FlexSys, FlexSys UVB EPL 308 nm, FlexSys UVB EPL lite 308 nm, FlexSys IR 1550 nm
K221623 · Gme German Medical Engineering GmbH · GEX · Mar 10, 2023 · General, Plastic Surgery
Device Facts
Record ID
K221623
Device Name
FlexSys, FlexSys UVB EPL 308 nm, FlexSys UVB EPL lite 308 nm, FlexSys IR 1550 nm
Applicant
Gme German Medical Engineering GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device FlexSys is a multi-modality skin surface treatment system. Depending on the chosen modules the intended use and the mode of operation varies. Modules UVB EPL 308 and UVB EPL 308 lite: The GME FlexSys System with module UVB EPL 308 or UVB EPL 308 lite is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only. Module Infrared 1550: The GME FlexSys System with Infrared 1550 module is indicated for use in dermatological procedures requiring the coagulation of soft tissue, as well as for skin resurfacing procedures. It is also indicated for treatment of dyschromia and cutaneous lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots), actinic keratosis, and melasma, and for treatment of periorbital wrinkles, acne scars and surgical scars.
Device Story
Multi-modality skin surface treatment system; base unit with touch-screen control; connects to interchangeable laser modules (UVB EPL 308, UVB EPL 308 lite, Infrared 1550) via electrical cables and optical fibers. Handpieces/scanners direct laser light onto skin via mirrors and lenses; spot size varies with distance. UVB modules modulate immune system/stimulate melanocytes; Infrared 1550 module heats water in skin to denature tissue and stimulate healing. Operated by clinicians in dermatological settings. Output affects tissue via coagulation or phototherapy; benefits include lesion removal, skin resurfacing, and treatment of dermatological conditions.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified through adherence to international safety standards for laser products (IEC 60825-1) and medical electrical equipment (IEC 60601-1, IEC 60601-2-22, IEC 60601-1-2).
Technological Characteristics
Multi-modality laser system; base unit with touch-screen interface; interchangeable handpieces/scanners; optical fiber delivery; 308nm UVB and 1550nm infrared wavelengths. Complies with IEC 60825-1 (laser safety), IEC 60601-2-22 (laser equipment safety), IEC 60601-1 (general safety), and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for treatment of psoriasis and vitiligo (UVB modules) on intact skin; dermatological procedures requiring soft tissue coagulation, skin resurfacing, treatment of dyschromia, cutaneous lesions (lentigos, solar lentigos, actinic keratosis, melasma), periorbital wrinkles, acne scars, and surgical scars (Infrared 1550 module).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Solta Medical Inc., Fraxel DUAL 1550/1927 Laser System (K130193)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
March 10, 2023
GME German Medical Engineering GmbH Dietmar Fischer Technical Director Dreikoenigstr. 6-8 Erlangen, Bavaria 91054 Germany
Re: K221623
Trade/Device Name: FlexSys, FlexSys UVB EPL 308 nm, FlexSys UVB EPL lite 308 nm, FlexSys IR 1550 nm Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, FTC Dated: December 6, 2022 Received: December 12, 2022
Dear Dietmar Fischer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221623
Device Name FlexSys
#### Indications for Use (Describe)
The device FlexSys is a multi-modality skin surface treatment system. Depending on the chosen modules the intended use and the mode of operation varies.
#### Modules UVB EPL 308 and UVB EPL 308 lite:
The GME FlexSys System with module UVB EPL 308 or UVB EPL 308 lite is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
#### Module Infrared 1550:
The GME FlexSys System with Infrared 1550 module is indicated for use in dermatological procedures requiring the coagulation of soft tissue, as well as for skin resurfacing procedures. It is also indicated for treatment of dyschromia and cutaneous lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots), actinic keratosis, and melasma, and for treatment of periorbital wrinkles, acne scars and surgical scars.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K221623
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.92. Summary preparation date 2023-03-08 [21 CFR 807.92(a)(1)].
#### A. Applicant Name and Address [21 CFR 807.92(a)(1)]
GME German Medical Engineering GmbH. Dreikoenigstrasse 6-8 91054 Erlangen Bavaria, Germany Tel: +49 9131 934159 10 Fax: +49 9131 934159 99
#### B. Contact Information
GME German Medical Engineering GmbH. Dreikoenigstrasse 6-8 91054 Erlangen Bavaria, Germany Contact person: Dr. Dietmar Fischer Dietmar.fischer@gmeonline.de
#### C. Device Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Trade Name: FlexSys Device Common Name: Laser Instrument for Dermatology Classification Name: Laser Instrument, Surgical Powered 21 CFR 878.4810 Product Code: GEX, FTC Device Classification: Class II
#### D. Predicate Devices [21 CFR 807.92(a)(3)]
The FleySys uses similar technology and physical output characteristics as the following predicate devices: K130193 Solta Medical Inc., Fraxel DUAL 1550/1927 Laser System K150752 GME ExSys 308 K171702, K191086 Clarteis Exciplex308nm
#### E. Device Description [21 CFR 807.92(a)(4)]
The device FlexSys consists of a base unit with up to two installed internal laser modules, handpieces and/or scanners for these modules and attachable applicators. All applicators, handpieces, and scanners are connected by an electrical cable.
The base unit contains most of the controls, such as the touch-screen, the on / off switch or the emergency stop button. In addition, the power cord, foot switch, and the door plug are connected there. The touch-screen serves as the main control panel.
{4}------------------------------------------------
The applicators, handpieces, and scanners direct the light onto the skin. Some applicators contain a hand switch on the front of the handle by which light emission can be triggered.
The scanners and handpieces of the internal modules are connected to the base unit by an electric signal cable and by an optical fiber which guides the aiming beam and working beam laser light to the handpiece/scanner in its core. The light is then guided onto the skin via several mirrors and lenses. Due to the round shape of the fiber core the resulting laser spot on the skin is circular. The size of the working beam spot corresponds to the nominal value when the distance tip touches the skin. For larger distances the spot size will increase, i.e. the laser beam is divergent.
During development and manufacturing applicable directives, legal requirements and standards for medical devices were considered. Our product complies to them.
# F. Indications for Use [21 CFR 807.92(a)(5)]
The device FlexSys is a multi-modality skin surface treatment system. Depending on the chosen modules the intended use and the mode of operation varies.
#### Modules EPL 308 and EPL 308 lite:
The GME Flex.Sys System with module EPL 308 lite is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
# Module Infrared 1550:
The GME Flex.Sys System with Infrared 1550 module is indicated for use in dermatological procedures requiring the coagulation of soft tissue, as well as for skin resurfacing procedures. It is also indicated for treatment of dyschromia and cutaneous lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots), actinic keratosis, and melasma, and for treatment of periorbital wrinkles, acne scars and surgical scars.
# G. Device Specifications and Comparison to Predicates [21 CFR 807.92(a)(6)]
The FlexSys Platform is a multi-modality skin treatment device that can be equipped with up to six different modules. Thus, the predicate devices will vary from module to module. Three of these modules have already been approved, see K203054. Thus this comparison to predicates only deals with the remaining three modules, the Infrared 1550mm module and the two UVB 308nm modules.
{5}------------------------------------------------
The predicate devices used to argue substantial equivalence are the GME ExSys 308 (K150752) and Clarteis Exciplex (K171702, K191086) for the UVB 308nm modules, and the Solta Fraxel DUAL (K130193) for the Infrared 1550nm module.
The equivalence comparisons for the UVB 308 nm wavelengths use the predicates GME ExSys 308 and Clarteis Exciplex.
The equivalence comparisons for the infrared 1550 nm laser wavelength use the predicate Solta Fraxel DUAL.
| Characteristic | GME FlexSys<br>System<br>Module EPL 308<br>"FlexSys EPL 308" | GME FlexSys<br>System<br>Module EPL 308 lite<br>"FlexSys EPL 308 lite" | GME ExSys 308<br>"ExSys 308" | Clarteis<br>Exciplex308nm<br>"Exciplex" |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Applicable 510(k)s | NA | NA | K150752 | K171702, K191086 |
| Panel/ | General and Plastic<br>Surgery | General and Plastic<br>Surgery | General and Plastic<br>Surgery | General and Plastic<br>Surgery |
| Product Code/ | FTC | FTC | FTC | FTC |
| Regulation<br>Number | 21 CFR 878.4630 | 21 CFR 878.4630 | 21 CFR 878.4630 | 21 CFR 878.4630 |
| Indications for Use<br>Statement | The GME FlexSys<br>System with module<br>EPL 308 is intended<br>to be used for the<br>treatment of<br>psoriasis and<br>vitiligo. It is to be<br>used on intact skin<br>only | The GME FlexSys<br>System with module<br>EPL 308 lite is<br>intended to be used<br>for the treatment of<br>psoriasis and<br>vitiligo. It is to be<br>used on intact skin<br>only | The GME ExSys<br>308 System is<br>intended to be used<br>for the treatment of<br>psoriasis and<br>vitiligo. It is to be<br>used on intact skin<br>only. | The Exciplex308nm<br>is intended to be<br>used for the<br>treatment of<br>psoriasis, vitiligo,<br>atopic dermatitis,<br>and leukoderma. |
| Classification | Class II | Class II | Class II | Class II |
| Common Name | Ultraviolet lamp for<br>dermatologic<br>disorders | Ultraviolet lamp for<br>dermatologic<br>disorders | Ultraviolet lamp for<br>dermatologic<br>disorders | Ultraviolet lamp for<br>dermatologic<br>disorders |
| Mechanism of<br>Action | UVB light<br>penetrates the skin<br>and modulates the<br>immune system and<br>stimulates<br>melanocytes to<br>produce melanin | UVB light<br>penetrates the skin<br>and modulates the<br>immune system and<br>stimulates<br>melanocytes to<br>produce melanin | UVB light<br>penetrates the skin<br>and modulates the<br>immune system and<br>stimulates<br>melanocytes to<br>produce melanin | UVB light<br>penetrates the skin<br>and modulates the<br>immune system and<br>stimulates<br>melanocytes to<br>produce melanin |
# 1. UVB modules
# 2. Infrared 1550 module
| Characteristic | GME FlexSys System | Solta Fraxel DUAL<br>1550/1927 Laser System |
|----------------|----------------------------|---------------------------------------------|
| - | Module Infrared 1550 | |
| - | "FlexSys Infrared<br>1550" | "Fraxel" |
{6}------------------------------------------------
| Applicable<br>510(k)s | NA | K130193 |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Panel/ | General and Plastic<br>Surgery | General and Plastic Surgery |
| Product Code/ | GEX | GEX |
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Indications for<br>Use Statement | The GME FlexSys<br>System with Infrared<br>1550 module is indicated<br>for use in dermatological<br>procedures requiring the<br>coagulation of soft tissue,<br>as well as for skin<br>resurfacing procedures. It<br>is also indicated for<br>treatment of dyschromia<br>and cutaneous lesions,<br>such as, but not limited to<br>lentigos (age spots), solar<br>lentigos (sun spots),<br>actinic keratosis, and<br>melasma, and for<br>treatment of periorbital<br>wrinkles, acne scars and<br>surgical scars. | The Fraxel 1550 nm laser is<br>indicated for use in<br>dermatological procedures<br>requiring the coagulation of<br>soft tissue, as well as for skin<br>resurfacing procedures. It is<br>also indicated for treatment of<br>dyschromia and cutaneous<br>lesions, such as, but not<br>limited to lentigos (age spots),<br>solar lentigos (sun spots),<br>actinic keratosis, and<br>melasma, and for treatment of<br>periorbital wrinkles, acne<br>scars and surgical scars. |
| Classification | Class II | Class II |
| Common Name | Laser surgical instrument<br>for use in general and<br>plastic surgery and in<br>dermatology | Laser surgical instrument for<br>use in general and plastic<br>surgery and in dermatology |
| Mechanism of<br>Action | Heats water in skin<br>through the absorption of<br>light. The heating<br>denatures the skin and<br>stimulates a healing<br>reaction by the skin. | Heats water in skin through<br>the absorption of light. The<br>heating denatures the skin and<br>stimulates a healing reaction<br>by the skin. |
#### H. Performance Data [21 CFR 807.92(b)(2)]
The Guidance Document, Laser Products – Conformance with IEC 60825-1 and IEC 60601-2-22 (Laser Notice 56) January 19, 2018 was used.
Testing reports were submitted for the following standards:
- IEC 60825-1 Edition 2.0 2007-03: Safety of Laser Products Part 1: Equipment -Classification, and Requirements
- IEC 60601-2-22 Edition 3.1 2012-10: Medical Electrical Equipment Part 2-22: -Particular Requirements for Basic Safety and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment
- -IEC 60601-1:2005, AMD1:2012: Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance
{7}------------------------------------------------
- -IEC 60601-1-2:2014: Medical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Compatibility – Requirements and Tests
- I. Conclusion [21 CFR 807.92(b)(3)]
The GME FlexSys System is substantially equivalent to the predicate devices; in terms of technology, function and indications for use. There are no new questions of safety or efficacy raised by the introduction of the FlexSys System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.