Bonafix TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
Device Story
Bonafix TiBase is a two-piece dental abutment system consisting of a titanium alloy base and a CAD/CAM-fabricated zirconia superstructure. The device is used by dental professionals to support single-unit or multi-unit prostheses. The titanium base connects to the dental implant, while the zirconia superstructure is bonded to the base using a dental cement (e.g., G-CEM LinkAce). The system is designed for use with specific compatible implant platforms. The restoration is designed using CAD/CAM software (EXOCAD) with locked parameters (e.g., 0° angulation, specific wall thicknesses) and manufactured at a validated milling center. The device provides a stable interface between the implant and the final dental restoration, benefiting patients by restoring function and aesthetics in edentulous areas.
Clinical Evidence
No clinical data was provided. Substantial equivalence was demonstrated through non-clinical bench testing, including dimensional analysis, compatibility verification, and MRI safety assessment based on scientific rationale and literature.
Technological Characteristics
Two-piece abutment: Ti-6Al-4V alloy (ASTM F136) base and zirconia (ISO 13356) superstructure. Internal connection. Dimensions: platform 3.0-5.7mm, post height 4.0-6.5mm, gingival height 0.7-5.0mm. Straight (0°) angulation. Provided non-sterile. CAD/CAM design process with locked parameters in EXOCAD software.
Indications for Use
Indicated for partially or fully edentulous patients requiring single-unit or multi-unit dental prostheses supported by dental implants in the maxillary or mandibular arch.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Zimmer Dental Tapered Screw Vent T Implant, HA Coated (K133339)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Zentek Medical LLC % Juan Tezak Consultant Compliance 4 Devices 118 W Prive Cr. Delray Beach, Florida 33445
Re: K221673
Trade/Device Name: Bonafix TiBase Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 15, 2023 Received: June 15, 2023
Dear Juan Tezak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221673
Device Name Bonafix TiBase abutment
#### Indications for Use (Describe)
Bonafix TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
| Compatible Implant System | Implant Body Diameter (mm) | Implant Platform Diameter (mm) |
|------------------------------------------|----------------------------------|--------------------------------|
| HIOSSEN ET III | 3.5<br> 4.0, 4.5, 5.0, 6.0, 7.0 | Mini<br> Regular |
| Nobel Active | 3.5<br> 4.3, 5.0 | NP<br> RP |
| Straumann Bone Level | 3.3<br> 4.1, 4.8 | NC<br> RC |
| Zimmer Screw-Vent/<br>Tapered Screw Vent | 3.3, 3.7, 4.1<br> 4.7<br> 6.0 | 3.5<br> 4.5<br> 5.7 |
| Bonafix 2 Plus Implants | 3.5, 3.75, 4.20, 5.0, 6.0 | 3.5 |
All digitally designed superstructures, and/or hybrid crowns for use with Titanium Base are to a Bonafix validated milling center for manufacture.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Bonafix TiBase (K221673) July 14th, 2023 ADMINISTRATIVE INFORMATION Applicant ZENTEK MEDICAL LLC 200 Craig Rd Ste 107 Manalapan, NJ 077268735 Phone: +1 732-2840545 Establishment Registration Number Official Contact Michael Vinnik Owner 200 Craig Rd Ste 107 Manalapan, NJ 077268735 +1 732-2840545 E-mail: mvinnik2@gmail.com Juan Tezak
510(k) Summary
Representative/Consultant
Carlos Marín Compliance4Devices 118 W Prive Cr. Delray Beach Fl, 33445 Phone: +1 561-789-2411 E-mail: compliance4devices@gmail.com
# DEVICE AND CLASSIFICATION NAME
Device Trade Name: Common Name: Classification Regulation: Classification Name: Device Classification: Classification Panel: Primary Product Code:
Bonafix TiBase Dental implant abutment 21 CFR 872.3630 Endosseous dental implant abutment Class II Dental Products Panel NHA
#### PREDICATE DEVICE INFORMATION
Primary predicate K212108, Dynamic Ti-base. Talladium España, SL
{4}------------------------------------------------
# Reference Devices
| K140934 | HIOSSEN ETIII SA. HIOSEEN® Implant |
|---------|--------------------------------------------------------------------------------------------------------------------------------|
| K153758 | STRAUMANN DENTAL IMPLANT SYSTEM. STRAUMANN USA |
| K061410 | Tapered Screw-Vent. Zimmer Dental |
| K142260 | Nobel Active conical. Nobel Biocare AB |
| K153332 | ETIII SA Fixture System (Ø3.2mm). HIOSEEN® Implant |
| K193352 | Abutment CAD. Exocad GmbH |
| K120243 | G-CEM LinkAce™. GC AMERICA INC |
| K133339 | Zimmer Dental Tapered Screw Vent T Implant, HA Coated; Zimmer Dental<br>Tapered Screw-Vent M Implant, HA Coated, Zimmer Dental |
### Intended Use
Bonafix TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
| COMPATIBLE IMPLANT SYSTEM | IMPLANT BODY DIAMETER<br>(mm) | IMPLANT PLATFORM DIAMETER<br>(mm) |
|------------------------------------------|-------------------------------|-----------------------------------|
| HIOSSEN ET III | 3.5 | Mini |
| | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular |
| Nobel Active | 3.5 | NP |
| | 4.3, 5.0 | RP |
| Straumann Bone Level | 3.3 | NC |
| | 4.1, 4.8 | RC |
| Zimmer Screw-Vent/<br>Tapered Screw-Vent | 3.3, 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| | 6.0 | 5.7 |
| Bonafix 2 Plus Implants | 3.5, 3.75, 4.20, 5.0, 6.0 | 3.5 |
All digitally designed superstructures, and/or hybrid crowns for use with Titanium Base are to be sent to a Bonafix validated milling center for manufacture.
# Device Description
The Bonafix TiBase abutment is composed of two-piece abutment that is a titanium base at the bottom and a zirconia superstructure (CAD/CAM patient specific superstructure) at the top. The dental restoration and mesostructure are fabricated using a CAD/CAM process. The subject device abutment platform diameters range from 3.0 mm to 5.7 mm, and the corresponding compatible implant body diameters also range from 3.3 mm to 7.0 mm. The apical end is prefabricated to match the compatible implant platform and is available with implant connections for crowns (engaging) or bridges (non-engaging). Each abutment is provided with a screw designed to match the compatible implant.
The titanium base abutment and screw are manufacturated from titanium alloy conforming to ASTM F136. The superstructure is to be manufactured from zirconia conforming to ISO 13356. The subject devices are provided non-sterile to the end user. All digitally designed superstructures, and/or hybrid crowns for use with Bonafix TiBase abutments are to be sent to a Zentek validated Milling center for manufacture. The zirconia superstructure in straight only and is not to be designed to provide an angle or divergence correction.
{5}------------------------------------------------
The design parameters for the fabrication of the restoration on Bonafix TiBase abutments, which are already locked in the EXOCAD CAD/CAM software (K193352), are as follows:
- Minimum wall thickness 0.43 mm
- Minimum abutment post height 4.0 mm
- Maximum post height 6.5 mm
- Maximum gingival height 5.0 mm
- Minimum gingival height 0.7 mm
- Angulation 0°
The recommended cement for bonding the zirconia superstructure to the Bonafix TiBase abutments to create the final two-piece abutment is G-CEM LinkAce™ cleared in K120243.
{6}------------------------------------------------
#### Equivalence to Marketed Device
The subject device is substantially equivalent in inciples to the primary predicate device. Below is a table comparing the indications for use and technological characteristics of the subject device and the primary predicate device.
The subject device is equivalent to the identified primary predicate, with the compatible implant systems for which the reference devices are included and supported by reverse engineering.
The Indications for Use Statement (IFUS) for the subject device is identified primary predicate K212108, with the differences being the compatible implant systems for which the reference devices K14034, K142260, K15332, and K13339 are included and which are supported by reverse engineering.
The other reference devices K120243 and K193352 serve as references for the cement and software used in the design of the Bonafix TiBase abutment restoration.
| COMPARISON | SUBJECT DEVICE | PREDICATE DEVICE | COMPARISON | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Device | Bonafix TiBase abutment (K221673)<br>Zentek Medical LLC | Dynamic Ti-base (K212108).<br>Talladium España, SL | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Intended Use | Bonafix TiBase abutments are intended for<br>use with dental implants as a support for<br>single-unit or multi-unit prostheses in the<br>maxillary or mandibular arch of a partially or<br>fully edentulous patient. | Dynamic TiBase abutments are intended for use<br>with dental implants as a support for single-unit or<br>multi-unit prostheses in the maxillary or mandibular<br>arch of a partially or fully edentulous patient. | Similar | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | COMPATIBLE IMPLANT SYSTEM IMPLANT BODY DIAMETER (mm) IMPLANT PLATFORM DIAMETER (mm) HIOSSEN ET III 3.5 Mini Nobel Active 4.0, 4.5, 5.0, 6.0, 7.0 Regular 3.5 NP Straumann Bone Level 4.3, 5.0 RP 3.3 NC Zimmer Screw-Vent/<br>Tapered Screw-Vent 4.1, 4.8 RC 3.3, 3.7, 4.1 3.5 4.7 4.5 6.0 5.7 Bonafix 2 Plus Implants 3.5, 3.75, 4.20, 5.0, 6.0 3.5 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Implant Compatibility Implant Body Diameter, mm Implant Platform, mm 2.7 3.5 SPI CONTACT Dental Implant 3.5 4.0 3.5 4.5 4.2 5.0 | | | | | | | | | | | | | | | | |
#### Table 1. Comparison with predicate device for Summary
{7}------------------------------------------------
| COMPARISON | SUBJECT DEVICE | PREDICATE DEVICE | COMPARISON |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | All digitally designed superstructures, and/or hybrid crowns for use with Titanium Base are to be sent to a Bonafix validated milling center for manufacture. | All digitally designed custom abutments for use with Dynamic TiBase abutments are to be sent to a Thommen Medical validated milling center for manufacture. | |
| Abutment<br>Designs | Titanium Base Engaging<br>Titanium Base Non-Engaging | Titanium Base Engaging<br>Titanium Base Non-Engaging | Same |
| Prosthesis | Cement-retained | Cement-retained | Same |
| Attachment | Screw-retained | Screw-retained | |
| Abutment/Implant Platform<br>Diameter (mm) | 3.0 – 5.7 | 3.5, 4.0, 4.5, 5.0 | Similar |
| Abutment<br>Angle | Straight (0°) | Up to 30° | Different |
| | CAD/CAM Abutment Design Parameters | | |
| Minimum post<br>height, mm | 4.0 | 4.0 | Same |
| Maximun post<br>heigth, mm | 6.5 | 9.0 | Different |
| Maximum<br>gingival height,<br>mm | 5.0 | 5.83 | Similar |
| Minimum<br>gingival height,<br>mm | 0.7 | 0.7 | Same |
| Wall Minimum<br>wall thickness | 0.43 | 0.43 | Same |
| COMPARISON | SUBJECT DEVICE | PREDICATE DEVICE | COMPARISON |
| Maximum<br>abutment<br>angulation | 0° | Up to 30° | Different |
| Abutment<br>attachment to<br>implant | Screw | Screw | Same |
| Prosthesis<br>attachment to<br>abutment | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Same |
| Restoration | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Same |
| Abutment/<br>Implant<br>Interface | Internal connection | Internal | Same |
| Materials | | | |
| Abutment | Ti-6Al-4V alloy (ASTM F136) | Titanium alloy, ASTM F136 | Same |
| Screw | Ti-6Al-4V alloy (ASTM F136) | Titanium alloy, ASTM F136 | Same |
| Superstructure | Zirconia - ISO 13356 | Zirconia - ISO 13356 | Same |
{8}------------------------------------------------
{9}------------------------------------------------
# Non-Clinical Testing Summary
Bonafix TiBase underwent evaluation tests based on the following standards:
| ISO 10993-5 | 2014 | Biological evaluation of medical devices - Part 5: Tests for in vitro<br>cytotoxicity. |
|--------------------------|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI/ISO<br>17665-1 | (2006/(R)2<br>013) | Sterilization of health care products — Moist heat — Part 1: Requirements<br>for the development, validation and routine control of a sterilization process<br>for medical devices, Annex D |
#### Table 2 Standards compliance for Summary
End User Sterilization Validation Test according to ISO 17665-1.
Engineering and dimensional analysis of original manufactures' components (abutments, implants & abutment screws) for determination of compatibility. A plan for ongoing monitoring to ensure continued compatibility is also implemented.
Additional, Non-clinical worst-case MRI review was performed to evaluate the metallic Bonafix TiBase device in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
# Clinical Testing Summary
Clinical testing was not required to demonstrate the substantial equivalence of the Bonafix TiBase to its predicate device.
# Conclusion
Based on the information presented in this 510(k), the Bonafix TiBase is considered substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.