm×200m | 0.55 | 0.53 | 0.50 |
| | R-105 | 8" | 200mm×200m | 0.60 | 0.58 | 0.55 |
| | R-106 | 10' | 250mm×200m | 1.06 | 1.05 | 0.62 |
| | R-107 | 12" | 300mm×200m | 0.60 | 0.58 | 0.55 |
| | R-108 | 14" | 350mm×200m | 1.06 | 1.05 | 0.62 |
| | R-109 | 16" | 400mm×200m | 1.08 | 1.07 | 0.64 |
| | R-110 | 17.7" | 450mm×200m | 1.10 | 1.09 | 0.66 |
| | R-111 | 20" | 500mm×200m | 1.10 | 1.09 | 0.66 |
| | GR-201 | 3" | 75mm×100m | 0.09 | 0.05 | 0.004 |
| | GR-202 | 4" | 100mm×100m | 0.14 | 0.13 | 0.11 |
| | GR-203 | 6" | 150mm×100m | 0.55 | 0.53 | 0.50 |
| | GR-204 | 8" | 200mm×100m | 0.60 | 0.58 | 0.55 |
| | GR-205 | 10" | 250mm×100m | 1.06 | 1.05 | 0.62 |
| | GR-206 | 12" | 300mm×100m | 0.60 | 0.58 | 0.55 |
| | GR-207 | 14" | 350mm×100m | 1.06 | 1.05 | 0.62 |
| | GR-208 | 16" | 400mm×100m | 1.08 | 1.07 | 0.64 |
(Model and Dimension and Content/Max. Load)
{4}------------------------------------------------
# 510(k) Summary K221875
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
| Name: | Qianjiang Kingphar Medical Packaging & Printing Co., Ltd. |
|----------|-----------------------------------------------------------------------------------|
| Address: | Dongjing Road 23, Yuanlin Office, 433100 Qianjiang,<br>PEOPLE'S REPUBLIC OF CHINA |
| Tel: | +86-13349816939 |
| Contact: | Zhao Fusong |
#### Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------|-----------------------------------------------------------------------|
| Name: | Shanghai Truthful Information Technology Co., Ltd. |
| Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai<br>200120 China |
| Tel: | +86-21-50313932 |
| Email: | Info@truthful.com.cn |
Date Submitted: Feb.11,2023
### 2.0 Device Information
| Trade/Device name: | Sterilization Pouch and Roll |
|------------------------------|--------------------------------------------------------------|
| Common name: | Sterilization Pouch and Roll |
| Regulation Name: | 1) Sterilization Wrap;<br>2) Sterilization process indicator |
| Classification Product Code: | 1) FRG; |
| Subsequent Product Code: | 2) JOJ |
| Regulation number: | 1) 21 CFR880.6850<br>2) 21 CFR880.2800 |
| Classification: | Class II |
| Panel: | General Hospital |
## 3.0 Predicate Device Information
#### Predicate Device:
{5}------------------------------------------------
| Manufacturer: | Sterileright Packaging Mfg., Inc |
|--------------------|-------------------------------------------|
| Trade/Device Name: | SterileRight Sterilization Pouch and Roll |
| 510(k) number: | K212338 |
#### Reference Device:
| Manufacturer: | Sigma Medical Supplies Corp. |
|--------------------|------------------------------------|
| Trade/Device Name: | SIGMA Sterilization Pouch and Roll |
| 510(k) number: | K180661 |
## 4.0 Device Description
The Sterilization pouch and roll is composed of medical grade paper(60g/m²) and medical compound film(52um), it is intended to be used to contain medical devices to be terminally sterilized by the EtO or Steam sterilization process. The recommended sterilization cycle parameter is:
Steam: 4 minutes at 132°C (270°F); 20 minutes dry time.
Ethylene oxide: 6 hours at 55 ℃; relative humidity between 40%- 90%; ethylene oxide concentration is 800 mg/L, 7 days aeration time at 20 ℃.
The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized for the EtO or Steam Sterilization Process. The heat-sealed pouch/roll are heat sealed prior to sterilization processing. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened.
The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 24 months post sterilization.
The Pouch/Roll is printed with a chemical indicator bar that changes from Pink to Yellow (EtO) or Blue to Dark grey (Steam) when exposed to EtO gas or Steam vapor during process. The EtO and Steam Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The EtO and Steam Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the EtO or Steam.
### 5.0 Indication for Use Statement
The Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
• Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
• Ethylene Oxide (EO) with a concentration of 800 mg/L at 55° C (131° F) and 40% to 90% relative humidity for 6 hour. Aeration time of 7 days at 20℃).
{6}------------------------------------------------
The sterilization pouch and roll are made with medical grade paper and medical compound film. The sterilization pouch and roll maintains the sterility of the enclosed devices for up to 24 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam). Ethylene oxide sterilization indicator color change from prink before sterilization to yellow after sterilization. The steam sterilization indicator color change from blue before sterilization to dark grey after sterilization.
The maximum validated pouch load is 1.1 lbs (0.5kg); The maximum validated roll load is 1.1 lbs (0.5kg).
| Model | Item<br>No. | Specification Size (WXL) | | Content/Max Load(lbs) | | |
|------------------------------------|-------------|--------------------------|------------|-----------------------|---------|--------------|
| | | Inch | mm*mm | Metal | Plastic | Gauze/Linens |
| | P-001 | 2.25"x3.5" | 57x90 | 0.02 | 0.01 | 0.003 |
| | P-002 | 2.25"x4" | 57x100 | 0.03 | 0.02 | 0.004 |
| | P-003 | 2.25"x5" | 57x130 | 0.04 | 0.03 | 0.005 |
| | P-004 | 2.75"x10" | 70x260 | 0.09 | 0.08 | 0.07 |
| | P-005 | 3.5"x5.25" | 90x135 | 0.10 | 0.09 | 0.08 |
| | P-006 | 3.5"x6.5" | 90x165 | 0.12 | 0.11 | 0.1 |
| | P-007 | 3.5"x10" | 90x260 | 0.14 | 0.13 | 0.11 |
| PLANE-01<br>Sterilization<br>Pouch | P-008 | 51/3"x10" | 135x260 | 0.48 | 0.46 | 0.45 |
| | P-009 | 51/3"x11 | 135x280 | 0.50 | 0.47 | 0.46 |
| | P-010 | 51/3"x13" | 135x330 | 0.55 | 0.53 | 0.50 |
| | P-011 | 7.5"x14" | 190x360 | 0.60 | 0.58 | 0.55 |
| | P-012 | 10"x14.5" | 250x370 | 1.06 | 1.05 | 0.62 |
| | P-013 | 10.25"x16" | 260x410 | 1.08 | 1.07 | 0.64 |
| | P-014 | 12"x17.7" | 300x450 | 1.10 | 1.09 | 0.66 |
| | R-101 | 2" | 50mm×200m | 0.02 | 0.01 | 0.003 |
| | R-102 | 3" | 75mm×200m | 0.09 | 0.05 | 0.004 |
| | R-103 | 4" | 100mm×200m | 0.14 | 0.13 | 0.11 |
| | R-104 | 6" | 150mm×200m | 0.55 | 0.53 | 0.50 |
| ROLL-01<br>Sterilization<br>Roll | R-105 | 8" | 200mm×200m | 0.60 | 0.58 | 0.55 |
| | R-106 | 10' | 250mm×200m | 1.06 | 1.05 | 0.62 |
| | R-107 | 12" | 300mm×200m | 0.60 | 0.58 | 0.55 |
| | R-108 | 14" | 350mm×200m | 1.06 | 1.05 | 0.62 |
| | R-109 | 16" | 400mm×200m | 1.08 | 1.07 | 0.64 |
| | R-110 | 17.7" | 450mm×200m | 1.10 | 1.09 | 0.66 |
(Model and Dimension and Content/Max. Load)
{7}------------------------------------------------
| R-111 | 20" | 500mm×200m | 1.10 | 1.09 | 0.66 |
|--------|-----|------------|------|------|-------|
| GR-201 | 3" | 75mm×100m | 0.09 | 0.05 | 0.004 |
| GR-202 | 4" | 100mm×100m | 0.14 | 0.13 | 0.11 |
| GR-203 | 6" | 150mm×100m | 0.55 | 0.53 | 0.50 |
| GR-204 | 8" | 200mm×100m | 0.60 | 0.58 | 0.55 |
| GR-205 | 10" | 250mm×100m | 1.06 | 1.05 | 0.62 |
| GR-206 | 12" | 300mm×100m | 0.60 | 0.58 | 0.55 |
| GR-207 | 14" | 350mm×100m | 1.06 | 1.05 | 0.62 |
| GR-208 | 16" | 400mm×100m | 1.08 | 1.07 | 0.64 |
# 6.0 Summary of Non-Clinical Testing
Summary of non-clinical and performance testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The test results demonstrated the subject device is confirmed to be safe and effective for the intended use.
| Test Item | Test Methodology | Acceptance Criteria or End Point | Results |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>Process Validation | ISO 11135:2014 | Ethylene oxide: 6 hours at 55 °C; relative<br>humidity between 40%- 90%;<br>ethylene oxide concentration<br>is 800 mg/L, 7days aeration<br>time at 20°C.<br>SAL=10-6 | SAL=10-6<br>Pass |
| | ISO 17665-1:2006;<br>ISO TS 17665-2:2009 | Steam; 8 minutes at 134°C; 8<br>minutes dry time.<br>SAL=10-6 | SAL=10-6<br>Pass |
| EO/ECH Residuals | ISO 10993-7:2008 | EO < 4mg<br>ECH < 9mg | Pass |
| Biocompatibility<br>testing | ISO 10993-5:2009 | Non-cytotoxic | Under conditions of the<br>study, did not show<br>potential toxicity to L-929<br>cells.<br>Pass |
| | ISO 10993-10:2010 | Non-irritating | Under the conditions of<br>the study, not an irritant.<br>Pass |
| | ISO 10993-10:2010 | Non-sensitizing | Under conditions of the<br>study, not a sensitizer.<br>Pass |
| Package Integrity /<br>Material<br>Compatibility<br>/Sterility<br>Maintenance | ASTM F 2251-13<br>ASTM F1140/F1140M-13 | 52μm±12%<br>Burst value > 3 Kpa or No Burst | Pass<br>Minimum of Burst pressure = 6.4 (kPa)<br>Pass |
| | ASTM F1929-15 | The dye solution is no any leakage across the seal width of sterile barrier system.<br>(No Infiltration) | No Infiltration<br>Pass |
| | ASTM F88/F88M-15 | Seal strength > 2.5 (N/15mm) | Minimum of Seal strength = 2.9<br>Pass |
| | ISO 1924-2:2008 | Tensile strength:<br>CD $≧$ 2.2KN/m,MD $≧$ 4.4KN/m | Pass |
| | Real time aging | 5 years real time aging<br>Incubation 24 months real time aging Incubation for post steam/EO sterilization storage | The test results of all samples are meet the requirements.<br>Pass |
| | ASTM F2096-11 | No Leakage | No Leakage<br>Pass |
| | DIN 58953-6 | CFU = 0 | CFU = 0<br>Pass |
| Chemical Indicator<br>Efficacy Testing | ISO 11140-1:2014 Steam:<br>The color of CI changes from Blue to Dark Grey, when exposed to Steam; | Steam<br>Change the color:<br>Color of indicator changes from blue to dark grey after Steam sterilization | Color of indicator changes from blue to dark grey under indicated condition and no changed under other condition.<br>Pass |
| | EO:<br>The color of CI changes from Pink to Yellow, when exposed to EO gas. | EO gas<br>Color of indicator changes from pink to yellow after EO sterilization | Color of indicator changes from pink to yellow under indicated condition and no changed under other condition.<br>Pass |
| 1. Remain stable<br>before use based on<br>its shelf life.<br><br>2. Maintain the<br>endpoint stability of the<br>color change after<br>being in the presence<br>of the sterilant. | All performance attributes<br>should maintain the original<br>color : 5 years shelf life | The real-time test was<br>carried out from Jan. 08,<br>2020 to Jan.20, 2022 that<br>demonstrates: the test<br>device which exposed to<br>Steam maintain the color<br>of Dark Grey, the test<br>group which exposed to<br>EO maintain the color of<br>Yellow, and the real-time<br>test was carried out on<br>the empty pouch that<br>demonstrates the test<br>device maintain the<br>original color from Jan.<br>08, 2017 to Jan.28, 2022. | |
#### Table 1: Performance testing summary – Bench
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
## 7.0 Technological Characteristic Comparison Table
| Item | Subject Device | Predicate Device | Reference Device | Comparison<br>Analysis |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name | Sterilization Pouch and Roll | SterileRight sterilization pouch<br>and roll | SIGMA Sterilization Pouch and<br>Roll | ---- |
| 510(k) No. | K221875 | K 212338 | K180661 | ---- |
| Product Code | FRG,JOJ | FRG,JOJ | FRG | Same |
| Regulation No. | 21 CFR 880.6850<br>21 CFR 880.2800 | 21 CFR 880.6850 | 21 CFR 880.6850 | Same |
| Class | II | II | II | Same |
| Intended Use | The Sterilization Pouch and Roll<br>are intended to provide health care<br>workers with an effective method<br>to enclose devices intended for<br>sterilization in the Steam or via<br>Ethylene Oxide (EO). The<br>recommended sterilization cycles<br>are as follows:<br>- Steam Sterilization at 132°C<br>(270°F) for 4 minutes; Drying time<br>of 30 minutes.<br>- Ethylene Oxide (EO) with a | The SterileRight Sterilization<br>Pouch and Roll are intended to<br>provide health care workers<br>with an effective method to<br>enclose devices intended for<br>sterilization in the Steam or via<br>Ethylene Oxide (EO). The<br>recommended sterilization cycles<br>are as follows:<br>- Gravity steam at 121°C (250°F)<br>for 30 minutes; Drying time of 25<br>minutes. | The SIGMA sterilization pouch<br>and roll are intended to provide<br>health care workers with an<br>effective method to enclose<br>devices intended for sterilization<br>in steam auto claves and via<br>Ethylene Oxide (EO). The<br>recommended steam sterilization<br>cycle parameters are 30 minutes<br>at 121°C. The recommended EO<br>sterilization cycle is 4 hours at<br>55°C with a relative humidity | Similar |
### Table 2- Comparison of Technology Characteristics
{11}------------------------------------------------
| concentration of 800 mg/L at 55°<br>C (131 ° F) and 40% to 90%<br>relative humidity for 6 hour.<br>Aeration time of 7 days at 20°C).<br><br>The sterilization pouch and roll are<br>made with medical grade paper<br>and medical compound film.The<br>sterilization pouch and roll<br>maintains the sterility of the<br>enclosed devices for up to 24<br>months post Steam or EO gas<br>sterilization, and before<br>sterilization has a maximum shelf<br>life of 5 years from the date of<br>manufacture.<br><br>The pouch's external chemical ink<br>indicators are designed to indicate<br>to the user that the pouch has<br>undergone either a steam or EO<br>sterilization process. Sterilization<br>indicator will change color<br>according to the sterilization<br>method (ethylene oxide, steam).<br>Ethylene oxide sterilization<br>indicator color change from Prink | • Pre-vacuum steam at 132°C<br>(270°F) for 4 minutes; Drying time<br>of 20 minutes.<br>• Ethylene Oxide (EO) with a<br>concentration of 735 mg/L at<br>55°C (131°F) and 50% to 80%<br>relative humidity for 60 minutes.<br>Aeration time of 8 hours at 60°C<br>(140°F).<br><br>The SterileRight provide<br>sterilization pouch and roll made<br>with Paper/Film or Tyvek® /Film.<br>The SterileRight sterilization<br>pouch and roll which are made<br>with Paper maintains the sterility<br>of the enclosed devices for up to<br>6 months post Steam or EO gas<br>sterilization, and before<br>sterilization has a maximum shelf<br>life of 3 years from the date of<br>manufacture. The pouch's<br>external chemical ink indicators<br>are designed to indicate to the<br>user that the pouch has<br>undergone either a steam or EO<br>sterilization process. | between 50%-85% and a sterilant<br>concentration of 600 mg/L.<br>Furthermore, the sterilization<br>pouch and roll maintains the<br>enclosed devices up until 3 years<br>post EO gas sterilization<br>and maintains the enclosed<br>devices up until 6 months post<br>Steam sterilization. Lastly, the<br>pouch's external chemical ink<br>indicators are designed to<br>indicate to the user that the pouch<br>has undergone either a steam or<br>EO sterilization process.<br><br>The SIGMA sterilization pouch<br>and roll is offered in the following<br>5 types:<br>• Self-sealing sterilization<br>pouches<br>• Sterilization pouches, Flat<br>• Sterilization pouches,<br>Gusseted<br>•Sterilization rolls, Flat<br>•Sterilization rolls, Gusseted |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{12}------------------------------------------------
| | before sterilization to Yellow after sterilization. The steam<br>sterilization indicator color change from Blue before sterilization to<br>Dark Grey after sterilization. | The SterileRight sterilization<br>pouch and roll which are made with Tyvek® is for EO gas<br>sterilization only. It also maintains the sterility of the enclosed<br>devices for up to 6 months post EO gas sterilization, and<br>before sterilization has a maximum shelf life of 3 years<br>from the date of manufacture. | | |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Material<br>Composition | Composed of medical grade paper<br>and medical compound film, EO<br>and Steam Process Indicator | Medical Grade Paper, CPP, PET,<br>adhesive,EO and Steam Process<br>Indicator, Print Ink.<br>Tyvek®, PET, PE, adhesive. | Medical Grade Paper, CPP, PET,<br>PU adhesive, EO and Steam<br>Process Indicator<br>Print Ink | Same |
| Sterilization<br>Cycles | • Steam Sterilization at 132°C<br>(270°F) for 4 minutes; Drying time<br>of 20 minutes.<br>• Ethylene Oxide (EO) with a<br>concentration of 800 mg/L at 55°<br>C (131 ° F) and 40% to 90%<br>relative humidity for 6 hour.<br>Aeration time of 7 days at 20°C). | • Gravity steam at 121°C (250°F)<br>for 30 minutes; Drying time of 25<br>minutes.<br>• Pre-vacuum steam at 132°C<br>(270°F) for4 minutes; Drying time<br>of 20 minutes.<br>• Ethylene Oxide (EO) with a<br>concentration of 735 mg/L at<br>55°C (131°F) and 50% to 80%<br>relative humidity for 60 minutes.<br>Aeration time of 8hours at 60°C<br>(140°F). | • Steam sterilization cycle<br>parameters are 30 minutes at<br>121°C.<br>• EO sterilization cycle is 4<br>hours at 55 °C with a relative<br>humidity between 50%-85%<br>and a sterilant concentration of<br>600 mg/L | Different |
| Configuration/<br>Dimension | Various Size, Heat Sealing | Various Size, Heat Sealing and<br>Self Sealing | Various Size, Heat Sealing | Same |
| Performance Testing | | | | |
| Sterilant<br>Penetration<br>Efficacy | The test meet the requirement of<br>SAL 10-6 | The test meet the requirement of<br>SAL 10-6 | The test meet the requirement of<br>SAL 10-6 | Same |
| Chemical<br>Indicator (CI)<br>Functionality<br>and Endpoint | The sterilant penetrated through<br>the pouch configuration and<br>affected the Cl color change to<br>the endpoint color | The sterilant penetrated through<br>the pouch configuration and<br>affected the CI color change to the<br>endpoint color | The sterilant penetrated through<br>the pouch configuration and<br>affected the Cl color change to the<br>endpoint color | Same |
| Device Design<br>of Steam CI | The color of Chemical Indicator<br>changes from Blue to<br>dark grey, when exposed to<br>Steam | The color of Chemical Indicator<br>changes from Pink to<br>Brown/Black, when exposed to<br>Steam | The color of Chemical Indicator<br>changes from Blue to<br>Greenish Black, when exposed to<br>Steam | Similar |
| Device Design<br>of EO gas CI | The color of Chemical Indicator<br>changes from Pink to Yellow,<br>when exposed to Steam | The color changes from Blue to<br>Yellow/Brown, when exposed to<br>EO gas | The color changes from - Red to<br>Yellow, when exposed to EO gas | Similar |
| Thickness<br>Variations (mm)<br>ASTM F 2251 | Passed | Passed | Passed | Same |
| Tensile strength<br>ISO 1924-2 | Passed | Passed | Passed | Same |
| Burst Strength<br>(kPa)<br>ASTM F1140 : | > 3.0 Kpa<br>Passed | Passed | Passed | Same |
| ISO 11607-1 | | | | |
| Bubble Leak<br>Test<br>ASTM D 3078<br>or ASTM-F<br>2096 | No Leakage<br><br>Passed | Passed | Passed | Same |
| Seal Peel Test<br>(N/15mm)<br>ASTM<br>F88/F88M ;<br>ISO 11607-1 | Min value= 2.08<br><br>Passed | Passed | Passed | Same |
| Dye penetration<br>Test ASTM<br>F1929 ;ISO<br>11607-1 | No Infiltration<br><br>Passed | Passed | Passed | Same |
| Microbial<br>Barrier Test<br>DIN 58953-6 | CFU = 0<br><br>Passed | Passed | Passed | Same |
| End point<br>stability<br>testing results | The color of chemical indicator for<br>EO sterilization indicator ink is<br>Pink, and the color of chemical<br>indicator for steam sterilization<br>indicator ink is Blue after 5 year<br>shelf life before sterilization. | The color of chemical indicator for<br>EO sterilization indicator ink is<br>Blue, and the color of chemical<br>indicator for steam sterilization<br>indicator ink is Pink after 2<br>year shelf life before sterilization. | The color of chemical indicator for<br>EO sterilization indicator ink is<br>Red, and the color of chemical<br>indicator for steam sterilization<br>indicator ink is Blue after 2 year<br>shelf life before sterilization. | Similar |
| | The color of chemical indicator for | The color of chemical indicator for | The color of chemical indicator for | Similar |
| | EO sterilization indicator ink is<br>Yellow, and the color of chemical<br>indicator for steam sterilization<br>indicator ink is Blue after EO<br>sterilized and 24 months shelf life. | EO sterilization indicator ink is<br>Yellow/Brown , and the color of<br>chemical indicator for steam<br>sterilization indicator ink is Pink<br>after EO sterilized and 6 months<br>shelf life | EO sterilization indicator ink is<br>yellow, and the color of chemical<br>indicator for steam sterilization<br>indicator ink is Blue after EO<br>sterilized and 3 years shelf life | |
| | The color of chemical indicator for<br>EO sterilization indicator ink is<br>Pink, and the color of chemical<br>indicator for steam sterilization<br>indicator ink is Dark Grey after<br>steam sterilized and 24 months<br>shelf life. | The color of chemical indicator for<br>EO sterilization indicator ink is<br>Blue, and the color of chemical<br>indicator for steam sterilization<br>indicator ink is Brown/Black after<br>steam sterilized and 6 months<br>shelf life. | The color of chemical indicator for<br>EO sterilization indicator ink is<br>Red, and the color of chemical<br>indicator for steam sterilization<br>indicator ink is Greenish Black<br>after steam sterilized and 6<br>months shelf life. | Similar |
| Maintenance of<br>Sterility | 24 months | 6 months | 3 years post EO gas sterilization<br>6 months post Steam sterilization | Different |
| Shelf Life | 5 years from date of manufacture<br>for EO and Steam Indicators | 3 years from date of manufacture<br>for EO and Steam Indicators | 3 years from date of manufacture<br>for EO and Steam Indicators | Different |
| Biocompatibility | Conform with ISO10993-1<br>(ISO10993-4, ISO10993-5,<br>ISO10993-10, ISO10993-11) | Conform with ANSI/AAMI/ISO<br>10993-10 | Conform with ISO 10993<br>standards | Same |
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device.
The comparison between the subject and predicate devices is based on the following:
- Same intended use
- · Same indications for use
- · Similar material types that meet ISO 10993 biocompatibility requirements
- · Same sterilization methods (EO and Steam sterilization process)
- Same fundamental technology/principal of operation/user interface
The sterilization parameters of the subject device is different with those of the predicate device, but the EO sterilization results demonstrate the subject device fully meet the requirements of ISO 11135 and the steam sterilization validation results fully meet the requirements of ISO 17665-1. The subjec…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.