K221792 · Smith & Nephew, Inc. · OSN · Aug 25, 2022 · Orthopedic
Device Facts
Record ID
K221792
Device Name
Acute QC Strut and Components
Applicant
Smith & Nephew, Inc.
Product Code
OSN · Orthopedic
Decision Date
Aug 25, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
External fixation devices are used on adults or pediatric patients as required. External fixation systems consist of various components that are used to build fixator assemblies unique to the patient's needs. These devices are modular; therefore, a multitude of different fixator frame configurations are possible. External fixation devices are used for the following indications below. 1. Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction 2. Open and closed fracture fixation 3. Pseudarthrosis of long bones 4. Limb lengthening by distraction 5. Correction of bony or soft tissue deformities 6. Joint arthrodesis 7. Infected fractures 8. Nonunions
Device Story
Acute QC Strut and Components are modular external fixation system line additions; used to build fixator assemblies for orthopedic trauma and deformity correction. Components include struts (various sizes), speed nuts, bands, and washers. System allows for acute multi-planar angular corrections. Operates via mechanical assembly; surgeons build circular frames to stabilize bone segments. Replaces universal joints of predicate with ball joints for locking mechanism; speed nuts feature knurled sections for tool-less advancement; washers include raised posts for half-pin stability. Used in clinical settings by surgeons. Output is a rigid external frame providing distraction, compression, or alignment of bone. Benefits include rapid assembly and stabilization of fractures or deformities. Device is MR Unsafe.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including fully reversed compressive fatigue loading, continuous compressive static loading, static compressive bending load on ball joint assembly, compressive fatigue loading of ball joint assembly, and torsional strength testing of half pin washers. Biological risk assessment performed per ISO 10993-1.
Technological Characteristics
Materials: aluminum, composite, stainless steel. Design: modular external fixation components (struts, nuts, bands, washers). Features: ball joint mechanism, knurled speed nuts, raised-post washers. Connectivity: none. Sterilization: not specified. Software: none (software from predicate SMART Taylor system is not used with these components).
Indications for Use
Indicated for adults and pediatric patients requiring external fixation for fractures and disease resulting in joint contractures or loss of range of motion, fractures requiring distraction, open/closed fracture fixation, pseudarthrosis of long bones, limb lengthening, correction of bony/soft tissue deformities, joint arthrodesis, infected fractures, and nonunions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
August 25, 2022
Smith & Nephew, Inc. Amanda Lammey Regulatory Affairs Specialist I 7135 Goodlett Farms Parkway Cordova, Tennessee 38016
Re: K221792
Trade/Device Name: Acute QC Strut and Components Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT, OSN, JDW Dated: June 17, 2022 Received: June 21, 2022
Dear Amanda Lammey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, Ph.D., R.A.C. For: Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K221792
Device Name Acute QC Strut and Components
#### Indications for Use (Describe)
1. Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
- 2. Open and closed fracture fixation
- 3. Pseudarthrosis of long bones
- 4. Limb lengthening by distraction
- 5. Correction of bony or soft tissue deformities
- 6. Joint arthrodesis
- 7. Infected fractures
- 8. Nonunions
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K221792 Acute QC Strut and Components
| 510(k) Summary Submitted by: | Smith & Nephew, Inc.<br>Advanced Surgical Devices Division<br>7135 Goodlett Farms Parkway<br>Cordova, Tennessee 38016 |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Date of Submission: | June 18, 2022 |
| Primary Contact Person: | Amanda Lammey, Regulatory Affairs Specialist I<br>M 901-833-1364<br>H 901-386-3453 |
| Secondary Contact Person: | Brad Sheals, Senior Manager Regulatory Affairs<br>M 901-288-7141<br>W 901-399-6897 |
| Name of Device: | Acute QC Strut and Components |
| Common Name: | External Fixation Systems |
| Device Classification Name and<br>Reference: | 21 CFR 888.3030 - Single/multiple component<br>metallic bone fixation appliances and accessories |
| Device Class: | Class II |
| Panel Code: | Orthopedics/87 |
| Product Code: | KTT, OSN, JDW |
| Predicate Device: | Primary Predicate – K210953 – SMART Taylor<br>Spatial Frame (S.E. 07/29/2021) |
| | Secondary Predicate – K994143 – ILIZAROVTM<br>External Fixation System (S.E. 02/18/2000) |
{4}------------------------------------------------
#### Device Description
The purpose of this Special 510(k) is to notify FDA of our intent to market the subject Smith & Nephew Acute QC Struts. The Smith & Nephew Acute QC Strut and Components consist of multiple sized struts (e.g., X Short, Short, Medium, and Long) and components such as nuts, bands, and washers to build an external fixation construct.
Subject of this premarket notification is the Acute QC Strut and Components. The Smith & Nephew Acute QC Strut and Components consist of multiple sized struts (e.g., X Short, Short, Medium, and Long) and components such as nuts, bands, and washers to build an external fixation construct. The proposed devices incorporate design features similar to those currently incorporated in previously cleared Smith & Nephew SMART Taylor Spatial Frame and ILIZAROV™ External Fixation System. The Acute QC Strut and Components will be manufactured from aluminum, composite, and stainless-steel material, which is identical to that of the predicate devices of the SMART Taylor Spatial Frame System (e.g., K210953, S.E. 07/29/2021).
The Smith & Nephew Acute QC Strut and Components are line additions to the predicate, SMART Taylor Spatial Frame System, which allows for acute multi-planer angular corrections to bone alignment due to a trauma event. The subject Acute QC Strut and Components and the predicate SMART Taylor Spatial Frame are identical in intended use and indications for use. The subject Acute QC Strut and Components are identical in indications for use also to the predicate ILIZAROV™ External Fixation System except for two statements: post-traumatic joint contracture which has resulted in loss of range of motion and management of comminuted intraarticular fractures of the distal radius. Both predicates are used in this submission because the subject Acute QC Strut and Components are line additions to the SMART Taylor Spatial Frame System which is based on the ILIZAROV™ Method. Therefore, the SMART Taylor Spatial
{5}------------------------------------------------
Frame System is the primary predicate and the ILIZAROV™ External Fixation System is the secondary predicate. The subject Acute QC Strut and Components: Acute QC Strut X Short, Acute QC Strut Short, Acute QC Strut Medium, Acute QC Strut Long, and Acute QC Strut Band are substantially equivalent to the cleared SMART Taylor Spatial Frame devices cleared via premarket notification K210953 (S.E. 07/29/2021). The subject Acute QC Strut and Components: Speed Nut, Half Pin Washer 2MM, and the Half Pin Washer 4MM are substantially equivalent to the previously cleared ILIZAROV™ External Fixation System devices: NUT, FIXATION BOLT WASHER 2MM and the FIXATION BOLT WASHER 4MM cleared via premarket notification K994143 (S.E. 02/18/2000).
#### Intended Use
External fixation devices are used on adults or pediatric patients as required. External fixation systems consist of various components that are used to build fixator assemblies unique to the patient's needs. These devices are modular; therefore, a multitude of different fixator frame configurations are possible. External fixation devices are used for the following indications below.
#### Indications for Use
The indications for the Smith & Nephew Acute QC Strut and Components are the following:
- 1. Fractures and disease which generally may result in joint contractures or loss of range
- of motion and fractures requiring distraction
- 2. Open and closed fracture fixation
- 3. Pseudarthrosis of long bones
- 4. Limb lengthening by distraction
- 5. Correction of bony or soft tissue deformities
{6}------------------------------------------------
- 6. Joint arthrodesis
- 7. Infected fractures
- 8. Nonunions
These indications are identical to the indications of the subject devices and the primary predicate devices of the SMART Taylor Spatial Frame System (e.g., K210953, S.E. 07/29/2021). All of the subject devices, primary predicate devices, and secondary predicate devices are intended for single use.
## Technological Characteristics
Device comparisons described in this premarket notification demonstrate that the subject devices, Acute QC Strut and Components, are substantially equivalent to legally marketed predicates with respect to intended use, indications for use, similar design, and performance characteristics. Clinical data was not needed to support the safety and effectiveness of the subject devices.
### Design Changes
The overall kinematics, intended use, material, and manufacturing process of the new device mimics those of the predicate devices. Design modifications have been introduced which provided a lock multiplanar mechanism to provide surgeons the ability to quickly build a circular frame to correct multiplanar deformities and fractures in trauma settings. The subject Acute QC Struts removed the universal joints that the primary predicate, SMART Taylor Spatial Frame (e.g., K210953, S.E. 07/29/2021), uses and replaced it with a ball joint. The subject Acute QC Strut Band removed the laseretch number from the predicate SMART FX ID Band (K210953) which was only used when utilizing the program software from the predicate SMART Taylor Spatial Frame System. The software is not used for the subject Acute QC Strut Band. The
{7}------------------------------------------------
subject Speed Nut modified the predicate, ILIZAROV™ External Fixation System (e.g., K994143 S.E. 01/18/2000), nut by adding knurled sections to the nut. The modifications were made by adding knurled sections to the nut to provide the ability to advance the nut without the use of instruments such as a wrench. The subject Half Pin Washers modified the predicate WASHER FIX BOLTS (K994143) by adding raised posts to provide additional stability to the 3sided half pins. The intended use of the subject devices did not change from the predicate devices. For detailed technology changes and visualizations of the subject and predicate devices, see Section 12 of this submission.
## Substantial Equivalence
The overall design, materials, and indications for use for the Acute QC Strut and Components are substantially equivalent to the predicate devices listed in the Table 6.1 below.
Table 6.1: Substantially Equivalent Predicate Devices
| | Manufacturer | Description | Submission Number | Pro Code | Clearance Date |
|------------------------|-------------------------|------------------------------------------|-------------------|-------------|-------------------|
| Primary<br>Predicate | Smith &<br>Nephew, Inc. | SMART Taylor<br>Spatial Frame | K210953 | KTT,<br>OSN | July 29, 2021 |
| Secondary<br>Predicate | Smith &<br>Nephew, Inc. | ILIZAROV™<br>External Fixation<br>System | K994143 | JDW | February 18, 2000 |
## Performance Testing:
To further support a determination of substantial equivalence, performance testing was
conducted on the subject, implantable devices in comparison against one or
more of the previously cleared predicate devices described in Table 6.1 above. A review
{8}------------------------------------------------
of the mechanical data in the submission indicates that the Acute QC Strut and Components are substantially equivalent to the previously cleared predicate devices. The following characteristics and performance testing were reviewed to determine the substantial equivalence.
- Fully Reversed Compressive Fatigue Loading of the Construct .
- o Continuous Compressive Static Loading of the Construct
- Static Compressive Bending Load on the Ball Joint Assembly ●
- Compressive Fatigue Loading of the Ball Joint Assembly
- o Torsional Strength Testing of the Subject Half Pin Washer
A rationale was provided for the subject Acute QC Strut and Components which states that the subject devices match the pyrogenicity information of the primary predicate, SMART Taylor Spatial Frame System (e.g., K210953, S.E. 07/29/2021).
A biological risk assessment of the devices was evaluated and completed in compliance with ISO 10993-1: 2018 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process to assess the risk of harm to health resulting from exposure to the device for the intended duration of contact. The subject Acute QC Struts are classified as surface-contacting devices that contact intact skin for prolonged durations (>24 hr to 30 days). The subject Acute QC Struts were determined to not present a greater biological risk than the cleared worst-case representative SMART FX STRUT (e.g., K210953, S.E. 07/29/2021). The Acute QC Strut Band, Speed Nut, and Half Pin Washers are all nonpatient contacting devices that are not required to have detailed biological endpoint evaluation per ISO 10993-1. All manufacture materials of these devices have long histories of safe use in medical devices.
An MR safety technical memo, which can be found in Exhibit 9 of this submission, was completed comparing the subject Acute QC Strut to external fixation devices of similar
{9}------------------------------------------------
indications which have been previously evaluated for MRI safety and compatibility. The material and devices associated with the subject Acute QC Strut and Components are the same as the previously evaluated Smith and Nephew, Inc. external fixation system, SMART Taylor Spatial Frame (e.g., K210953, S.E. 07/29/2021). The subject Acute QC Strut is considered at least similar in maximum mass, maximum length, material, surface finish, overall shape, and location/orientation as compared to the corresponding tests in the predicate studies. Thus, the subject Acute QC Strut would be expected to have at least similar responses in consideration of magnetically induced RF heating, magnetically induced displacement force, magnetically induced torque, and the MR image artifact as the previously cleared premarket notification tests determined. This technical memo matches the MR parameters of the cleared primary predicate, SMART Taylor Spatial Frame (e.g., K210953, S.E. 07/29/2021) and the subject Acute QC Strut and Components were determined to be MR Unsafe.
The packaging and labeling, including the instructions for use (Exhibit 4) and package insert (Exhibit 5) which was previously submitted to the FDA via premarket notification K210953 (S.E. 07/29/2021) for the SMART Taylor Spatial Frame System, also contains the MR Unsafe icons and/or parameters as specified in ASTM F2503, as appropriate.
#### Conclusion
In summary, this Special 510(k) premarket notification is being submitted to request clearance for the subject Acute QC Strut and Components. Based on similarities to the predicate SMART Taylor Spatial Frame and ILIZAROV™ External Fixation System, the subject line additions introduce no new risks related to the safety and effectiveness of the devices and Smith&Nephew, Inc believes the subject devices are substantially equivalent to the commercially available predicate devices listed in Table 6.1.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.