The Wearable Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
Device Story
Wearable Breast Pump (model 1203) is a battery-powered, hands-free, electric breast pump for home use by lactating women. Device consists of a pump motor unit, breast shield, collection bottle, and accessories. Rotary electric pump generates negative pressure to express milk into the integrated bottle. User controls suction via physical interface or mobile companion app. App provides real-time milk monitoring, battery status, elapsed time, and pumping history. Device is worn inside a nursing bra. Benefits include hands-free operation and portability. Healthcare providers do not operate the device; it is for patient self-use.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1, 60601-1-2, 60601-1-11), battery safety (IEC 62133), biocompatibility (ISO 10993-5, 10993-10), and performance testing (vacuum pressure, cycle rate, backflow, use life, and battery status).
Technological Characteristics
Rotary electric pump; Li-Ion battery powered; 12-253 mmHg suction range; 8 expression levels; two-phase expression. Food-grade materials compliant with 21 CFR 174-179. Connectivity via mobile app. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133, ISO 10993.
Indications for Use
Indicated for lactating women to express and collect milk from their breasts. Intended for single-user, over-the-counter use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
K212180 — Wearable Breast Pump (Model S12) · Shenzhen TPH Technology Co., Ltd. · Feb 11, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 9, 2022
Mayborn (UK) Limited % Shelley Li Offical Correspondent Landlink Healthcare Technology (Shanghai) Co., Ltd. Room 1308, Baohua International Plaza, 555 West Guangzhong Road Shanghai, 200072 China
K221598 Trade/Device Name: Wearable Breast Pump Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: October 31, 2022 Received: October 31, 2022
Dear Shellev Li:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K221598
Device Name Wearable Breast Pump
Indications for Use (Describe)
The Wearable Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) summary - K221598
#### l. Submitter
MAYBORN (UK) LIMITED MAYBORN HOUSE BALLIOL BUSINESS PARK, NEWCASTLE UPON TYNE NE12 8EW UNITED KINGDOM
#### Manufacturer:
Jiangsu Xinbei Electrical Appliances Co., Ltd. No. 115 Xinjin Road, Xinwu District, Wuxi City, Jiangsu Province
#### Contact person
Sean Neasham Group Quality Manager Phone: +44 191 250 4509 E-mail: sneasham@mayborngroup.com
Date of preparation: December 7, 2022
#### II. Proposed Device
| Trade Name: | Wearable Breast Pump |
|--------------------|----------------------------|
| Common name: | Powered breast pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulation Name | Powered breast pump |
| Regulatory Class: | Class II |
| Product code: | HGX (Pump, Breast, Powered |
| Review Panel | Obstetrics/Gynecology |
#### III. Predicate Device
| 510(k) Number: | K181863 |
|-----------------|------------------------------|
| Trade name: | Elvie Pump |
| Common name: | Powered breast pump |
| Classification: | Class II |
| Product Code: | HGX |
| Manufacturer | Jabil Circuit (Shanghai) LTD |
The predicate device has not been subject to a design related recall.
{4}------------------------------------------------
#### IV. Device description
The Wearable Breast Pump (model: 1203) is an electric breast pump system comprised of a single or double portable unit that integrates the pump body and milk collection bottle. It consists of a pump (motor unit) and includes a breast shield, bottle, seal, valve, spout, and bra adjuster. All components (minus the pump) are reusable and may be manually cleaned.
The Wearable Breast Pump includes rotary electric pump technology which generates negative pressure on the nipple to express milk, which is collected in the integrated milk collection bottle. It is designed to work in the user's nursing bra and has a rechargeable battery so it can be used hands-free without external power cords or milk collection tubes.
The Wearable Breast Pump is a battery-powered electro-mechanical device that can be controlled through the physical interface on the device or through a mobile companion app, which also provides real-time milk monitoring, pump battery life, pumping time elapsed, and pumping history information.
All milk contacting components are constructed out of food grade materials that are compliant with 21 CFR 174-179..
#### V. Indications for use
The Wearable Breast Pump is a powered breast pump to be used by lactating women to express and collect mills from their breasts. It is intended for a single user.
#### VI. Comparison of technological characteristics with the predicate device
The intended use and key technological characteristics of the subject and predicate device are compared in the table below.
| Item | Proposed device | Predicate device<br>(K181863) | Comparison |
|---------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------|
| Trade name | Wearable Breast<br>Pump | Elvie Pump | NA |
| Product Code | HGX | HGX | Same |
| Regulation<br>No. | 21 CFR 884.5160 | 21 CFR 884.5160 | Same |
| Classification | Class II | Class II | Same |
| Indications<br>for use | The Wearable Breast<br>Pump is a powered<br>breast pump to be<br>used by lactating | The Elvie Pump is a<br>powered breast pump<br>to be used by<br>lactating women to | Same |
| | women to express<br>and collect milk from<br>their breasts. It is<br>intended for a single<br>user. | express and collect<br>milk from their<br>breasts.<br>The Elvie Pump is<br>intended for a single<br>user. | |
| Intended Use | Express milk from<br>breast | Express milk from<br>breast | Same |
| Environments<br>of Use | Home | Home | Same |
| Single User<br>device | Yes | Yes | Same |
| Single /<br>Double<br>Pumping | Both | Both | Same |
| Patient<br>Population | Lactating women | Lactating women | Same |
| Over the<br>Counter | Yes | Yes | Same |
| Power<br>Source | Power adapter<br>(battery charger) and<br>Li-Ion Battery | USB cable (battery<br>charger) and Li-Ion<br>Battery | Similar |
| Adjustable<br>Suction<br>Levels | Yes | Yes | Same |
| Backflow<br>Protection | Yes | Yes | Same |
| Suction<br>Strength | 12-253 mmHg | 40-220 mmHg | Different |
| Maximum<br>Suction<br>Strength | 253 mmHg | 220 mmHg | Different |
| Expression<br>levels | 8 | 7 | Different |
| Pump type | Rotary Electric Pump | Piezoelectric Pump | Same |
| Two-phase<br>Expression | Yes | Yes | Same |
| User Controls | On pump body and/or<br>through app | On pump body and/or<br>through app | Same |
| Control<br>Mechanism | Microcontroller | Microcontroller | Same |
| Mobile App | Yes | Yes | Same |
{5}------------------------------------------------
{6}------------------------------------------------
The subject and predicate device have identical indications for use statements and have the same intended use. The subject and predicate device have different technological characteristics, including different pump types, suction specifications, maximum suction pressure specifications, and power sources. However, the different technological characteristics do not raise different safety and effectiveness questions.
#### VII. Non-Clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications and that the subject device is substantially equivalent to the predicate device, as follows.
- Electrical Safety, Electromagnetic Compatibility, and Wireless Technology a)
- IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical . equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020 (second edition) ● Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance-Collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home
{7}------------------------------------------------
healthcare environments
- . IEC 62133:2012 – Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
- Documentation provided in accordance with the FDA Guidance document, . " Radio Frequency Wireless Technology in Medical Devices" dated August 14, 2013.
### b) Biocompatibility
Biocompatibility testing in accordance with the FDA guidance "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" dated September 4, 2020. Testing included the following assessments:
- Cytotoxicity per ISO 10993-5:2009 Biological evaluation of medical devices -. Part 5: Tests for in vitro cytotoxicity
- Sensitization per ISO 10993-10:2010 Biological evaluation of medical devices -. Part 10: Tests for irritation and skin sensitization
- Irritation per ISO 10993-10:2010 Biological evaluation of medical devices Part . 10: Tests for irritation and skin sensitization
The user-contacting materials were shown to be non-cytotoxic, non-irritating, and non-sensitizing.
#### c) Software Validation
Software validation testing was conducted for a "moderate level of concern" as recommended in the 2005 FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
#### d) Performance Testing
- Vacuum pressure and cycle rate testing was conducted at all settings and . demonstrated that the device met its specifications.
- Backflow testing was conducted to demonstrate that liquid does not backflow . into the tubing/pump.
- Use life testing was conducted to demonstrate that the device maintains its . specifications throughout its proposed use life.
- Battery performance testing was conducted to demonstrate that the battery ●
{8}------------------------------------------------
remains functional during its stated battery use-life.
- . Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
### IX. Conclusion
The performance testing described above demonstrates that the subject device is as safe and effective as the predicate device and supports a determination of substantial equivalence to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.