K221652 · Renal Care Dialysis Solutions, S.A. DE C.V. · KPO · Mar 3, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K221652
Device Name
NIKKICART
Applicant
Renal Care Dialysis Solutions, S.A. DE C.V.
Product Code
KPO · Gastroenterology, Urology
Decision Date
Mar 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NikkiCart cartridge is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute renal failure, chronic renal failure, or acute intoxication with dialyzable substances. The NikkiCart cartridge is intended to be used as one component in the preparation of dialysate according to a physician's prescription in a 3-stream proportioning machine on the Nikkiso DBB-06 dialysis system equipped with a compatible bicarbonate cartridge holder.
Device Story
NikkiCart is a single-use polypropylene cartridge containing dry sodium bicarbonate powder; used as a component in hemodialysis. When attached to a Nikkiso DBB-06 dialysis system, the machine draws water through the cartridge to create a saturated sodium bicarbonate solution. This solution is mixed with acid concentrate and water by the machine to produce bicarbonate-based dialysis fluid. Used in clinical settings by healthcare professionals; output is dialysate fluid for extracorporeal treatment. Benefits include enabling on-line preparation of dialysis fluid tailored to physician prescription for renal replacement therapy.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility), chemical characterization (chronic systemic toxicity, mutagenicity/carcinogenicity), stability testing (24-month shelf life), and shipping verification per ASTM D4169-16.
Technological Characteristics
Single-use polypropylene (PP) cartridge; contains USP/European Pharmacopeia grade sodium bicarbonate powder. Operates via water flow through the cartridge to create a saturated solution. Designed for 1:44 (45X) acid dilution ratio. Compatible with Nikkiso DBB-06 dialysis system. Sterilization not specified. No electronic or software components.
Indications for Use
Indicated for patients with acute renal failure, chronic renal failure, or acute intoxication with dialyzable substances undergoing extracorporeal bicarbonate hemodialysis.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 3, 2023
Renal Care Dialysis Solutions, S.A. DE C.V. % Brittany Valdez Nava Head of Quality Healthcare Innovations Catalysts 7811 Montrose Road, Suite 215 Potomac, Maryland 20854
Re: K221652
Trade/Device Name: Nikkicart Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: KPO Dated: January 30, 2023 Received: February 1, 2023
Dear Brittany Valdez Nava:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Gema Gonzalez -S
Gema Gonzalez, MS Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221652
Device Name NIKKICART
#### Indications for Use (Describe)
The NikkiCart cartridge is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute renal failure, chronic renal failure, or acute intoxication with dialyzable substances. The NikkiCart cartridge is intended to be used as one component in the preparation of dialysate according to a physician's prescription in a 3-stream proportioning machine on the Nikkiso DBB-06 dialysis system equipped with a compatible bicarbonate cartridge holder.
Type of Use (Select one or both, as applicable)
| <div> <span style="text-decoration: overline;">☒</span> <span>Transitioning To A STEP-NJ Science Program</span> </div> | <div> <span>☐</span> <span>For The Current Year In A STEP-NJ Science Program</span> </div> |
|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for RCDS Renal Care Dialysis Solutions SA de CV. The logo features a stylized image of two kidneys on the left, followed by the letters RCDS in a sans-serif font. The text "Renal Care Dialysis Solutions SA de CV" is written in a smaller font below the letters RCDS.
#### 510(k) Summary
Company Name, Address: a. RENAL CARE DIALYSIS SOLUTIONS, S.A. DE C.V. Carretera a los Cues K.M. 2.2., Lote 23, Bodegas 5 y 6, Parque Tecnològico Innovaciòn Querètaro El Marquès, Querètaro 76246 Mexico
- b. Contact:
Mr. José Enrique Aguirre Valadés General Manager RENAL CARE DIALYSIS SOLUTIONS, S.A. DE C.V. Carretera a los Cues K.M. 2.2., Lote 23, Bodegas 5 y 6, Parque Tecnològico Innovaciòn Querètaro El Marquès, Querètaro 76246 Mexico Email:jeav@rcds.mx Phone: 521-5536558830
- Official Correspondent: ﻥ Brittany Valdez Nava Head of Quality Healthcare Innovations Catalysts 7811 Montrose Road, Suite 215 Potomac, Maryland 20854
- d. Date prepared June 6, 2022
- e. Date revised January 26, 2023
- f. Subject Device
Device Name: Device Classification Name: Regulation Number: Common Name: Device Class: Classification Product Code: Regulation Medical Specialty: 510(k) Review Panel:
NikkiCart Hemodialysis system and accessories 21 CFR 876.5820 Sodium Bicarbonate for Hemodialysis Class II KPO Gastroenterology/Urology Gastroenterology/Urology
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for RCDS Renal Care Dialysis Solutions S.A. de C.V. The logo features a stylized image of two kidneys to the left of the letters "RCDS" in a blue sans-serif font. Below the letters is the text "Renal Care Dialysis Solutions S.A. de C.V." in a smaller font, also in blue. The overall design is clean and professional, suggesting a focus on healthcare and medical solutions.
## Predicate Device
The NikkiCart cartridge is substantially equivalent to:
| 510(k) Number: | K013724 |
|-------------------------------|--------------------------------------------------------------|
| Device Name: | Gambro BiCart® |
| Applicant: | GAMBRO Renal Products |
| Device Classification Name: | Dialysate Concentrate For Hemodialysis (Liquid Or<br>Powder) |
| Regulation Number: | 21 CFR 876.5820 |
| Common Name: | Sodium Bicarbonate for Hemodialysis |
| Device Class: | Class II |
| Classification Product Code: | KPO |
| Regulation Medical Specialty: | Gastroenterology/Urology |
| 510(k) Review Panel: | Gastroenterology/Urology |
# h. Device Description
The NikkiCart is a single use polypropylene (PP) cartridge containing dry sodium bicarbonate powder, which enables on-line preparation of saturated Sodium Bicarbonate solution that, in conjunction with appropriate acid concentrate solution and dialysis water, creates a bicarbonate -based dialysis fluid that meets the Association for the Advancement of Medical Instrumentation (AAMI) guidelines.
When the NikkiCart is attached to the Nikkiso DBB-06 dialysis machine with a compatible bicarbonate cartridge holder, water is drawn by the dialysis machine through the cartridge, producing a saturated solution of sodium bicarbonate. The dialysis machine mixes the sodium bicarbonate solution with water and the acid concentrate to produce a bicarbonate-based dialysis fluid. The acid concentrate must have a dilution ratio of 1:44 also known as 45X to be used with the NikkiCart Sodium Bicarbonate cartridge.
- i. Intended Use
The NikkiCart cartridge is intended for preparation of hemodialysis solutions on theNikkiso dialysis systems. The NikkiCart must always be used with a suitable acid concentrate.
# Indications for Use
The NikkiCart cartridge is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute renal failure, chronic renal failure, or acute intoxication with dialyzable substances. The NikkiCart cartridge is intended to be used as one component in the preparation of dialysate according to a physician's prescription in a 3-stream proportioning machine on the Nikkiso DBB-06 dialysis system equipped with a compatible bicarbonate cartridge holder.
- j. Statement of Substantial Equivalence
The NikkiCart is substantially equivalent to the predicate GAMBRO BiCart® (K013724) regarding the following:
- Indications for use
- Technological characteristics ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for RCDS Renal Care Dialysis Solutions SA de CV. The logo features a stylized image of two kidneys on the left, followed by the letters "RCDS" in a large, sans-serif font. Below the letters is the text "Renal Care Dialysis Solutions SA de CV" in a smaller font.
- Chemical composition ●
- Materials of construction ●
Table 1 below present a comparison of the NikkiCart cartridge to the predicate BiCart (K013724) cartridge. The only functional difference between BiCart and the NikkiCart is the quantity of sodium bicarbonate contained in each cartridge. The different quantities of sodium bicarbonate provide for different lengths of treatment; the resulting saturated sodium bicarbonate solution is used in the same way.
- k. Comparison Table
Table 1: Comparison table between BiCart (K013724) and the NikkiCart
| | Predicate Device:<br>BiCart<br>(K013724) | Proposed Device: NikkiCart | Discussion |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Fill Weight | 720g and 1250g | 720g and 900g | Substantially<br>Equivalent |
| Dimensional/Geo<br>metrical<br>Measurements | Total Height - 236.01mm<br>BODY Nipple Height - 10.27 mm<br>BODY Nipple Diameter – 9.90mm<br>CAP Nipple Height - 10.30mm<br>CAP Nipple Diameter – 9.89mm | Total Height → 235.26mm<br>BODY Nipple Height → 10.04mm<br>BODY Nipple Diameter → 9.98mm<br>CAP Nipple Height → 10.02mm<br>Cap Nipple Diameter → 9.82mm | Substantially<br>Equivalent |
| Machine<br>Interface<br>Dimensions | No data available | Lower Arm O-ring(interface<br>with nipple):<br>● Distance: 8.52mm<br>● Diameter: 9.49mm<br>Upper Arm O-ring (interface with<br>nipple):<br>● Distance to bottom - O-<br>ring: 4.37mm<br>● Diameter 1 (interface with<br>nipple): 9.48mm<br>● Distance: 10.06mm<br>Diameter 2 (Seal Upper /Lower<br>area when closed): 16.19mm | |
| Duration of Use | 720g:<br>● 500ml/min dialysis flow<br>rate: 6h 45min<br>● 700ml/min dialysis flow<br>rate: 4h 50 min<br>1250g:<br>● 500ml/min dialysis flow<br>rate: 12h 26min<br>● 600ml/min dialysis flow<br>rate: 10h 22min<br>● 700ml/min dialysis flow<br>rate: 8 h 53 min | 720g:<br>● 500ml/min dialysis flow<br>rate: 6h 45min<br>● 600ml/min dialysis flow<br>rate: 5h 37 min<br>900g:<br>● 500ml/min dialysis flow<br>rate: 8h 41min<br>● 800ml/min dialysis flow<br>rate: 5h 25min | Substantially<br>Equivalent |
| Acid Dilution<br>Ratio | 1:44 | 1:44 | Identical |
| Final Dialysate<br>Conductivity | Final dialysate Conductivity<br>13.7mS/cm with Centrisol Acid<br>concentrate SB-111. | Final dialysate Conductivity<br>13.7mS/cm with Centrisol Acid<br>concentrate SB-111 | Identical |
| Fill Water<br>Temperature | 33.0°C to 40.0°C | 34.0°C to 40.0°C | Substantially<br>Equivalent |
| Compatible<br>Hemodialysis<br>Models | Baxter Phoenix and AK98<br>hemodialysis models | Nikkiso DBB-06 hemodialysis<br>model | |
| The variation in the sizes offered does not raise any questions of the safety and effectiveness of our device<br>because these sizes will accommodate the Nephrologist prescription for the patient's hemodialysis treatment,<br>e.g., Treatment time dialysate flow rate and Sodium Bicarbonate. | | | |
| Indications for<br>Use | To be used in bicarbonate<br>hemodialysis treatment for<br>patients suffering from acute<br>renal failure, chronic renal<br>failure, or acute intoxication<br>with dialyzable substances. | To be used in a bicarbonate<br>dialysis treatment for patients<br>suffering from acuterenal<br>failure, chronic renal failure, or<br>acute intoxicationwith<br>dialyzable substances. | Identical |
| Disposable | Yes | Yes | Identical |
| Sodium<br>Bicarbonate<br>Grade | USP & European<br>Pharmacopeia | USP & European<br>Pharmacopeia | Identical |
| Bicarbonate<br>Concentration | 38 mEq/L | 37 mEq/L | Substantially<br>Equivalent |
| The difference in the bicarbonate concentration does not raise new issues of safety and effectiveness because<br>the dialysis machine Sodium Bicarbonate is prescribed by the Nephrologist to be specific to the patient<br>medical status. | | | |
| Storage<br>Condition | Store below +40 °C (+104<br>°F) | Store below +30 °C (+86<br>°F) | Substantially<br>Equivalent |
| The difference in storage conditions does not raise new issues of safety and effectiveness because the bench<br>testing data showed that bicarbonates integrity was not compromised. | | | |
| Housing<br>Material | Polypropylene | Polypropylene | Identical |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for RCDS Renal Care Dialysis Solutions S.A. de C.V. The logo features a stylized drawing of two kidneys on the left, followed by the letters "RCDS" in a large, sans-serif font. The words "Renal Care Dialysis Solutions S.A. de C.V." are printed in a smaller font below the letters.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for RCDS Renal Care Dialysis Solutions SA de CV. The logo features a stylized image of two kidneys in blue, followed by the letters "RCDS" in a similar blue color. Below the letters, the full name of the company, "Renal Care Dialysis Solutions SA de CV," is written in a smaller font.
## Performance Data
The proposed device has been subjected to biocompatibility testing, functional performance testing, and stability testing to support safety and effectiveness. No clinical testing was performed.
The following endpoints were assessed to support the biological safety of the NikkiCart
- Cytotoxicity ●
- Sensitization ●
- Irritation ●
- Acute Systemic toxicity ●
- Material mediated pyrogenicity ●
- Hemocompatibility ●
Stability evaluations were conducted for the NikkiCart in support of the 24-month (2-year) shelf life. The 24-month shelf life is supported by real time stability evaluations.
Additionally, a chemical characterization assessment was conducted to assess the following endpoints to support the biological safety of the NikkiCart:
- Chronic systemic toxicity ●
- Mutagenicity/Carcinogenicity ●
Shipping and distribution verification testing was performed for the NikkiCart in accordance with ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems. Results support that the product's packaging can withstand the distribution environment.
There are no differences between the predicate device (K013724) and the proposed device that raise new issues of safety and effectiveness.
- m. Conclusion
The intended use, technological characteristics, chemical composition, and materials of construction of the NikkiCart sodium bicarbonate cartridge are substantially equivalent to the BiCart (K013724) bicarbonate cartridge the predicate device. The non-clinical data support the safety of the NikkiCart and demonstrate the device performs as intended. Differences between the NikkiCart Bicarbonate cartridge and the predicate do not raise new or modified issues of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.