K153332 · Hiossen, Inc. · DZE · Oct 27, 2016 · Dental
Device Facts
Record ID
K153332
Device Name
ETIII SA Fixture System (O3.2mm)
Applicant
Hiossen, Inc.
Product Code
DZE · Dental
Decision Date
Oct 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ETIII SA Fixture System (03.2mm) is indicated for use in mandibular and maxillary lateral and central incisor, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading.
Device Story
ETIII SA Fixture System (Ø3.2mm) is an endosseous dental implant system; surgically placed in mandibular or maxillary bone to support prosthetic restorations. Device consists of titanium alloy (Ti-6Al-4V) fixture, mount, and pure titanium cover screw. Features internal hex, bone-level design, straight body, and 3-sided self-tapping cutting edge. Surface treated via sandblasting and acid etching (SA). Supplied sterile via radiation. Used by dental professionals in clinical settings. Provides stable foundation for artificial teeth; restores chewing function in partially edentulous patients.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including static and dynamic fatigue testing, rotation angle/gap accuracy, and biocompatibility testing (ISO 10993 series: subchronic toxicity, genotoxicity, cytotoxicity, sensitization, irritation, systemic toxicity, and pyrogenicity).
Indicated for patients requiring dental implants in mandibular and maxillary lateral and central incisor regions to support single or multiple-unit restorations (cemented, screw-retained, or overdenture) and fixed bridgework. Intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2016
HiOSSEN Inc. David Kim Manager 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030
Re: K153332
Trade/Device Name: ETIII SA Fixture System (Ø 3.2mm) Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 27, 2016 Received: September 28, 2016
Dear David Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
SECTION 007
Image /page/2/Picture/1 description: The image shows the logo and contact information for Hiossen Inc. The logo features the company name in a stylized font, with the "H" in red and the rest of the name in gray. Below the logo is the company name, address (85 Ben Fairless Dr. Fairless Hills, PA 19030), phone number (1-888-678-0001), fax number (1-267-759-7004), and website address (www.hiossen.com).
# Indications for Use Statement
510(k) Number K ______________________________________________________________________________________________________________________________________________________________
Device Name : The ETIII Fixture System (Ø3.2mm)
Indication for use :
The ETIII SA Fixture System (03.2mm) is indicated for use in mandibular and maxillary lateral and central incisor, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading.
Prescription Use X (Per 21CFR801 Subpart D) OR Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Section 008
Image /page/3/Picture/1 description: The image shows the logo for Hiossen. The logo is in two colors, red and gray. The letters "HI" are in red, and the rest of the letters are in gray.
Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com
## 510(k) Summary
This summary of 510(k) substantial equivalence information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date : Sept. 27, 2016
1. Company and Correspondent making the submission:
| - Submitter's Name : | HiOSSEN Inc. |
|----------------------|------------------------------------------------|
| - Address : | 85 Ben Fairless Dr.<br>Fairless Hills PA 19030 |
| - Telephone No. | 888 678 0001 |
| - Contact : | Mr. David Kim |
#### 2. Device :
| Trade or (Proprietary) Name : | ETIII SA Fixture System (Ø3.2) |
|-------------------------------|--------------------------------|
| Common or usual name : | Endosseous Dental Implant |
| Regulation Number: | 21 CFR 872.3640 |
| Regulation Name: | Endosseous Dental Implant |
| Regulatory Class: | Class II |
| Product Code: | DZE, NHA |
#### 3. Predicate Device:
Primary Predicate: HIOSSEN IMPLANT SYSTEM (HIOSSEN Inc., K140934) Referenced Predicate #1: MS SA IMPLANT SYSTEM (HIOSSEN Inc., K122171) Referenced Predicate #2: NobelActive 3.0 ( NobelBiocare, K102436 )
#### 4. Description :
- 1) The ETIII SA Fixture System (Ø3.2) is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.
- 2) The ETIII SA Fixture System (Ø3.2mm) is made of titanium alloy (Ti 6Al 4V) for Fixtures and Simple Mount and pure titanium for Cover Screw. The ETIII SA Fixture System (Ø3.2mm) is indicated for use in mandibular and maxillary lateral and central incisor, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for HIOSSEN. The letters "HI" are in red, and the rest of the letters are in gray. There is a red swoosh above the "HI".
### Section 008
Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com
#### 3) Fixture
| Device | Intended to be surgically placed in the bone of the upper or lower jaw arches. | | |
|-------------|--------------------------------------------------------------------------------|-----------------------------|------------|
| Description | Fixture is supplied sterile. | | |
| Material | Alloyed Titanium (ASTM F136) | | |
| Surface | SA surface treatment (Sand blasted and Acid etched ) | | |
| | Diameter (mm) | Length (mm) | Connection |
| Dimension | 3.2 | 8.7, 10.2, 11.7, 13.2, 15.2 | Mini |
#### 4) Component
| Mount Set | |
|---------------|----------------------------------------------------------------------------------|
| Description | To be used for easy implanting of the Fixture |
| Material | Alloyed Titanium (ASTM F136) |
| Diameter (mm) | 4.8 |
| Length (mm) | 16 |
| Connection | Mini |
| Cover Screw | |
| Description | To be used to protect the exposed platform of the implant during healing period. |
| Material | Pure Titanium (ASTM F 67) |
| Diameter (mm) | 3.1 |
| Length (mm) | 4.75 |
| Connection | Mini |
#### Substantial Equivalence
| | Proposed device | Primary<br>predicate device | Reference<br>predicate device | Reference<br>predicate device |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ETIII SA Fixture<br>System (Ø3.2mm) | Hiossen Implant<br>System | MS SA Implant<br>System | NobelActive 3.0 |
| Design | Image: ETIII SA Fixture System (Ø3.2mm) | Image: Hiossen Implant System | Image: MS SA Implant System | Image: NobelActive 3.0 |
| 510(K) No. | N/A | K140934 | K122171 | K102436 |
| www.hiossen.com | | | | |
| Indication<br>for use | The ETIII SA Fixture<br>System (Ø3.2mm) is<br>indicated for use in<br>mandibular and<br>maxillary lateral and<br>central incisor, in<br>support of single or<br>multiple-unit<br>restorations including;<br>cemented retained,<br>screw retained, or<br>overdenture<br>restorations, and final<br>or temporary<br>abutment support for<br>fixed bridgework. It is<br>intended for delayed<br>loading. | The HIOSSEN Implant<br>System is indicated for<br>use in partially or fully<br>edentulous mandibles<br>and maxillae, in support<br>of single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is intended<br>for delayed loading.<br>ETIII SA Ultra-Wide<br>Fixture is intended to be<br>used in the molar region. | The MS SA Implant<br>(Narrow Ridge) is<br>intended to use in the<br>treatment of missing<br>mandibular central and<br>lateral incisors to support<br>prosthetic device, such as<br>artificial teeth, in order to<br>restore chewing function<br>in partially edentulous<br>patients. The MS SA<br>Implant (Narrow Ridge)<br>is intended for single use<br>only. It is intended for<br>delayed loading. | The NobelActive<br>3.0mm implant is<br>indicated for use in the<br>treatment of missing<br>maxillary lateral incisors<br>or the mandibular<br>central and lateral<br>incisors to support<br>prosthetic devices, such<br>as artificial teeth,in<br>order to restore chewing<br>function in partially<br>edentulous patients. The<br>NobelActive 3.0<br>implants may be put into<br>immediate function<br>provided that stability<br>requirements detailed in<br>the manual are satisfied. |
| Surgery type | One or two stage<br>Surgery | One and two stage<br>Surgery | One stage<br>Surgery | One and two stage<br>Surgery |
| Structure | - Internal Hex<br>- Bone Level Implant<br>- Straight body shape<br>- 3 sided cutting edge<br>with self-tapping | - Internal Hex<br>- Bone Level Implant<br>- Tapered body shape<br>- 3 sided cutting edge<br>with self-tapping | - One Body Implant<br>- Straight body shape<br>- Threaded body<br>design<br>- 2 sided cutting edge | -Internal Hex<br>- Bone Level Implant |
| Body<br>Diameter | 3.2 mm | 3.5, 3.75, 3.77, 4.2, 4.25,<br>4.45, 4.6, 4.63, 4.65, 4.9,<br>5.0, 5.1, 5.05, 5.08, 5.92,<br>5.95, 6.8 mm | 2.5, 2.9 mm | 3.0 mm |
| Length | 8.7, 10.2, 11.7, 13.2,<br>15.2 mm | 6.2, 7.2, 8.2, 8.7, 9.7,<br>10.2, 11.2, 11.7, 12.7,<br>13.2, 15.2, 18.2 mm | 8.5, 10, 11.5, 13 mm | 10, 11.5, 13, 15 mm |
| Material of<br>Fixture | Titanium Alloy<br>Ti-6Al-4V<br>(ASTM F136) | Pure Titanium Grade 4<br>(ASTM F67) | Titanium Alloy<br>Ti-6Al-4V<br>(ASTM F136) | Pure Titanium Grade 4<br>(ASTM F67) |
| Surface | SA(Sandblasting<br>& Acid etching) | SA(Sandblasting<br>& Acid etching) | SA(Sandblasting<br>& Acid etching) | TiUnite |
| Steriliza-tion | Radiation Sterile | Radiation Sterile | Radiation Sterile | Radiation Sterile |
| Shelf life | 8years | 8years | 8 years | n/a |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for HIOSSEN. The logo is in two colors, red and gray. The letters "HI" are in red, and the rest of the word "OSSEN" is in gray. There is a red swoosh above the "HI".
Hiossen Inc.
85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com
# Section 008
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Hiossen. The letters "HI" are in red, with a red swoosh above the "H". The rest of the word, "OSSEN", is in gray. The logo is simple and modern.
Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com
| Differences | The indications for use:<br>The indication for use statement for the proposed device is not identical to the predicate device(K140934);<br>however, the difference does not alter the general intended use of dental implant. Both the subject and<br>predicate devices have the same intended use for the replacement of missing teeth. And the indication for<br>use of proposed device is very similar with predicate device, K102436.<br>Material:<br>The proposed device is made with alloyed titanium (Ti-6Al 4V) and the predicate device, K140934 is made<br>with pure titanium (Ti CP4). The difference in material is insignificant chemical composition. And another<br>predicate device, K122171 is made with alloyed titanium(Ti-6Al-4V) |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarities<br>And<br>S.E | Proposed device, ETIII SA Fixture System (Ø3.2mm) is same with predicate devices in intended<br>use (K140934 / K102436), material (K122171 ), and surface treatment (K140934 ). |
- 5. Indication for use :
The ETII SA Fixture System (Ø3.2mm) is indicated for use in mandibular and maxillary lateral and central incisor, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading.
- 6. Summary of nonclinical testing
Hiossen predicate K140934 was relied upon for sterility, shelf-life and biocompatibility. Additionally LAL testing per USP <85> was conducted for the subject device.
Non-clinical testing data submitted to demonstrate substantial equivalence includes adaptation accuracy for rotation angle and gap, static and dynamic fatigue test according to "Guidance for industry and FDA staff Class Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" All tests were conducted in accordance with relative standards without any discrepancies.
| Test Items | Standard |
|-----------------------------------------------------|----------------------------|
| Subchronic Toxicity test ( ISO 90 Days, Bone ) | ISO 10993-11 , ISO 10993-6 |
| Bacterial reverse mutation test (Genotoxicity test) | ISO 10993-3, ISO 10993-12 |
| Chromosome aberration test (Genotoxicity test) | ISO 10993-3, ISO 10993-12 |
| Cytotoxicity test | ISO 10993-5, ISO 10993-12 |
| Sterility test | USP 36 <71> |
| Skin sensitization test (GPMT) | ISO 10993-10 |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "HIOSSEN" in a stylized font. The "HI" is in red, and the rest of the word is in gray. There is a red swoosh above the "I" in "HI".
Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com
| Oral Mucosa Irritation test | ISO 10993-10 |
|------------------------------|----------------------------|
| Acute Systemic Toxicity test | ISO 10993-11 |
| Pyrogen test | ISO 10993-11, USP 36 <151> |
7. Summary of clinical testing No clinical studies are submitted.
#### 8. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, HIOSSEN Inc. concludes that the The ETIII SA Fixture System (Ø3.2) is substantially equivalent to the predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.